Fitbone Trochanteric is indicated for limb lengthening of the femur. Fitbone Trochanteric is indicated for adult and pediatric (greater than 12 through 21 years of age) patients.
Device Story
Fully implantable intramedullary lengthening nail for femur; implanted by healthcare professionals. System includes nail, locking screws, trial nails, K-wire, and subcutaneous receiver connected via bipolar feed line. External control set (electronics station and transmitter) induces electromagnetic field to power hermetically enclosed internal motor; motor converts field to DC voltage to drive telescope distraction. Used in clinic or home settings. Output is mechanical limb lengthening; provider monitors progress to inform clinical decisions. Benefits include controlled, non-invasive distraction post-surgery.
Clinical Evidence
Bench testing only. Performed static and fatigue 4-point bending (ASTM F1264-16), axial load and vibration tests, torsional strength, insertion/removal torque, self-tapping properties, and pull-out tests (ASTM F543-17). Compliance with ISO 14708-1:2014 for active implantable medical devices.
Technological Characteristics
Implant grade stainless steel 1.4441 (ASTM F138-13) and silicone (NuSil MED series). Intramedullary fixation rod with internal motor; electro-magnetic induction power source. Dimensions: 225-365mm length, 11-13mm diameter. Sterilization: Gas Plasma.
Indications for Use
Indicated for limb lengthening of the femur in adult and pediatric patients (ages >12 to 21).
Regulatory Classification
Identification
An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.
Orthofix® External fixation Screw (Pin) with Hydroxyapatite Coating (K974186)
Submission Summary (Full Text)
{0}------------------------------------------------
June 18, 2024
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".
Orthofix Srl Elvira Taccarelli Regulatory Affairs Manager Via delle Nazioni, 9 Bussolengo, 37012 Italy
## Re: K233867
Trade/Device Name: Fitbone Trochanteric Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary fixation rod Regulatory Class: Class II Product Code: HSB Dated: May 22, 2024 Received: May 22, 2024
Dear Elvira Taccarelli:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerelv.
Joseph P. Russell - Digitally signed by Joseph P. Russell -S S Date: 2024.06.18 10:51:53 -04'00"
for: Farzana Sharmin, PhD Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
Submission Number (if known)
K233867
Device Name
Fitbone Trochanteric
Indications for Use (Describe)
Fitbone Trochanteric is indicated for limb lengthening of the femur. Fitbone Trochanteric is indicated for adult and pediatric (greater than 12 through 21 years of age) patients.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# ORTHOFIX®
# 510(K) SUMMARY
# ORTHOFIX SRL FITBONE™ TROCHANTERIC
## Submitter information
| Company Name: | Orthofix S.r.l. |
|---------------|-------------------------------------------------------|
| Address | Via Delle Nazioni, 9<br>37012 Bussolengo (VR) - Italy |
| Telephone | +39 045 6719000 |
| Fax | +39 045 6719380 |
| Contact Person | Elvira Taccarelli<br>Regulatory Affairs Manager |
|----------------|-------------------------------------------------------|
| Address | Via Delle Nazioni, 9<br>37012 Bussolengo (VR) - Italy |
| Telephone | +39 045 6719000 |
| Fax | +39 045 6719380 |
| Email address | elvirataccarelli@orthofix.it |
| Date prepared | 2024, June 17 |
### Trade Name, Common Name, Classification
| Trade Name: | FITBONE® TROCHANTERIC |
|----------------------|---------------------------------|
| Common Name: | Intramedullary lengthening nail |
| Classification Name: | Intramedullary fixation rod |
| Requlation Number: | 21 CFR 888.3020 |
| Product Code: | HSB |
| Classification: | Class II |
| Panel code: | Orthopedic |
#### Primary predicate and additional reference devices
| Primary Predicate | 510(k) Number | Manufacturer |
|------------------------------------------------------------------------|---------------|------------------------------------------|
| FITBONE® TAA | K203399 | Orthofix s.r.l. |
| Reference Devices | 510(k) Number | Manufacturer |
| PRECICE® Intramedullary Limb<br>Lengthening System | K220234 | NuVasive Specialized<br>Orthopedics Inc. |
| Orthofix Modulsystem | K955848 | Orthofix s.r.l. |
| RIVAL View Plating Systems and<br>Reduce Fracture Plating Systems | K190388 | |
| JPS JuniOrtho Plating System | K200246 | |
| Orthofix® External fixation Screw<br>(Pin) with Hydroxyapatite Coating | K974186 | |
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the logo for Orthofix. The logo consists of a blue abstract shape on the left, followed by the word "ORTHOFIX" in black, sans-serif font. There is a registered trademark symbol to the right of the word "ORTHOFIX".
