AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K233968 · Mar 13, 2024
CINA-iPE
Avicenna.Ai
Retrospective clinical anonymized CT scan cases from multiple U.S. and OUS clinical sites
The retrospective clinical dataset was used to evaluate the standalone performance (sensitivity and specificity) of the CINA-iPE software in identifying incidental pulmonary embolisms in a real-world clinical setting.
Retrospective study; Standalone performance; Clinical anonymized cases; Multinational data
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Retrospective, multinational and blinded study; Retrospective, blinded, standalone performance evaluation
Patients undergoing contrast-enhanced CT scans (not dedicated CTPA protocol) including at least a part of the lung; 53.5% Male, 46.7% Female; Mean age 64.5 ± 15.8 years; Sample Size: 381 clinical anonymized cases; Number of Sites: Multiple U.S. and OUS clinical sites
Not applicable for this study
Sensitivity and Specificity for incidental pulmonary embolism (iPE) detection; Time-to-notification
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Incidental Pulmonary Embolism Detection
Deep learning model
Sensitivity and Specificity > 80%
Sensitivity 87.8%, Specificity 92.0%
5,429 cases acquired from US and France
—
Retrospective, multinational and blinded study of 381 clinical anonymized cases
3 (US-board-certified expert radiologists)
Indications for Use
CINA-iPE is a radiological computer-aided triage and notification software indicated for use in patients undergoing contrast-enhanced CT scans (not dedicated CTPA protocol) for other clinical indications than pulmonary embolism suspicion, including at least a part of the lung. The device is intended to assist hospital networks and trained radiologists in workflow triage by flagging and communicating suspected positive findings for incidental Pulmonary Embolism (iPE). The device is indicated for adults and transitional adolescents (18 to 21 years old but treated as adults). CINA-iPE uses an artificial intelligence algorithm to analyze images and highlight cases with detected incidental PE on a standalone application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected incidental PE findings. The device is not designed to detect PE in subsegmental arteries. Notifications include compressed preview images that are meant for informational purposes only and are not intended for diagnostic use beyond notification. The device does not alter the original medical image, and it is not intended to be used as a diagnostic device. The results of CINA-iPE are intended to be used in conjunction with other patient information and based on professional judgment to assist with triage/prioritization of medical images. Notified clinicians are ultimately responsible for reviewing full images per the standard of care.
Device Story
CINA-iPE is a radiological computer-aided triage and notification software. It receives contrast-enhanced CT scans (non-dedicated CTPA) via a medical image communications platform (e.g., CINA Platform). Deep learning algorithms process images to identify suspected incidental pulmonary embolism (iPE). The device segments lung parenchyma and locates the aorta/main pulmonary artery to assess contrast timing. It outputs notifications to a worklist application, including unannotated, compressed preview images for informational purposes. Used in hospital radiology departments; operated by radiologists. The device functions in parallel to standard-of-care workflows without altering original images or removing cases from the standard queue. By flagging suspected iPE cases, it enables worklist prioritization, potentially facilitating earlier clinical review and intervention.
Clinical Evidence
Retrospective, multinational, blinded study of 381 clinical cases (181 positive, 200 negative). Primary endpoints: global sensitivity and specificity. Results: Sensitivity 87.8% [95% CI: 82.2% - 92.2%]; Specificity 92.0% [95% CI: 87.3% - 95.4%]. Subgroup analysis provided for main, interlobar, lobar, and segmental arteries. Mean time-to-notification was 1.5 minutes. Performance met the 80% study goal and was comparable to the predicate device.
Technological Characteristics
Software-only device running on off-the-shelf server/workstation. Uses deep learning AI algorithms for image analysis. Interoperability via DICOM protocol. Integrates with PACS/worklist systems. No hardware components. Complies with NEMA PS 3.1-3.20 (DICOM).
Indications for Use
Indicated for adults and transitional adolescents (18-21 years) undergoing contrast-enhanced CT scans (non-dedicated CTPA) for indications other than suspected pulmonary embolism, where the scan includes at least part of the lung.
