← Product Code [JAK](/productcode/JAK) · K233767

# OmniTom Elite (K233767)

_Neurologica Corporation · JAK · Jun 10, 2024 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K233767

## Device Facts

- **Applicant:** Neurologica Corporation
- **Product Code:** [JAK](/productcode/JAK.md)
- **Decision Date:** Jun 10, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1750
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

The OmniTom Elite computed tomography (CT) system is intended to be used for X-ray computed tomography applications for anatomy that can be imaged in the 40cm aperture, primarily the head and neck. The CT system is intended to be used for both pediatric and adult imaging and as such has preset dose settings based upon weight and age The CT images can be obtained either with or without contrast. OmniTom Elite with photon counting detectors (PCD) configuration has multi-energy CT functionality with spectral capability for material decomposition and virtual monoenergetic images (VMI). OmniTom Elite with PCD is supported for adult imaging of the head and neck.

## Device Story

OmniTom Elite is a mobile, translating CT scanner featuring a 40cm bore and 30cm field of view. It utilizes a rotating gantry with a solid-state X-ray generator, collimator, and photon counting detectors (PCD). The system captures multi-energy CT data to enable material decomposition and virtual monoenergetic imaging (VMI). Operated by clinicians in clinical settings, the device provides images for diagnostic review. The system includes an MD Plus algorithm for bone removal and virtual non-contrast (VNC) processing. It supports helical scanning and ultra-high-resolution modes. Output images assist radiologists in clinical decision-making by providing detailed anatomical information. Benefits include high-resolution imaging and spectral capabilities for tissue characterization, facilitating diagnostic assessment in a portable format.

## Clinical Evidence

Clinical evaluation involved volunteer CT scans of the head (skull, midbrain, sinuses). Images were reviewed by independent board-certified radiologists, who confirmed diagnostic quality equivalent to the predicate device. Bench testing included software verification/validation (IEC 62304), electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and phantom-based image quality metrics (noise, slice thickness, resolution, MTF).

## Technological Characteristics

Mobile CT system with 40cm aperture; 16-row detector array. PCD configuration supports multi-energy spectral imaging. Materials/components: X-ray generator, collimator, slip-ring, brushless DC servo drive. Connectivity: DICOM. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-44, NEMA XR 25/26/28/29. Software: MD Plus algorithm for material decomposition. Sterilization: N/A (non-invasive).

## Regulatory Identification

A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- OmniTom Elite ([K223447](/device/K223447.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

June 10, 2024

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

NeuroLogica Corporation % Ninad Gujar Vice President 14 Electronics Avenue DANVERS, MA 01923

Re: K233767

Trade/Device Name: OmniTom Elite Regulation Number: 21 CFR 892.1750 Regulation Name: Computed Tomography X-Ray System Regulatory Class: Class II Product Code: JAK Dated: November 16, 2023 Received: May 6, 2024

Dear Ninad Gujar:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

approvals in the device master record (21 CFR 820.181).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

1

for

Lu Jiang Assistant Director DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K233767

Device Name OmniTom Elite

#### Indications for Use (Describe)

The OmniTom Elite computed tomography (CT) system is intended to be used for X-ray computed tomography applications for anatomy that can be imaged in the 40cm aperture, primarily the head and neck.

The CT system is intended to be used for both pediatric and adult imaging and as such has preset dose settings based upon weight and age The CT images can be obtained either with or without contrast.

OmniTom Elite with photon counting detectors (PCD) configuration has multi-energy CT functionality with spectral capability for material decomposition and virtual monoenergetic images (VMI). OmniTom Elite with PCD is supported for adult imaging of the head and neck.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "NeuroLogica" in a bold, sans-serif font. The "O" in "NeuroLogica" is replaced with a blue circle. Below the word "NeuroLogica" is the text "A Subsidiary of Samsung Electronics Co., Ltd" in a smaller font.