| Device description | The FITBONE® Trochanteric is a fully implantable intramedullary lengthening nail<br>and accessories.<br>The subject FITBONE® Trochanteric consists of the implantable intramedullary<br>lengthening nail and accessories (Locking screws, Trial nails, K-wire and<br>Convenience kits). The Subject device is implanted into the medullary canal of the<br>femur and connected to the primary predicate intracutaneous Receiver (K163368)<br>by a bipolar feed line. The external FITBONE Control Set is the same as previously<br>cleared for the primary predicate Fitbone TAA device (K203399) and consists of a<br>control electronics station and transmitter. There are no changes to the previously |
|--------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | cleared Control Sets and Receiver as a result of this submission.<br>The power required for the distraction process is controlled by hermetically enclosed<br>motor which draws the telescope apart. The electro-magnetic field sent from the<br>Transmitter to the Receiver is converted in the Receiver into DC-Voltage to supply<br>the motor of the subject Fitbone Trochanteric Nails with voltage, when actioned.<br>The subject Fitbone Trochanteric Nails are available in two different diameter<br>models (D09mm, D11mm), different lengths and lengthening capabilities.<br>The subject nail is anchored to the bone by subject locking screws. The locking |
| | screws are available in two variants (standard locking screws and revision locking<br>screws), in two diameters, D4.5mm and D4mm, and in multiple lengths.<br>The energy needed for the distraction process is transmitted from the outside by<br>placing the external transmitter over the implanted receiver, which is placed in the<br>subcutaneous tissue during surgery. |
| | There is no transcutaneous contact between the implanted intramedullary nail and<br>the outer surface of the patient's body. |
| | The subject trial nails accessories are available for each variant of the Fitbone<br>Trochanteric nails and are used to simulate the shape of the implant.<br>The subject Fitbone Trochanteric nail and K-wire are provided in sterile conditions<br>only. The trial nails are provided in non-sterile version only. The bone screws are<br>available in both sterile and non-sterile versions. |
| | The subject Fitbone Trochanteric Nails and their accessories are made from, as<br>follows:<br>Nail: implant grade stainless steel 1.4441, according to ASTM F138-13<br>"Standard Specification for Wrought 18Chromium-14Nickel-<br>2.5Molybdenum Stainless Steel Bar and Wire for Surgical Implants (UNS<br>S31673)", and Silicone Nusilmed (NuSil MED-4870, NuSil MED-1511,<br>Nusil MED 4750, NUSIL MED1-161, NUSIL MED2-4502). Trial nail: implant grade stainless steel 1.4441, according to ASTM F138-<br>13 "Standard Specification for Wrought 18Chromium-14Nickel-<br>2.5Molybdenum Stainless Steel Bar and Wire for Surgical Implants (UNS<br>S31673)" and ASTM F899-20 Standard Specification for Wrought<br>Stainless Steels for Surgical Instruments. Locking screws: implant grade stainless steel 1.4441, according to ASTM<br>F138-13 "Standard Specification for Wrought 18Chromium-14Nickel-<br>2.5Molybdenum Stainless Steel Bar and Wire for Surgical Implants (UNS<br>S31673)" K-wire: implant grade stainless steel 1.4441, according to ASTM F138-13<br>"Standard Specification for Wrought 18Chromium-14Nickel-<br>2.5Molybdenum Stainless Steel Bar and Wire for Surgical Implants (UNS<br>S31673)" |
| | The Subject, as the primary predicate, will be implanted only by Healthcare<br>Professionals (HCP), with full awareness of the appropriate orthopedic procedures |
{5}------------------------------------------------
Image /page/5/Picture/2 description: The image shows the logo for Orthofix. The logo consists of a blue abstract shape on the left, followed by the word "ORTHOFIX" in black, sans-serif font. A small registration mark is present to the right of the "X" in "ORTHOFIX".
