K233954 · B&E Korea Co., Ltd. · ELW · Feb 13, 2024 · Dental
Device Facts
Record ID
K233954
Device Name
Chemi-SiL (HB, MB, LB, LBS); Chemi-SiL Bite-Blu
Applicant
B&E Korea Co., Ltd.
Product Code
ELW · Dental
Decision Date
Feb 13, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3660
Device Class
Class 2
Indications for Use
Chemi-SiL (HB, MB, LB, LBS) - Crown and bridge impression, inlay and onlay impressions Chemi-SiL Bite-Blu - Impression of the occlusal surface
Device Story
Chemi-SiL is a hydrophilic, addition-type polyvinylsiloxane (PVS) elastomeric impression material. It is supplied as a two-part base and catalyst system in 50 mL cartridges. The device is used by dental professionals to capture precise impressions of teeth, gingiva, or occlusal surfaces. The material is self-curing upon mixing. It is a single-patient, single-use, non-sterile device. The clinician uses the resulting impression to create dental models for restorative procedures like crowns, bridges, inlays, and onlays. The material's physical properties, such as consistency, working time, and dimensional stability, ensure accurate reproduction of oral structures, facilitating the fabrication of well-fitting dental restorations.
Clinical Evidence
No clinical data. Bench testing only. Testing included ISO 4823 (appearance, consistency, working/setting time, detail reproduction, gypsum compatibility, dimensional change, elastic recovery, strain in compression) and ISO 10993 (cytotoxicity, skin sensitization, oral mucosa irritation, acute systemic toxicity). All results met preset criteria.
Technological Characteristics
Addition-type polyvinylsiloxane (PVS) elastomeric impression material. Components: poly(dimethylsiloxane), vinyl-terminated siloxanes, methyl hydrogen polysiloxanes, synthetic amorphous silica, hydrophilic fumed silica, platinum complex catalyst, pigments, and surfactant. Supplied in 50 mL dual-cartridge systems. Self-curing. Conforms to ISO 4823 and DIN 13903 standards.
Indications for Use
Indicated for dental impression procedures, specifically crown and bridge, inlay, and onlay impressions (Chemi-SiL HB, MB, LB, LBS) and occlusal surface impressions (Chemi-SiL Bite-Blu).
Regulatory Classification
Identification
Impression material is a device composed of materials such as alginate or polysulfide intended to be placed on a preformed impression tray and used to reproduce the structure of a patient's teeth and gums. The device is intended to provide models for study and for production of restorative prosthetic devices, such as gold inlays and dentures.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
February 13, 2024
B&E Korea Co., Ltd. Jayong Shin RA Representative 995-16. Baran-ro. Jeongnam-myeon Hwaseong-si, Gyeonggi-do 18515 SOUTH KOREA
Re: K233954
Trade/Device Name: Chemi-SiL (HB, MB, LB, LBS) and Chemi-SiL Bite-Blu Regulation Number: 21 CFR 872.3660 Regulation Name: Impression Material Regulatory Class: Class II Product Code: ELW Dated: December 15, 2023 Received: December 15, 2023
Dear Jayong Shin:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Michael E. Adjodha -S
Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Submission Number (if known)
K233954
Device Name
Chemi-SiL (HB, MB, LB, LBS) Chemi-SiL Bite-Blu
Indications for Use (Describe)
Chemi-SiL (HB, MB, LB, LBS) - Crown and bridge impression, inlay and onlay impressions
Chemi-SiL Bite-Blu
- Impression of the occlusal surface
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## K233954 510(k) SUMMARY
Date: Dec 15, 2023
### 1. SUBMITTER
B&E KOREA Co., Ltd. 995-16, Baran-ro, Jeongnam-myeon, Hwaseong-si, Gyeonggi-do, Republic of Korea TEL : +82-31-8077-2364 FAX : +82-31-8077-2365 Contact Name: Shin Jayong Email: jayong@bnekorea.com
## 2. DEVICE
·Trade Name: Chemi-SiL (HB, MB, LB, LBS), Chemi-SiL Bite-Blu
·Common Name: Impression Material
·Classification Name: Material, Impression
·Regulation Number: 872.3660
·Class: 2
·Classification Product Code: ELW
## 3. PREDICATE DEVICE
Chemi-SiL (HB, MB, LB, LBS) Vonflex STM, K152615
Vericom Co. Ltd
Chemi-SiL Bite-Blu
K170736 Hysil Impression Materials HySil Bite, OSSTEM Implant Co., Ltd.
