Ringer perfusion balloon catheter, 2.00 x 20mm (5881); Ringer perfusion balloon catheter, 2.50 x 20mm (5882); Ringer perfusion balloon catheter, 2.50 x 30mm (5883); Ringer perfusion balloon catheter, 3.00 x 20mm (5884); Ringer perfusion balloon catheter, 3.00 x 30mm (5885); Ringer perfusion balloon catheter, 3.50 x 20mm (5886); Ringer perfusion balloon catheter, 3.50 x 30mm (5887); Ringer perfusion balloon catheter, 4.00 x 20mm (5888); Ringer perfusion balloon catheter, 4.00 x 30mm (58
Applicant
Vascular Solutions, LLC
Product Code
LOX · Cardiovascular
Decision Date
May 31, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5100
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Ringer Perfusion Balloon Catheter is indicated for balloon dilatation of coronary artery or coronary bypass graft stenoses where the physician desires distal blood perfusion during balloon inflation for the purpose of improving myocardial perfusion.
Device Story
Rapid exchange PTCA catheter with spiral-shaped inflatable balloon; distal wire lumen for ≤ 0.014" guidewire. Used in clinical settings by physicians for coronary artery or bypass graft stenosis dilatation. Spiral balloon design allows distal blood perfusion during inflation, maintaining TIMI 2/3 flow to distal coronary bed; enables prolonged balloon inflation (≥ 1 minute). Radiopaque markers and shaft positioning marks facilitate placement. Output is mechanical dilatation of stenotic vessels. Benefits include improved myocardial perfusion during procedure and ability to sustain inflation without ischemia.
Clinical Evidence
Prospective clinical study (G200321, 'Ringer PTCA Study') of 60 subjects. Primary efficacy (procedural success, no perforation/dissection, TIMI 3 flow) met in 95% (57/60). Primary safety (freedom from MACE at discharge) met in 95% (57/60). Secondary efficacy (successful PCI) met in 93% (56/60). Secondary safety (tolerance of ≥1 min inflation) met in 83.3% (50/60). 100% of cases maintained TIMI II/III flow during inflation.
Technological Characteristics
Rapid exchange catheter; spiral-shaped inflatable balloon; ≤ 0.014" guidewire compatible. Features radiopaque markers and shaft positioning marks. Sterilized via ethylene oxide. Biocompatibility per ISO 10993-1. Mechanical operation; no electronic or software components.
Indications for Use
Indicated for balloon dilatation of coronary artery or coronary bypass graft stenoses in patients requiring distal blood perfusion during inflation to improve myocardial perfusion.
Regulatory Classification
Identification
Standard PTCA Catheter: A PTCA catheter is a device that operates on the principle of hydraulic pressurization applied through an inflatable balloon attached to the distal end. A PTCA balloon catheter has a single or double lumen shaft. The catheter features a balloon of appropriate compliance for the clinical application, constructed from a polymer. The balloon is designed to uniformly expand to a specified diameter and length at a specific pressure as labeled, with well characterized rates of inflation and deflation and a defined burst pressure. The device generally features a type of radiographic marker to facilitate fluoroscopic visualization of the balloon during use. A PTCA catheter is intended for balloon dilatation of a hemodynamically significant coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion. A PTCA catheter may also be intended for the treatment of acute myocardial infarction; treatment of in-stent restenosis (ISR) and/or post-deployment stent expansion. Cutting/scoring PTCA Catheter: A cutting/scoring PTCA catheter is a balloon-tipped catheter with cutting/scoring elements attached, which is used in those circumstances where a high pressure balloon resistant lesion is encountered. A cutting/scoring PTCA catheter is intended for the treatment of hemodynamically significant coronary artery stenosis for the purpose of improving myocardial perfusion. A cutting/scoring PTCA catheter may also be indicated for use in complex type C lesions or for the treatment of in-stent restenosis.
Special Controls
*Classification.* Class II (special controls). The special control for this device is “Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters.” See § 870.1(e) for the availability of this guidance document.(b)
*Cutting/scoring PTCA Catheter* —(1)*Identification.* A cutting/scoring PTCA catheter is a balloon-tipped catheter with cutting/scoring elements attached, which is used in those circumstances where a high pressure balloon resistant lesion is encountered. A cutting/scoring PTCA catheter is intended for the treatment of hemodynamically significant coronary artery stenosis for the purpose of improving myocardial perfusion. A cutting/scoring PTCA catheter may also be indicated for use in complex type C lesions or for the treatment of in-stent restenosis.(2)
*Classification.* Class III (premarket approval). As of May 28, 1976, an approval under section 515 of the act is required before this device may be commercially distributed. See § 870.3.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text on the right. The symbol on the left is a stylized representation of a human figure. The text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters. The words "U.S. FOOD & DRUG" are on the top line, and the word "ADMINISTRATION" is on the bottom line.
