SP NEWTRON CAN-A; MODULE NEWTRON CAN-A EMB; SUPPORT XINETIC; NWT CAN SLIM LED CORD G3 SHORT; NWT CAN SLIM LED CORD G3 LONG
K233922 · Satelec - Acteon Group · ELC · Mar 11, 2024 · Dental
Device Facts
Record ID
K233922
Device Name
SP NEWTRON CAN-A; MODULE NEWTRON CAN-A EMB; SUPPORT XINETIC; NWT CAN SLIM LED CORD G3 SHORT; NWT CAN SLIM LED CORD G3 LONG
Applicant
Satelec - Acteon Group
Product Code
ELC · Dental
Decision Date
Mar 11, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Scaling: Interdental junction treatment Tooth neck and subgingival treatment Treatment of large deposits Treatment of coating and tobacco stains Interproximal treatment Prosthesis conservative/restorative: Inlay/onlay condensation Amalgam plugging Loosening prostheses (bridge, crown, post, pivot ... ) Endodontics: Canal preparation Canal cleaning Canal filling Gutta percha condensation Treatment resumption Retro surgery Micro retro surgery Surface smoothing after burring Periodontics: Root planning Initial therapy Treatment of periodontal pockets Treatment of furcations Maintenance therapy
Device Story
The NEWTRON CAN A Module is an ultrasonic control module integrated into A-DEC dental delivery systems. It receives digital power and light commands from the delivery system's touchpad. The module converts low-voltage power into an ultrasonic electrical signal, which is transmitted to a SATELEC dental ultrasonic handpiece. The handpiece uses a piezoelectric ceramic transducer to convert this signal into mechanical micro-vibrations, which are delivered to a dental tip for clinical use. The device features a feedback system that automatically adjusts frequency and amplitude based on the resistance encountered by the tip. It is operated by dental professionals in a clinical setting. The output affects clinical decision-making by enabling precise removal of calculus, canal cleaning, and restorative procedures. Benefits include efficient, controlled ultrasonic treatment for various dental applications.
Clinical Evidence
Bench testing only. Performance testing compared the subject device to the predicate, confirming similar output frequencies (28-36 KHz), power output (10 mW to 12 W), and electrical safety/EMC compliance. Biocompatibility testing for irrigation tubing was conducted per ISO 10993 standards.
Technological Characteristics
Ultrasonic control module; piezoelectric transducer; 28-36 KHz frequency; 10 mW-12 W power output. Casing: UL94V-0 thermoplastic. Irrigation tubing: polyurethane/silicone. Connectivity: CAN bus digital interface for integration with dental delivery systems. Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), ISO 10993 (biocompatibility).
Indications for Use
Indicated for dental cleaning and periodontal therapy, including scaling, endodontic procedures, and restorative/prosthetic maintenance in patients requiring professional dental care.
Regulatory Classification
Identification
An ultrasonic scaler is a device intended for use during dental cleaning and periodontal (gum) therapy to remove calculus deposits from teeth by application of an ultrasonic vibrating scaler tip to the teeth.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health and Human Services and the U.S. Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.
