K233996 · Ulthera, Inc. · OHV · Feb 22, 2024 · General, Plastic Surgery
Device Facts
Record ID
K233996
Device Name
Ulthera System (UC-1 Control Unit PRIME)
Applicant
Ulthera, Inc.
Product Code
OHV · General, Plastic Surgery
Decision Date
Feb 22, 2024
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4590
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Ulthera System is intended to apply focused ultrasound energy to the body to achieve temporary changes in the physical appearance of the skin. The Ulthera System is indicated for use as a non-invasive dermatological aesthetic treatment to: • Lift the eyebrow • Lift lax submental (beneath the chin) and neck tissue, which can also affect the appearance of lax tissue in the submental and neck regions • Improve lines and wrinkles of the décolleté The Ulthera® System, in conjunction with the Ulthera® DeepSEE® transducer, allows for ultrasonic visualization of depths up to 8 mm below the surface of the skin. The indicated use of the imaging is to visualize the dermal and subdermal layers of tissue to: • Ensure proper coupling of the transducer to the skin • Confirm appropriate depth of treatment such as to avoid bone
Device Story
Ulthera System uses micro-focused ultrasound energy to induce thermal coagulation in dermal/subdermal tissue; results in skin contraction, lifting, and wrinkle reduction. System comprises control console, handpiece, and interchangeable transducers. Operator uses integrated touchscreen and supplemental ultrasound imaging to visualize tissue layers up to 8 mm deep; ensures proper transducer coupling and avoids bone contact. Used in clinical settings by trained medical professionals. Subject device features updated industrial design, modernized GUI, and hardware upgrades for electromedical compliance. Output is visual feedback for clinician and therapeutic ultrasound energy for patient. Benefits include non-invasive aesthetic improvement of skin laxity and wrinkles.
Clinical Evidence
Bench testing only. No clinical data provided. Performance verified via software validation, usability testing (IEC 62366-1), mechanical cyclic testing, packaging/transit testing (ASTM D4169/D4332), and acoustic reliability testing. Electrical safety and EMC verified per IEC 60601 series.
Technological Characteristics
Focused ultrasound stimulator; console-based system with handpiece and interchangeable transducers. 18.5" touchscreen (1920x1080). Windows 10 OS. Biocompatible patient-contacting surfaces. Compliant with IEC 60601-1, 60601-1-2, 60601-1-6, 60601-2-37, and 60601-2-62. Power: 100-240 VAC, 50/60 Hz.
Indications for Use
Indicated for adult patients as a non-invasive dermatological aesthetic treatment to lift the eyebrow, lift lax submental and neck tissue, and improve lines and wrinkles of the décolleté. Includes ultrasonic visualization of dermal/subdermal layers up to 8 mm to ensure transducer coupling and confirm treatment depth to avoid bone.
Regulatory Classification
Identification
A Focused Ultrasound Stimulator System for Aesthetic Use is a device using focused ultrasound to produce localized, mechanical motion within tissues and cells for the purpose of producing either localized heating for tissue coagulation or for mechanical cellular membrane disruption intended for noninvasive aesthetic use.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Focused Ultrasound Stimulator System for Aesthetic Use.” See § 878.1(e) for the availability of this guidance document.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out as "U.S. Food & Drug Administration".
February 22, 2024
Ulthera, Inc. Kim Kelly Sr. Director, Regulatory Affairs 1840 S. Stapley Dr. Suite 200 Mesa, Arizona 85204
Re: K233996
Trade/Device Name: Ulthera System (UC-1 Control Unit PRIME) Regulation Number: 21 CFR 878.4590 Regulation Name: Focused Ultrasound Stimulator System For Aesthetic Use Regulatory Class: Class II Product Code: OHV Dated: January 23, 2024 Received: January 23, 2024
Dear Kim Kelly:
We have reviewed vour section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{2}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Mark Mark Trumbore -S Trumbore -S Date: 2024.02.22 15:01:25 -05'00'
Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices
{3}------------------------------------------------
Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
#### 510(k) Number (if known) K233996
Device Name
Ulthera System (UC-1 Control Unit PRIME)
Indications for Use (Describe)
The Ulthera System is intended to apply focused ultrasound energy to the body to achieve temporary changes in the physical appearance of the skin.