Regulatory Classification
Identification
Radiological computer aided triage and notification software is an image processing prescription device intended to aid in prioritization and triage of radiological medical images. The device notifies a designated list of clinicians of the availability of time sensitive radiological medical images for review based on computer aided image analysis of those images performed by the device. The device does not mark, highlight, or direct users' attention to a specific location in the original image. The device does not remove cases from a reading queue. The device operates in parallel with the standard of care, which remains the default option for all cases.
Special Controls
Radiological computer aided triage and notification software must comply with the following special controls: 1. Design verification and validation must include: i. A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations. ii. A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (e.g., improved time to review of prioritized images for pre-specified clinicians). iii. Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (e.g., subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment. iv. Standalone performance testing protocols and results of the device. v. Appropriate software documentation (e.g., device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results). 2. Labeling must include the following: i. A detailed description of the patient population for which the device is indicated for use. ii. A detailed description of the intended user and user training that addresses appropriate use protocols for the device. iii. Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (e.g., poor image quality for certain subpopulations), as applicable. iv. A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images. v. Device operating instructions. vi. A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (e.g., improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (e.g., confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians).(iii) Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (
*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment.(iv) Stand-alone performance testing protocols and results of the device.
(v) Appropriate software documentation (
*e.g.,* device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use;
(ii) A detailed description of the intended user and user training that addresses appropriate use protocols for the device;
(iii) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality for certain subpopulations), as applicable;(iv) A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images;
(v) Device operating instructions; and
(vi) A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (*e.g.,* confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Avicenna.AI % John Smith Partner Hogan Lovells US LLP 555 13th St. NW Washington, District of Columbia 20005
Re: K233968
March 13, 2024
Trade/Device Name: CINA-iPE Regulation Number: 21 CFR 892.2080 Regulation Name: Radiological computer aided triage and notification software Regulatory Class: Class II Product Code: QAS Dated: December 15, 2023 Received: December 15, 2023
Dear John Smith:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Jessica Lamb. Ph.D. Assistant Director Imaging Software Review DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K233968
Device Name CINA-iPE
#### Indications for Use (Describe)
CINA-iPE is a radiological computer-aided triage and notification software indicated for use in patients undergoing contrast-enhanced CT scans (not dedicated CTPA protocol) for other clinical indications than pulmonary embolism suspicion, including at least a part of the lung. The device is intended to assist hospital networks and appropriately trained radiologists in workflow triage by flagging and communicating suspected positive findings for incidental Pulmonary Embolism (iPE). The device is indicated for adults and transitional adolescents (18 to 21 years old but treated as adults).
CINA-iPE uses an artificial intelligence algorithm to analyze images and highlight cases with detected incidental PE on a standalone application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected incidental PE findings. The designed to detect PE in subsegmental arteries.
Notifications include compressed preview images that are meant for informational purposes only and are not intended for diagnostic use beyond notification. The device does not alter the original medical image, and it is not intended to be used as a diagnostic device.
The results of CINA-iPE are intended to be used in conjunction with other patient information and based on their professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are ultimately responsible for reviewing full images per the standard of care.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------|-----------------------------------------------|
| <span style="font-size: larger;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDEDThis section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K233968
#### 510(K) SUMMARY
#### AVICENNA.AI's CINA-IPE
## Submitter
#### Applicant:
AVICENNA.AI ZI Athelia, 297 Av. du Mistral Bât. A, 13600 La Ciotat France
#### Contact Person:
Stephane Berger Regulatory Manager Phone: +33 6 12 12 28 13 E-mail: stephane.berger@avicenna.ai
Date prepared: March 12, 2024
#### Device Identification
| Name of Device: | CINA-IPE |
|-----------------------|-----------------------------------------------------------------|
| Classification Name: | Radiological computer aided triage and<br>notification software |
| Regulation No: | 21 CFR § 892.2080 |
| Product Code: | QAS |
| Regulatory Class: | Class II |
| Classification Panel: | Radiology |
#### Predicate Device
The CINA-iPE device is substantially equivalent to the following predicate device with regard to indications for use, performance, and technological characteristics:
| 510(k): | K213886 |
|----------------------|--------------------------------------------------------------|
| Trade Name: | BriefCase |
| Manufacturer: | Aidoc Medical, Ltd. |
| Classification Name: | Radiological computer aided triage and notification software |
| Regulation No: | 21 CFR § 892.2080 |
| Product Code: | QAS |
| Regulatory Class: | Class II |
{4}------------------------------------------------
#### Device Description
CINA-iPE is a radiological computer aided triage and notification software device.