March 11, 2024
SATELEC-ACTEON GROUP Noelle Zuccarello Quality Manager 124 Gaither Drive Suite # 140 Mt. Laurel, New Jersey 08054
Re: K233922
Trade/Device Name: SP NEWTRON CAN-A; MODULE NEWTRON CAN-A EMB; SUPPORT XINETIC; NWT CAN SLIM LED CORD G3 SHORT; NWT CAN SLIM LED CORD G3 LONG Regulation Number: 21 CFR 872.4850 Regulation Name: Ultrasonic Scaler Regulatory Class: Class II Product Code: ELC Dated: December 5, 2023 Received: December 13, 2023
Dear Noelle Zuccarello:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Bobak Shirmohammadi -S
For Michael E. Adjodha, M. ChE., RAC, CQIA Assistant Director
{2}------------------------------------------------
DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Submission Number (if known)
#### K233922
Device Name
- · SP NEWTRON CAN-A
- · MODULE NEWTRON CAN-A EMB
- · SUPPORT XINETIC
- · NWT CAN SLIM LED CORD G3 SHORT
- · NWT CAN SLIM LED CORD G3 LONG
#### Indications for Use (Describe)
Scaling:
- · Interdental junction treatment
- · Tooth neck and subgingival treatment
- · Treatment of large deposits
- · Treatment of coating and tobacco stains
- · Interproximal treatment
- · Prosthesis conservative/restorative:
- o Inlay/onlay condensation
- o Amalgam plugging
- o Loosening prostheses (bridge, crown, post, pivot ... )
### Endodontics:
- · Canal preparation
- · Canal cleaning
- · Canal filling
- · Gutta percha condensation
- · Treatment resumption
- · Retro surgerv
- · Micro retro surgery
- · Surface smoothing after burring
Periodontics:
- · Root planning
- · Initial therapy
- · Treatment of periodontal pockets
- · Treatment of furcations
- · Maintenance therapy
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
{4}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reductiori Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
| 510(k) Summary - K233922<br>[As Required by 21 CFR 807.92.c] | | |
|--------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|
| I - SUBMITTER | | |
| Device Submitter: | SATELEC-ACTEON GROUP<br>Z.I DU PHARE<br>17, AVENUE GUSTAVE EIFFEL MERIGNAC<br>Gironde, FR 33708 | |
| | Phone:<br>Fax: | 011-33-556-340-607<br>011-33-556-349-292 |
| Contact Person: | Noelle ZUCCARELLO<br>SATELEC<br>c/o ACTEON, Inc.<br>124 Gaither Drive, Suite 140 Mt.<br>Laurel, NJ 08054 | |
| | Telephone:<br>Fax:<br>E-Mail: | 856-222-9988 Ext. 280<br>856-222-4726<br>noelle.zuccarello@acteongroup.com |
| Date Prepared: | March 11, 2024 | |
| II - DEVICE | | |
| Name of Device: | SP NEWTRON CAN-A<br>MODULE NEWTRON CAN-A EMB<br>SUPPORT XINETIC<br>NWT CAN SLIM LED CORD G3 SHORT<br>NWT CAN SLIM LED CORD G3 LONG | |
| Common or Usual Name: | Ultrasonic Scaler | |
| Classification Name: | Ultrasonic Scaler (21 CFR 872.4850) | |
| Regulatory Class: | II | |
| Product Code: | ELC | |
## III – PREDICATE DEVICE
| Name of Device | Manufacturer | Product<br>Code | 510(k) number | Date Cleared |
|-------------------|--------------|-----------------|---------------|----------------|
| SP NEWTRON Module | SATELEC | ELC | K033764 | March 01, 2004 |
SP NEWTRON Module Predicate Device has not been subject to a design-related recall.
No reference devices were used in this submission.
Image /page/5/Picture/6 description: The image shows the Acteon logo. The logo consists of a blue triangle shape on the left and the word "acteon" in gray on the right. Below the word "acteon" is the phrase "Caring People. Improving Lives" in gray.
{6}------------------------------------------------
## IV - DEVICE DESCRIPTION
Key Device components included in this Submission:
- NEWTRON CAN A Module . (reference: X55650 and F55660)
- Mounting Bracket (reference: E55072)
- Handpiece Tubing 3 75" (refence: F01212)
- Handpiece Tubing 3 89" ● (refence: F01213)
#### Device Characteristics:
The NEWTRON CAN A Module is an Ultrasonic Control Module intended to be integrated in a Dental Delivery System. The NEWTRON CAN A Module contains an Embedded Software.
#### Environment of Use:
The NEWTRON CAN A Module is intended to be used in Dental Offices (installed in a Dental Delivery System).
#### Description of the Device:
The NEWTRON CAN A Module is an Ultrasonic Control Module intended to be integrated in a Dental Delivery System manufactured by the American Company named A-DEC. The A-DEC Dental Delivery System provides different function to the NEWTRON CAN A Module such as Ultrasonic Power Setting, Light Command through a communication frame. The communication frame is designed and imposed by A-DEC, the Manufacturer of the Dental Delivery System. The values of Ultrasonic Power Setting and the Light Command are coded and communication frame. Ultrasonic Power Setting and the Light Command are driven by the user with a touch pad embedded in the A-DEC Dental Delivery System.