The Ulthera System is indicated for use as a non-invasive dermatological aesthetic treatment to:
· Lift the eyebrow
• Lift lax submental (beneath the chin) and neck tissue. which can also affect the appearance of lax tissue in the submental and neck regions
· Improve lines and wrinkles of the décolleté
The Ulthera® System, in conjunction with the Ulthera® DeepSEE® transducer, allows for ultrasonic visualization of depths up to 8 mm below the surface of the skin. The indicated use of the imaging is to visualize the dermal and subdermal layers of tissue to:
· Ensure proper coupling of the transducer to the skin
- · Confirm appropriate depth of treatment such as to avoid bone
Type of Use (Select one or both, as applicable)
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
|----------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|
Description 3.0 (Part 2: Support & Support 2)
I he-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# 510(k) Summary
#### APPLICANT
| Company's Name: | Ulthera, Inc. |
|--------------------|-------------------------------------------------------|
| Company's Address: | 1820 South Stapley Drive. Suite 200<br>Mesa, AZ 85204 |
| Telephone: | (984) 286-9190 |
| Contact Person: | Kim Kelly, Senior Director, Regulatory Affairs |
| Date Prepared: | February 22, 2024 |
#### DEVICE
| Device Name: | Ulthera System (UC-1 Control Unit PRIME) |
|-----------------------|--------------------------------------------------------------------------------|
| Classification Name: | Focused Ultrasound Stimulator System for Aesthetic Use |
| Classification Name: | 21 C.F.R § 878.4590, Focused Ultrasound Stimulator Use |
| Regulatory Class: | Class II |
| Product Codes: | OHV, IYO |
| Applicable Guidances: | Focused Ultrasound Stimulator System for Aesthetic Use |
| Applicable Guidances: | Marketing Clearance of Diagnostic Ultrasound Systems and<br>Transducers |
| Applicable Guidances: | Content of Premarket Submissions for Software Contained<br>in Medical Devices. |
# PREDICATE DEVICE
Ulthera System, Ulthera, Inc., K180623
{6}------------------------------------------------
#### DEVICE DESCRIPTION
The Ulthera® System consists of the Ulthera® Control Unit (with system software), a handpiece with cable, and interchangeable transducers. The device produces controlled tissue coagulation below the skin surface (epidermis) within the first few millimeters of tissue (dermis) using highly focused, low-energy ultrasound deposition. The Ulthera® System directs micro-focused acoustic waves to the treatment area at desired depths without affecting or requiring a secondary action to protect the skin surface. The operator may also use the device's supplemental imaging capability to visualize the treatment area and aid in assuring full/proper skin contact of the Ulthera® System transducer to the skin in the target area.
#### INTENDED USE / INDICATIONS FOR USE
The Ulthera System is intended to apply focused ultrasound energy to the body to achieve temporary changes in the physical appearance of the skin.
The Ulthera System is indicated for use as a non-invasive dermatological aesthetic treatment to:
- Lift the eyebrow .
- . Lift lax submental (beneath the chin) and neck tissue, which can also affect the appearance of lax tissue in the submental and neck regions
- Improve lines and wrinkles of the décolleté ●
The Ulthera® System, in conjunction with the Ulthera® DeepSEE® transducer, allows for ultrasonic visualization of depths up to 8 mm below the surface of the skin. The indicated use of the imaging is to visualize the dermal and subdermal layers of tissue to:
- Ensure proper coupling of the transducer to the skin
- . Confirm appropriate depth of treatment such as to avoid bone
## SUMMARY OF TECHNOLOGICAL CHARACTERISTICS
The purpose of this 510(k) notification relates to modifications to the console of the Ulthera System. Low-intensity, highly focused ultrasound is the main technological principle of both the subject and the predicate device. In both devices, focused ultrasound energy is delivered below the skin and produces discrete points of thermal coagulation that results in contraction of the skin, which produces a lifting or tightening effect and improves appearance of lax tissue, lines, and wrinkles. Both the subject and predicate device share the following similar technological elements:
- · Identical ultrasound signal energy and ultrasound treatment guidelines
{7}------------------------------------------------
- · Same main components, including a console with integrated touchscreen, connected handpiece, and interchangeable transducers
- · An imaging mode of operation to visualize the treatment area and aid in assuring full/proper skin contact of the transducer to the skin
The following differences between the subject and predicate device do not significantly affect clinical functionality or performance specifications of the device and have been verified and tested:
- · Updates to the console's industrial design and modernization of the graphics within the graphical user interface (GUI)
- Hardware updates to comply with the latest electromedical standards and address end-of-life components
Comparison between the subject and predicate device are summarized in Table 1 below:
| | Ulthera Modified Console | Ulthera Console K180623 |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Subject Device | (Predicate) |
| Intended Use | The Ulthera System is intended to apply<br>focused ultrasound energy to the body to<br>achieve temporary changes in the<br>physical appearance of the skin. | The Ulthera System is intended to apply<br>focused ultrasound energy to the body to<br>achieve temporary changes in the<br>physical appearance of the skin. |