| Indications for use | | FITBONE® Trochanteric is indicated for limb lengthening of the femur.<br>FITBONE® Trochanteric is indicated for adult and pediatric (greater than 12 through<br>21 years of age) patients. | |
|----------------------------------|--------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technological<br>Characteristics | | The following table provides a comparison of technological characteristic of the<br>subject and primary predicate device.<br>Any differences have been demonstrated not to raise different questions of safety<br>and effectiveness. | |
| | Technological<br>Characteristic | Subject Device<br>(Fitbone Trochanteric) | Predicate Device<br>(Fitbone TAA, K203399) |
| 1. | Intended use and<br>indications for<br>Use | Fitbone™ Trochanteric is indicated for<br>limb lengthening of the femur.<br>Fitbone Trochanteric is indicated for<br>adult and pediatric (greater than 12<br>through 21 years of age) patients.<br>Assessment: The Subject device is indicated for limb lengthening of the femur.<br>This anatomical site is a subset of the indications of the primary predicate. | Intended for limb lengthening of the<br>femur and tibia.<br>Indicated for adult and pediatric<br>(greater than 12 through 21 years of<br>age) patients<br> |
| 2. | Anatomical Sites | Femur | Femur, tibia |
| | | Assessment: The anatomical sites for use are a subset of primary predicate. | |
| 3. | Intended<br>environment | Clinic or Home environment | Clinic or Home environment |
| | | Assessment: The intended environment is the same as the primary predicate. | |
| 4. | Nail Material | Implant Grade Stainless Steel (1.4441,<br>316LVM) | Implant Grade Stainless Steel<br>(1.4441, 316LVM) |
| | | Assessment: Subject material identical to primary predicate. | |
| 5. | Nail Size Range | 225-365mm in length; 11 and 13mm<br>diameters in varying configurations | 200-245mm in length; 9, 11, and<br>13mm diameters in varying<br>configurations |
| | | Assessment: Diameter size ranges are similar to primary predicate.<br>Potential performance differences arising from variations in size offerings was<br>analyzed through bench testing, performed on subject and predicate device<br>Further mitigation is provided through weight bearing precautions in labeling. | |
| 6. | Maximum<br>distraction<br>possible | From 40mm (with nail length 225mm)<br>to 80mm (with longer nails) | From 40mm (with nail length 200mm)<br>to 80mm (with longer nails)) |
| | | Assessment: The maximum distraction possible of the Subject device is<br>equivalent to the primary predicate. The worst case scenario (maximum<br>distraction possible of 80 mm) is identical to the primary predicate device. | |
| 7. | Method of<br>Distraction/Energy<br>Source | Internal motor electro-magnetically<br>induced by an external transmitter with<br>signal received through a receiver<br>placed just under skin | Internal motor electro-magnetically<br>induced by an external transmitter<br>with signal received through a<br>receiver placed just under skin |
| | | Assessment: Subject distraction methodology is identical to primary predicate. | |
| 8. | Sterilization<br>Method | Gas Plasma<br>Gas Plasma<br>Assessment: The sterilization method is identical to the primary predicate. | |
| | Performance<br>Analysis | Subject devices have similar configuration, material, sizes, and design to the primary<br>predicate FITBONE® TAA (K203399) and the additional reference devices. The<br>following mechanical testing was performed:<br>Implantable nails and screws:<br>Static and fatigue 4-point bending test for subject Fitbone Trochanteric nails performed according to ASTM F1264-16 - "Standard Specification and Test Methods for Intramedullary Fixation Devices", 4-point bending test for subject locking screws per ASTM F1264-16 - "Standard Specification and Test Methods for Intramedullary Fixation Devices", Axial Load and vibration tests of subject Fitbone Trochanteric nails per custom test protocol, Testing of subject Fitbone Trochanteric nails per ISO 14708-1:2014 - "Implants for surgery - Active implantable medical devices - Part 1: General requirements for safety, marking and for information to be provided by the manufacturer", Torsional strength, insertion/removal torque, self-tapping properties, and pull out test performed according to ASTM F543-17. | |
| Conclusion | | Based upon equivalences in: intended use, patient population, site of application,<br>conditions of use, operating principles, and the non-clinical performance data, the<br>subject FITBONE® Trochanteric nail have been shown to be safe and effective and<br>to perform equivalently as compared to the legally marketed predicate device<br>(K203399). | |
{6}------------------------------------------------
Image /page/6/Picture/2 description: The image shows the word "ORTHOFIX" in black font. To the left of the word is a blue symbol that looks like three curved shapes stacked on top of each other. The symbol is a stylized version of the letter O. There is a registered trademark symbol to the right of the word.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.