K233732 · Zeltiq Aesthetics, Inc. (Acquired by Allergan Aesthetics · OOK · Sep 12, 2024 · General, Plastic Surgery
Device Facts
Record ID
K233732
Device Name
CoolSculpting Elite System
Applicant
Zeltiq Aesthetics, Inc. (Acquired by Allergan Aesthetics
Product Code
OOK · General, Plastic Surgery
Decision Date
Sep 12, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4340
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CoolSculpting Elite System is a skin cooling or heating device. It can be used in cooling or heating mode. Cooling Mode • Indicated for cold-assisted lipolysis (breakdown of fat) of the upper arm, bra fat, banana roll, thigh, abdomen, and flank (or "love handles") in individuals with a Body Mass Index (BMI) of 30 or less. • Intended for cold-assisted lipolysis of the submental and submandibular areas in individuals with a BMI up to 46.2. · Intended to affect the appearance of visible fat bulges in the upper arm, bra fat, banana roll, submental and submandibular areas, thigh, abdomen, and flank. · When used for cold-assisted lipolysis of the submental area, the device can also affect the appearance of lax tissue in the submental area. - · Can be used to minimize pain and thermal injury during laser and dermatological treatments. - · Can be used as a local anesthetic for procedures that induce minor local discomfort. Heating or Cooling Mode - · Can be used to minimize pain post-trauma and post-surgery. - · Can be used to provide temporary relief of minor aches, pains, and muscle spasms. • The ZELTIQ Pretreatment Skin Wipe and gel pad facilitate thermal contact of the device with a patient's skin by mitigating minor variances in device-to-skin contact.
Device Story
Portable thermoelectric cooling/heating device; applies controlled thermal energy to treatment sites. Comprises control unit, detachable applicators (including new S180), and accessories (gel pads, wipes). Operates by cooling/heating skin to target temperatures (down to -11°C); automatically stops if interface temperature deviates >1°C from target when <5°C. Used in clinical settings by trained operators. Output includes thermal treatment of adipose tissue or pain relief. Software update adds Wi-Fi for cloud-based data transfer (logs, diagnostics, patches); Wi-Fi does not affect therapeutic function. Healthcare providers use device to manage fat bulges and provide localized pain relief. Benefits include non-invasive fat reduction and symptomatic relief of minor pain/spasms.
Clinical Evidence
Bench testing only. Non-clinical testing confirmed the modified system functions as intended with the software update and S180 applicator, demonstrating equivalence to the predicate device.
Technological Characteristics
Thermoelectric cooling/heating system; portable control unit with detachable applicators. Target temperature down to -11°C (±0.5°C accuracy). Includes Wi-Fi connectivity for cloud-based diagnostics and log management. Accessories include gel pads and pretreatment wipes to facilitate thermal contact. Software-controlled operation with automated safety shut-off.
Indications for Use
Indicated for cold-assisted lipolysis of upper arm, bra fat, back fat, banana roll, thigh, abdomen, and flank in individuals with BMI ≤ 30; submental and submandibular areas in individuals with BMI ≤ 46.2. Also indicated to affect appearance of visible fat bulges and submental lax tissue; minimize pain/thermal injury during laser/dermatological treatments; act as local anesthetic for minor discomfort; minimize post-trauma/post-surgery pain; and provide temporary relief of minor aches, pains, and muscle spasms.
Regulatory Classification
Identification
A contact cooling system for aesthetic use is a device that is a combination of a cooling pad associated with a vacuum or mechanical massager intended for the disruption of adipocyte cells intended for non-invasive aesthetic use.
Special Controls
*Classification.* Class II (special controls). The special controls for this device is FDA's “Guidance for Industry and FDA Staff; Class II Special Controls Guidance Document: Contact Cooling System for Aesthetic Use.” See § 878.1(e) for the availability of this guidance document.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.