| Indications for<br>Use | The Ulthera® System is indicated for use<br>as a non-invasive dermatological<br>aesthetic treatment to: | The Ulthera® System is indicated for<br>use as a non-invasive dermatological<br>aesthetic treatment to: |
| | ● lift the eyebrow | ● lift the eyebrow |
| | ● lift lax submental (beneath the<br>chin) and neck tissue; which can<br>also affect the appearance of lax<br>tissue in the submental and neck<br>regions | ● lift lax submental (beneath the<br>chin) and neck tissue; which<br>can also affect the appearance<br>of lax tissue in the submental<br>and neck regions |
| | ● improve lines and wrinkles of<br>the décolleté. | ● improve lines and wrinkles of<br>the décolleté. |
| | The Ulthera® System in conjunction<br>with the Ulthera® DeepSEE transducer<br>allows for ultrasonic visualization of<br>depths up to 8 mm below the surface of<br>the skin. The indicated use of the<br>imaging is to visualize the dermal and<br>subdermal layers of tissue to: | The Ulthera® System in conjunction<br>with the Ulthera® DeepSEE transducer<br>allows for ultrasonic visualization of<br>depths up to 8 mm below the surface of<br>the skin. The indicated use of the<br>imaging is to visualize the dermal and<br>subdermal layers of tissue to: |
| | ● ensure proper coupling of the<br>transducer to the skin | ● ensure proper coupling of the<br>transducer to the skin |
| | ● confirm appropriate depth of<br>treatment such as to avoid bone | ● confirm appropriate depth of<br>treatment such as to avoid bone |
| | | Table 1: Substantial Equivalence Table Comparing Subject & Predicate Devices | | | | |
|--|--|------------------------------------------------------------------------------|--|--|--|--|
{8}------------------------------------------------
| User Population | Treatment of adult patient population by trained medical professional | Treatment of adult patient population by trained medical professional |
|----------------------------------------------|------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| Main System Components | • Control console<br>• Handpiece<br>• Transducers | • Control console<br>• Handpiece<br>• Transducers |
| Additional Components Required for Operation | • ACLF Power Cord with pigtail adapter<br>• USB Access Key | • Power cord<br>• USB Access Key |
| Dimensions<br>(l x w x h) | Height: <16.69" (424 mm)<br>Width: 19.4" (493.6 mm)<br>Depth: 13.1" (333 mm) | Height: 15.3" (389 mm)<br>Width: 16.5" (419 mm)<br>Depth: 13.0" (330 mm) |
| Weight | Weight: ≤27 lbs (12.2 kg) | Weight: 22 lbs (10 kg) |
| Display | 18.5" screen; aspect ratio of 16:9 with<br>1920 x 1080 resolution | 15" screen; aspect ratio of 4:3 with 1024<br>x 768 resolution |
| Power Source | 100-240 VAC, 50/60 Hz, 3A max<br>Fuse: (2) 5x20mm, 6.3A fast acting,<br>250V | 100-240 VAC, 50/60 Hz, 3A max<br>Fuse: (2) 5x20mm, 6.3A fast acting,<br>250V |
| Biocompatibility | Biocompatible user contacting surfaces | Biocompatible user contacting surfaces |
| Software | version 2.1.2030 | version 1700 |
| Operating<br>System | Windows 10 | Windows XP |
| Electrical<br>Safety & EMC<br>standards | Compliant with relevant IEC standards<br>for console and ACLF power cord | Compliant with relevant IEC standards<br>for console |
## PERFORMANCE DATA
The following non-clinical data were provided to support the substantial equivalence of the subject device to the predicate device. In all instances, the subject device functioned as intended.
#### Biocompatibility
Biocompatibility of the user-contacting components of the device was verified in accordance with ISO 10993-1.
## Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC of the subject device were verified in accordance with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-37, and IEC 60601-2-62.
#### Software Verification and Validation Testing
{9}------------------------------------------------
Software verification and validation testing were conducted, and documentation was provided as recommended by FDA Guidance, "Content of Premarket Submissions Device Software Functions." A basic documentation level was used, as a failure or flaw of any device software function(s) would not present a hazardous situation with a probable risk of death or serious injury prior to the implementation of risk control measures.
# Usability
Usability testing was conducted in accordance with IEC 62366-1 and FDA Guidance, "Applving Human Factors and Usability Engineering to Medical Devices." The console's portability, GUI visibility and layout, and display responsiveness were also verified. Testing demonstrated that clinicians were able to use the device in a representative environment and use conditions. No new risks were identified during the stimulated use study.
## Mechanical Testing
The display hinge of the console was verified via mechanical cyclic testing.
## Packaging, Transit, and Environmental Testing
Packaging and transit testing were conducted per ASTM D4169 and ASTM D4332. Environmental conditioning testing was conducted to verify device functionality at defined operating and storage conditions.
## Ultrasound Output Testing
Acoustic reliability was verified using simulated treatment over the device's service life. Power output and imaging were verified to be within specification.
## CONCLUSIONS
The modified console has undergone robust performance testing, including software, mechanical, electrical safety, electromagnetic compatibility, packaging, environmental, and usability testing. Imaging and acoustical testing were also conducted to verify the ultrasound signal. These non-clinical performance tests of the Ulthera System demonstrated the modified console continues to operate as intended and that it is as safe and effective as its predicate for the proposed indications for use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.