CarboClear® X Pedicle Screw System

K233793 · CarboFix Orthopedics , Ltd. · NKB · Aug 1, 2024 · Orthopedic

Device Facts

Record IDK233793
Device NameCarboClear® X Pedicle Screw System
ApplicantCarboFix Orthopedics , Ltd.
Product CodeNKB · Orthopedic
Decision DateAug 1, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3070
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The CarboClear® X Pedicle Screw System is intended to provide rigid immobilization and stabilization of lumbar and/or sacral segments as an adjunct to fusion in patients with degenerative disc disease (DDD) at up to three levels from L2 to S1, with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Patients should be skeletally mature and have at least six months of non-operative treatment. The CarboClear® X Pedicle Screw System is intended to be used with intervertebral body fusion device/s implanted at the same spinal level/s with autogenous and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft.

Device Story

CarboClear® X Pedicle Screw System provides spinal stabilization; used as adjunct to fusion in lumbar/sacral segments. System comprises pedicle screws, rods, and locking set screws; implants constructed from carbon fiber-reinforced PEEK (CFR-PEEK) with thin titanium shell on threaded screw portions; titanium alloy rods also available; tantalum markers included. Used by surgeons in open or minimally invasive (MIS) procedures. Implants provide rigid immobilization; intended for single use; supplied sterile. Benefits include stabilization of spinal segments in patients with DDD and spondylolisthesis.

Clinical Evidence

Bench testing only. Static and dynamic testing performed per ASTM F1717; static flexion-extension, axial gripping capacity, and torsional gripping capacity testing performed per ASTM F1798. Results demonstrate performance comparable to predicate devices.

Technological Characteristics

Materials: Carbon fiber-reinforced PEEK (CFR-PEEK), titanium alloy, tantalum markers. Components: Pedicle screws, rods, locking set screws. Standards: ASTM F1717 (static/dynamic testing), ASTM F1798 (gripping capacity). Sterilization: Supplied sterile. Usage: Single-use. Approach: Open or MIS.

Indications for Use

Indicated for skeletally mature patients with degenerative disc disease (DDD) at up to three levels from L2 to S1, with up to Grade I spondylolisthesis, requiring rigid immobilization and stabilization as an adjunct to fusion. Requires at least six months of prior non-operative treatment.

