The VasoGuard devices are intended for use in the non-invasive evaluation of peripheral vascular pathology in patients. The devices are not intended to replace other means of evaluating vital testing, are not intended to be used in neonatal applications, and are not intended to be used inside the sterile field. The intended use by qualified and trained medical professionals in hospitals, clinics, and physician offices by prescription or on the order of a medical doctor.
Device Story
VasoGuard is a family of non-invasive peripheral vascular diagnostic systems (V2, V4, V6, V8, V10) used in hospitals, clinics, and physician offices by trained medical professionals. The device inputs include physiologic signals from pneumatic pressure cuffs (segmental pressures/PVR), PPG sensors, and 4MHz/8MHz CW Doppler probes. It also incorporates a touchless infrared skin thermometer. The system processes these signals to calculate Ankle Brachial Index (ABI), segmental blood pressures, and record blood flow waveforms (Doppler, PPG, PVR). The device connects to a Windows-based PC via USB, where software manages patient databases, exam protocols, and displays waveforms and quantitative reports. Clinicians use these outputs to assess vascular pathology, such as claudication or occlusions, aiding in clinical decision-making for peripheral vascular disease. The system benefits patients by providing non-invasive diagnostic data to guide treatment.
Clinical Evidence
Bench testing only. No clinical data. Performance was validated using simulators and live signals to compare waveform quality, sensitivity, and accuracy against the predicate device. Compliance with IEC 60601, biocompatibility, acoustic output, EMC, and environmental standards was verified.
Technological Characteristics
Aluminum enclosure; pneumatic pressure channels (2-10); PPG sensors; 4MHz/8MHz CW Doppler probes (piezoelectric); touchless IR thermometer. Connectivity: USB to Windows-based PC. Software: Windows 10 Enterprise LTSC. Sterilization: Not applicable (non-sterile). Standards: IEC 60601.
Indications for Use
Indicated for non-invasive evaluation of peripheral vascular pathology in patients. Not for neonatal use or use within sterile fields. For use by trained medical professionals in clinical settings via prescription.
Regulatory Classification
Identification
An ultrasonic transducer is a device applied to the skin to transmit and receive ultrasonic energy that is used in conjunction with an echocardiograph to provide imaging of cardiovascular structures. This device includes phased arrays and two-dimensional scanning transducers.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
July 19, 2024
CorVascular Diagnostics, LLC % Paul Dryden President ProMedic Consulting, LLC 131 Bay Point Dr. NE St. Petersburg, Florida 33704
Re: K233976
Trade/Device Name: VasoGuard (V10, V8, V6, V4, V2) Regulation Number: 21 CFR 870.2880 Regulation Name: Ultrasonic transducer Regulatory Class: Class II Product Code: JOP Dated: June 19, 2024 Received: June 20, 2024
Dear Paul Dryden:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, for Hetal B. Odobasic -S
LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K233976
Device Name VasoGuard (V10, V8, V6, V4, V2)
#### Indications for Use (Describe)
The VasoGuard devices are intended for use in the non-invasive evaluation of peripheral vascular pathology in patients. The devices are not intended to replace other means of evaluating vital testing, are not intended to be used in neonatal applications, and are not intended to be used inside the sterile field. The intended use by qualified and trained medical professionals in hospitals, clinics, and physician offices by prescription or on the order of a medical doctor.
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# Submission Type
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with 21CFR Part 807.92.