September 12, 2024
ZELTIQ Aesthetics, Inc.(acquired by Allergan Aesthetics and now, an AbbVie, Inc. company) Victoria Montemayor Manager, Regulatory Affairs 4410 Rosewood Drive Pleasanton, California 94588
Re: K233732
Trade/Device Name: CoolSculpting Elite System Regulation Number: 21 CFR 878.4340 Regulation Name: Contact Cooling System For Aesthetic Use Regulatory Class: Class II Product Code: OOK Dated: August 16, 2024 Received: August 16, 2024
Dear Victoria Montemayor:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{2}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{3}------------------------------------------------
Sincerely,
Digitally signed by Long H. Chen Long H. Chen-S Date: 2024.09.12 15:18:34 -04'00'
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
Submission Number (if known)
K233732
Device Name
CoolSculpting Elite System (N/A)
Indications for Use (Describe)
The CoolSculpting Elite System is a skin cooling or heating device. It can be used in cooling or heating mode.
Cooling Mode
• Indicated for cold-assisted lipolysis (breakdown of fat) of the upper arm, bra fat, banana roll, thigh, abdomen, and flank (or "love handles") in individuals with a Body Mass Index (BMI) of 30 or less.
• Intended for cold-assisted lipolysis of the submental and submandibular areas in individuals with a BMI up to 46.2.
· Intended to affect the appearance of visible fat bulges in the upper arm, bra fat, banana roll, submental and submandibular areas, thigh, abdomen, and flank.
· When used for cold-assisted lipolysis of the submental area, the device can also affect the appearance of lax tissue in the submental area.
- · Can be used to minimize pain and thermal injury during laser and dermatological treatments.
- · Can be used as a local anesthetic for procedures that induce minor local discomfort.
Heating or Cooling Mode
- · Can be used to minimize pain post-trauma and post-surgery.
- · Can be used to provide temporary relief of minor aches, pains, and muscle spasms.
• The ZELTIQ Pretreatment Skin Wipe and gel pad facilitate thermal contact of the device with a
- patient's skin by mitigating minor variances in device-to-skin contact.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
## K233732
Image /page/4/Picture/1 description: The image shows the Abbvie logo on the left side. To the right of the logo is the text "CoolSculpting Elite System" followed by "Software Update 2 and S180 Applicator - Traditional 510(k) Application". The text appears to be describing the CoolSculpting Elite System software update.
# 510(K) SUMMARY
| I. SUBMITTER | | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| 510(k) Submitter: | ZELTIQ Aesthetics, Inc.<br>4410 Rosewood Drive<br>Pleasanton, CA 94588 USA | |
| Contact Person(s): | Primary Contact | Secondary Contact |
| | Victoria Montemayor<br>Manager, Regulatory Affairs<br>Phone: (949) 761-0747<br>Email: victoria.montemayor@abbvie.com | Bryan McMahon<br>Director, Global Regulatory Affairs<br>Phone: (720) 425-5661<br>Email: bryan.mcmahon@abbvie.com |
| Date Prepared: | April 30, 2024 | |
| II. DEVICE | | |
| Device Trade Name: | CoolSculpting Elite System | |
| Medical Specialty | General and Plastic Surgery | |
| Regulation | 21 CFR 878.4340<br>Class II Special Control, Contact Cooling System for Aesthetic Use | |
| Regulatory Class | Class II | |
| Product Code | OOK<br>Dermal Cooling Pack/Vacuum/Massager | |
| III. PREDICATE DEVICE | | |
| Predicate Device | CoolSculpting Elite System<br>K212707 (cleared on November 5, 2021) | |
| IV. DEVICE DESCRIPTION | | |
| The CoolSculpting Elite System is a portable thermoelectric cooling device that applies controlled cooling to a<br>treatment site. The CoolSculpting Elite System comprises of a control unit, detachable applicators, and<br>accessories such as cycle cards, CoolAdhesive gelpads, gel traps, pretreatment skin wipes, liners, foam borders,<br>and comfort straps. The device treats a target temperature down to -11°C with an accuracy of ±0.5°C. The device<br>will automatically stop the treatment if the interface temperature goes past the target temperature by more than<br>1°C when treating below 5°C | | |
{6}------------------------------------------------
### K233732
Image /page/5/Picture/1 description: The image contains the Abbvie logo and information about the CoolSculpting Elite System. The Abbvie logo is on the left side of the image. To the right of the logo, the text reads "CoolSculpting Elite System" and "Software Update 2 and S180 Applicator – Traditional 510(k) Application."
