K233767 · Neurologica Corporation · JAK · Jun 10, 2024 · Radiology
Device Facts
Record ID
K233767
Device Name
OmniTom Elite
Applicant
Neurologica Corporation
Product Code
JAK · Radiology
Decision Date
Jun 10, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The OmniTom Elite computed tomography (CT) system is intended to be used for X-ray computed tomography applications for anatomy that can be imaged in the 40cm aperture, primarily the head and neck. The CT system is intended to be used for both pediatric and adult imaging and as such has preset dose settings based upon weight and age The CT images can be obtained either with or without contrast. OmniTom Elite with photon counting detectors (PCD) configuration has multi-energy CT functionality with spectral capability for material decomposition and virtual monoenergetic images (VMI). OmniTom Elite with PCD is supported for adult imaging of the head and neck.
Device Story
OmniTom Elite is a mobile, translating CT scanner featuring a 40cm bore and 30cm field of view. It utilizes a rotating gantry with a solid-state X-ray generator, collimator, and photon counting detectors (PCD). The system captures multi-energy CT data to enable material decomposition and virtual monoenergetic imaging (VMI). Operated by clinicians in clinical settings, the device provides images for diagnostic review. The system includes an MD Plus algorithm for bone removal and virtual non-contrast (VNC) processing. It supports helical scanning and ultra-high-resolution modes. Output images assist radiologists in clinical decision-making by providing detailed anatomical information. Benefits include high-resolution imaging and spectral capabilities for tissue characterization, facilitating diagnostic assessment in a portable format.
Clinical Evidence
Clinical evaluation involved volunteer CT scans of the head (skull, midbrain, sinuses). Images were reviewed by independent board-certified radiologists, who confirmed diagnostic quality equivalent to the predicate device. Bench testing included software verification/validation (IEC 62304), electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and phantom-based image quality metrics (noise, slice thickness, resolution, MTF).
Technological Characteristics
Mobile CT system with 40cm aperture; 16-row detector array. PCD configuration supports multi-energy spectral imaging. Materials/components: X-ray generator, collimator, slip-ring, brushless DC servo drive. Connectivity: DICOM. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-44, NEMA XR 25/26/28/29. Software: MD Plus algorithm for material decomposition. Sterilization: N/A (non-invasive).
Indications for Use
Indicated for X-ray CT imaging of anatomy within a 40cm aperture, primarily head and neck. Supports both pediatric and adult patients with weight/age-based dose presets. PCD configuration is indicated for adult head and neck imaging.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
June 10, 2024
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
NeuroLogica Corporation % Ninad Gujar Vice President 14 Electronics Avenue DANVERS, MA 01923
Re: K233767
Trade/Device Name: OmniTom Elite Regulation Number: 21 CFR 892.1750 Regulation Name: Computed Tomography X-Ray System Regulatory Class: Class II Product Code: JAK Dated: November 16, 2023 Received: May 6, 2024
Dear Ninad Gujar:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
approvals in the device master record (21 CFR 820.181).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
1
for
Lu Jiang Assistant Director DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K233767
Device Name OmniTom Elite
#### Indications for Use (Describe)
The OmniTom Elite computed tomography (CT) system is intended to be used for X-ray computed tomography applications for anatomy that can be imaged in the 40cm aperture, primarily the head and neck.
The CT system is intended to be used for both pediatric and adult imaging and as such has preset dose settings based upon weight and age The CT images can be obtained either with or without contrast.
OmniTom Elite with photon counting detectors (PCD) configuration has multi-energy CT functionality with spectral capability for material decomposition and virtual monoenergetic images (VMI). OmniTom Elite with PCD is supported for adult imaging of the head and neck.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "NeuroLogica" in a bold, sans-serif font. The "O" in "NeuroLogica" is replaced with a blue circle. Below the word "NeuroLogica" is the text "A Subsidiary of Samsung Electronics Co., Ltd" in a smaller font.