CINA-iPE runs on a standard "off the shelf" server/workstation and consists of an Image Processing Application, which can be integrated, deployed, and used with the CINA Platform (cleared under K200855) or other medical image communications devices. CINA-iPE receives contrast-enhanced CT scans (not dedicated CTPA protocol) including at least a part of the lung identified by the CINA Platform or other medical image communications device, processes them using deep learning algorithms involving the execution of multiple computational steps to identify the suspected presence of an incidental pulmonary embolism and generates results files to be transferred by CINA Platform or a similar medical image communications device for output to a PACS system or worklist prioritization.
To identify the suspected presence of pulmonary embolisms, the device uses a deep learning model trained end-to-end on 5.429 cases acquired from US and France, representing a distribution of PE sizes, locations and acquisition protocols, including multiple scanner models from Siemens, Philips, GE and Canon/Toshiba. Additional models are used to locate the aorta and main pulmonary artery, enabling assessment of the contrast timing. The lung's parenchyma is segmented to evaluate both the presence of the lungs in the field of view and to limit the region of interest for detecting the presence of pulmonary embolisms.
DICOM images are received, recorded and filtered before processing. The series are processed chronologically by running algorithms on each series to detect suspected positive findings of an incidental Pulmonary Embolism (iPE), then active notifications on the flagged series are sent to the Worklist Application.
The Worklist Application displays the active notification of new studies with suspected findings when they come in. All the contrast-enhanced CT studies received by CINA-iPE device are displayed in the worklist and those on which the algorithms have detected finding are marked with an icon (i.e., passive notification). In addition, a compressed, grayscale, unannotated image that is marked "not for diagnostic use" is displayed as a preview function. This compressed preview is meant for informational purposes only, does not contain any marking of the findings, and is not intended for diagnostic use beyond notification.
Presenting the radiologist with notification facilitates earlier triage by allowing prioritization of images in the PACS. Thus, the suspect case receives attention earlier than would have been the case in the standard of care image interpretation practice alone.
The CINA platform is an example of medical image communications platform for integrating and deploying the CINA-iPE image processing application. The medical image communications device (i.e., the technical platform) provides the necessary requirements for interoperability based on the standardized DICOM protocol and services to communicate with existing systems in the hospital radiology department such as CT modalities or other DICOM nodes (DICOM router or PACS for example). It is responsible for transferring, storing, converting formats, notifying of suspected findings and displaying medical device data such as radiological data. The medical image communications server includes the Worklist client application in which notifications from the CINA-iPE Image Processing application are received.
{5}------------------------------------------------
#### Intended Use / Indications for Use
CINA-iPE is a radiological computer-aided triage and notification software indicated for use in patients underqoing contrast-enhanced CT scans (not dedicated CTPA protocol) for other clinical indications than pulmonary embolism suspicion, including at least a part of the device is intended to assist hospital networks and trained radiologists in workflow triage by flagging and communicating suspected positive findings for incidental Pulmonary Embolism (iPE). The device is indicated for adults and transitional adolescents (18 to 21 years old but treated as adults).
CINA-iPE uses an artificial intelligence algorithm to analyze images and highlight cases with detected incidental PE on a standalone application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected incidental PE findings. The device is not designed to detect PE in subsegmental arteries.
