K5 is intended to supply power to and serve as a base for dental devices and accessories. This product includes a dental chair. The dental treatment unit is intended for use in the dental clinic environment and is used by trained dentists and/or dental assistants.
Device Story
K5 is an AC-powered dental operative unit and chair system; provides base and power for dental instruments. Inputs include user commands via touch panels and foot controls; sensors detect cup presence and suction handpiece removal. System uses electronic control circuits and solenoid valves to manage chair positioning, water/air flow, and suction. Outputs include mechanical chair movement, air/water delivery via 3-way syringe, and suction. Used in dental clinics by trained personnel. Benefits include ergonomic patient positioning and integrated instrument support. Device features include adjustable chair, dentist/assistant elements, and water sanitation system. Clinical decision-making is supported by providing a stable, functional platform for dental procedures.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing, including biocompatibility (ISO 10993-5), cleaning/disinfection validation (ISO 16954), electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and usability (IEC 60601-1-6, IEC 62366).
Indicated for use in dental clinics by trained dentists or dental assistants to provide power and a base for dental devices and accessories during treatment of children and adults.
Regulatory Classification
Identification
A dental operative unit and accessories is an AC-powered device that is intended to supply power to and serve as a base for other dental devices, such as a dental handpiece, a dental operating light, an air or water syringe unit, and oral cavity evacuator, a suction operative unit, and other dental devices and accessories. The device may be attached to a dental chair.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 3, 2024
OSSTEM Implant Co., Ltd. Chair Business % Peter Lee Sr. Manager QA/RA Production HIOSSEN Inc. 85 Ben Fairless Dr. Fairless Hills, Pennsylvania 19030
Re: K233805
Trade/Device Name: K5 Regulation Number: 21 CFR 872.6640 Regulation Name: Dental Operative Unit And Accessories Regulatory Class: Class I, reserved Product Code: EIA, KLC Dated: August 14, 2024 Received: August 14, 2024
Dear Peter Lee:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Michael E. Adjodha -S
Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Submission Number (if known)
| Submission Number (if known) | K233805 |
|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | K5 |
| Indications for Use (Describe) | K5 is intended to supply power to and serve as a base for dental devices and accessories. This product includes a dental chair. The dental treatment unit is intended for use in the dental clinic environment and is used by trained dentists and/or dental assistants. |
| Type of Use (Select one or both, as applicable) | <div> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </div> <div> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
{4}------------------------------------------------
192, Haebong-ro, Danwon-gu, Ansan-si, Gyeonggi-do, 15428, Republic of Korea Tel: +82 2 2016 7000 Fax: +82 2 2016 7001 www.osstem.com
#### 510(k) Summary K233805
### Date: September 2, 2024
## 1. Company and Correspondent making the submission
| - Submitter's Name | : Osstem Implant Co., Ltd. Chiar Business |
|------------------------|---------------------------------------------------------------------------------|
| - Address | : 192, Haebong-ro, Danwon-gu, Ansan-si, Gyeonggi-do<br>15428, Republic of Korea |
| - Contact | : Mr. Sangyong Lee |
| - Phone | : +82-70-4626-0881 |
| - Correspondent's Name | : Hiossen Inc. |
| - Address | : 85 Ben Fairless Dr. Fairless Hills, PA 19030 |
| - Contact | : Mr. Peter Lee |
| - Phone | : +1-267-759-7031 |
### 2. Proposed Device
| - Trade or (Proprietary) Name | : K5 |
|-------------------------------|-----------------------------------------|
| - Classification Name | : Dental operative unit and accessories |
| - Regulation Number | : 21 CFR872.6640 |
| - Devce Classification | : Class I |
| - Classification Product Code | : EIA |
## 3. Predicated Device
- K3(K183347) -
## 4. Description
- A Dental Unit and Chair K5 is designed for dental treatment of children and adults. It is a dental treatment unit in accordance with IEC 80601-2-60. The 3-way syringe is a dental instrument in accordance with EN 1639. It aids the dental application in the mouth of the patient by supplying air, water or spray. This product is designed for use in dentistry only and may only be used by trained medical personnel. K5 is similar to other commercially available products based on the intended use, the technology used, the claims, the electrical power and performance characteristics. It is substantially equivalent in design, function and intended use to the predicate device.