## K233767 510(k) Summary
Date Prepared: May 4, 2024
## Submitter
NeuroLogica Corp., a subsidiary of Samsung Electronics Co., Ltd 14 Electronics Avenue, Danvers, MA 01923
## Establishment Registration
3004938766
#### Manufacturing Site
NeuroLogica Corp., a subsidiary of Samsung Electronics Co., Ltd 14 Electronics Avenue, Danvers, MA 01923
#### Official Correspondent & Contact Person
Dr. Ninad Gujar Vice President, Regulatory Affairs & Quality Assurance Telephone: 978-564-8632 E-mail: ngujar@neurologica.com
## Device Name:
| Trade Name: | OmniTom Elite |
|------------------------|----------------------------------|
| Device Model: | NL5000 |
| Classification Name: | Computed Tomography X-ray System |
| Product Code: | JAK |
| Device Classification: | Class II (per 21 CFR § 892.1750) |
#### Predicate Device:
| 510(k): | K223447 |
|------------------------|----------------------------------|
| Trade Name: | OmniTom Elite |
| Device Model: | NL5000 |
| Classification Name: | Computed Tomography X-ray System |
| Product Code: | JAK |
| Device Classification: | Class II (per 21 CFR § 892.1750) |
{4}------------------------------------------------
#### Device Description:
The subject OmniTom Elite Computed Tomography (CT) system provides the same functionality as the previous version of the device, OmniTom Elite (K223447). Both computed tomography systems are identical in terms of the high resolution, 16 row, 40 cm bore, and 30 cm field of view. The lightweight translating gantry consists of a rotating disk with a solid-state x-ray generator, collimator, control computer, communications link, power slip-ring, data acquisition system, reconstruction computer, power system, brushless DC servo drive system (disk rotation) and an internal drive system (translation). The power system consists of batteries which provide system power while unplugged from the charging outlet. The system has the necessary safety features such as the emergency stop switch, x-ray indicators, interlocks, patient alignment laser and 110% x-ray timer. The gantry has omni-directional wheels that allow for robust diagonal, lateral, and rotational 360-degree movement and electrical drive system so the system can be moved easily to different locations.
The ability to upgrade the OmniTom Elite system with photon counting detector (PCD) remains the same. PCD provides the ability to capture CT data in multiple energy bands that can provide information on material composition of different tissues and contrast media. The multiple sets of CT data are acquired at the same time with configurable energy thresholds without any cross talk between images.
#### Indications for Use:
The OmniTom Elite computed tomography (CT) system is intended to be used for xray computed tomography applications for anatomy that can be imaged in the 40 cm aperture, primarily head and neck. The CT system is intended to be used for both pediatric and adult imaging and as such has preset dose settings based upon weight and age. The CT images can be obtained either with or without contrast.
OmniTom Elite with photon counting detectors (PCD) configuration has multi-energy CT functionality with spectral capability for material decomposition and virtual monoenergetic images (VMI). OmniTom Elite with PCD is supported for adult imaging of the head and neck.
#### Comparison of Technological Characteristics with the Predicate Device:
We modified the cleared OmniTom Elite (K223447) within our design controls to include technology improvements that include ultra-high resolution modes for PCD along with CT features based on material decomposition algorithm (MD Plus) for bone removal and virtual non contrast (VNC). Helical scanning capability was added along with increasing the field of view (FOV) and operating temperature.
{5}------------------------------------------------
The indications for use between the subject device and the predicate device (K223447) only vary slightly, in that the subject device offers an additional resolution mode for the photon counting detector. The subject device still contains the same technology and basic functionality as the predicate device (K223447), and therefore, safety and efficacy are not of a concern. Please find both the subject and predicate indications for use below.