## 510(K) SUMMARY
#### V. Indications for Use
The CoolSculpting Elite System is a skin cooling or heating device. It can be used in cooling or heating mode. A list of indications for use is provided below.
| Mode | Indications |
|--------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Cooling | Indicated for cold-assisted lipolysis (breakdown of fat) of the upper arm, bra fat, back fat, banana roll, thigh, abdomen, and flank (or "love handles") in individuals with a Body Mass Index (BMI) of 30 or less. |
| | Intended for cold-assisted lipolysis of the submental and submandibular areas in individuals with a BMI up to 46.2. |
| | Intended to affect the appearance of visible fat bulges in the upper arm, bra fat, back fat, banana roll, submental and submandibular areas, thigh, abdomen, and flank. |
| | When used for cold-assisted lipolysis of the submental area, the device can also affect the appearance of lax tissue in the submental area. |
| | Can be used to minimize pain and thermal injury during laser and dermatological treatments. |
| | Can be used as a local anesthetic for procedures that induce minor local discomfort. |
| Heating or cooling | Can be used to minimize pain-trauma and post-surgery. |
| | Can be used to provide temporary relief of minor aches, pains, and muscle spasms. |
| | The ZELTIQ Pretreatment Skin Wipe and gel pad facilitate thermal contact of the device with a patient's skin by mitigating minor variances in device-to-skin contact. |
{7}------------------------------------------------
### K233732
Image /page/6/Picture/1 description: The image contains the logo for Abbvie, along with the text "CoolSculpting Elite System". The text "Software Update 2 and S180 Applicator - Traditional 510(k) Application" is also present in the image. The image appears to be related to a medical device or software update for the CoolSculpting Elite System, developed by Abbvie. The text also mentions a 510(k) application, which is a premarket submission made to the FDA to demonstrate that the device is safe and effective.
## 510(K) Summary
#### VI. Comparison to Predicate Device
The purpose of this Traditional 510(k) submission is to introduce a software update, which notably includes Wi-Fi functionality, and new surface applicator, S180. The modified CoolSculpting Elite System is identical to the predicate device in terms of principle of operation, mechanical features, performance specifications, software, hardware, algorithm, and treatment workflow. There are minor differences to the indications for use and user interface.
The addition of Wi-Fi to the subject device is intended to allow data transfer between the device and cloud; the data consists of logs for usage monitoring, errors trending, engineering diagnostics, and debugging as well as patch deployment and remains unchanged; the Wi-Fi functionality is not used to achieve the device's intended use. The addition of the S180 applicator is intended to expand options when selecting an appropriate treatment plan for patients; the treatment sites targeted by the surface applicator and overall design remain unchanged.
The differences between the modified CoolSculpting Elite System and predicate device do not raise new questions of safety and effectiveness, and these devices are substantially equivalent based on the criteria described in 21 CFR §807.100.
#### V. Performance Testing
Non-clinical testing was successfully completed to demonstrate that the CoolSculpting Elite System, with the software update and new S180 applicator, is equivalent to the predicate device. Testing results confirmed that the modified CoolSculpting Elite System functions as intended, and no new issues of safety and effectiveness were identified.
#### VI. SOFTWARE AND CYBERSECURITY
An evaluation of the CoolSculpting Elite System's software and firmware) documentation was performed for cybersecurity considerations. The evaluation results demonstrated that the CoolSculpting Elite System complies with the requirements set forth in Section 524B, Ensuring Cybersecurity of Medical Devices, of the FD&C Act.
#### VII. Conclusion
The information presented in this Traditional Premarket Notification demonstrates that the modified CoolSculpting Elite System is substantially equivalent to the predicate device and will perform as safely and effectively within the intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.