Notifications include compressed preview images that are meant for informational purposes only and are not intended for diagnostic use beyond notification. The device does not alter the original medical image, and it is not intended to be used as a diagnostic device.
The results of CINA-iPE are intended to be used in conjunction with other patient information and based on professional judgment to assist with triage/prioritization of medical images. Notified clinicians are ultimately responsible for reviewing full images per the standard of care.
### Summary of Performance Data
The following performance data were provided in support of the substantial equivalence determination.
### Software Verification and Validation Testing
CINA-IPE complies with DICOM (Digital Imaging and Communications in Medicine) - Developed by the American College of Radiology and the National Electrical Manufacturers Association. NEMA PS 3.1 - 3.20.
Avicenna.Al conducted extensive performance validation testing and software verification and validation testing of the CINA-IPE device as standalone software. CINA-IPE is tested against its user needs and intended use by the successful execution of planned software verification and validation testing included in this submission.
Software performance, validation and verification testing demonstrated that the CINA-IPE met all design requirements and specifications associated with the intended use of the software.
#### Standalone Performance Testing
Avicenna. Al conducted a retrospective, multinational and blinded study with the CINA-IPE application with the primary endpoint to evaluate the software's performance in identifying incidental pulmonary embolisms (iPE) on contrast-enhanced CT images performed for another clinication than for pulmonary embolism (PE) evaluation (i.e. not dedicated CTPA protocol), in 381 clinical anonymized cases. The dataset included 53.5% Male and 46.7% Female. Mean ± SD age of patients included in the study was 64.5 ± 15.8 yo (range: 18 - 99 yo).
{6}------------------------------------------------
Multiple subgroups of interest were considered in the analysis. In fact, the validation data was acquired from different regions across the U.S. to account for race/ethnicity in the intended U.S. patient population. Additional scanner parameters considered were slice thickness, number of detector rows, and kVp ranges. Detailed subgroup analysis were reported in the labeling.
The data was provided from multiple U.S. and OUS clinical sites, with 56.4% (215) cases coming from U.S. clinical sources. The independence of the standalone validation dataset from the training data was ensured using data from independent sites and different time periods. There were 181 (47.5%) positive iPE (images with PE) cases and 200 (52.5%) negative iPE cases included in the analyses. The data was acquired primarily by 4 different scanner makers (GE-31.5%, Philips-28.3%, Siemens-26%, and Canon-13.9%) and 39 different scanner models.
Device Sensitivity [95% Cl] and Specificity [95% Cl] were computed against the groundtruth established by consensus of three US-board-certified expert radiologists.
As a primary endpoint, the global Sensitivity and Specificity were found to be 87.8% [95%C1: 82.2% -92.2%] and 92.0% [95%C1: 87.3% - 95.4%], respectively. These findings achieved the 80% performance goal of the study and are similar to the results reported for the predicate BriefCase device (Aidoc Medical) which demonstrated Sensitivity and Specificity of 89.7% (95%C1: 80.8% - 95.5%) and 90.1% [95%Cl: 81.5% - 95.6%], respectively.
Additionally, subgroup analysis was reported in terms of sensitivity 195%CII for each lesion location (main, interlobar, lobar and segmental) as presented in the following Table:
| Arterial Segment | Sensitivity<br>[95% CI] |
|------------------------|--------------------------|
| Main<br>(N = 55) | 96.3%<br>[87.5% - 99.6%] |
| Interlobar<br>(N = 73) | 94.5%<br>[86.6% - 98.5%] |
| Lobar<br>(N = 127) | 92.9%<br>[87.0% - 96.7%] |
| Segmental<br>(N = 179) | 88.3%<br>[82.6% - 92.6%] |
Table 1: Stratified statistical analysis (Sensitivity [95%Cl]) regarding arterial segments.
