3D Endoscopic Fluorescence Camera System (DPM-ENDOCAM-06PF)
K233892 · Zhuhai Dipu Medical Technology Co., Ltd. · GCJ · Jun 13, 2024 · Gastroenterology, Urology
Device Facts
Record ID
K233892
Device Name
3D Endoscopic Fluorescence Camera System (DPM-ENDOCAM-06PF)
Applicant
Zhuhai Dipu Medical Technology Co., Ltd.
Product Code
GCJ · Gastroenterology, Urology
Decision Date
Jun 13, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Indications for Use
Upon intravenous administration and use of an ICG consistent with its approved label, the 3D Endoscopic Fluorescence Camera System is intended to provide real-time endoscopic visible and near-infrared fluorescence visualization. The System enables surgeons to perform minimally invasve surgery using standard endoscope visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts, using near-infrared imaging. Upon intravenous administration and use of an ICG consistent with its approved label, the 3D Endoscopic Fluorescence Camera System is intended to be used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes.
Device Story
3D Endoscopic Fluorescence Camera System (DPM-ENDOCAM-06PF) provides real-time visible and near-infrared (NIR) fluorescence imaging during minimally invasive surgery. System components include 3D 4K video processor, 3D 4K camera head, endoscope, cold light source, and light guide cable. Device illuminates surgical site with visible light for standard endoscopy and NIR excitation light for fluorescence imaging. When used with ICG imaging agent, system captures fluorescence response, processes signal, and displays output on video monitor. Operated by physicians in healthcare facilities. Output assists surgeons in visual assessment of vessels, blood flow, tissue perfusion, bile ducts, and lymphatic structures. Benefits include enhanced intraoperative visualization of anatomical structures and perfusion, aiding clinical decision-making during surgery.
Clinical Evidence
No clinical tests were performed. Evidence consists of bench testing (electrical safety, electromagnetic compatibility, photobiological safety, optical performance per ISO 8600 series) and an in vivo animal model study to verify imaging performance and safety.
Technological Characteristics
System includes 3D 4K video processor, 3D 4K camera head, rigid endoscopes (0/30 degree), LED visible light source, and NIR laser diode (Class 3R). Connectivity via DVI, HDMI, SDI. Complies with IEC 60601-1 (Class I, CF-type), IEC 60601-1-2, IEC 60601-2-18, IEC 60825-1, IEC 62471, and ISO 8600 series. Non-sterile, reusable.
Indications for Use
Indicated for patients undergoing minimally invasive surgery requiring real-time endoscopic visible and near-infrared fluorescence visualization of vessels, blood flow, tissue perfusion, extra-hepatic bile ducts, and the lymphatic system (vessels and nodes) following intravenous administration of ICG.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
Predicate Devices
da Vinci® Fluorescence Imaging Vision System and da Vinci® Firefly Imaging System (K213710)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Zhuhai Dipu Medical Technology Co., Ltd. % Joyce Yang Consultant Shenzhen Joyantech Consulting Co., Ltd 1713A, 17th Floor, Block A. Zhongguan Times Square Nanshan District Shenzhen, Guangdong 518000 China
Re: K233892
Trade/Device Name: 3D Endoscopic Fluorescence Camera System (DPM-ENDOCAM-06PF) Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: GCJ, NAY, IZI Dated: May 13, 2024 Received: May 13, 2024
Dear Joyce Yang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Yan Fu -
Digitally signed by Yan Fu -S
Date: 2024.06.13 20:45:36 -04'00'
Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical & Infection Control Devices, Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K233892
Device Name
3D Endoscopic Fluorescence Camera System
Indications for Use (Describe)
Upon intravenous administration and use of an ICG consistent with its approved label, the 3D Endoscopic Fluorescence Camera System is intended to provide real-time endoscopic visible and near-infrared fluorescence visualization. The System enables surgeons to perform minimally invasve surgery using standard endoscope visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts, using near-infrared imaging.
Upon intravenous administration and use of an ICG consistent with its approved label, the 3D Endoscopic Fluorescence Camera System is intended to be used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary K233892
Date of Summary prepare: June 12, 2024
#### 1. Submission Sponsor
| Applicant Name: | Zhuhai Dipu Medical Technology Co., Ltd. |
|-----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | 1F, Building 13, GMP Production Workshop,<br>Zhuhai International Health Harbor, No. 628 Airport<br>West Road, Sanzao Town, Jinwan District, Zhuhai,<br>China |
| Contact person: | WANG HUAN |
| Phone: | +8613521289729 |
| Email: | huan.wang@digipmc.com |
#### 2. Submission correspondent
| Name: | Shenzhen Joyantech Consulting Co., Ltd |
|-----------------|---------------------------------------------------------------------------------------|
| Address: | 1713A, 17th Floor, Block A, Zhongguan Times Square, Nanshan District, Shenzhen 518000 |