## K233767 510(k) Summary

Date Prepared: May 4, 2024

## Submitter

NeuroLogica Corp., a subsidiary of Samsung Electronics Co., Ltd 14 Electronics Avenue, Danvers, MA 01923

## Establishment Registration

3004938766

#### Manufacturing Site

NeuroLogica Corp., a subsidiary of Samsung Electronics Co., Ltd 14 Electronics Avenue, Danvers, MA 01923

#### Official Correspondent & Contact Person

Dr. Ninad Gujar Vice President, Regulatory Affairs & Quality Assurance Telephone: 978-564-8632 E-mail: ngujar@neurologica.com

## Device Name:

| Trade Name:            | OmniTom Elite                    |
|------------------------|----------------------------------|
| Device Model:          | NL5000                           |
| Classification Name:   | Computed Tomography X-ray System |
| Product Code:          | JAK                              |
| Device Classification: | Class II (per 21 CFR § 892.1750) |

#### Predicate Device:

| 510(k):                | K223447                          |
|------------------------|----------------------------------|
| Trade Name:            | OmniTom Elite                    |
| Device Model:          | NL5000                           |
| Classification Name:   | Computed Tomography X-ray System |
| Product Code:          | JAK                              |
| Device Classification: | Class II (per 21 CFR § 892.1750) |

{4}------------------------------------------------

#### Device Description:

The subject OmniTom Elite Computed Tomography (CT) system provides the same functionality as the previous version of the device, OmniTom Elite (K223447). Both computed tomography systems are identical in terms of the high resolution, 16 row, 40 cm bore, and 30 cm field of view. The lightweight translating gantry consists of a rotating disk with a solid-state x-ray generator, collimator, control computer, communications link, power slip-ring, data acquisition system, reconstruction computer, power system, brushless DC servo drive system (disk rotation) and an internal drive system (translation). The power system consists of batteries which provide system power while unplugged from the charging outlet. The system has the necessary safety features such as the emergency stop switch, x-ray indicators, interlocks, patient alignment laser and 110% x-ray timer. The gantry has omni-directional wheels that allow for robust diagonal, lateral, and rotational 360-degree movement and electrical drive system so the system can be moved easily to different locations.

The ability to upgrade the OmniTom Elite system with photon counting detector (PCD) remains the same. PCD provides the ability to capture CT data in multiple energy bands that can provide information on material composition of different tissues and contrast media. The multiple sets of CT data are acquired at the same time with configurable energy thresholds without any cross talk between images.

#### Indications for Use:

The OmniTom Elite computed tomography (CT) system is intended to be used for xray computed tomography applications for anatomy that can be imaged in the 40 cm aperture, primarily head and neck. The CT system is intended to be used for both pediatric and adult imaging and as such has preset dose settings based upon weight and age. The CT images can be obtained either with or without contrast.

OmniTom Elite with photon counting detectors (PCD) configuration has multi-energy CT functionality with spectral capability for material decomposition and virtual monoenergetic images (VMI). OmniTom Elite with PCD is supported for adult imaging of the head and neck.

#### Comparison of Technological Characteristics with the Predicate Device:

We modified the cleared OmniTom Elite (K223447) within our design controls to include technology improvements that include ultra-high resolution modes for PCD along with CT features based on material decomposition algorithm (MD Plus) for bone removal and virtual non contrast (VNC). Helical scanning capability was added along with increasing the field of view (FOV) and operating temperature.

{5}------------------------------------------------

The indications for use between the subject device and the predicate device (K223447) only vary slightly, in that the subject device offers an additional resolution mode for the photon counting detector. The subject device still contains the same technology and basic functionality as the predicate device (K223447), and therefore, safety and efficacy are not of a concern. Please find both the subject and predicate indications for use below.