The only parameters of the NEWTRON CAN A Module that can be modified by the A-DEC Dental Delivery System are the values of Ultrasonic Power Setting and the Light Command. The NEWTRON CAN A Module converts a low voltage Power Supply provided by the A-DEC Dental Delivery System into an Ultrasonic Electric Signal.
The NEWTRON CAN A Module is intended to be used in combination with a SATELEC Dental Ultrasonic Handpiece and a SATELEC range of Dental Tips. The electrical signal emitted by the NEWTRON CAN A Module supplies the Dental Ultrasonic Handpiece equipped with SATELEC Dental Tip. The amplitude of the ultrasonic micro-vibrations depends on the electrical Signal provided by the NEWTRON CAN A Module (amplitude). The SATELEC Dental Ultrasonic Handpiece is connected to the Dental Delivery System via a Handpiece cord. Mechanical micro- vibrations are transmitted to a Dental tip attached to the end of the SATELEC Dental Ultrasonic Handpiece.
The NEWTRON CAN A Module is intended to be used in combination with SATELEC Ultrasonic Handpieces previously cleared [NEWTRON Handpiece (K050895 – Apr 20, 2005, K113430 – Feb 23, 2012, K131997 and Nov 26, 2013), NEWTRON SLIM Handpiece (K131997 - Nov 26, 2013 and K132267 -Mar 10, 2014), NEWTRON SLIM B.LED Handpiece (K132267 - Mar 10, 2014 and K132322 - Mar 11, 2014), NEWTRON LED Handpiece (K071424 – Aug 24, 2007, K091331 – Dec 11, 2009 and K091252 – Jul 22, 2009)].
#### Scientific Principles:
The NEWTRON CAN A Module uses a control system of amplitude or "feedback system", according to the resistance encountered by the dental tip and the setting mode selected. Likewise, its "frequency tuning capability" finds and constantly adapts the oscillating frequency to supply to the dental ultrasonic handpiece.
Image /page/6/Picture/18 description: The image shows the Acteon logo. The logo consists of a blue triangle shape on the left and the word "acteon" in black on the right. Below the word "acteon" is the phrase "Caring People. Improving Lives" in a smaller, gray font.
{7}------------------------------------------------
### Principles of Operation:
The NEWTRON CAN A Module emits an electrical signal to the Dental Ultrasonic Handpiece and adapts it permanently in accordance to the effort and the encountered impedance of the Dental Ultrasonic Handpiece and the user settings. The Dental Ultrasonic Handpiece comprises a piezoelectric ceramic transducer, which converts the electrical signal into ultrasonic micro- vibrations. The amplitude of the ultrasonic micro-vibrations depends on parameters of the electrical Signal provided by the NEWTRON CAN A Module (amplitude, frequency). Mechanical micro- vibrations are transmitted to a Tip attached on the distal part of the ultrasonic Handpiece. The working part of the Tip moves on the clinical site.
#### Material of Use:
The casing of the NEWTRON CAN A Module is made of thermoplastic materials protected against the fire propagation (self-extinguishing). The thermoplastic materials are certified by the Underwriters Laboratories according to the UL94 standard. The thermoplastic materials are classified UL94V-0. Additionally, the Printed Circuit Boards (PCB) are classified UL 94V-0. Duration and type of contact are not applicable for NEWTRON CAN A Module.
The Handpiece Tubing Cod is made of Silicone and the Irrigation Tubing contained in the Handpiece Tubing is made of Polyurethane. Irrigation Tubing present an indirect contact less than 24 hours.
Key Performance Specifications / Characteristics of the Device:
| Output Frequency (KHz) : | 28 KHz to 36 KHz |
|--------------------------------------------------|--------------------------|
| Output Power in the Ultrasonic handpiece: | 10 mW to 12 W |
| Output Current in the Ultrasonic handpiece (mA): | 8 ± 10% to 100 +20% -10% |
Image /page/7/Picture/8 description: The image shows the Acteon logo. The logo consists of a blue triangle shape on the left and the word "acteon" in gray on the right. Below the word "acteon" is the phrase "Caring People. Improving Lives" in a smaller font size.