# Device Name
VasoGuard® V10, V8, V6, V4, V2
# Owner/Submitter
807
| 7.92(a)(1) | |
|-------------------|-----------------------------------------|
| Company Name: | CorVascular Diagnostics, LLC |
| | 600 Twelve Oaks Center Drive, Suite 102 |
| | Wayzata, MN 55391 |
| Telephone: | (763) 333-1540 |
| Contact Person: | Spencer Lien |
| Contact Title: | Managing Director |
| Preparation Date: | December 8, 2023 |
### Device Name
| 807.92(a)(2) | |
|----------------------|-------------------------------|
| Device Trade Name: | VasoGuard V10, V8, V6, V4, V2 |
| Common Name: | CW Doppler, PPG, VPR / PVR |
| Classification Name: | Transducer, Ultrasonic |
| Regulation Number: | 21 CFR 870.2880 |
| Product Code: | JOP |
| Class: | Class II |
| Review Panel: | Cardiovascular |
## Substantial Equivalence
807.92(a)(3)
| 510(k): | K111416 |
|---------------|--------------------------------------|
| Trade Name: | Falcon/Pro, Falcon/Quad, Falcon/ABI+ |
| Company Name: | Viasonix Ltd |
{4}------------------------------------------------
# Device Description
## 807.92(a)(4)
The VasoGuard is a family of products designed for non-invasive peripheral vascular diagnostic testing. It uses Doppler probes and photoplethysmography (PPG) sensors positioned on the body to measure physiologic signals and report data to the interpreting clinician. The system consists of up to 10 independent pneumatic pressure channels, up to five PPG ports, up to three Doppler ports that support 4MHz and 8MHz continuous wave (CW) Doppler probes, and a touchless temperature sensor. The V series indicates the products are all made from the same parts with the only differences being that certain parts are not installed when assembled during manufacturing or some features are not enabled.
The VasoGuard family consists of five different configurations of the same device, each varying in the number of pressure channels and sensor ports accessible through one of two available enclosures. The Full-Size enclosure supports Models V6, V8 and V10, and the Mini enclosure supports Models V2 and V4:
| Model V10: | 10 Pressure Channels / 5 PPG / 3 Doppler |
|------------|------------------------------------------|
| Model V8: | 8 Pressure Channels / 4 PPG / 2 Doppler |
| Model V6: | 6 Pressure Channels / 4 PPG / 1 Doppler |
| Model V4: | 4 Pressure Channels / 4 PPG / 1 Doppler |
| Model V2: | 2 Pressure Channels / 2 PPG / 0 Doppler |
All VasoGuard models contain the same main printed circuit board (PCB), manifold PCB(s), built-in power supply, built-in USB hub, and connect to a dedicated Windows-based medical grade PC via USB. Internally the models all utilize the same manifold assemblies and only differ in the quantity installed. The models include physical connections for control of 10 BP cuffs simultaneously, up to five PPG sensors simultaneously, one of up to three Doppler probes (4 MHz or 8 MHz), one USB camera, one IR remote control, and one touch-free infrared skin thermometer.
The two enclosures are made of the same aluminum material and share a common design. The Mini enclosure is approximately 20% smaller footprint. The Mini enclosure differs from the Full-Size enclosure only by the pressure channel connections on the front of the device instead of the back. Each model differs only in the number of pressure, Doppler and PPG ports on front and back panels of the device.
The VasoGuard software is pre-installed on the Windows PC. It controls all the models and automatically recognizes which model is connected thereby exposing only software capabilities only available on that model.
Each model includes certain components and accessories in addition to the VasoGuard device, including:
- Medical Grade Touchscreen PC
- . Windows® 10 Enterprise LTSC
- . Washable Keyboard with Trackpad
- . Mobile Cart with Height-adjustable Rolling Cart
- . Set of Pneumatic Hoses with Articulating Support Arm
{5}------------------------------------------------
- . Set of Blood Pressure Cuffs (Shenzhen Vistar Medical Supplies Co., Ltd. K152468)
- . Set of PPG Sensors
- . Doppler Probes (4 MHz and/or 8 MHz)
- . USB Camera
- . Infrared Remote Control and Receiver
- Touch-free Infrared Skin Thermometer (Tecnimed SRL - 510(k) K122412)
#### VasoGuard V10 System
Image /page/5/Picture/9 description: The image shows a medical diagnostic machine on a mobile cart. The machine has a large monitor at the top, a keyboard and other controls on the middle shelf, and a diagnostic unit with wires and connectors on the bottom shelf. The cart is white and gray and has wheels for easy movement.