May 31, 2024
Vascular Solutions LLC Becky Astrup Pr. Regulatory Product Specialist 6464 Sycamore Court North Minneapolis, Minnesota 55369
#### Re: K233729
Dear Becky Astrup:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{2}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product: and 21 CFR 820.100. Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{3}------------------------------------------------
Sincerely,
# Brian D. Pullin -S
for Lydia Glaw Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### Indications for Use
510(k) Number (if known) K233729
Device Name Ringer perfusion balloon catheter
Indications for Use (Describe)
The Ringer Perfusion Balloon Catheter is indicated for balloon dilatation of coronary bypass graft stenoses where the physician desires distal blood perfusion during balloon inflation for the purpose of improving myocardial perfusion.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| <span style="font-size:10pt">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:10pt">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image is a logo for Vascular Solutions. The logo features a tan square with a design of black squares that decrease in size as they move upward, creating a sense of movement or flow. Below the square, the words "Vascular" and "Solutions" are written in a serif font, with "Vascular" in a larger, bolder font than "Solutions."
## 510(k) Summary
[As required by 21 CFR 807.92]
Date Prepared: May 1, 2024
510(k) Number: K233729
#### Submitter's Name / Contact Person
Manufacturer Vascular Solutions LLC 6464 Sycamore Court North Minneapolis, MN 55369 USA Establishment Registration # 2134812
Contact Person Becky Astrup Pr. Regulatory Product Specialist Tel: 763-762-2601
#### General Information Trade Name Ringer perfusion balloon catheter Common / Usual Name Catheter Classification Name 21 CFR 870.5100, LOX, Percutaneous Transluminal Coronary Angioplasty Catheters, Class II Predicate Device K153682, Chocolate XD PTCA Balloon Catheter (Vascular Solutions) The Chocolate XD PTCA balloon catheter has not been subject to a design related recall.
#### Device Description
The Ringer perfusion balloon catheter is a rapid exchange catheter with a spiral-shaped inflatable balloon on the distal end and a wire lumen for delivery over a ≤ 0.014" guidewire. The inflated spiral balloon approximates a hollow cylinder. Radiopaque markers at the proximal and distal end of the balloon facilitate visualization and intravascular placement of the catheter prior to inflation. The catheter shaft has positioning marks located at 95cm (single mark) and 105cm (double mark) from the distal tip. The Ringer perfusion balloon catheter has been sterilized with ethylene oxide.
{6}------------------------------------------------
#### Indications for Use
The Ringer Perfusion Balloon Catheter is indicated for balloon dilatation of coronary artery or coronary bypass graft stenoses where the physician desires distal blood perfusion during balloon inflation for the purpose of improving myocardial perfusion.
The Indications for Use statement for the Ringer Perfusion Balloon Catheter is not identical to the predicate device; however, the differences do not alter the intended therapeutic use of the device, nor do they affect the safety and effectiveness of the device relative to the predicate. Both the subject and predicate devices have the same intended use for the balloon dilatation of a hemodynamically significant coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion.
#### Comparison of Technological Characteristics with the Predicate Device
The subject device and predicate device have the same intended use and the indications for use for the subject device fall within the intended use of the predicate device. Differences between the subject device indications for use and predicate device indications for use are supported by benchtop, animal, and clinical tests.
The Ringer perfusion balloon catheter and the predicate device have the same technological characteristics with respect to:
- . Conditions of Use
- . General Principles of Operation
- . Cutting/Scoring Features (none)
- Materials ●
- Catheter Shaft Design and Configuration
- Guidewire and Guide Catheter Compatibility ●
- . Balloon Length
The following technological differences exist between the subject and predicate device:
- . Ancillary Device Delivery
- Balloon Design .
- Distal Perfusion Mechanism ●
- Crossing Profile ●
- Balloon Inflation Time
- Radiopaque Markers ●
- Catheter Length ●
- Packaging
- Sterilization ●
- Shelf Life ●
- Balloon Diameter
- Nominal Pressure
- Rated Burst Pressure ●
- Balloon Deflation Time ●
Results of benchtop, animal, and clinical tests demonstrate that the technological characteristics of the subject device do not raise different questions of safety and effectiveness and that the subject device is substantially equivalent to the predicate device.
#### Performance Data
The following performance data were provided in support of substantial equivalence.