{8}------------------------------------------------
## V - INDICATION FOR USE
### Scaling:
- Interdental junction treatment ●
- Tooth neck and subgingival treatment ●
- Treatment of large deposits ●
- . Treatment of coating and tobacco stains
- Interproximal treatment ●
- Prosthesis conservative/restorative:
- Inlay/onlay condensation o
- Amalgam plugging O
- Loosening prostheses (bridge, crown, post, pivot ... ) o
## Endodontics:
- . Canal preparation
- Canal cleaning ●
- Canal filling .
- Gutta percha condensation ●
- Treatment resumption
- Retro surgery ●
- . Micro retro surgery
- Surface smoothing after burring
### Periodontics:
- Root planning ●
- Initial therapy
- Treatment of periodontal pockets
- Treatment of furcations ●
- . Maintenance therapy
Image /page/8/Picture/27 description: The image shows the Acteon logo. The logo consists of a blue triangle shape on the left and the word "acteon" in black on the right. Below the word "acteon" is the phrase "Caring People. Improving Lives" in a smaller font size.
{9}------------------------------------------------
# VI - COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
A comparison Between SATELEC NEWTRON CAN A and the Predicate Device is present in Table 01.
### Table 01 – Comparison Between SATELEC NEWTRON CAN A and the Predicate Device
| Trade /<br>Device<br>Name | New Device | Predicate Device | Result / Impact of<br>the differences on<br>Safety /<br>Effectiveness |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|
| Knumber | NEWTRON CAN A<br>K233922 | SP NEWTRON Module<br>K033764 | Identical |
| Intended Use | Scaling:<br>Interdental junction treatment Tooth neck and subgingival treatment Treatment of large deposits Treatment of coating and tobacco stains Interproximal treatment Prosthesis conservative/restorative:<br>Inlay/onlay condensation Amalgam plugging Loosening prostheses (bridge, crown, post, pivot...) Endodontics:<br>Canal preparation Canal cleaning Canal filling Gutta percha condensation Treatment resumption Retro surgery Micro retro surgery Surface smoothing after burring Periodontics:<br>Root planning Initial therapy Treatment of periodontal pockets Treatment of furcations Maintenance therapy | Scaling:<br>Interdental junction treatment Tooth neck and subgingival treatment Treatment of large deposits Treatment of coating and tobacco stains Interproximal treatment Prosthesis conservative/restorative:<br>Inlay/onlay condensation Amalgam plugging Loosening prostheses (bridge, crown, post, pivot...) Endodontics:<br>Canal preparation Canal cleaning Canal filling Gutta percha condensation Treatment resumption Retro surgery Micro retro surgery Surface smoothing after burring Periodontics:<br>Root planning Initial therapy Treatment of periodontal pockets Treatment of furcations Maintenance therapy | |
SATELEC ■ A company of ACTEON Group 17, avenue Gustave Eiffel = Zone Industrielle du Phare = 33700 MERIGNAC = FRANCE Tel. +33 (0) 556 340 607 ■ Fax. +33 (0) 556 349 292 = info@acteongroup.com
Image /page/9/Picture/6 description: The image shows the Acteon logo. The logo consists of a blue triangle shape on the left and the word "acteon" in black on the right. Below the word "acteon" is the phrase "Caring People. Improving Lives" in gray.
