K233834 · Draeger Medical Systems, Inc. · MSX · Jul 12, 2024 · Cardiovascular
Device Facts
Record ID
K233834
Device Name
Infinity Gateway Suite
Applicant
Draeger Medical Systems, Inc.
Product Code
MSX · Cardiovascular
Decision Date
Jul 12, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2300
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The Infinity Gateway software applications are intended to provide clinicians with the capability of viewing patient data remotely via the Infinity Network and for the data exchange of select clinical and administrative information between the Infinity Network and the hospital network.
Device Story
Infinity Gateway Suite is a client-server software application facilitating data exchange between Infinity Network and hospital LAN. Server runs on Windows Server; connects to hospital network and Infinity Network. Enables remote viewing of patient data; supports HL7 interface, ASTM Stat Lab interface, pager interface, alarm history database, time master functions, and 12-lead ECG export. Licensed via physical dongle with option passwords. Developer tools (WinAccess API) allow custom application development for research or downstream systems. Deployed via virtual machine technology. Provides clinicians with integrated patient data for decision-making and continuity-of-care support. Not for primary monitoring or independent diagnosis.
Clinical Evidence
Bench testing only. No clinical data. Verification and validation testing confirmed that the software performs according to intended use, risk mitigations function correctly, and all test cases met requirements.
Technological Characteristics
Software-only device; no physical form. Operates on Windows Server 2022 with Microsoft SQL 2022. Connectivity via hospital LAN and Infinity Network. Complies with ANSI AAMI ISO 14971:2019 (risk management), ANSI AAMI IEC 62304:2015 (software life cycle), and ANSI AAMI IEC 62366-1:2020 (usability).
Indications for Use
Indicated for clinicians requiring remote viewing of patient data and exchange of clinical/administrative information between Infinity Network and hospital networks. Not intended for primary patient monitoring or independent diagnosis.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name, "U.S. Food & Drug Administration," written out next to it.
July 12, 2024
Draeger Medical Systems, Inc. Megan Brewster Regulatory Affairs Specialist 6 Tech Dr Andover, Massachusetts 01810
Re: K233834
Trade/Device Name: Infinity Gateway Suite Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer And Rate Alarm) Regulatory Class: Class II Product Code: MSX Dated: June 7, 2024 Received: June 10, 2024
Dear Megan Brewster:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
This clearance applies to the version of the device that is intended to be used with Infinity Central Station Wide version VG5.0.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Kimberly N. Crowley -S
For:
Jennifer Kozen Assistant Director
{3}------------------------------------------------
Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K233834
Device Name Infinity Gateway Suite
Indications for Use (Describe)
The Infinity Gateway software applications are intended to provide clinicians with the capability of viewing patient data remotely via the Infinity Network and for the data exchange of select clinical and administrative information between the Infinity Network and the hospital network.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: Arial, sans-serif;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "Dräger" in a bold, sans-serif font. The word is in blue color. The two dots above the letter "a" are also in blue.
# Draeger Medical Systems, Inc. 6 Tech Drive Andover, MA 01810-2434
# 510(k) SUMMARY
#### I. Submitter:
Draeger Medical Systems, Inc. 6 Tech Drive Andover, MA 01810, U.S.A.
Contact Person: Megan Brewster Senior Regulatory Affairs Specialist Draeger Medical Systems, Inc. Phone: 978-447-2410 E-mail: megan.brewster(a)draeger.com July 10, 2024
#### II. Device
## Names / Common Names / Classification Names:
| Common Name: | Computers and Software, Medical |
|----------------------|---------------------------------------------------------|
| Trade Name: | Infinity® Gateway Suite |
| Classification Name: | System Network and Communication Physiological Monitors |
| Product Code: | MSX |
| Subsequent Code: | LNX |
| Regulatory Class: | II |
| Regulation Number: | §870.2300 |
#### III. Predicate Device
The Infinity® Gateway Suite, VF9.0, cleared via K203579 on June 7, 2022.
{6}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the word "Dräger" in blue font. The font is bold and sans-serif. The two dots above the "a" are also blue. The word is likely a logo for the Dräger company, which is a German company that makes medical and safety technology.