Pneumatic Hose and Articulating Support Arm
Image /page/5/Figure/11 description: The image shows two different views of a pneumatic tube system. The left side of the image shows a technical drawing of the system, with labels pointing to the pneumatic tube, locking levers, retaining clips, junction box, and mounting bolt. The right side of the image shows the system installed on a wall, with the pneumatic tube connected to a junction box and a monitor.
VasoGuard V10 Device – Front
#### VasoGuard V10 Device – Back
VasoGuard V4 Device - Back
g
Image /page/5/Picture/14 description: The image shows the front and back of a device labeled as "VasoGuard V10 Device". The front of the device has several ports labeled "Doppler" and "PPG", each with multiple connection points. The back of the device features multiple USB ports, a power input, and other connection interfaces.
VasoGuard Device V10, V8, V6 Dimensions: 4″H x 16″W x 13″D
Image /page/5/Picture/16 description: The image shows the front of a VasoGuard V4 device. The device has several ports labeled "Doppler", "PPG", and "Pneumatic". Each port has multiple inputs with different colored rings around them. There are also small icons of a person near some of the ports.
VasoGuard Device V4, V2 Dimensions: 4"H x 11"W x 12"D
Image /page/5/Picture/18 description: The image shows the text "CorVascular | VasoGuard 510k Submission". The text is arranged horizontally, with "CorVascular" on the left, followed by a vertical line, and then "VasoGuard 510k Submission" on the right. The text is in a simple, sans-serif font and is presented against a plain white background.
{6}------------------------------------------------
### VasoGuard Accessories
Image /page/6/Picture/3 description: The image shows a white tray filled with medical equipment. There is a white handheld device with a screen, a silver remote control, two black cylindrical objects, two black cables with blue and black connectors, and four black cables with silver clips. The equipment is arranged neatly in the tray, which has a black foam lining.
8 MHz Doppler 4 MHz Doppler PPG (Red, Green, Yellow, Blue, Gray) Remote Control Temperature Meter
Image /page/6/Picture/5 description: The image shows a blue water reservoir bag with a tube coming out of it. The tube is connected to a series of colorful resistance bands. The bands are arranged in a curved pattern, with the ends of the bands facing the viewer.
Blood Pressure Cuffs and Tubing
The VasoGuard software features include patient database management; patient and exam search; importing and exporting of exams; facility management; importing of settings; exam protocol management, custom segment and vessel naming; customizable testing screens, reports, and graphs; backup and restore database; and keyboard shortcuts.
The quantitative measurements are the same for all VasoGuard models. One of the primary measurements is the Ankle Brachial Index (ABI). The ABI uses the Doppler probe to determine the ratio of the highest systolic pressure at the arm to the systolic pressure at the ankle. Systolic pressures are taken at both arms, and at the posterior tibial and dorsalis pedis arteries on each leg by inflating pressure cuffs past the point where Doppler sounds cease, then slowly deflating the cuffs until Doppler sounds return. The ABI is calculated by dividing the ankle pressure by the highest arm pressure.
Another primary measurement of the VasoGuard is segmental blood pressures. Segmental pressure studies are performed using the same method as an ABI but incorporate additional pressure cuffs wrapped at various positions on the leg. The software automatically places a cursor at the time location which is suspected as being the systolic pressure, yet it is the responsibility of the medical staff to modify the cursor location to define the correct segmental pressure.
Another main measurement of the VasoGuard is recording a waveform representing blood flow for each heartbeat. The system is capable of recording Doppler, PPG (photoplethysmography) and PVR (Pulse Volume Recording) waveforms. Each corresponds to the phases of the cardiac cycle, with a brisk upstroke and sharp peak that occur during systole, followed by a gradual downslope that occurs in diastole. A reflective wave, or dicrotic notch, represents reflected blood flow.
