K233815 · Inari Medical, Inc. · QEW · Dec 18, 2023 · Cardiovascular
Device Facts
Record ID
K233815
Device Name
ClotTriever Sheath
Applicant
Inari Medical, Inc.
Product Code
QEW · Cardiovascular
Decision Date
Dec 18, 2023
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.5150
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ClotTriever Thrombectomy System is indicated for: • The non-surgical removal of thrombi and emboli from blood vessels. • Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel. The ClotTriever Thrombectomy System is intended for use in the peripheral vasculature including deep vein thrombosis (DVT). The ClotTriever Sheaths are indicated for use as a conduit for the insertion of endovascular devices into the vasculature while minimizing blood loss associated with such insertions.
Device Story
ClotTriever Sheath is a single-use, over-the-wire, catheter-based system for peripheral vasculature procedures. It features a polymeric shaft, distal self-expanding nitinol mesh funnel, and proximal hub with a hemostasis valve. The mesh funnel acts as a backstop for clot removal. The device is inserted over a 0.035" guidewire under fluoroscopic guidance. The funnel is deployed proximal to the thrombus; the dilator is then removed. The sheath serves as a conduit for advancing other endovascular devices to the treatment site and facilitates their withdrawal. The integrated hemostasis valve minimizes blood loss during device exchange. The system includes a sideport for aspiration of clot via a 60 cc syringe. Used by physicians in clinical settings to treat thromboemboli, the device benefits patients by providing a minimally invasive method for clot removal and safe endovascular access.
Clinical Evidence
No clinical data; clinical testing was not required to support substantial equivalence. Verification testing from previous submissions (K180329, K192036) remains applicable.
Technological Characteristics
Polymeric shaft, distal self-expanding nitinol mesh funnel, proximal hub with hemostasis valve, sideport tubing. Guidewire compatibility: 0.035". Sterilization: EtO (SAL 10^-6). Shelf-life: 2 years. Mechanical operation via manual slide actuator for funnel deployment.
Indications for Use
Indicated for non-surgical removal of thrombi/emboli from peripheral vasculature (including DVT) and for injection/infusion/aspiration of fluids. Also indicated as a conduit for endovascular device insertion while minimizing blood loss.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with an abstract design. To the right of the seal, there is the FDA logo in blue, followed by the words "U.S. FOOD & DRUG" in a larger font and "ADMINISTRATION" in a smaller font. The logo is simple and professional, and it is easily recognizable.
December 18, 2023
Inari Medical, Inc. Tosan Eweka Principal Regulatory Affairs Specialist 6001 Oak Canyon, Suite 100 Irvine. California 92618
### Re: K233815
Trade/Device Name: ClotTriever Sheath Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: QEW, KRA, DYB Dated: November 30, 2023 Received: November 30, 2023
Dear Tosan Eweka:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/6 description: This image shows the name Gregory W. O'connell -S. The image also shows that it was digitally signed by Gregory W. O'connell -S on 2023.12.18. The time stamp is 11:55:40 -05'00'.
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K233815
Device Name ClotTriever Sheath
#### Indications for Use (Describe)
The ClotTriever Thrombectomy System is indicated for:
- · The non-surgical removal of thrombi and emboli from blood vessels.
- · Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel.
The ClotTriever Thrombectorny System is in the peripheral vasculature including deep vein thrombosis (DVT).
The ClotTriever Sheaths are indicated for use as a conduit for the insertion of endovascular devices into the vasculature while minimizing blood loss associated with such insertions.
| | Type of Use (Select one or both, as applicable) | |
|--|-------------------------------------------------|--|
| | | |
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
| Date prepared | November 30, 2023 |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | Inari Medical, Inc.<br>6001 Oak Canyon, Suite 100<br>Irvine, CA 92618<br>949.600.8433 |
| Contact person | Tosan Eweka<br>Principal Regulatory Affairs Specialist |
| Trade Name | ClotTriever Sheath |
| Device Name | ClotTriever Sheath |
| Common name | Embolectomy catheter |
| Regulation name | Embolectomy catheter |
| Classification<br>number | 21 CFR 870.5150 |
| Product code | QEW, KRA, DYB |
| Regulatory class | II |
| Predicate device | Protrieve Sheath (K230331) |
| Description | The ClotTriever Thrombectomy System ("ClotTriever System") is a single-use over-<br>the-wire catheter-based system used for the minimally invasive treatment of<br>thromboemboli in the peripheral vasculature. The system consists of ClotTriever<br>Catheters, ClotTriever Sheaths and Protrieve Sheath, each packaged separately.<br>The ClotTriever Sheath consist of a polymeric shaft with a distal self-expanding<br>nitinol mesh funnel, a proximal hub with a hemostasis valve and a sideport tubing.<br>The mesh funnel is deployed by retracting the slide actuator back until it snaps into<br>place. The mesh funnel self-expands to the diameter of the vessel and serves as a<br>backstop as clot is removed through the lumen of the sheath. A hemostasis valve is<br>integrated into the proximal hub of the sheath to prevent blood loss from devices<br>passing through it. The sideport tubing has a large-bore stopcock and a terminal quick-<br>disconnect coupling for connection to the Large Bore 60 cc Syringe for the aspiration<br>of clot. A pre-dilator is provided with the ClotTriever Sheaths, to expand the target<br>vessel prior to sheath insertion. A dilator is provided to aid insertion and positioning<br>of the sheath, while the loading tool facilitates insertion of the mesh funnel into the<br>dilator. |