#### IV. Device Description
The Infinity Gateway Suite is a suite of software applications that are intended to provide clinicians with the capability of viewing patient data remotely via the Infinity Network and for the data exchange of select clinical and administrative information between the Infinity Network and the hospital network.
Infinity Gateway Suite is designed to support the unique needs of hospitals, and provide a complete range of options for information connectivity, including:
- Server Software Health Level seven (HL7) Interface Software Options -
- -American society for testing and materials (ASTM) Stat Lab interface
- -Developer Tools
- Pager Interface -
- -Alarm history database
- -Time master functions
- -12-lead electrocardiogram (ECG) export
Infinity Gateway Suite is designed as a client-server application that provides a way for users to view patient information on a hospital workstation that is connected via the hospital network. The server portion runs on a Windows server that is connected directly to the hospital Local Area Network (LAN) and to the Infinity Network. Data access methods are available as options that customers can purchase and enable independently once they install the basic Infinity Gateway application.
The Infinity Gateway Suite facilitates the exchange of important clinical information between the Infinity protocol and existing hospital and patient care systems. The Infinity Gateway Suite is designed to provide flexibility by using common healthcare protocols and data format standards for managing communications between multiple disparate systems. The user may select one or more Infinity Gateway Developer Tools and/or Interface Options to create a seamless flow of information tailored to support clinical workflow.
Infinity Gateway Developer Tools and Interface Options are licensed or unlocked by using option passwords associated with a dongle. An option password is the electronic proof of purchase for the Infinity Gateway software. During the at-hospital setup procedure, the user will be asked for an option or license password which is a unique number associated with the licensing dongle issued. The licenses are always associated with the physical dongle. The Infinity Gateway Developer Tools (such as WinAccess API) enable the development of custom applications to support customers' homegrown applications to support clinical research projects or downstream information systems. Finally, Infinity Gateway provides flexible deployment opportunities by leveraging the virtual machine technology, which facilitates software deployments.
Infinity Gateway also promotes patient safety with efficient workflows for timely decisionmaking by integrating patient data and providing continuity-of-care support. Furthermore, it makes comprehensive clinical data available at the point-of-care by facilitating the exchange of lab reports and admission information between the Infinity Network and other hospital systems.
The subject device is compatible with Infinity Central Station Wide version VG5.0
{7}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the word "Dräger" in blue font. The font is bold and sans-serif. The two dots above the "a" are also blue.
#### V. Intended Use
The Infinity Gateway software applications are intended to provide clinicians with the capability of viewing patient data remotely via the Infinity Network and for the data exchange of select clinical and administrative information between the Infinity Network and the hospital network.
The Infinity Gateway and its options are not intended to be used for primary patient monitoring or independent diagnosis.
#### VI. Indications for Use
The Infinity Gateway software applications are intended to provide clinicians with the capability of viewing patient data remotely via the Infinity Network and for the data exchange of select clinical and administrative information between the Infinity Network and the hospital network.
#### VII. Comparison of Technological Characteristics with Predicate Device
The Infinity Gateway Suite VF9.1.1 device, like its predicate, is a software only device; it has no physical form. The intended use, performance and technological characteristics of the subject device are substantially equivalent to the predicate device.
Some device modifications were made to mitigate vulnerabilities by eliminating device features as outlined in the "Significant Modifications Introduced in this Submission" section below. These modifications have no negative impact on the safety and effectiveness of the Gateway VF9.1.1 device. See the Table below for the comparison between the predicate device, Infinity Gateway VF9.0, and the subject device, Infinity Gateway VF9.1.1.