{10}------------------------------------------------
| | New Device | Predicate Device | Result / Impact<br>of the<br>differences on<br>Safety /<br>Effectiveness |
|---------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------|
| Trade / Device<br>Name | NEWTRON CAN A | SP NEWTRON Module | Identical |
| Knumber | K233922 | K033764 | Identical |
| Product Code | ELC | ELC | Identical |
| Common or Usual Name | Ultrasonic Scaler | Ultrasonic Scaler | Identical |
| Classification Name: | Ultrasonic Scaler<br>21 CFR 872.4850 | Ultrasonic Scaler<br>21 CFR 872.4850 | Identical |
| Regulation<br>Identification | An ultrasonic scaler is a<br>device intended for use<br>during dental cleaning and<br>periodontal (gum) therapy<br>to remove calculus deposits<br>from teeth by application of<br>an ultrasonic vibrating<br>scaler tip to the teeth | An ultrasonic scaler is a<br>device intended for use<br>during dental cleaning and<br>periodontal (gum) therapy to<br>remove calculus deposits<br>from teeth by application of<br>an ultrasonic vibrating<br>scaler tip to the teeth | Identical |
| Regulatory Class | II | II | Identical |
| | Principle of operation | | |
| Principles of Operation | The Device emits an electrical signal to a Dental Ultrasonic<br>Handpiece and adapts it permanently in accordance to the<br>effort and the encountered impedance of the handpiece<br>and the user settings. | | |
| | Dimensions | | |
| Height (mm) | 36.6 | 33 | No Impact<br>(#01) |
| Depth (mm) | 59.5 | 50 | No Impact<br>(#02) |
| Width (mm) | 85.95 | 60 | No Impact<br>(#03) |
| Weight (g) | 160 | 130 | No Impact<br>(#04) |
| Mounting Unit | Embedded in Dental<br>Delivery System | Embedded in Dental<br>Delivery System | Identical |
| | Power Supply | | |
| Typical Voltage (VAC) | 24 | 24 | Identical |
| Minimum<br>Voltage<br>(VAC) | 18.8 | 21.5 | No Impact<br>(#05) |
| Maximum<br>Voltage<br>(VAC) | 28.5 | 27.6 | No Impact<br>(#06) |
| Input Frequency (Hz) | 50 or 60 | 50 or 60 | Identical |
| Maximum Current<br>consumption (AC) | 1.2 | 1 | No Impact<br>(#07) |
| Maximum<br>Power<br>Consumption (VA AC) | 29 | 26 | No Impact<br>(#08) |
| Maximum Idle Current<br>Consumption (mA AC) | 100 | 100 | Identical |
| Maximum Idle Power<br>Consumption (VA AC) | 2 | 2 | Identical |
| Pictures | | | |
| | New Device | Predicate Device | Result / Impact<br>of the<br>differences on<br>Safety /<br>Effectiveness |
| Trade / Device<br>Name | NEWTRON CAN A | SP NEWTRON Module | |
| Knumber | K233922 | K033764 | |
| Picture | Image: NEWTRON CAN A | Image: SP NEWTRON Module | No Impact<br>(#09) |
| Electrical Safety Classification | | | |
| Electrical Safety Class | B type | B type | Identical |
| External constitution | | | |
| Fire aspects (for<br>Casing) | UL94-V0 (self-<br>extinguishing material) | UL94-V0 (self-extinguishing<br>material) | Identical |
| Ultrasonic Function Performances | | | |
| Electrical Technology | NEWTRON Technology | NEWTRON Technology | Identical |
| Minimum<br>Output<br>Frequency (KHz) | 28 | 28 | Identical |
| Maximum<br>Output<br>Frequency (KHz) | 36 | 36 | Identical |
| Minimum Output Power<br>in the Ultrasonic<br>handpiece (mW) | 10 | 10 | Identical |
| Maximum<br>Output<br>Power in the Ultrasonic<br>handpiece (W) | 12 | 12 | Identical |
| Minimum<br>Output<br>Current<br>in the<br>Ultrasonic<br>handpiece<br>(mA) | $8 \pm 10\%$ | $8 \pm 10\%$ | Identical |
| Maximum<br>Output<br>Current<br>in the<br>Ultrasonic<br>handpiece<br>(mA) | $100 +20\% -10\%$ | $100 +20\% -10\%$ | Identical |
| Power Factor | More than 0.65 | More than 0.65 | Identical |
| Function to power Light in the Ultrasonic Handpiece | | | |
| Standard LED Ring<br>Output Current (mA<br>DC) | 45 | | No Impact<br>(#10) |
| Low Voltage LED Ring<br>Output Current (mA<br>DC) | $135 \pm 15$ | 800 | No Impact<br>(#11) |
| Maximum<br>Open<br>Voltage (V DC) | 14 | 5 | No Impact<br>(#12) |
| Maximum Short -circuit<br>Current (mA DC) | 180 | 1000 | No Impact<br>(#13) |
| User Interface | | | |
| Quantity of Connectors | 3 | 2 | No Impact<br>(#14) |
| Power Supply<br>Connector | 1 | 1 | No Impact<br>(#15) |
| Command Connector | 1 | 1 (one the same connector) | No Impact<br>(#16) |