| OmniTom Elite (Subject Device)                                                                                                                                                                                                                                                                | OmniTom Elite (Predicate – K223447)                                                                                                                                                                                                                                                                   |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| The OmniTom Elite CT system is intended<br>to be used for x-ray computed tomography<br>applications for anatomy that can be imaged<br>in the 40cm aperture, primarily head and<br>neck.                                                                                                       | The OmniTom Elite computed tomography<br>(CT) system is intended to be used for x-ray<br>computed tomography applications for<br>anatomy that can be imaged in the 40cm<br>aperture, primarily head and neck.                                                                                         |
| The CT system is intended to be used for<br>both pediatric and adult imaging and as such<br>has preset dose settings based upon weight<br>and age. The CT images can be obtained<br>either with or without contrast.                                                                          | The CT system is intended to be used for<br>both pediatric and adult imaging and as such<br>has preset dose settings based upon weight<br>and age. The CT images can be obtained<br>either with or without contrast.                                                                                  |
| OmniTom Elite with photon counting<br>detectors (PCD) configuration has multi-<br>energy CT functionality with spectral<br>capability for material decomposition and<br>virtual monoenergetic images (VMI).<br>OmniTom Elite with PCD is supported for<br>adult imaging of the head and neck. | OmniTom Elite with photon counting<br>detectors (PCD) configuration has multi-<br>energy CT functionality with spectral<br>capability for material decomposition and<br>virtual monoenergetic images (VMI).<br>OmniTom Elite with PCD is only supported<br>for adult axial imaging for head and neck. |

The optional upgrade of the photon counting detector (PCD) remains the same from a hardware standpoint with the exception of additional fans for temperature regulation. OmniTom Elite with the optional PCD kit for its intended use, is of comparable type in design, material, functionality, and technology and is substantially equivalent to the cleared predicate device - OmniTom Elite (K223447).

Similarities

- 트 Design: The OmniTom Elite is identical in design characteristics to its previous version and shares all of the control system designs and features of the cleared predicate device.

{6}------------------------------------------------

- 트 Components: The OmniTom Elite uses similar components as the predicate device (OmniTom Elite) such as x-ray generator, collimator, slip ring and power system.
- . Software: The OmniTom Elite, with the exclusion of PCD kit, uses the same software as the predicate (K223447). Refer to the differences listed below for software and hardware differences with PCD.

#### Differences

The following differences exist between the subject device (OmniTom Elite) and its previously cleared version, predicate device (K223447).

- . The PCD configuration now includes the following:
	- o Utilizing the current MD Plus algorithm, post reconstruction features of VNC and bone removal are added.
	- Expanding scope of PCD scanning to the full 30 cm field of view о
	- Helical scanning O
	- Operating environment ambient temperature is now increased to O 27°C/80.6°F
	- With the addition of ultra-high-resolution mode, detector coverage is the o following:
		- 16 x 0.707 mm for PCD (standard resolution) O
		- 26 X 0.424 mm for PCD (high resolution) o
		- 80 X 0.141 (ultra high resolution) o

For scanning with PCD configuration, scan voltage is 120 kV and scan current is selectable from 5 to 20 mA. Scanning of the head and neck using the PCD configuration is only applicable for adults.

#### General Safety and Effectiveness:

All components of the subject OmniTom Elite CT system that are subject to Federal Diagnostic Equipment Performance Standard and applicable regulations of 21 CFR \$1020.30 and §1020.33 are certified to meet those requirements. To minimize electrical, mechanical and radiation hazards, NeuroLogica adheres to recognized and established industry practices.

OmniTom Elite CT system is designed and manufactured to comply with the FDA Quality System Regulations (21 CFR part 820) and ISO 13485:2016 requirements. The device is in conformance with all applicable parts of the following FDA recognized consensus standards:

