Clinical validation study: 46 pre- and post-operative MRIs of patients diagnosed with glioblastoma multiforme from 19 centers in the United States.
3 (US board certified neuroradiologists)
Indications for Use
MRIMath i2contour is intended for the semi-automatic labeling, visualization, and volumetric quantification of WHO grade 4 glioblastoma (GBM) from a set of standard MRI images of male or female patients 18 years of age or older who are known to have pathologically proven glioblastoma. Volumetric measurements may be compared to past measurements if available. MRIMath i2contour is not to be used for primary diagnosis and is not intended to be the sole diagnostic metric.
Device Story
Web-based software platform for contouring/segmentation of T1c and FLAIR MRI sequences in patients with diagnosed GBM. Inputs: 2D MRI slices (T1c or FLAIR). Operation: Two independent AI models generate initial tumor contours; user interface allows trained medical professionals to review, manually edit, and approve contours. Output: Volumetric measurements of tumor/subcomponents, segmented images, and PDF report. Used in clinical settings by trained professionals. Benefits: Provides standardized volumetric quantification and aids in tumor contouring process; enables comparison to prior scans. Does not perform primary diagnosis or tumor detection.
Clinical Evidence
Performance evaluated using 46 pre- and post-operative MRIs from 19 clinical centers. AI contours compared against manual segmentations by three board-certified neuroradiologists. Results: T1c AI mean DSC 0.95 (95% CI 0.93-0.96), sensitivity 92.7%, specificity 97.2%. FLAIR AI mean DSC 0.92 (95% CI 0.90-0.94), sensitivity 93.4%, specificity 98.6%. Z-test confirmed AI performance significantly exceeded predicate threshold (p<0.001). Inter-user variability <5% (T1c) and <10% (FLAIR).
Technological Characteristics
Web-based software platform. AI-powered segmentation using two independent models (one for T1c, one for FLAIR). Processes 2D MRI slices. DICOM input/output. No skull stripping or registration required. Standalone software application.
Indications for Use
Indicated for semi-automatic labeling, visualization, and volumetric quantification of WHO grade 4 glioblastoma (GBM) in male or female patients ≥18 years with pathologically proven glioblastoma. Not for primary diagnosis; not sole diagnostic metric.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
August 8, 2024
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
MRIMath LLC % Paul Dryden Consultant ProMedic Consulting LLC 131 Bay Point Dr NE Saint Petersburg, Florida 33704
Re: K233822
Trade/Device Name: i2Contour Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: QIH Dated: July 9, 2024 Received: July 10, 2024
Dear Paul Dryden:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Ningzhi Li-S
for
Daniel Krainak, Ph.D. Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Enclosure
{3}------------------------------------------------
# Indications for Use
| Submission Number ( <i>if known</i> ) | |
|---------------------------------------|--|
|---------------------------------------|--|
K233822
Device Name
i2Contour
Indications for Use (Describe)
MRIMath i2contour is intended for the semi-automatic labeling, visualization, and volumetric quantification of WHO grade 4 glioblastoma (GBM) from a set of standard MRI images of male or female patients 18 years of age or older who are known to have pathologically proven glioblastoma. Volumetric measurements may be compared to past measurements if available. MRIMath i2contour is not to be used for primary diagnosis and is not intended to be the sole diagnostic metric.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 510(k) Summary Page 1 of 5
7-Aug-24
MRIMath LLC 3473 Birchwood Lane Birmingham, AL 35243 Phone: 773-484-8461
Sponsor Contact:
Hayat Rahal - RA
Submission Contact:
Paul Dryden ProMedic, LLC 131 Bay Point Dr NE St. Petersburg, FL 33704
#### Subject Device
Name of Device: i2Contour Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management and Processing System Product Code: QIH Class: II
#### Predicate Device
Neosoma, Inc. NS-HGlio - K221738 Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management and Processing System Product Code: QIH Class: II
#### Device Description
The MRIMath i2Contour is a web-based software platform designed for the contouring and segmentation of the T1c and FLAIR sequences of the MRIs of patients already diagnosed with GBM. It combines AI with a user interface (UI) for review, manual contouring, and approval. The software is intended to be used by trained medical professionals as an aid in the tumor contouring process. Review by a trained professional is a requirement for completion.