| | New Device | Predicate Device | Result / Impact<br>of the<br>differences on<br>Safety /<br>Effectiveness |
| Trade / Device<br>Name | NEWTRON CAN A | SP NEWTRON Module | Identical |
| Knumber | K233922 | K033764 | Identical |
| Handpiece Connector | 1 | 1 | Identical |
| Type of Command /<br>Input Setting | Digital Values<br>0 to 100<br>Contained in the<br>communication frame | Digital Signal stepped in<br>Volts<br>0 to 5 Volts | No Impact<br>(#17) |
| | Environmental | | |
| Where used | Dental Office (installed in a<br>Dental Delivery System) | Dental Office (installed in a<br>Dental Delivery System) | Identical |
| | Temperature | | |
| Thermal Cooling | Natural Convection | Natural Convection | Identical |
| Operating Temperature | +10°C to +30°C | +10°C to +30°C | Identical |
| Storage Temperature | 0°C to +50°C | -20°C to +70°C | No Impact |
| | Humidity | | |
| Operating Humidity | 30% to 75% | 30% to 75% | Identical |
| Storage Humidity | 10% to 95% | 10% to 100% | Identical |
| | Pressure / Altitude | | |
| Operating Pressure | 800 hPa to 1060 hPa | 800 hPa to 1060 hPa | Identical |
| Storage Pressure | 500 hPa to 1060 hPa | 500 hPa to 1060 hPa | Identical |
| Maximum Operating<br>Altitude | Less than 2000 m | Less than 2000 m | Identical |
| | Safety Standards | | |
| Safety Standard | IEC 60601-1 | IEC 60601-1 | Identical |
| EMC Standard | IEC 60601-1-2 | IEC 60601-1-2 | Identical |
| | Installation Method | | |
| Location of Fixation | Fixed in the Dental<br>Delivery System | Fixed in the Dental<br>Delivery System | Identical |
| Principles of fixation | Mounting bracket fixed in<br>the Dental Delivery System<br>with 2 screws | Mounting bracket fixed in<br>the Dental Delivery System<br>with 2 screws | Identical |
Image /page/10/Picture/2 description: The image shows the logo for Acteon. The logo consists of a blue triangle shape on the left and the word "acteon" in black letters on the right. Below the word "acteon" is the phrase "Caring People. Improving Lives" in a smaller font size.
{11}------------------------------------------------
Image /page/11/Picture/2 description: The image shows the contact information for SATELEC, a company of ACTEON Group. The address is 17, avenue Gustave Eiffel, Zone Industrielle du Phare, 33700 MERIGNAC, FRANCE. The telephone number is +33 (0) 556 340 607, and the fax number is +33 (0) 556 349 292. The email address is info@acteongroup.com.
Image /page/11/Picture/3 description: The image shows the logo for Acteon. The logo consists of a blue triangle shape on the left, followed by the word "acteon" in a dark gray sans-serif font. Below the word "acteon" is the tagline "Caring People. Improving Lives" in a smaller, dark gray font. The logo is clean and modern, with a focus on the company's name and its mission.
{12}------------------------------------------------
A lot of features and characteristics of NEWTRON CAN A are identical to the Predicate Device. It exists differences between NEWTRON CAN A and the Predicate Device.
Explanation on Differences #01, #02, #03, #04: The SATELEC New Device is embedded in a Dental Delivery System and it is not in contact to the Patient. Differences of Height, Depth, Width and Weight between SATELEC New Device and the Predicate Device has no impact on the Safety and Effectiveness.
Explanation on Differences #05, #06: The Power Supply of the SATELEC New Device is adapted to the Dental Delivery System where it is embedded. The Difference of Minimum and Maximum Voltage (VAC) between SATELEC New Device and the Predicate Device has no impact on the Safety and Effectiveness.
Explanation on Differences #07, #08: The differences of the Current consumption and the Power Consumption (VA AC) are due to Microprocessor of the SATELEC New Device that is permanently activated to drive the SATELEC New Device. Difference of Maximum Current consumption (AC) between SATELEC New Device and the Predicate Device has no impact on the Safety and Effectiveness.
{13}------------------------------------------------
Explanation on Difference #09: Difference of Shape / Presentation between SATELEC New Device and the Predicate Device has no impact on the Safety and Effectiveness.