## (1) Design Feature
K5, dental operative unit, has features of ergonomic and optimal position for treatment. It consists of adjusted touch panel, 3-way syringe, assistant touch panel, easy opening
{5}------------------------------------------------
192, Haebong-ro, Danwon-gu, Ansan-si, Gyeonggi-do, 15428, Republic of Korea Tel: +82 2 2016 7000 Fax: +82 2 2016 7001 www.osstem.com
and closing arm rest.
#### (2) Main Types
There are a couple of types for this device –K5 Mount and K5 Cart which consists of chair, unit, table, seat, stool, 3-way syringe, monitor arms, fool control and console.
#### (3) Technological Characteristics
K5 is an AC-powered dental operative unit with accessories, intended to supply power to and serve as a base for other dental devices. It includes a treatment chair, dentist element, assistant element and a dental light as offering several additional options and electronically-controlled chair movements with software and water unit functions.
#### 5. Indication for use
K.5 is intended to supply power to and serve as a base for dental devices and accessories. This product includes a dental chair. The dental treatment unit is intended for use in the dental clinic environment and is used by trained dentists and/or dental assistants.
#### 6. Summary of Technological Characteristics
The K5 functions in a manner similar to and is intended for the same use to the predicate device. Primarily, K5 is substantially equivalent to the K3 (K183347) marketed by OSSTEM IMPLANT Co., Ltd and does have some different technological characteristics and slightly different external design. However, these differences do not raise new concerns of substantial equivalence as the performance data and testing of the K5 demonstrate that the devices are deemed to be substantially equivalent as described in a following comparison table:
| Description | Subject Device | Predicate Device |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | K5 is intended to supply power to<br>and serve as a base for dental<br>devices and accessories. This<br>product includes a dental chair.<br>The dental treatment unit is<br>intended for use in the dental<br>clinic environment and is used by<br>trained dentists and/or dental<br>assistants. | K3 is intended to supply power to<br>and serve as a base for dental<br>devices and accessories. This<br>product includes a dental chair.<br>The dental treatment unit is<br>intended for use in the dental<br>clinic environment and is used by<br>trained dentists and/or dental<br>assistants. |
| Product Name | K5 | K3 |
| Code | EIA | EIA |
| Power & Utility<br>Supply | AC 100-120/220-240V,<br>50/60Hz, compressed air and<br>water | AC 100-120/220-240V,<br>50/60Hz, compressed air and<br>water |
| Main Components | Chair, Unit, Table, Seat, Stool, | Chair, Unit, Table, Seat, Stool, |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Osstem Implant. The logo is orange and features the word "OSSTEM" in a bold, sans-serif font. Below the word "OSSTEM" is the word "IMPLANT" in a smaller, italicized font. The logo is simple and modern, and the orange color is eye-catching.
92, Haebong-ro, Danwon-gu, Ansan-si, Gyeonggi-do, 15428, Republic of Korea el: +82 2 2016 7000 Fax: +82 2 2016 7001 www.osstem.com
| | Monitor Arm, Hanaro Console | Monitor Arm*, Hanaro Console* (Note: K3 Cart* model applied ONLY) |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|
| Syringe | 3-way syringe | 3-way syringe |
| Control of water and air | Uses pneumatically controlled vales to water control the flow of air and water. On/off and intensity controlled by foot pedal. | Uses pneumatically controlled vales to water control the flow of air and water. On/off and intensity controlled by foot pedal. |
| Air Pressure | 550kPa(min)/750kPa(max) | 500kPa(min)/750kPa(max) |
| Water Pressure | 250kPa(min)/600 kPa(max) | 250kPa(min)/600 kPa(max) |
| Water System | City water supply | City water supply |
| Water Sanitation System | Distilled water container added | Distilled water container added |
| Suction | HVE (High volume evacuator) Saliva Ejectors | HVE (High volume evacuator) Saliva Ejectors |
| Patient Load | Max. 150kg | Max. 135kg |
| Chair Height | Max. 840±30mm, Min. 440±30mm | Max. 795±10mm, Min. 365±10mm |
| Back Rest | $0°±5°$ to $70°±5°$ | $0°±5°$ to $67°±5°$ |
| Head Rest | -90° to 70° | -10° to 45° |
| Lift Motor | Electromotor | Hydraulic electromotor |
| Electrical Safety | Complied with IEC 60601-1 | Complied with IEC 60601-1 |
| Electromagnetic compatibility | Complied with IEC 60601-1-2 | Complied with IEC 60601-1-2 |
[Table 1] Comparison between K3 and its predicate device
#### 7. Principle of Operation
#### Raising and lowering the dental chair
The K5 is an electronically controlled, electrically powered chair. The footswitch is used to position the chair and program auto-positioning functions into the chair. The chair hydraulic system is controlled by the electronics using electronic switching components and electrically powered solenoid valves.