| OmniTom Elite (Subject Device) | OmniTom Elite (Predicate – K223447) |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| The OmniTom Elite CT system is intended<br>to be used for x-ray computed tomography<br>applications for anatomy that can be imaged<br>in the 40cm aperture, primarily head and<br>neck. | The OmniTom Elite computed tomography<br>(CT) system is intended to be used for x-ray<br>computed tomography applications for<br>anatomy that can be imaged in the 40cm<br>aperture, primarily head and neck. |
| The CT system is intended to be used for<br>both pediatric and adult imaging and as such<br>has preset dose settings based upon weight<br>and age. The CT images can be obtained<br>either with or without contrast. | The CT system is intended to be used for<br>both pediatric and adult imaging and as such<br>has preset dose settings based upon weight<br>and age. The CT images can be obtained<br>either with or without contrast. |
| OmniTom Elite with photon counting<br>detectors (PCD) configuration has multi-<br>energy CT functionality with spectral<br>capability for material decomposition and<br>virtual monoenergetic images (VMI).<br>OmniTom Elite with PCD is supported for<br>adult imaging of the head and neck. | OmniTom Elite with photon counting<br>detectors (PCD) configuration has multi-<br>energy CT functionality with spectral<br>capability for material decomposition and<br>virtual monoenergetic images (VMI).<br>OmniTom Elite with PCD is only supported<br>for adult axial imaging for head and neck. |
The optional upgrade of the photon counting detector (PCD) remains the same from a hardware standpoint with the exception of additional fans for temperature regulation. OmniTom Elite with the optional PCD kit for its intended use, is of comparable type in design, material, functionality, and technology and is substantially equivalent to the cleared predicate device - OmniTom Elite (K223447).
Similarities
- 트 Design: The OmniTom Elite is identical in design characteristics to its previous version and shares all of the control system designs and features of the cleared predicate device.
{6}------------------------------------------------
- 트 Components: The OmniTom Elite uses similar components as the predicate device (OmniTom Elite) such as x-ray generator, collimator, slip ring and power system.
- . Software: The OmniTom Elite, with the exclusion of PCD kit, uses the same software as the predicate (K223447). Refer to the differences listed below for software and hardware differences with PCD.
#### Differences
The following differences exist between the subject device (OmniTom Elite) and its previously cleared version, predicate device (K223447).
- . The PCD configuration now includes the following:
- o Utilizing the current MD Plus algorithm, post reconstruction features of VNC and bone removal are added.
- Expanding scope of PCD scanning to the full 30 cm field of view о
- Helical scanning O
- Operating environment ambient temperature is now increased to O 27°C/80.6°F
- With the addition of ultra-high-resolution mode, detector coverage is the o following:
- 16 x 0.707 mm for PCD (standard resolution) O
- 26 X 0.424 mm for PCD (high resolution) o
- 80 X 0.141 (ultra high resolution) o
For scanning with PCD configuration, scan voltage is 120 kV and scan current is selectable from 5 to 20 mA. Scanning of the head and neck using the PCD configuration is only applicable for adults.
#### General Safety and Effectiveness:
All components of the subject OmniTom Elite CT system that are subject to Federal Diagnostic Equipment Performance Standard and applicable regulations of 21 CFR \$1020.30 and §1020.33 are certified to meet those requirements. To minimize electrical, mechanical and radiation hazards, NeuroLogica adheres to recognized and established industry practices.