{7}------------------------------------------------

#### NeurOLog Co., Ltd

| FDA<br>Recognition<br>Number          | Standard                  | Description                                                                                                                                                                                          | Version |
|---------------------------------------|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| 19-4                                  | AAMI / ANSI<br>ES 60601-1 | Medical Electrical Equipment --<br>Part 1: General Requirements For<br>Basic Safety And Essential<br>Performance                                                                                     | 2012    |
| 19-8                                  | IEC 60601-1-2             | Medical electrical equipment -<br>Part 1-2: General Requirements<br>for Basic Safety and Essential<br>Performance - Collateral Standard:<br>Electromagnetic Compatibility.<br>Requirements and Tests | 2014    |
| 12-269                                | IEC 60601-1-3             | Medical Electrical Equipment -<br>Part 1-3: General Requirements<br>For Basic Safety And Essential<br>Performance - Collateral Standard:<br>Radiation Protection In Diagnostic<br>X-Ray Equipment    | 2013    |
| 5-89                                  | IEC 60601-1-6             | Medical Electrical Equipment -<br>Part 1-6: General Requirements<br>For Basic Safety And Essential<br>Performance - Collateral Standard:<br>Usability                                                | 2013    |
| 12-302                                | IEC 60601-2-44            | Medical Electrical Equipment-<br>Part 2-44: Particular Requirements<br>for Basic Safety and Essential<br>Performance of X-Ray Equipment<br>for Computed Tomography                                   | 2016    |
| 12-273                                | IEC 60825-1               | Safety Of Laser Products - Part 1:<br>Equipment Classification, And<br>Requirements                                                                                                                  | 2007    |
| 5-125                                 | ISO 14971                 | Medical Devices - Application Of<br>Risk Management To Medical<br>Devices                                                                                                                            | 2019    |
| FDA<br>Recognition<br>Number          | Standard                  | Description                                                                                                                                                                                          | Version |
| 13-79                                 | IEC 62304                 | Medical Device Software - Software Life Cycle Processes                                                                                                                                              | 2015    |
| 5-134                                 | ISO 15223-1               | Medical devices - Symbols to be used with information to be supplied by the manufacturer - Part 1: General requirements                                                                              | 2016    |
| 2-258                                 | ISO 10993-1               | Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process                                                                                           | 2018    |
| 2-245                                 | ISO 10993-5               | Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity                                                                                                                   | 2009    |
| 2-174                                 | 10993-10                  | Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization                                                                                                      | 2010    |
| 12-325                                | NEMA XR 25                | Computed Tomography Dose Check                                                                                                                                                                       | 2019    |
| NR                                    | NEMA XR 26                | Access Controls for Computer Tomography: Identification, Interlocks, and Logs                                                                                                                        | 2012    |
| FDA<br>Recognition<br>Number          | Standard                  | Description                                                                                                                                                                                          | Version |
| 12-330                                | NEMA XR 28                | Supplemental Requirements for<br>User Information and System<br>Function Related to Dose in CT                                                                                                       | 2018    |
| NR                                    | NEMA XR 29                | Standard Attributes on Computed<br>Tomography (CT) Equipment<br>Related to Dose Optimization and<br>Management                                                                                       | 2013    |
| 12-349                                | NEMA PS 3.1 -<br>3.20     | Digital Imaging and<br>Communications in Medicine<br>(DICOM) Set                                                                                                                                     | 2022    |
| 21 CFR<br>subchapter J<br>\\$ 1020.30 | FDA                       | Performance Standards for<br>Ionizing Radiation Emitting<br>Products: Diagnostic x-ray<br>systems and their major<br>components                                                                      | 2023    |
| 21 CFR<br>subchapter J<br>\\$ 1020.33 | FDA                       | Performance Standards for<br>Ionizing Radiation Emitting<br>Products: Computed tomography<br>(CT) equipment                                                                                          | 2023    |

{8}------------------------------------------------

#### NeurOLo . . Ltd

{9}------------------------------------------------

# NeurOLoc

The OmniTom Elite was designed in accordance with the following FDA Guidance documents:

- Content of Premarket Submissions for Device Software Functions, June 2023
- Cybersecurity in Medical Devices: Quality System Considerations and Content of ■ Premarket Submissions, September 2023
- Recommendations for the Use of Clinical Data in Premarket Notification [510(k)] I Submissions, September 2023
- . Informed Consent: Guidance for IRBs, Clinical Investigators, and Sponsors, August 2023

{10}------------------------------------------------

- . Electronic Submission Template for Medical Device 510(k) Submissions, October 2023
- . Medical Device Accessories – Describing Accessories and Classification Pathways, December 2017
- . Computer Software Assurance for Production and Quality System Software, September 2022
- . Guidance for Postmarket Management of Cybersecurity in Medical Devices, December 2016
- . Off-The-Shelf Software Use in Medical Devices, August 2023
- . Guidance for Deciding When to Submit a 510(k) for a Change to an Existing Device, October 2017
- . Deciding When to Submit a 510(k) for a Software Change to an Existing Device, October 2017
- . Guidance for Medical X-Ray Imaging Devices Conformance with IEC Standards, Mav 2019
- . Information Disclosure by Manufacturers to Assemblers for Diagnostic X-ray Systems, September 2003
- . Unique Device Identification System: Form and Content of the Unique Device Identifier (UDI), July 2021
- . Premarket Assessment of Pediatric Medical Devices, March 2014
- . Pediatric Information for X-ray Imaging Device Premarket Notifications, November 2017
- . The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)], July 28, 2014
- . Guidance for Industry and FDA Staff: Use of Standards in Substantial Equivalence Determinations, March 12, 2000
- Laser Products Conformance with IEC 60825-1 and IEC 60601-2-22; (Laser . Notice No. 50), June 2007
- . Radiofrequency Wireless Technology in Medical Devices, August 2013