| Indications for Use | The ClotTriever Thrombectomy System is indicated for:<br>• The non-surgical removal of thrombi and emboli from blood vessels.<br>• Injection, infusion, and/or aspiration of contrast media and other fluids into<br>or from a blood vessel.<br>The ClotTriever Thrombectomy System is intended for use in the peripheral<br>vasculature including deep vein thrombosis (DVT). |
{4}------------------------------------------------
| Device<br>Modifications | The ClotTriever Sheaths are indicated for use as a conduit for the insertion of<br>endovascular devices into the vasculature while minimizing blood loss associated<br>with such insertions. | | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device<br>Modifications | The purpose of this submission is to modify the indications for use statement to<br>include use of the ClotTriever Sheaths as conduits for the insertion of endovascular<br>devices into the vasculature while minimizing blood loss associated with such<br>insertions. | | |
| Summary of<br>substantial<br>equivalence | The subject device and predicate device have the same intended use and indications<br>for use. Both devices are intended to be used as conduits for the insertion of<br>endovascular devices into the vasculature while minimizing blood loss associated<br>with such insertions. The subject device has the same principle of operation and<br>similar technological characteristics as the predicate device. A tabular comparison<br>between the predicate and subject device is provided below: | | |
| | ClotTriever Sheaths<br>(Subject Device) | Protrieve Sheath<br>(Predicate Device) | |
| Manufacturer | Inari Medical, Inc. | Inari Medical, Inc. | |
| Model<br>Designators | 50-101 (13 Fr ClotTriever<br>Sheath)<br>51-101 (16 Fr ClotTriever<br>Sheath) | 60-101 (20 Fr Protrieve<br>Sheath) | |
| Product Code | QEW, KRA, DYB | QEW, KRA, DYB | |
| Indications for<br>Use | The ClotTriever<br>Thrombectomy System is<br>indicated for:<br>• The non-surgical removal<br>of thrombi and emboli from<br>blood vessels.<br>• Injection, infusion, and/or<br>aspiration of contrast media<br>and other fluids into or from<br>a blood vessel.<br>The ClotTriever<br>Thrombectomy System is<br>intended for use in the<br>peripheral vasculature<br>including deep vein<br>thrombosis (DVT).<br>The ClotTriever Sheaths are<br>indicated for use as a conduit | The ClotTriever<br>Thrombectomy System is<br>indicated for:<br>• The non-surgical removal<br>of thrombi and emboli from<br>blood vessels.<br>• Injection, infusion, and/or<br>aspiration of contrast media<br>and other fluids into or from<br>a blood vessel.<br>The ClotTriever<br>Thrombectomy System is<br>intended for use in the<br>peripheral vasculature<br>including deep vein<br>thrombosis (DVT).<br>The Protrieve Sheath is<br>indicated for use as a conduit | |
| | | for the insertion of<br>endovascular devices into<br>the vasculature while<br>minimizing blood loss<br>associated with such<br>insertions. | for the insertion of<br>endovascular devices into<br>the vasculature while<br>minimizing blood loss<br>associated with such<br>insertions. |
| Principles of<br>Operation | | The ClotTriever Sheath and<br>dilator are inserted over a<br>pre-placed 0.035" guidewire<br>into the vessel. Under<br>fluoroscopic guidance, the<br>Sheath's mesh funnel is<br>deployed proximal to the<br>thrombus targeted for<br>extraction. After<br>deployment of the funnel,<br>the dilator is withdrawn<br>through the Sheath and from<br>the patient entirely. The<br>catheter or other<br>endovascular device is then<br>advanced over the guidewire<br>through the ClotTriever<br>Sheath to the targeted<br>treatment site. Following the<br>diagnostic or therapeutic<br>procedure, the endovascular<br>device is retracted through<br>the ClotTriever Sheath and<br>removed from the patient.<br>The user-actuated<br>hemostasis valve integrated<br>into the Sheath hub allows<br>for Insertion and withdrawal<br>of endovascular devices. | The Protrieve Sheath and<br>dilator are inserted over a<br>pre-placed 0.035" guidewire<br>into the vessel. Under<br>fluoroscopic guidance, the<br>Sheath's mesh funnel is<br>deployed proximal to the<br>thrombus targeted for<br>extraction. After deployment<br>of the funnel, the dilator is<br>withdrawn through the<br>Sheath and from the patient<br>entirely. The catheter or<br>other endovascular device is<br>then advanced over the<br>guidewire through the<br>Protrieve Sheath to the<br>targeted treatment site.<br>Following the diagnostic or<br>therapeutic procedure, the<br>endovascular device is<br>retracted through the<br>Protrieve Sheath and<br>removed from the patient.<br>The user-actuated hemostasis<br>valve integrated into the<br>Sheath hub allows for<br>Insertion and withdrawal of<br>endovascular devices. |
| Target Vessel | | Peripheral vessels | Peripheral vessels |
| Guidewire<br>compatibility | | 0.035" | 0.035" |
| Sheath shaft<br>ID/OD | | 13 Fr ID - 0.180"<br>16 Fr ID - 0.215"<br>13 Fr OD - 0.210"<br>16 Fr OD - 0.248" | ID - 0.270"<br>OD - 0.345" |
| Sheath length<br>(deployed) | | 15 cm | 32 cm |
| Dilator OD | 13 Fr Sheath - 0.178" | 0.264" | |
| | 16 Fr Sheath - 0.206" | | |
| Dilator Length | 10" | 25" | |
| Sterilization | SAL 10-6, EtO | SAL 10-6, EtO | |
| Shelf-Life | 2 years | 2 years | |
| Non-Clinical Testing | | | |
| Verification testing provided in K180329 and K192036 remains applicable to support<br>use of the ClotTriever Sheaths as conduits for the insertion of endovascular<br>devices into the peripheral vasculature. | |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.