#### Biocompatibility Tests:
The biocompatibility evaluation for the Ringer perfusion balloon catheter device was conducted in accordance with the FDA Guidance Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process, September 2023
{7}------------------------------------------------
and International Standard ISO 10993-1 Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process, as recognized by FDA. The battery of testing included the following tests:
- . Cvtotoxicitv
- Sensitization ●
- Irritation
- Acute Systemic Toxicity ●
- Materia Mediated Pyrogenicity
- Hemolysis ●
- Complement Activation ●
- Thrombogenicity ●
The Ringer perfusion balloon catheter is considered an external communicating device contacting circulating blood for a limited duration (<24 hours).
#### Bench Tests:
- . Dimensional and Visual Verification
- . Balloon Preparation. Deployment and Retraction
- Balloon Rated Burst Pressure .
- Balloon Fatigue ●
- Balloon Compliance ●
- . Balloon Inflation and Deflation Time
- Catheter Bond Strength .
- . Tip Pull Test
- Flexibility and Kink .
- Torque Strength ●
- Radiopacity ●
- ISO 80369 Hub Verification .
#### Animal Tests:
The safety of the Ringer perfusion balloon catheter was evaluated during simulated clinical use in a swine model. The testing was performed in accordance with 21 CFR Part 58 for Good Laboratory Practice (GLP) for Non-Clinical Laboratory Studies.
The primary objective of demonstrating safety of the Ringer perfusion balloon catheter was exhibited in this study in terms of the measured adverse events, procedural data, gross necropsy, histopathology, and clinical observations. Normal, TIMI 3 perfusion was maintained to all vasculature distal to the test article for the duration of inflation. Thrombogenicity scores demonstrated no to minimal thrombus formation. Additionally, no significant visual changes observed from baseline to terminal time points through angiographic evaluation.
#### Clinical Tests:
Clinical data was collected for the Ringer perfusion balloon catheter to support substantial equivalence (G200321, 'Ringer PTCA Study'). The objective of the Ringer PTCA study was to demonstrate performance of the Ringer Perfusion Balloon Catheter for dilatation of coronary stenoses during percutaneous coronary intervention (PCI). The secondary objective was to demonstrate maintenance of coronary blood flow distal to the inflated Ringer device thereby enabling prolonged balloon inflation (balloon inflation equal to or greater than one minute).
#### Primary Effectiveness Endpoint
The primary efficacy endpoint was device procedural success consisting of successful delivery, inflation, deflation, and removal of the study device: no vessel perforation. flow-limiting vessel dissection, no reduction in TIMI flow from baseline, and no clinically significant arrhythmias requiring medical treatment or device intervention following dilatation with the study device; achievement of final TIMI
{8}------------------------------------------------
flow grade of 3 at the conclusion of the PCI procedure for the target lesion. The primary efficacy endpoint was met in 57/60 (95%) of the subjects.
#### Primary Safety Endpoint
The study primary safety endpoint of occurrence of clinically relevant events during prolonged balloon inflations and freedom from MACE at hospital discharge was met in 95% (57/60) of the subjects.
#### Secondary Efficacy Endpoint
The secondary efficacy endpoint for the study was successful PCI defined as final residual stenosis ≤20% diameter stenosis (in-stent) or ≤ 50% diameter stenosis (stand-alone balloon) by visual angiographic estimation and freedom from MACE at hospital discharge; maintenance of TIMI-2 or -3 flow into distal coronary bed during study device inflation. The secondary efficacy endpoint of successful PCI was met in 93% (56/60) of the subjects.
#### Secondary Efficacy Endpoint
The secondary safety endpoint evaluated tolerance of at least one study device inflation equal to or greater than one minute (prolonged inflation) at target site. The secondary safety endpoint of tolerance of at least one Ringer perfusion balloon catheter prolonged inflation at target site was met in 50/60 (83.3%) of subjects.
#### Summary
Procedural success was achieved in 95% of cases with no serious adverse events attributable to the subject device. Maintenance of TIMI II or III flow into distal coronary bed during Ringer perfusion balloon inflation was achieved in 100% of cases.
This clinical data demonstrates that the subject device is appropriate for use in PTCA procedures. The clinical data raised no new questions of safety and effectiveness compared to the predicate device, supporting that the subject device is substantially equivalent to the predicate device.
#### Conclusions
The subject and predicate devices share the same intended use and fundamental scientific technology, including principle of operation. Differences in technological characteristics between the subject and predicate devices do not raise different questions of safety and effectiveness as demonstrated by bench. animal, and clinical tests. Non-clinical and clinical performance and safety evaluations support substantial equivalence of the Ringer perfusion balloon catheter to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.