A high-level overview of continuous-wave (CW) Doppler is as follows. A CW Doppler transducer contains two piezoelectric elements: one to transmit the sound waves of constant frequency continuously and one to receive the echoes continuously. A single-element transducer cannot send
{7}------------------------------------------------
and receive at the same time. The sampling volume is restricted by the transmitted ultrasonic field (dependent on the frequency and focal properties of the sound beam) and the geometric arrangement of the elements. For the detection of a moving reflector located along the path of the transmitted beam, the resulting echo must strike the receiving crystal. The sensitive volume, or zone of sensitivity, is defined by the intersection of the transmitted ultrasound field and the reception zone. In essence then, each two-element transducer is focused to a particular depth. The two elements are tilted slightly to allow overlap between their respective fields of view (transmission and reception). The Doppler signal is amplified, filtered to remove unwanted low-frequency components caused by slow-moving structures such as vessel walls, and then routed to the speaker. The pitch of the audio output corresponds to the frequency shift between the transmitted and received sound waves and indicates the flow velocity within the vessel. As flow velocity becomes greater, a higher pitch is heard. A typical audio Doppler display for an artery exhibits a rhythmic rise and fall in the audible frequency due to the acceleration and deceleration of blood with systole and diastole.
A high-level overview of photoplethysmography (PPG) is as follows. PPG makes uses of low-intensity infrared (IR) light. When light travels through biological tissues it is absorbed by bones, skin pigments and both venous and arterial blood. Since light is more strongly absorbed by blood than the surrounding tissues, the changes in blood flow can be detected by PPG sensors as changes in the intensity of light. The voltage signal from PPG is proportional to the quantity of blood flowing through the blood vessels. The PPG waveform comprises a pulsatile ('AC') physiological waveform attributed to cardiac synchronous changes in the blood volume with each heartbeat, and is superimposed on a slowly varying ('DC') baseline with various lower frequency components attributed to respiration, sympathetic nervous system activity and thermoregulation.
A high-level overview of pulse volume recording (PVR) is as follows. A PVR is a graph of the pulsatile change in limb volume from blood flow using constant standard pressure. Users acquire these tracings using the same pressure cuffs used for segmental limb pressure measurement. Normal PVRs consist of a rapid upstroke with a sharp peak, a dicrotic notch, and a concave-up late diastolic component. A reduced amplitude from trough to peak has been used as a criterion for diagnosing vascular claudication. Abrupt changes in amplitude and contour can indicate occlusion between the two levels.
### Indications for Use
#### 807.92(a)(5)
The VasoGuard devices are intended for use in the non-invasive evaluation of peripheral vascular pathology in patients. The devices are not intended to replace other means of evaluating vital patient physiological testing, are not intended to be used in neonatal applications, and are not intended to be used inside the sterile field. The intended use by qualified and trained medical professionals in hospitals, clinics, and physician offices by prescription or on the order of a medical doctor.