{5}------------------------------------------------
### 4. DEVICE DESCRIPTION
Chemi-SiL of silicone impression materials are hydrophilic impression materials based on polyvinylsiloxane and it is used for impression of teeth or gingiva. It consists of three different consistencies (heavy body, light body, bite registration material) according to ISO 4823. And it is supplied base and catalyst in contrasting color. Each product is consisted of 1:1 base and catalyst component, packaged in two 50 mL catridges.
#### 5. INDICATIONS FOR USE
Chemi-SiL (HB, MB, LB, LBS)
- · Crown and bridge impression, inlay and onlay impressions
Chemi-SiL Bite-Blu
- Impression of the occlusal surface
#### 6. NON-CLINICAL TESTING
The following test articles were tested based on the referenced standard. All the test results met the preset test criteria.
- ISO 4823 Appearance and Component colors, Consistency, Working time, Detail reproduction, Setting time, Compatibility with gypsum, Linear dimensional change, Elastic recovery, Strain in compression
- ISO 7405 Cytotoxicity .
- ISO 10993-10 - Skin sensitization, Oral mucosa irritation
- ISO 10993-11 - Acute systemic toxicity
### 7. SUBSTANITAL EQUIVALENCE
| Subject device | Predicate Device | Discuss/Justify the Differences | |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | | |
| 510(k) Number | New | K152615 | - |
| Trade Name | Chemi-SiL (HB, MB, LB,<br>LBS) | Vonflex STM | - |
| Manufacturer | B&E KOREA Co., Ltd. | Vericom Co. Ltd | - |
| Common Name | Impression material | Impression material | Same |
| Device Class | 2 | 2 | Same |
| Product Code | ELW | ELW | Same |
| Indications for use | • Crown and bridge<br>impression, inlay and onlay<br>impressions | • Impression material in dual<br>phase impression technique<br>• Precise duplication of models<br>• Capturing multiple unit<br>impressions<br>• Impression of inlay, crown,<br>bridge and partial denture etc. | Same<br>The subject and<br>predicate device<br>have the same<br>intended use. Both<br>devices are used for<br>dental impression<br>techniques according<br>to their consistencies. |
| Physical State | Two product lines are provided<br>with different viscosity. (Heavy<br>body, Light body) | Various pastes with various<br>viscosity | Similar<br>Both subject and<br>predicate devices are<br>provided as putties<br>with various<br>elasticities.<br>The physical states<br>are similar, as<br>compared in physical<br>properties. |
| Structure | Addition type silicone based<br>elastomeric impression<br>materials | Addition type silicone based<br>elastomeric impression<br>materials | Same |
| Packaging | Primary packaging: Cartridges<br>Secondary packaging: Carton<br>box | Primary packaging: Cartridges<br>Secondary packaging: Carton<br>box | Same |
| Usage | Single patient, single use.<br>Not reusable | Single patient, single use.<br>Not reusable | Same |
| Sterility | Non-sterile | Non-sterile | Same |
| Handling System | Two part base/catalyst system | Two part base/catalyst system | Same |
| Type of Curing | Self-curing after mixing of base and catalyst part. | Self-curing after mixing of base and catalyst part. | Same |
| Physical properties | See the table below | See the table below | Similar<br>Individual values of<br>physical properties<br>are same or similar.<br>Every physical<br>property of predicate<br>and subject device<br>conforms to the<br>related standard. |
| Chemical composition | Poly(dimethylsiloxane), vinyl<br>terminated<br>Siloxanes and Silicones, di-ME,<br>ME Hydrogen, Hydrogen-<br>terminated Methyl hydrogen<br>polysiloxanes<br>Synthetic amorphous Silica<br>Hydrophilic Fumed Silica<br>Platinum complex<br>(3 % Pt by weight)<br>Pigments<br>surfactant | Vinyl siloxanes<br>Hydrogen polysiloxane<br>Silica<br>Fumed silica<br>Organo platinum complex<br>Pigments<br>Surfactant | Similar<br>Both device are<br>composed of vinyl<br>terminated siloxane,<br>cross-linking<br>siloxane, Filler,<br>platinum catalyst,<br>pigments and<br>surfactant. |