| Infinity® Gateway Suite VF9.0 and VF9.1.1 Comparison | | | |
|-------------------------------------------------------------|-------------------------------------------------------------|-----------------------------------------------------|---------------------------------------------------------------|
| Device Attributes | Infinity® Gateway Suite VF9.0<br>Predicate Device (K203579) | Infinity® Gateway Suite<br>VF9.1.1 Subject Device | Explanation of Changes |
| Microsoft SQL<br>Compatible | 2017 | 2022 | Migrate to latest Microsoft SQL to<br>stay current. |
| Microsoft Server | 2016 | 2022 | Migrate to latest Microsoft Server<br>to stay current. |
| Service-Oriented<br>Device Connectivity<br>(SDC) – OUS only | Includes SDC functionality | SDC functionality removed | This was a non-standard<br>implementation of SDC. |
| Infinity<br>Unicast/Multicast<br>encryption | N/A | Includes unicast and multicast<br>encryption | Increases cybersecurity of the<br>Gateway device. |
| Passwords | No password complexity<br>requirements | Password NIST SP800-63B<br>compliant | Increases cybersecurity of the<br>Gateway device. |
| Network allow listing | N/A | Communication between<br>authenticated devices only | Increases cybersecurity of the<br>Gateway device. |
| PatientWatch | PatientWatch support | PatientWatch functionality<br>removed | PatientWatch used ActiveX, which<br>is no longer supported by |
{8}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the word "Dräger" in blue font. The font is bold and sans-serif. The two dots above the "a" are also blue and are slightly larger than the rest of the letters. The word is slightly slanted to the right.
| Infinity® Gateway Suite VF9.0 and VF9.1.1 Comparison | | | |
|------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|
| Device Attributes | Infinity® Gateway Suite VF9.0<br>Predicate Device (K203579) | Infinity® Gateway Suite<br>VF9.1.1 Subject Device | Explanation of Changes |
| | | | Microsoft and the removal<br>increases cybersecurity of the<br>Gateway device. Alarm (Visual)<br>removed as part of PatientWatch<br>removal. |
| Alarm Grade / State | PatientWatch, Pager, WinAccess<br>API, FileAccess, and HL7 ACM<br>allow for accessing alarm grade<br>and state. | Pager, WinAccess API, and<br>FileAccess allow for accessing<br>alarm grade and state. | PatientWatch and HL7 ACM<br>removed from VF9.1.1. Alarm<br>Grade/State still available through<br>Pager, WinAccess API, and<br>FileAccess |
#### VIII. Significant Modifications Introduced in this Submission
Cybersecurity updates and other software improvements in Infinity Gateway VF9.1.1. See the Table above for detailed information on the modifications introduced in Infinity Gateway VF9.1.1.
#### IX. Performance Data
# Verification and Validation Testing:
The verification and validation testing conducted, confirms that the Infinity Gateway Suite VF9.1.1 product performs and functions according to its intended use with no adverse effects upon other medical devices in the Infinity system to which it is connected. Testing confirmed that identified product risk mitigations functions with the new code. All test cases were determined to meet their respective requirements.
Also, all known use-related hazards pertaining to usability have been reviewed, including those beyond the primary operating functions. As a result of this review, all risk control measures identified in the risk management process are deemed adequate to further mitigate and control the risks at an acceptable level under consideration of the intended use. The results of the Verification and Validation testing confirm that the modified software is substantially equivalent to that of the predicate device.
Biocompatibility: The proposed Infinity Gateway Suite VF9.1.1 is a software only device and has no physical form. Therefore, biocompatibility requirements are not applicable.
Sterilization: Sterilization and Shelf-Life: The proposed Infinity Gateway Suite VF9.1.1 is a software only device and has no physical form. Therefore, sterilization and shelf-life requirements are not applicable.
## Standards/Compliance Testing:
The Infinity Gateway Suite software modifications have been tested and developed in compliance with the following standards:
{9}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the word "Dräger" in a bold, sans-serif font. The word is blue. The two dots above the "a" are also blue.
- ANSI AAMI ISO 14971: 2019 Medical devices Applications of risk management to ● medical devices
- . ANSI AAMI IEC 62304:2015
- Medical device software Software life cycle processes [Including Amendment 1 . (2016)]
- ANSI AAMI IEC 62366-1:2020 Medical devices Part 1: Application of usability . engineering to medical devices
#### X. Conclusion
The Infinity Gateway Suite VF9.1.1 modifications have been demonstrated to be substantially equivalent to the predicate device, Infinity Gateway Suite VF9.0, which was cleared under K203579 on June 7, 2022. The intended use of the Infinity Gateway Suite as described in the product labeling, although slightly modified, does not change the actual intended use of the device. The intended use statement has been modified for clarification purposes only, not as a result of the proposed device modifications. Verification and Validation results for the proposed modifications support substantial equivalence to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.