{8}------------------------------------------------
# Summary of Technological Characteristics Compared to Predicate
807.92(a)(6)
| Feature | Falcon/Pro, Falcon/Quad, Falcon/ABI+ | VasoGuard<br>V2, V4, V6, V8, V10 | Diff |
|-------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
| 510(k) Number | K111416 | This application | |
| Classification | JOP | JOP | |
| Date | July 15, 2011 | Pending | |
| Manufacturer | Viasonix Ltd | CorVascular Diagnostics, LLC | |
| Indications for use | The Falcon/Pro, Falcon/Quad,<br>Falcon/ABI+ are intended for use in<br>the noninvasive evaluation of<br>peripheral vascular pathology in<br>patients.<br><br>The devices are not intended to<br>replace other means of evaluating vital<br>patient physiological processes, are<br>not intended to be used in fetal<br>applications, and are not intended to<br>be used inside the sterile field.<br><br>They are to be used by trained medical<br>personnel in hospitals, clinics and<br>physicians offices by prescription or<br>doctor's orders. | The VasoGuard devices are intended for<br>use in the non-invasive evaluation of<br>peripheral vascular pathology in<br>patients.<br><br>The devices are not intended to replace<br>other means of evaluating vital patient<br>physiological testing, are not intended to<br>be used in neonatal applications, and are<br>not intended to be used inside the sterile<br>field.<br><br>The intended use is attended use by<br>trained medical professionals in<br>hospitals, clinics, and physician offices by<br>prescription or on the order of a medical<br>doctor. | |
| Auto cuff inflation | Yes | Yes | |
| Pressure channels | 4, 4, 10 | 2, 4, 6, 8, 10 | (1) |
| Simultaneous PVR<br>measurements | 4, 4, 10 | 2, 4, 6, 8, 10 | (2) |
| PPG sensors | 4, 4, 5 | 2, 4, 4, 4, 5 | |
| Doppler probes | 4, 8 and 10 MHz CW | 4 and 8 MHz CW | (3) |
| Doppler ports | 0, 3, 3 | 0, 1, 1, 2, 3 | |
| Skin temperature<br>measurement | Yes, wired (auto) | Yes, touchless (manual) | (4) |
| Remote control operation | Yes, wired | Yes, (infrared) | (5) |
| Touchscreen controls | Yes | Yes | |
| Foot switch controls | Yes | No | (6) |
| Mouse and keyboard<br>controls | Yes | Yes | |
| Control of target inflation<br>pressures | Yes | Yes | |
| Control of pressures<br>deflation rate | Yes | Yes | |
| Acoustic testing | Track 1 | Track 1 | |
| Max Ispta.α 4 MHz | 691 mW/cm² | 366 mW/cm² | (7) |
| Max Ispta.α 8 MHz | 662 mW/cm² | 306 mW/cm² | (7) |
| Doppler spectral analysis | 256 FFT | 256 FFT | |
| Bidirectional Doppler<br>invert function | Yes | Yes | |
| Doppler volume controls | Yes | Yes | |
| Doppler envelope | Upper, lower, both, none | Upper, lower, both, none | (8) |
| Doppler, PVR, segmental<br>pressure, exercise, venous<br>reflux, Raynauds, TOS<br>testing | Yes | Yes | |
| Complete System Tested<br>to IEC 60601 | No | Yes | (9) |
{9}------------------------------------------------
The technical characteristics differences listed in the "Diff" column in the table above were compared to determine if any different issues of safety and/or efficacy exist. Upon review it was determined no safety or efficacy issues exist and no different questions of safety and effectiveness were raised. The differences listed do not constitute a new intended use.
### Non-Clinical Testing
#### 807.92(b)(1)
The VasoGuard (V10, V8, V6, V4, V2) range of non-invasive peripheral vascular diagnostic systems have been subjected to Biocompatibility, Acoustic Output, EMC Emissions and Immunity, Environmental Exposure, RFID Immunity, Electrical Safety, Performance, and Mechanical Safety testing by certified laboratories to ensure the devices comply with applicable industry and safety standards. Internally the VasoGuard has been subjected to unit, design verification, performance, accelerated aging, and validation testing to ensure the devices meet all their functional requirements and specifications.
Bench testing with the predicate device using simulators and live signals was conducted to ensure comparable waveform quality, sensitivity, and accuracy of reported values. The VasoGuard was found to be substantially equivalent to the predicate in all functions and parameters evaluated.
### Conclusions
#### 807.92(b)(3)
The conclusion drawn from these tests is that the VasoGuard V10, V8, V6, V2 range of non-invasive peripheral vascular diagnostic systems are equivalent in safety and efficacy to the predicate devices listed in the comparison table above.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.