OmniTom Elite CT system is designed and manufactured to comply with the FDA Quality System Regulations (21 CFR part 820) and ISO 13485:2016 requirements. The device is in conformance with all applicable parts of the following FDA recognized consensus standards:
{7}------------------------------------------------
#### NeurOLog Co., Ltd
| FDA<br>Recognition<br>Number | Standard | Description | Version |
|---------------------------------------|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| 19-4 | AAMI / ANSI<br>ES 60601-1 | Medical Electrical Equipment --<br>Part 1: General Requirements For<br>Basic Safety And Essential<br>Performance | 2012 |
| 19-8 | IEC 60601-1-2 | Medical electrical equipment -<br>Part 1-2: General Requirements<br>for Basic Safety and Essential<br>Performance - Collateral Standard:<br>Electromagnetic Compatibility.<br>Requirements and Tests | 2014 |
| 12-269 | IEC 60601-1-3 | Medical Electrical Equipment -<br>Part 1-3: General Requirements<br>For Basic Safety And Essential<br>Performance - Collateral Standard:<br>Radiation Protection In Diagnostic<br>X-Ray Equipment | 2013 |
| 5-89 | IEC 60601-1-6 | Medical Electrical Equipment -<br>Part 1-6: General Requirements<br>For Basic Safety And Essential<br>Performance - Collateral Standard:<br>Usability | 2013 |
| 12-302 | IEC 60601-2-44 | Medical Electrical Equipment-<br>Part 2-44: Particular Requirements<br>for Basic Safety and Essential<br>Performance of X-Ray Equipment<br>for Computed Tomography | 2016 |
| 12-273 | IEC 60825-1 | Safety Of Laser Products - Part 1:<br>Equipment Classification, And<br>Requirements | 2007 |
| 5-125 | ISO 14971 | Medical Devices - Application Of<br>Risk Management To Medical<br>Devices | 2019 |
| FDA<br>Recognition<br>Number | Standard | Description | Version |
| 13-79 | IEC 62304 | Medical Device Software - Software Life Cycle Processes | 2015 |
| 5-134 | ISO 15223-1 | Medical devices - Symbols to be used with information to be supplied by the manufacturer - Part 1: General requirements | 2016 |
| 2-258 | ISO 10993-1 | Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process | 2018 |
| 2-245 | ISO 10993-5 | Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity | 2009 |
| 2-174 | 10993-10 | Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization | 2010 |
| 12-325 | NEMA XR 25 | Computed Tomography Dose Check | 2019 |
| NR | NEMA XR 26 | Access Controls for Computer Tomography: Identification, Interlocks, and Logs | 2012 |
| FDA<br>Recognition<br>Number | Standard | Description | Version |
| 12-330 | NEMA XR 28 | Supplemental Requirements for<br>User Information and System<br>Function Related to Dose in CT | 2018 |
| NR | NEMA XR 29 | Standard Attributes on Computed<br>Tomography (CT) Equipment<br>Related to Dose Optimization and<br>Management | 2013 |
| 12-349 | NEMA PS 3.1 -<br>3.20 | Digital Imaging and<br>Communications in Medicine<br>(DICOM) Set | 2022 |
| 21 CFR<br>subchapter J<br>\\$ 1020.30 | FDA | Performance Standards for<br>Ionizing Radiation Emitting<br>Products: Diagnostic x-ray<br>systems and their major<br>components | 2023 |
| 21 CFR<br>subchapter J<br>\\$ 1020.33 | FDA | Performance Standards for<br>Ionizing Radiation Emitting<br>Products: Computed tomography<br>(CT) equipment | 2023 |
{8}------------------------------------------------
#### NeurOLo . . Ltd
{9}------------------------------------------------
# NeurOLoc
The OmniTom Elite was designed in accordance with the following FDA Guidance documents:
- Content of Premarket Submissions for Device Software Functions, June 2023
- Cybersecurity in Medical Devices: Quality System Considerations and Content of ■ Premarket Submissions, September 2023
- Recommendations for the Use of Clinical Data in Premarket Notification [510(k)] I Submissions, September 2023
- . Informed Consent: Guidance for IRBs, Clinical Investigators, and Sponsors, August 2023
{10}------------------------------------------------
- . Electronic Submission Template for Medical Device 510(k) Submissions, October 2023
- . Medical Device Accessories – Describing Accessories and Classification Pathways, December 2017
- . Computer Software Assurance for Production and Quality System Software, September 2022
- . Guidance for Postmarket Management of Cybersecurity in Medical Devices, December 2016
- . Off-The-Shelf Software Use in Medical Devices, August 2023
- . Guidance for Deciding When to Submit a 510(k) for a Change to an Existing Device, October 2017
- . Deciding When to Submit a 510(k) for a Software Change to an Existing Device, October 2017
- . Guidance for Medical X-Ray Imaging Devices Conformance with IEC Standards, Mav 2019
- . Information Disclosure by Manufacturers to Assemblers for Diagnostic X-ray Systems, September 2003
- . Unique Device Identification System: Form and Content of the Unique Device Identifier (UDI), July 2021
- . Premarket Assessment of Pediatric Medical Devices, March 2014
- . Pediatric Information for X-ray Imaging Device Premarket Notifications, November 2017
- . The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)], July 28, 2014
- . Guidance for Industry and FDA Staff: Use of Standards in Substantial Equivalence Determinations, March 12, 2000
- Laser Products Conformance with IEC 60825-1 and IEC 60601-2-22; (Laser . Notice No. 50), June 2007
- . Radiofrequency Wireless Technology in Medical Devices, August 2013
In addition to conformance to the above harmonized standards, OmniTom Elite quality assurance activities include the following:
- . Risk analysis and mitigation
- . Software verification and validation testing
- . System verification and validation testing
- Image quality tests
- . Testing at unit level
{11}------------------------------------------------
#### Performance Data:
The risk analysis, verification and validation activities and testing of product safety and EMC / EMI was completed successfully. The differences noted raise no new issues based on all testing performed and establish that specifications for the device have been met. Below a summary has been provided for the testing conducted.