Please note that cases with more than one PE location were included in more than one category (e.g., a case with a main and a lobar PE was included in the "main" category and in the "lobar" category) Finally, the CINA-iPE prioritization and triage effectiveness were evaluated by the standalone percase processing time of the device (time-to-notification). The results are presented in Table 2 below:
{7}------------------------------------------------
| Time-to-Notification<br>(minutes) | MEAN ± SD | MEDIAN | 95% Cl | MIN | MAX |
|----------------------------------------------|-----------|--------|-------------|-----|-----|
| CINA-iPE<br>All cases<br>(N = 381) | 1.5 ± 0.5 | 1.4 | [1.4 - 1.5] | 0.3 | 2.7 |
| CINA-iPE<br>True Positive cases<br>(N = 159) | 1.5 ± 0.4 | 1.5 | [1.4 - 1.6] | 0.7 | 3.1 |
Table 2: Time-to-Notification for CINA-iPE Image Processing Applications
The mean [95% C1] time-to-notification for all included cases (n = 381) was estimated to be 1.5 [95% Cl: 1.4 - 1.5] minutes for CINA-iPE. When taking into account only true positive cases (n = 159), the mean [95% Cl] time-to-notification was 1.5 [95% Cl: 1.4 - 1.6] minutes for CINA-iPE and 4.7 [95% Cl: 4.4 - 5.1] minutes for the predicate device.
The performance testing of the CINA-iPE device establishes that the subject device is as safe and effective as the predicate device and compatible with the same clinical use, since the performance demonstrated the clinical effectiveness of the subject device and its ability to provide effective prioritization and triage that was substantially equivalent to that of the predicate device. This establishes that the CINA-iPE device achieves its intended use and is substantially equivalent to the predicate BriefCase device.
#### Substantial Equivalence
The subject CINA-iPE and the predicate BriefCase device are both intended to aid in prioritization and triage of radiological images of time sensitive findings for patient detection and diagnosis (i.e., incidental Pulmonary Embolism) based on the analysis of medical images acquired from radiological signal acquisition systems. The CINA device provides the CINA Platform in which the subject CINAiPE prioritization and triage application can be integrated, deployed and used. The labeling of the subject and the predicate devices clearly states that the devices are not for diagnostic use. All devices are software packages with similar technological characteristics and principles of operation, and incorporate deep learning Al algorithms that process images, and software to send notifications and to display unannotated preview images. In all three devices, the labeling instructs the user to further evaluate and diagnose based only on the original images in the local PACS.
The subject CINA-iPE and the predicate BriefCase device operate in parallel to the standard of care workflow in the sense that they do not change the original image, do not provide any marking on the output preview, and do not remove images from the standard of care FIFO queue; thus, not disturbing standard interpretation of the images by the attending radiologists. The subject and predicate devices achieve performance of the time-to-notification metric in similar ranges of time, and thus contribute similarly to effective triage and early involvement of the radiologist in evaluating suspected images of incidental pulmonary embolism.
The standalone performance and effectiveness assessment studies demonstrated that the CINA-iPE device performs as intended and substantially equivalent to the BriefCase predicate device.
The table below compares the key features of the subject and the predicate devices.