| Biocompatibility | Meets ISO 10993-1 for a<br>surface device contacting<br>mucosal membrane and dentin<br>for a short-term contact<br>duration (>24h) | Conforms with ISO 10993-1 | Same |
| Standards conformed | ISO 4823<br>ISO 10993-1 | ISO 4823<br>ISO 10993-1 | Same |
{6}------------------------------------------------
{7}------------------------------------------------
Physical properties
| - | Chemi-SiL (HB, MB, LB, LBS) | Vonflex STM | Standard | Discuss/Justify the | |
|------------------------------------------------------|-----------------------------|---------------------------|--------------------------------|------------------------------------------------------------------------|---------------------------------------------|
| | | | | | |
| | | | | | Differences |
| Working time | MB | 90 s | 2'15"(Heavy,<br>normal) | [ISO<br>4823:2021<br>7.3]<br>suggested<br>value of the<br>manufacturer | Equivalent |
| | HB | | 1'30"(Heavy,<br>Fast) | | |
| | LB | 90 s | 2'30"(Light,<br>normal) | | Equivalent |
| | LBS | | 1'30"(Light,<br>Fast) | | |
| Setting reaction time | MB | 210 s | 4'00" (Heavy,<br>normal) | - | Equivalent |
| | HB | | 2'30" (Heavy,<br>Fast) | | |
| | LB | 210 s | 4'00"(Light,<br>normal) | | Equivalent |
| | LBS | | 2'30"(Light,<br>Fast) | | |
| Consistency | MB | 32.75 mm | 34.76 mm<br>(Heavy,<br>normal) | [ISO<br>4823:2021<br>7.2] | Equivalent |
| | HB | | 34.78 mm<br>(Heavy, Fast) | <= 35 mm | |
| | LB | 48.83 mm | 43.43 mm<br>(Light, normal) | [ISO<br>4823:2021<br>7.2] | Equivalent |
| | LBS | | 42.02 mm<br>(Light, Fast) | > =36 mm | |
| Compatibility with<br>gypsum | MB | Reproduced for<br>0.05 mm | Reproduced for<br>0.05 mm | [ISO<br>4823:2021<br>7.6] | Equivalent |
| | HB | | | 50 µm should<br>be reproduced<br>without<br>interruption. | |
| | LB | Reproduced for<br>0.05 mm | Reproduced for<br>0.05 mm | | Equivalent |
| | LBS | | | | |
| Strain in<br>compression (Curve<br>of the shrinkage) | MB | 1.2 % | 2.56 % (Heavy,<br>normal) | [ISO 4823:<br>2021 7.8]<br>it should be<br>within<br>0.8~20 % | Similar<br>Both conforms<br>to the standard |
| | HB | | 2.38 % (Heavy,<br>Fast) | | |
| | LB | 3.1 % | 7.71 % (Light,<br>normal) | [ISO 4823:<br>2021 7.8]<br>within<br>2.0~20 % | Similar<br>Both conforms<br>to the standard |
| | LBS | | 5.24 % (Light,<br>Fast) | | |
| Linear Dimensional<br>Change | MB | 0.12 % | 0.11 % (Heavy,<br>normal) | [ISO<br>4823:2021<br>7.5]<br><= 1.5 % | Similar<br>Both conforms<br>to the standard |
| | HB | | 0.07 % (Heavy,<br>Fast) | | |
| | LB | 0.04 % | 0.08 % (Light,<br>normal) | | Similar<br>Both conforms<br>to the standard |
| | LBS | | 0.09 % (Light,<br>Fast) | | |
{8}------------------------------------------------
{9}------------------------------------------------
510(k) Summary of Vonflex STM is provided in Appendix A_K152615
| | Subject device | Predicate Device | Discuss/Justify the<br>Differences | |
|-------------------------|------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|
| 510(k) Number | New | K170736 | - | |
| Trade Name | Chemi-SiL Bite-Blu | Hysil Impression Materials<br>HySil Bite | - | |
| Manufacturer | B&E KOREA Co., Ltd. | OSSTEM Implant Co., Ltd. | - | |
| Common Name | Impression material | Impression material | Same | |
| Device Class | 2 | 2 | Same | |
| Product Code | ELW | ELW | Same | |
| Indications for use | • Impression of the occlusal<br>surface | HySil Bite is used for<br>impression as below.<br>- Taking occlusal surfaces<br>- Confirming occlusal surfaces<br>- Recording after putting the | Same | |
| | | | articulator | |
| | Physical State | Bite registration materials | Bite registration materials | Same |
| Structure | | Addition type silicone based<br>elastomeric impression<br>materials | Addition type silicone based<br>elastomeric impression<br>materials | Same |