#### Bench testing
The software contained in the proposed device has been developed & tested in accordance with IEC 62304, and the FDA guidance for Content of Premarket Submissions for Device Software Functions. Software is critical to the operation of the OmniTom Elite CT system and a malfunction or design flaw in the software could result in delay in delivery of appropriate medical care. As such, the risk management analysis identified potential hazards which were controlled and mitigated during development of OmniTom Elite. The verification/validation testing ensured substantial equivalence of OmniTom Elite.
Design verification and design validation testing was performed to confirm all design and user requirements were met. The proposed OmniTom Elite device demonstrated that the new features did not exhibit any negative effects on the requirements in place, as well as they did not exhibit any concerns.
Software verification and software validation testing was executed to confirm all software requirements were met. The proposed OmniTom Elite device was shown to meet all requirements and to not have any impact on imaging.
The OmniTom Elite with PCD underwent Electrical Safety and Electromagnetic Compatibility testing and proved to be in compliance with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-44.
#### Image quality evaluation
Image quality metrics such as noise, slice thickness, low and high contrast resolution, radiation metrics, and modulation transfer function were measured utilizing phantom image quality tests in accordance with the equipment performance standards for diagnostic x-ray systems administered by the FDA. Imaging metrics successfully demonstrated that the proposed device with PCD has comparable image quality with its previous version, predicate device (K223447) and meets all the image quality criteria that are used for testing.
#### Clinical evaluation
Clinical study was conducted for evaluation of OmniTom Elite with PCD images. All study protocols were performed in accordance with the relevant regulations and the guidelines. Volunteers underwent CT scanning for a full head scan, the top of the skull, the midbrain, and through the sinuses. These images were reviewed by independent board-certified
{12}------------------------------------------------
practicing radiologists and confirmed diagnostic quality of the images demonstrating the effectiveness of the subject device for clinical scanning.
The clinical scans indicate the ability of the PCD configuration of the OmniTom Elite device to generate diagnostically acceptable images, which are equivalent to those produced by predicate device. The subject device clinical benefits outweigh the risks and as such the device should be considered safe for its intended use and performs like other CT devices currently on the market. The clinical evaluation demonstrates performance of the OmniTom Elite system with PCD, when used under the conditions and for the purposes intended.
#### Conclusion
Both the proposed device (OmniTom Elite) with the optional PCD upgrade and the predicate device (K223447) are CT systems that are used for similar CT imaging purposes. The overall design of the CT system and basic functionality that it provides to the end user are the same. The differences in technological characteristics through the additional software features adding applications utilizing the multi-energy CT functionality for ultrahigh-resolution mode, helical scanning and full FOV do not raise additional questions and the system demonstrates substantial equivalence.
The results of the performance testing and conformance to the harmonized standards demonstrate that the subject device operates in accordance with specifications and meets user needs and intended use. The proposed OmniTom Elite CT system performs as well in its intended use as similar CT devices currently on the market.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.