{8}------------------------------------------------
| Ltd) | | |
|-----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Subject device: CINA-IPE Software | Predicate device: BriefCase Software (K213886) | |
| Intended Use / Indications for Use | CINA-iPE is a radiological computer-aided triage and notification software indicated for use in patients undergoing contrast-enhanced CT scans (not dedicated CTPA protocol) for other clinical indications than pulmonary embolism suspicion, including at least a part of the lung. The device is intended to assist hospital networks and trained radiologists in workflow triage by flagging and communicating suspected positive findings for incidental Pulmonary Embolism (IPE). The device is indicated for adults and transitional adolescents (18 to 21 years old but treated as adults).<br>CINA-IPE uses an artificial intelligence algorithm to analyze images and highlight cases with detected incidental PE on a standalone application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected incidental PE findings. The device is not designed to detect PE in subsegmental arteries.<br>Notifications include compressed preview images that are meant for informational purposes only and are not intended for diagnostic use beyond notification. The device does not alter the original medical image, and it is not intended to be used as a diagnostic device.<br>The results of CINA-iPE are intended to be used in conjunction with other patient information and based on professional judgment to assist with triage/prioritization of medical | BriefCase is a radiological computer aided triage and notification software indicated for use in the analysis of contrast-enhanced chest CTs (not dedicated CTPA protocol) in adults or transitional adolescents age 18 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communication of suspect cases of incidental Pulmonary Embolism (IPE) pathologies. The device is intended to be used on single energy exams only.<br>BriefCase uses an artificial intelligence algorithm to analyze images and flag suspect cases on a standalone desktop application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for suspect cases. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device.<br>The results of BriefCase are intended to be used in conjunction with other patient information and based on their professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care. |
| | Subject device: CINA-IPE Software | Predicate device: BriefCase Software<br>(K213886) |
| | images. Notified clinicians are<br>ultimately responsible for reviewing<br>full images per the standard of care. | |
| User<br>population | Radiologist | Appropriately trained medical specialists |
| Anatomical<br>region of<br>interest | Chest or exams including at least a<br>part of the lung | Chest |
| Data<br>acquisition<br>protocol | Contrast-enhanced CT scans (not<br>dedicated CTPA protocol) containing<br>at least a part of the lung | Contrast-enhanced chest CTs (but not<br>dedicated CTPA protocol) |
| View DICOM<br>data | DICOM information about the<br>patient, study and current image | DICOM information about the patient,<br>study and current image |
| Segmentatio<br>n of region of<br>interest | No; device does not mark, highlight,<br>or direct users' attention to a specific<br>location in the original image | No; device does not mark, highlight, or<br>direct users' attention to a specific<br>location in the original image |
| Algorithm | Artificial intelligence algorithm with<br>database of images | Artificial intelligence algorithm with<br>database of images |
| Notification /<br>Prioritization | Yes | Yes |
| Preview<br>images | Presentation of a compressed,<br>grayscale, unannotated image that is<br>marked "not for diagnostic use" is<br>displayed as a preview function. | Presentation of a low-quality,<br>compressed grayscale preview image<br>that is captioned "Not for diagnostic use". |
| Alteration of<br>original<br>image | No | No |
| Removal of<br>cases from<br>worklist<br>queue | No. The device operates in parallel<br>with the standard of care, which<br>remains the default option for all<br>cases. | No. The device operates in parallel with<br>the standard of care, which remains the<br>default option for all cases. |
| Structure | -iPE image processing application<br>- Compatibility of use with the CINA<br>Platform device (worklist and Image<br>Viewer) or other medical image<br>communications device, | -AHS module/Orchestrator (image<br>acquisition).<br>- ACS module (image processing).<br>- Aidoc Worklist application for workflow<br>integration (worklist and non-diagnostic<br>basic Image Viewer) |
Table 3: Comparison of key features between CINA-iPE and the predicate device (Aidoc Medical
{9}------------------------------------------------
{10}------------------------------------------------
#### Conclusion
CINA-iPE and BriefCase have the same intended use and substantially similar indications, technological characteristics, and principles of operation. The following is a summary of the substantial equivalence comparison:
- -The predicate device is legally marketed.
- CINA-iPE has the same intended use as the predicate device, i.e., Aidoc Medical, Ltd.'s - BriefCase, and therefore it may be found substantially equivalent.
- -CINA-iPE and the predicate device have very similar indications for use.
- CINA-iPE has similar technological characteristics as the predicate and reference devices, i.e., the deployment of an artificial intelligence algorithm with a database of images.
- -Standalone performance study demonstrates that the CINA-iPE and the predicate BriefCase raise the same types of safety and effectiveness questions, namely, accurate detection of findings within the processed study.
Accordingly, the subject CINA-iPE device is substantially equivalent to the predicate BriefCase device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.