The AI algorithm within MRIMath i2Contour generates an initial tumor contour, which serves as a starting point for medical professionals to complete the contouring process manually. It is important to note that the software does not alter the original MRI images and is not intended for turnor detection or diagnostic purposes. MRIMath i2Contour is specifically designed to generate turnor volume contours for GBM. It is not intended for use with images of other brain tumor types.
## Indications for Use
MRIMath i2contour is intended for the semi-automatic labeling, visualization, and volumetric quantification of WHO grade 4 glioblastoma (GBM) from a set of standard MRI images of male or female patients 18 years of age or older who are known to have pathologically proven glioblastoma.
Volumetric measurements may be compared to past measurements if available. MRIMath i2contour is not to be used for primary diagnosis and is not intended to be the sole diagnostic metric.
## Comparison to Predicate Device
We have selected the Neosoma - NS-HGlio, K221738 as the predicate. We have compared the features and performance in the table below.
{5}------------------------------------------------
## 510(k) Summary Page 2 of 5
| | Subject Device | Predicate | Comparison |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|
| | i2Contour | Neosoma, Inc. - NS-HGlio | |
| K# | | K221738 | |
| Product Code | QIH | QIH | |
| CFR | 21 CFR 892.2050 | 21 CFR 892.2050 | |
| Classification | Medical image management and processing system | Medical image management and processing system | |
| Indications for Use | MRIMath i2contour is intended for the semi-automatic<br>labeling, visualization, and volumetric quantification<br>of WHO grade 4 glioblastoma (GBM) from a set of<br>standard MRI images of male or female patients 18<br>years of age or older who are known to have<br>pathologically proven glioblastoma.<br>Volumetric measurements may be compared to past<br>measurements if available. MRIMath i2contour is not<br>to be used for primary diagnosis and is not intended to<br>be the sole diagnostic metric. | NS-HGlio is intended for the semi-automatic<br>labeling, visualization, and volumetric<br>quantification of high-grade brain glioma (WHO<br>grade 3 astrocytoma, WHO grade 4 astrocytoma and<br>WHO grade 4 glioblastoma) from a set of standard<br>MRI images of male or female patients 18 years of<br>age or older who are known to have pathologically<br>proven high-grade glioma.<br>Volumetric measurements may be compared to past<br>measurements if available. NS-HGlio is not to be<br>used for primary diagnosis, and is intended to be<br>used by qualified clinical personnel as an additional<br>source of information and is not intended to be the<br>sole diagnostic metric. | Similar except the subject<br>device is specific for<br>glioblastoma (GBM) |
| Patients | Male or female patients 18 years of age or older who<br>are known to have pathologically proven glioblastoma | Male or female patients 18 years of age or older who<br>are known to have pathologically proven high-grade<br>glioma | Similar with the subject device<br>is limited to GBM |
| Type of Scans Used | MRI:<br>Acquired using two different MRI sequences either in<br>2D or 3D using a specified protocol: T1 post-contrast<br>(T1c) or FLAIR. | MRI:<br>Acquired using four different MRI sequences either<br>in 2D or 3D using a specified protocol: T1, T2, T1<br>post-contrast (T1c) and FLAIR. | Different |
| Intended Anatomy | Brain | Brain | Similar |
| Lesion Review | 2D and 3D | 2D and 3D | Similar |
| Segmentation | Semi-automatic and manual segmentation of<br>glioblastoma | Semi-automatic and manual segmentation of high-<br>grade glioma | Similar |
| Quantification | Volumetric measurements of the combination of the<br>enhancing and necrosis subcomponents of the T1c, and<br>the combination of edema, tumor, and necrosis<br>subcomponents in the FLAIR images of glioblastoma | Volumetric measurement of the edema, necrosis, and<br>enhancing sub-components of high-grade glioma | Different |