Regulatory Classification

Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards; (ii) Compliance with mechanical testing standards; (iii) Compliance with biocompatibility standards; and (iv) Labeling that contains these two statements in addition to other appropriate labeling information: “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.” “Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.” (2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls: (i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use. (ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant. (iii) Device components must be demonstrated to be biocompatible. (iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments. (v) Labeling must include the following: (A) A clear description of the technological features of the device including identification of device materials and the principles of device operation; (B) Intended use and indications for use, including levels of fixation; (C) Identification of magnetic resonance (MR) compatibility status; (D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and (E) Detailed instructions of each surgical step, including device removal. (3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls: (i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate. (ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ August 1, 2024 Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. CarboFix Orthopedics Ltd. Hila Wachsler-Avrahami Regulatory Affairs 11 Ha'Hoshlim Street Herzeliya, 4672411 Israel Re: K233793 Trade/Device Name: CarboClear® X Pedicle Screw System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral Pedicle Screw System Regulatory Class: Class II Product Code: NKB Dated: June 2, 2024 Received: June 3, 2024 Dear Hila Wachsler-Avrahami: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). {1}------------------------------------------------ Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Image /page/1/Picture/6 description: The image shows a digital signature. The signature is from Eileen Cadel. The date of the signature is August 1st, 2024. The time of the signature is 08:59:46 -04'00'. Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health {2}------------------------------------------------ # Indications for Use Submission Number (if known) K233793 Device Name CarboClear® X Pedicle Screw System Indications for Use (Describe) CarboClear® X Pedicle Screw System (DDD) The CarboClear® X Pedicle Screw System is intended to provide rigid immobilization and stabilization of lumbar and/or sacral segments as an adjunct to fusion in patients with degenerative disc disease (DDD) at up to three levels from L2 to S1, with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Patients should be skeletally mature and have at least six months of nonoperative treatment. The CarboClear® X Pedicle Screw System is intended to be used with intervertebral body fusion device/s implanted at the same spinal level/s with autogenous and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(k) Summary # CarboFix Orthopedics Ltd. # CarboClear® X Pedicle Screw System # Applicant Name CarboFix Orthopedics Ltd. 11 Ha'hoshlim St., Herzeliya 4672411, Israel Tel: +972 9 9511511, Fax: +972 9 9548939 ## Contact Person Hila Wachsler-Avrahami CarboFix Orthopedics Ltd. 11 Ha'hoshlim St., Herzeliya 4672411, Israel Tel: +972 9 9511511, Fax: +972 9 9548939 # Date Prepared June 2, 2024 ### Trade/Proprietary Name CarboClear® X Pedicle Screw System #### Common Name Pedicle Screw System ### Regulation Number and Device Class Class II; 21 CFR §888.3070 ## Product Code, Regulatory Description and Review Panel NKB; Thoracolumbosacral Pedicle Screw System; Orthopedic {4}------------------------------------------------ ## Predicate Devices ### Primary Predicate Device: CarboClear® X Pedicle Screw System (CarboFix Orthopedics Ltd.; K233944, K240846) ### Additional Predicate Devices: - " CarboClear® Pedicle Screw System (CarboFix Orthopedics Ltd.; K232341) - CarboClear® X Pedicle Screw System (CarboFix Orthopedics Ltd.; K210716, K231280) - · CD HORIZON® Spinal System (Medtronic Sofamor Danek; K082236 and more) - · S4® Spinal System (Aesculap Spine; K130291 and more) # Intended Use/Indications for Use ### CarboClear® X Pedicle Screw System The CarboClear® X Pedicle Screw System is intended to provide rigid immobilization and stabilization of lumbar and/or sacral segments as an adjunct to fusion in patients with degenerative disc disease (DDD) at up to three levels from L2 to S1, with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Patients should be skeletally mature and have at least six months of non-operative treatment. The CarboClear® X Pedicle Screw System is intended to be used with intervertebral body fusion device/s implanted at the same spinal level/s with autogenous and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft. ### System Description The CarboClear® X Pedicle Screw System is composed of implants in various dimensions, used to build a spinal construct; and of a set of instruments, intended to assist in the insertion and placement of the implants. The implants include pedicle screws, rods, and a locking element (set screw). They are made of carbon fiber-reinforced polyetheretherketone (CFR-PEEK). The threaded portion of the pedicle screws is encased within a thin titanium shell. The implants may include tantalum markers. Titanium alloy rod is also available. The implants are supplied sterile, and are intended for single use. The CarboClear® X implants may be implanted in an open approach or using a minimally invasive surgery (MIS) approach. {5}------------------------------------------------ ### Purpose of Submission Expansion of indications for use. #### Comparison of Technological Characteristics with Predicate Devices Compared to the primary predicate device, the subject CarboClear® X System includes longer rods. Otherwise, the subject CarboClear® X implant components are the same as the corresponding implant components of the cleared CarboClear X Pedicle Screw System (K233944, K240846). #### Performance Data Performance characteristics of the CarboClear® X Pedicle Screw System included static and dynamic tests according to ASTM F1717, as well as static flexion-extension test, axial gripping capacity test, and torsional gripping capacity test according to ASTM F1798. The results of the tests are comparable to those of the predicate devices, demonstrating substantially equivalent performance of the subject and predicate devices. #### Substantial Equivalence The intended use, design, dimensions, materials, technological characteristics, principles of operation, and performance of the subject CarboClear® X Pedicle Screw System are substantially equivalent to those of the predicate devices. ### Conclusion Based on the information provided in this 510(k) Premarket Notification, the subject CarboClear® X Pedicle Screw System is substantially equivalent to its predicate devices.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...