Explanation on Differences #10, #11: The Predicate Device is intended to power a Bulb whereas the SATELEC New Device powers a LED Ring (Light Diode Emitting). The current value provided by the SATELEC New Device is lower than the Predicate Device due to the characteristic differences between a bulb and LED (LEDs consume a lower current). The Bulb and the LED Ring are parts of the Ultrasonic Handpiece. The Predicate Device powers a Bulb contained in the Ultrasonic Handpiece. The SATELEC New Device powers a LED Ring contained in the NEWTRON SLIM B.LED Handpiece (K132267 - Mar 10, 2014 and K132322 - Mar 11, 2014), NEWTRON LED Handpiece (K071424 - Aug 24, 2007, K091331 - Dec 11, 2009 and K091252 - Jul 22, 2009). Difference of Standard LED Ring Output Current (mA DC) and Low Voltage LED Ring Output Current (mA DC) between SATELEC New Device and the Predicate Device has no impact on the Safety and Effectiveness.
Explanation on Difference #12: The Predicate Device is intended to power a Bulb whereas the SATELEC New Device powers a LED Ring (Light Diode Emitting). The Voltage value provided by the Predicate Device is lower than the SATELEC New Device due to the characteristic differences between a bulb and LED (Bulb needs a low voltage). The Bulb and the LED Ring are parts of the Ultrasonic Handpiece. The Predicate Device powers a Bulb contained in the Ultrasonic Handpiece. The SATELEC New Device powers a LED Ring contained in the NEWTRON SLIM B.LED Handpiece (K132267 – Mar 10, 2014 and K132322 - Mar 11, 2014), NEWTRON LED Handpiece (K071424 - Aug 24, 2007, K091331 - Dec 11, 2009 and K091252 - Jul 22, 2009). Difference of Maximum Open Voltage (V DC) between SATELEC New Device and the Predicate Device has no impact on the Safety and Effectiveness.
Explanation on Difference #13: The Predicate Device is intended to power a Bulb whereas the SATELEC New Device powers a LED Ring (Light Diode Emitting). The current value provided by the SATELEC New Device is lower than the Predicate Device due to the characteristic differences between a bulb and LED (LEDs consume a lower current). Because the SATELEC New Device provides less output current for LED Ring than the Predicate Device, the Maximum Short-circuit Current (mA DC) is reduced.
The Bulb and the LED Ring are parts of the Ultrasonic Handpiece. The Predicate Device powers a Bulb contained in the Ultrasonic Handpiece. The SATELEC New Device powers a LED Ring contained in the NEWTRON SLIM B.LED Handpiece (K132267 - Mar 10, 2014 and K132322 - Mar 11, 2014), NEWTRON LED Handpiece (K071424 - Aug 24, 2007, K091331 - Dec 11, 2009 and K091252 - Jul 22, 2009). Difference of Maximum Short -circuit Current (mA DC) between SATELEC New Device and the Predicate Device has no impact on the Safety and Effectiveness.
Discussion #14: Difference of Quantity of Connectors between SATELEC New Device and the Predicate Device has no impact on the Safety and Effectiveness.
Discussion #15: The Predicate Device includes the Power Supply Connector and the Command connector in the same connector whereas the SATELEC New Device is equipped by two dedicated connectors. Difference of Power Supply Connector between SATELEC New Device and the Predicate Device has no impact on the Safety and Effectiveness.
Discussion #16: The Predicate Device includes the Power Supply Connector and the Command connector in the same connector whereas the SATELEC New Device is equipped by two dedicated connectors. Difference of Command Connector between SATELEC New Device and the Predicate Device has no impact on the Safety and Effectiveness.
Discussion #17: The Type of Command / Input Setting is designed and adapted only for the Dental Delivery System manufactured by A-DEC. The SATELEC New Device receives digital information (Digital Values) provided by the Delivery System. The Command / Setting of the SATELEC New Device translates digital information into analogic setting used by the analogic structure contained in the Device. The method of setting of the SATELEC New Device has slightly changed from the Predicate Device. The A-DEC Dental Delivery System uses a digitally stepped signal to transfer output power setting (input setting comprised between 0 and 5 Volts by steps) to the Predicate Device, the A-DEC
Image /page/13/Picture/11 description: The…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.