| Contact person: | Joyce Yang |
| Phone: | +86-755-86069197 |
| Email: | joyce@cefda.com |
#### 3. Device Identification
| Trade Name: | 3D Endoscopic Fluorescence Camera System |
|-----------------------|------------------------------------------|
| Common or Usual Name: | Endoscopic Imaging System |
| Model: | DPM-ENDOCAM-06PF |
| Classification name: | Endoscope and accessories |
| Review Panel: | Gastroenterology/Urology |
| Product Code: | GCJ, NAY, IZI |
| Device Class: | Class II |
| Regulation Number: | 21 CFR § 876.1500 |
#### Legally Marketed Predicate Device 4.
| Trade Name | da Vinci® Fluorescence Imaging Vision System and da Vinci® Firefly Imaging System |
|-------------------|-----------------------------------------------------------------------------------|
| Regulation number | 21 CFR § 876.1500 |
| Regulation class | Class II |
| Regulation name | Endoscope and accessories |
| 510(k) Number | K213710 |
| Product Code | NAY, GCJ, IZI |
| Manufacturer | Intuitive Surgical, Inc. |
{4}------------------------------------------------
## 5. Device Description
The proposed system, 3D Endoscopic Fluorescence Camera System (DPM-ENDOCAM-06PF) is comprised of a 3D 4K video processor (DPM-ENDOCAM-06P-MF), a 3D 4K camera head (DPM-ENDOCAM- 06P-CAM), Endoscope (DPM-ENDOSCOPE-3D01/DPM-ENDOSCOPE- 3D02), a cold light source (DPM-LIGHT-03) and a light guide cable (DPM-FB-01).
3D Endoscopic Fluorescence Camera System is capable of providing real-time endoscopic visible and near-infrared fluorescence imaging.
During surgical procedures, the system may be operated to provide visualization similar to that provided by conventional imaging systems used in surgical endoscopy. The area of interest is illuminated with visible light from the light source and the resulting reflecting light is imaged by the camera and displayed on the video monitor. When used with the VIS-only laparoscopes, the System is only capable of the conventional mode of visualization described herein.
To provide NIR fluorescence imaging, the system is used with the imaging agent, indocyanine green (ICG). The patient is injected with ICG imaging agent. The ICG fluoresces when illuminated through the laparoscope with NIR excitation light from the light source, and the fluorescence response is then imaged with the camera, processed and displayed on a video monitor.
### 6. Intended Use/ Indications for Use
Upon intravenous administration and use of an ICG consistent with its approved label, the 3D Endoscopic Fluorescence Camera System is intended to provide real-time endoscopic visible and near-infrared fluorescence visualization. The System enables surgeons to perform minimally invasive surgery using standard endoscope visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts, using near-infrared imaging.
Upon intravenous administration and use of an ICG consistent with its approved label, the 3D Endoscopic Fluorescence Camera System is intended to be used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes.
## 7. Technological characteristics comparison
{5}------------------------------------------------
| Comparison<br>item | Subject Device(K233892) | Predicate Device(K213710) | Comments |
|-------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Product<br>Name | 3D Endoscopic Fluorescence<br>Camera System | <i>da</i> Vinci® Fluorescence /<br>Imaging Vision System and <i>da</i><br>Vinci® Firefly Imaging System | |
| Product<br>Code | GCJ, NAY, IZI | GCJ, NAY, IZI | Same |
| Regulation<br>Number | 21 CFR § 876.1500 | 21 CFR § 876.1500 | Same |
| Classification | Class II | Class II | Same |
| Type of use | Prescription Use | Prescription Use | Same |
| Intended use<br>& Indications<br>for Use | Upon intravenous<br>administration and use of an<br>ICG consistent with its<br>approved label, the 3D<br>Endoscopic Fluorescence<br>Camera System is intended to<br>provide real-time endoscopic<br>visible and near-infrared<br>fluorescence visualization. The<br>System enables surgeons to<br>perform minimally invasive<br>surgery using standard<br>endoscope visible light as well<br>as visual assessment of<br>vessels, blood flow and related<br>tissue perfusion, and at least<br>one of the major extra-hepatic<br>bile ducts, using near-infrared<br>imaging.