#### Automatic positioning
Operating the 'Automatic' S/W on the Table Panel commands the chair's control device to automatically move and position the chair and seat to a predetermined position.
#### Hand-Piece Operation Circuit
Micro S/W, closing the supply air valve and opening the rubber film of the Master Block. Pressing the controller (Foot Control) Pedal at this time opens the supply air valve inside the Table, and the air of the compressor is supplied to the removed Hand-Piece through the
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Osstem Implant. The logo is orange and features the word "OSSTEM" in a bold, sans-serif font. Below the word "OSSTEM" is the word "IMPLANT" in a smaller, lighter font. There is a small circle in the upper right corner of the logo.
192, Haebong-ro, Danwon-gu, Ansan-si, Gyeonggi-do, 15428, Republic of Korea Tel: +82 2 2016 7000 Fax: +82 2 2016 7001 www.osstem.com
distribution block, thereby rotating the Hand-Piece Turbine. The air used to rotate the Hand-Piece Turbine is discharged to the atmosphere through the exhaust pipe.
#### Vacuum Operation
Removing the Suction from the Holder activates the Sensor mounted inside the Holder, opening the Air Solenoid Valve and supplying air to the suction device to enable suction.
#### 3-Way Syringe
Pressing the button on the 3-way syringe , water or air is discharged.
#### Water Supply System
Upon detection of the cup by the Water Supply Port Sensor, PCB's Volume controls water volume.
#### Water Tube Cleaning
Pressing the Water Supply Button and Film Viewer Button together after supplying 1.2 L of cleaning solution (2~5% of hydrogen peroxide) to the Water Tube Cleaning/Distilled Water Container activates the Water Tube Cleaning System.
#### 8. Non-clinical Test Data
#### Biocompatibility Testing
The biocompatibility evaluation for K5 components was conducted in accordance with the FDA Guidance Document and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices -Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The waterlines and airlines were tested for cytotoxicity (ISO 10993-5). The components of the K5 are considered external communicating device for a duration of less than 24 hours.
#### Cleaning Validation
Cleaning/disinfection validation was conducted on the waterlines of the subject device. Validation was conducted using the following standards:
O 16954:2015 Dentistry - Test methods for dental unit waterline biofilm treatment -Sampling for microbiological analysis
- Essential characteristics of test methods for the evaluation of treatment methods intended to improve or maintain the microbiological quality of dental unit procedural water. The microbiological simulation test and physical & chemical test are performed to ensure that the Dental chair waterline is effectively cleaned.
Reprocessing validation was conducted per FDA Guidance, "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling - Guidance for Industry and Food and Drug Administration Staff (fda.gov)" including cleaning, disinfection, and sterilization validation for the components of the K5 device.
{8}------------------------------------------------
OSSTEM
IMPLANT
192, Haebong-ro, Danwon-gu, Ansan-si, Gyeonggi-do, 15428, Republic of Korea Tel: +82 2 2016 7000 Fax: +82 2 2016 7001 www.osstem.com
#### Electrical Safety and Electromagnetic compatibility(EMC)
The Electrical Safety and Electromagnetic compatibility tests were performed in accordance with the following standards. Comprehensive performance testing has been conducted on the K5 in accordance FDA recognized standards. EMC testing was conducted in accordance with Standard IEC 60601-1-2. Electrical, mechanical, and environmental safety testing according to Standard IEC 60601-1was performed. Usability testing was conducted in accordance with Standard IEC 60601-1-6 and IEC 62366.
#### Software and System Verification and Validation
Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern.
#### 9. Clinical Test Data
No clinical studies are submitted.
No clinical studies are needed to characterize its performance and establish substantial equivalence.
#### 10. Conclusion
Based upon the above mentioned data and comparison table, the K5 is substantially
equivalent to the predicate device as described herein.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.