| | Subject Device<br>i2Contour | Predicate<br>Neosoma, Inc. - NS-HGlio | Comparison |
| Output | - Provides volumetric measurements of glioblastoma<br>and the tumor + edema + necrosis subcomponents in<br>FLAIR and the enhancing + necrosis subcomponents<br>in T1c series<br>- Includes segmentation of the tumor + edema +<br>necrosis subcomponents in FLAIR and the enhancing<br>+ necrosis subcomponents in T1c series<br>- Automatically compares results to prior scans when<br>available<br>- Provides PDF Report of output data | - Provides volumetric measurements of glioblastoma<br>and the enhancing, necrosis, and edema sub-components<br>- Includes segmentation of sub-components<br>- Automatically compares results to prior scans when<br>available<br>- Provides PDF Report of output data | Different |
| Image Format | DICOM | DICOM | Similar |
| Input | FLAIR or T1c Series | T1, T2, FLAIR and T1c Series | Different |
| Registration | NO | YES | Different |
| Skull Stripping | NO | YES | Different |
| Number of AIs | Two | One | Different |
| Report | YES | YES | Similar |
| Indications | Grade 4 GBM | Grade 3 astrocytoma, grade 4 astrocytoma, grade 4 GBM | Different |
| Evaluation of Accuracy | Using three US board certified neuroradiologists with expertise in measuring GBM | Using three US board certified neuroradiologists with expertise in measuring high grade gliomas | Similar |
# Table 1 - Comparison of Subject vs. Predicate
{6}------------------------------------------------
{7}------------------------------------------------
#### 510(k) Summary Page 4 of 5
## Comparison of Technological Characteristics with the Predicate Device
AI-powered segmentation of the magnetic resonance images (MRI) of patients diagnosed with glioblastoma multiforme is the technological principle for both the subject and predicate devices. At a high level, the subject and predicate devices are based on the following same technological elements:
- The input consists of MRI of patients diagnosed with GBM. ●
- AI-powered prediction of the pixels that correspond to the tumor.
- Computing a tumor volume.
- Review and manual revisions, if needed
The following technological differences exist between the subject and predicate device:
- The subject device includes two independent AIs, one for T1c and the other for the FLAIR ● series; the predicate device consists of a single AI.
- The subject device processes individual 2D slices. The predicate device requires a 3D set of images.
- The predicate device is semi-automated as it requires skull stripping and registration. The subject device is fully automated as it does need registration nor skull stripping.
- The predicate device requires all the four series, T1, T1c, FLAIR and T2 for a single AI. The subject device has two independent AIs for T1c and FLAIR: it does not require the T1 and T2 series.
- The subject device Flair AI output is equivalent to the sum of all the sub-components of the predicate device.
- . The subject device T1c AI output is equivalent to the sum the enhancing lesion and necrosis subcomponents of the predicate device.
The Indications for use for the devices are as follows:
- Neosoma: grade 3 astrocytoma, grade 4 astrocytoma, and grade 4 GBM ●
- MRIMath: grade 4 GBM.
#### Performance Data
Like the predicate, the MRIMath i2Contour evaluated the accuracy of the subject device in the same manner.
The predicate evaluated using 33 subjects and 132 MRIs used for the machine learning model performance of males than females within the age range of 18 to 79. Three US board certified neuroradiologists with expertise in measuring high grade gliomas were used. The device a mean DSC of 0.88 with 95% CI of 0.86-0.90 on preoperative imaging and 0.80 with 95% CI of 0.77-0.83 on postoperative imaging which is higher than the mean DSC of the average of the three experts for the same task, which was 0.84 on preoperative imaging and 0.74 for postoperative imaging respectively.