| | Packaging | Primary packaging: Cartridge<br>Secondary packaging: Carton<br>box | Primary packaging: Cartridge<br>Secondary packaging: Carton<br>box | Same |
| Usage | | Single patient, single use.<br>Not reusable | Single patient, single use.<br>Not reusable | Same |
| | Sterility | Non-sterile | Non-sterile | Same |
| | Handling System | Two part base/catalyst system | Two part base/catalyst system | Same |
| | Type of Curing | Self-curing after Hand-<br>kneaded mixes | Self-curing after Hand-<br>kneaded mixes | Same |
| Physical properties | Colours | Pass | Pass | Same<br>The classification of<br>the subject and<br>predicate device is<br>complied with DIN<br>13903 |
| | Working time | 55 s | Pass | |
| | Minimum<br>period spent in<br>the mouth | 95 s | 1 min. 30 sec | |
| | Recovery after<br>Deformation | 0.0 mm | more than 0.1 mm | |
| | Load<br>endurance<br>during<br>bending(Flexu<br>ral strength) | 8.2 N(20 s) | Pass | |
| | Hardness<br>degree(HD) | 50 HD | 50 HD | |
| | Linear<br>dimensional<br>change | 0.3 % | -0.16 % | |
| Chemical<br>composition | Poly(dimethylsiloxane), vinyl<br>terminated | Vinylpolysiloxane | Similar | |
| | Hydrophilic fumed Silica<br>Synthetic amorphous Silica | | Both device are<br>composed of vinyl<br>terminated siloxane,<br>Filler, platinum<br>catalyst and<br>pigment. | |
| | Pigment | | | |
| | Platinum complex | | | |
| Biocompatibility | Meets ISO 10993-1 for a<br>surface device contacting<br>mucosal membrane and dentin<br>for a short-term contact<br>duration (>24h) | Conforms with ISO 10993-1 | Same | |
| Standards<br>conformed | DIN 13903<br>ISO 10993-1 | DIN 13903<br>ISO 10993-1 | Same | |
{10}------------------------------------------------
{11}------------------------------------------------
510(k) Summary of Hysil Impression Materials HySil Bite is provided in Appendix B K170736
#### 8. CONCLUSION
Subject devices have the same intended use and the principle of operation as the predicate devices. They are intended to perform as impression materials which meet the requirements according to ISO 4823. Both subject and predicate devices are classified under the same product code (ELW) and have equivalent indications of use (crown and bridge impressions, inlay and onlay impressions for Chemi-SiL (HB, MB, LB, LBS), and impression of the occlusal surface for Chemi-SiL Bite-Blu).
Both devices demonstrate similar physical properties and biocompatibilities with comparable performance specifications to the predicate devices. Information about working time, setting reaction time, consistency, compatibility with gypsum, strain in compression, and linear dimensional change is compared between Chemi-SiL (HB, MB, LB, LBS) and Vonflex STM. Colour, working time, minimum period spent in the mouth, recovery after deformation, load endurance (flexural strength), hardness degree (HD), and linear dimensional change are compared between Chemi-SiL Bite-Blu and Hysil Impression Materials HySil Bite. All subject
{12}------------------------------------------------
devices have shown better or equivalent performance compared to the predicate device, or have shown compliance with the device-specific standard (ISO 4823) at each endpoint.
In terms of chemical equivalence, the chemical constituents of both devices are analyzed and classified into comparable molecular subgroups. Even though the exact materials do not necessarily coincide, the essential composition of the same; vinyl-terminated siloxane, cross-linking siloxane, silica filler, platinum catalyst, pigments, and surfactant.
In terms of biological equivalence, the bench and biocompatibility testing performed demonstrate that any differences in their technological characteristics do not raise any new questions as to safety and effectiveness. Therefore, it is concluded that the subject devices are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.