<br>Upon intravenous<br>administration and use of an<br>ICG consistent with its<br>approved label, the 3D<br>Endoscopic Fluorescence<br>Camera System is intended to<br>be used to perform<br>intraoperative fluorescence<br>imaging and visualization of the<br>lymphatic system, including<br>lymphatic vessels and lymph<br>nodes. | <i>da</i> Vinci Fluorescence Imaging Vision<br>System<br>Upon intravenous administration and<br>use of an ICG drug product<br>consistent with its approved label, the<br><i>da</i> Vinci Fluorescence Imaging Vision<br>System is intended to provide real-<br>time endoscopic visible and near-<br>infrared fluorescence imaging. The<br><i>da</i> Vinci Fluorescence Imaging Vision<br>System enables surgeons to perform<br>minimally invasive surgery using<br>standard endoscopic visible light as<br>well as visual assessment of vessels,<br>blood flow and related tissue<br>perfusion, and at least one of the<br>major extrahepatic bile ducts (cystic<br>duct, common bile duct or common<br>hepatic duct), using near infrared<br>imaging.<br>Fluorescence imaging of biliary ducts<br>with the <i>da</i> Vinci fluorescence<br>Imaging Vision System is intended<br>for use with standard of care white<br>light and, when indicated,<br>intraoperative cholangiography. The<br>device is not intended for standalone<br>use for biliary duct visualization.<br>Upon intravenous administration and<br>use of an ICG drug product | Same |
| Comparison<br>item | Subject Device(K233892) | Predicate Device(K213710) | Comme<br>nts |
| | | consistent with its approved label, the<br>da Vinci Fluorescence Imaging Vision<br>System is used to perform<br>intraoperative fluorescence imaging<br>and visualization of the lymphatic<br>system, including lymphatic vessels<br>and lymph nodes. | |
| | | da Vinci Firefly Imaging System | |
| | | Upon intravenous administration and<br>use of an ICG drug product<br>consistent with its approved label, the<br>da Vinci Firefly Imaging System is<br>intended to provide real-time<br>endoscopic visible and near-infrared<br>fluorescence imaging. The da Vinci<br>Firefly Imaging System enables<br>surgeons to perform minimally<br>invasive surgery using standard<br>endoscopic visible light as well as<br>visual assessment of vessels, blood<br>flow and related tissue perfusion, and<br>at least one of the major extrahepatic<br>bile ducts (cystic duct, common bile<br>duct or common hepatic duct), using<br>near infrared imaging. | |
| | | Fluorescence imaging of biliary ducts<br>with the da Vinci Firefly Imaging<br>System is intended for use with<br>standard of care white light and,<br>when indicated, intraoperative<br>cholangiography. The device is not<br>intended for standalone use for biliary<br>duct visualization. | |
| | | Upon intravenous administration and<br>use of an ICG drug product<br>consistent with its approved label, the<br>da Vinci Firefly Imaging System is used to<br>perform intraoperative fluorescence<br>imaging and visualization of the<br>lymphatic system, including lymphatic<br>vessels and lymph nodes. | |
| Applicable<br>user | Physicians | Physicians | Same |
| Environment<br>of use | Healthcare facility/hospital | Healthcare facility/hospital | Same |
| Single use / | Reusable | Reusable | Same |
| Comparison<br>item | Subject Device(K233892) | Predicate Device(K213710) | Comme<br>nts |
| Reusable | | | |
| Sterile /non-<br>sterile | Marketed as non-sterile | Marketed as non-sterile | Same |
| Device<br>System<br>components | - 3D 4K Video Processor<br>- 3D 4K Camera Head<br>- Endoscope (0 degree and 30 degree)<br>- Light Source<br>- Light Guide Cable | - Video Processor<br>- Endoscope (0 degree and 30 degree)<br>- 3D HD Camera head<br>- Light Source<br>- Light guide cable | Same |
| Video output<br>signals | DVI, HDMI, SDI | HD-SDI, DVI | Similar |
| Video output<br>resolution | 4096×2160p;<br>3840×2160p;<br>1920×1080p | 1280×1024;<br>1920×1080 | Similar |
| Excitation<br>Wavelength | 785 nm ± 5 nm | 802-805 nm | Same |
| Endoscope<br>Diameter | 10.1 mm | 8.5 mm, 12 mm | Similar |
| Endoscope<br>length | 328 mm, 330 mm | 296.85 - 543.6 mm | Similar |
| Image<br>sensors | 4K CMOS sensor assembly | 3 chip RGB CCD;<br>Front and back side illuminated<br>CMOS HD sensor assembly | Similar |
| Light<br>sources | - Visible (VIS): Light-emitting diode<br>array<br>- Near infrared (NIR): NIR laser<br>diode | - Visible (VIS): Light-emitting diode<br>array<br>- Near infrared (NIR): NIR laser diode | Same |
| Light<br>guide<br>cable | Transmission spectrum: Visible<br>+<br>NIR | Transmission spectrum: Visible + NIR | Same |
| Imaging<br>agent | SPY AGENT GREEN | SPY AGENT GREEN | Same |
| Comparison<br>item | Subject Device(K233892) | Predicate Device(K213710) | Comments |
| Type of<br>protection<br>against<br>electric<br>shock<br>(as per IEC<br>60601-1) | Class I | Class I…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.