In addition to conformance to the above harmonized standards, OmniTom Elite quality assurance activities include the following:

- . Risk analysis and mitigation
- . Software verification and validation testing
- . System verification and validation testing
- Image quality tests
- . Testing at unit level

{11}------------------------------------------------

#### Performance Data:

The risk analysis, verification and validation activities and testing of product safety and EMC / EMI was completed successfully. The differences noted raise no new issues based on all testing performed and establish that specifications for the device have been met. Below a summary has been provided for the testing conducted.

#### Bench testing

The software contained in the proposed device has been developed & tested in accordance with IEC 62304, and the FDA guidance for Content of Premarket Submissions for Device Software Functions. Software is critical to the operation of the OmniTom Elite CT system and a malfunction or design flaw in the software could result in delay in delivery of appropriate medical care. As such, the risk management analysis identified potential hazards which were controlled and mitigated during development of OmniTom Elite. The verification/validation testing ensured substantial equivalence of OmniTom Elite.

Design verification and design validation testing was performed to confirm all design and user requirements were met. The proposed OmniTom Elite device demonstrated that the new features did not exhibit any negative effects on the requirements in place, as well as they did not exhibit any concerns.

Software verification and software validation testing was executed to confirm all software requirements were met. The proposed OmniTom Elite device was shown to meet all requirements and to not have any impact on imaging.

The OmniTom Elite with PCD underwent Electrical Safety and Electromagnetic Compatibility testing and proved to be in compliance with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-44.

#### Image quality evaluation

Image quality metrics such as noise, slice thickness, low and high contrast resolution, radiation metrics, and modulation transfer function were measured utilizing phantom image quality tests in accordance with the equipment performance standards for diagnostic x-ray systems administered by the FDA. Imaging metrics successfully demonstrated that the proposed device with PCD has comparable image quality with its previous version, predicate device (K223447) and meets all the image quality criteria that are used for testing.

#### Clinical evaluation

Clinical study was conducted for evaluation of OmniTom Elite with PCD images. All study protocols were performed in accordance with the relevant regulations and the guidelines. Volunteers underwent CT scanning for a full head scan, the top of the skull, the midbrain, and through the sinuses. These images were reviewed by independent board-certified 

{12}------------------------------------------------

practicing radiologists and confirmed diagnostic quality of the images demonstrating the effectiveness of the subject device for clinical scanning.

The clinical scans indicate the ability of the PCD configuration of the OmniTom Elite device to generate diagnostically acceptable images, which are equivalent to those produced by predicate device. The subject device clinical benefits outweigh the risks and as such the device should be considered safe for its intended use and performs like other CT devices currently on the market. The clinical evaluation demonstrates performance of the OmniTom Elite system with PCD, when used under the conditions and for the purposes intended.

#### Conclusion

Both the proposed device (OmniTom Elite) with the optional PCD upgrade and the predicate device (K223447) are CT systems that are used for similar CT imaging purposes. The overall design of the CT system and basic functionality that it provides to the end user are the same. The differences in technological characteristics through the additional software features adding applications utilizing the multi-energy CT functionality for ultrahigh-resolution mode, helical scanning and full FOV do not raise additional questions and the system demonstrates substantial equivalence.

The results of the performance testing and conformance to the harmonized standards demonstrate that the subject device operates in accordance with specifications and meets user needs and intended use. The proposed OmniTom Elite CT system performs as well in its intended use as similar CT devices currently on the market.

---

**Source:** [https://fda.innolitics.com/device/K233767](https://fda.innolitics.com/device/K233767)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