To evaluate the accuracy of the MRIMath i2Contour FLAIR and T1c AI contours, we compare their outputs with the manual segmentations by three board certified neuro-radiologists of 46 pre- and post-operative MRIs of patients diagnosed with glioblastoma multiforme. The manual contours were performed using the MRIMath smart manual contouring platform. The test MRIs were obtained at 19 centers in the United States located at 13 community hospitals and clinics, 4 imaging centers, and two university hospitals and clinics. The details are as follows: University of Alabama at Birmingham Hospital and Clinics, Birmingham, AL (n=25), MD Anderson Cancer Center, Houston, TX (n = 1), St Vincent Hospital, Birmingham, AL (n =2), Southwest Diagnostic Imaging Center, Dallas, TX (n = 1), Thomas Medical Center, Fairhope, AL (n=1), Carmichael Imaging Center, Montgomery, Alabama (n=1), East Alabama Medical Center, Opelika, AL (n=1), St Dominic, Jackson, MS (n=1), Mobile Infirmary, Mobile, AL (n=1), North Mississippi Medical Center Tupelo, MS (n=1), Sacred Heart Airport Medical Center, Pensacola, FL
{8}------------------------------------------------
#### 510(k) Summary Page 5 of 5
(n=1), SHHP (n=2), LX DCH, Tuscaloosa, AL (n=1), Leeds Imaging Center, Leeds, AL (n=1), Black Warrior Medical Center, Tuscaloosa, AL (n=1), American Health Imaging, Birmingham, AL (n=1), Main (n=1), Floyd Medical Center, Rome, GA (n=1), Trinity Medical Center, Birmingham, AL (n=1), Jackson Hospital, Jackson, MS (n=1).
We evaluate the AI models developed by MRIMath for GBM T1c and fluid attenuation inversion recovery (FLAIR) images, by comparing their contours to three neuro-radiologists, who used the MRIMath smart manual contouring platform. We test the hypothesis that the proportion of overall AI DICE score (DSC) measurements that exceed the designated threshold of po-0.88 is different from 50%. The designated threshold is the best mean DSC achieved by the predicate device. The two-sided, one-sample Z-test shows that:
- · For the FLAIR AI, the DSC proportions exceed p0, 85% of the time, with a confidence interval (CI) of (72%, 92%) and p-value of <0.001, implying that our proportion is significantly different than 50%.
- · For T1C, the DSC proportions exceed pg, 93% of the time, with a CI of (82%, 98%) and p-value of <0.001 of the time, implying that our proportion is significantly different than 50%.
The mean overall DICE scores for the post-contrast T1 (T1c) AI were 0.95 with a 95% confidence interval (C.I) of (93%. 96%), closely matching the radiologists' scores. For true positive T1c images, AI segmentation scored a mean DSC of 83%, versus radiologists ranging from 76% to 86%. Sensitivity and specificity for T1c AI were 92.7% and 97.2%, respectively. The FLAIR AI mean DSC was 92% with a 95% CI interval of (90%, 94%), also matching the radiologists scores. The AI also achieved a mean DICE score of 80% for true positive FLAIR slices, against the radiologists' 75%-83%, and exhibited a median sensitivity and specificity of 93.4% and 98.6%, respectively. The T1C and FLAIR AI models also produced mean Hausdorff distances (< 5 mm), volume measurements, kappa scores, and Bland-Altman differences that align closely with measurements by radiologists. Finally, the inter-user variability between radiologists was < 5% and < 10% for the T1c and FLAIR images, respectively. These results highlight the low inter-user variability of the MRIMath smart manual contouring platform and the high accuracy of its T1c and FLAIR AI models.
# Substantial Equivalence Conclusion
Based on the information submitted in this application, and based on the indications for use, technological characteristics, and performance testing, i2Contour raises no new questions of safety or effectiveness and is substantially equivalent to the predicate device in terms of safety, efficacy, and performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.