K233728 · Ge Healthcare (Tianjin) Company Limited · LNH · Jan 19, 2024 · Radiology
Device Facts
Record ID
K233728
Device Name
SIGNA Champion
Applicant
Ge Healthcare (Tianjin) Company Limited
Product Code
LNH · Radiology
Decision Date
Jan 19, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Attributes
AI/ML
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Magnetic Resonance Imaging
—
—
—
—
—
—
—
Indications for Use
The SIGNA™ Champion is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used. The images produced by SIGNA™ Champion reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
Device Story
SIGNA™ Champion is a 1.5T superconducting whole-body MR scanner; utilizes time-varying magnetic gradients and RF transmissions to generate diagnostic images (axial, sagittal, coronal, oblique, spectroscopic, parametric maps). Operated by trained clinicians in clinical settings. System features 70cm bore; integrates AIR Recon DL and Sonic DL software for image reconstruction/enhancement. Data input via RF receive chain (Digitize-Per-Pin architecture); output interpreted by physicians to assist in diagnosis. Supports wireless gating for respiratory/cardiac/peripheral applications. Benefits include high resolution, high signal-to-noise ratio, and reduced scan times.
Clinical Evidence
No clinical studies required. Substantial equivalence supported by bench testing and sample clinical images demonstrating diagnostic image performance equivalent to the predicate device per FDA guidance.
Technological Characteristics
1.5T superconducting magnet with active shielding; 70cm bore; water-cooled gradient coil; Digitize-Per-Pin (DPP) RF receive architecture; integrated body coil; wireless gating option. Complies with ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-2-33, IEC 62304, ISO 10993-1, NEMA MS standards, and NEMA PS3 (DICOM).
Indications for Use
Indicated for use as a diagnostic imaging device for whole-body MR scanning (head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, musculoskeletal) in patients requiring axial, sagittal, coronal, oblique, or spectroscopic imaging, parametric maps, or dynamic imaging. Contrast agents may be used.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 19, 2024
GE Healthcare (Tianjin) Company Limited % Glen Sabin Director - Regulatory Affairs, MR Strategy GE Medical Systems, LLC 3200 N Grandview Blvd. Waukesha, Wisconsin 53188
### Re: K233728
Trade/Device Name: SIGNA™ Champion Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: LNH, LNI Dated: November 21, 2023 Received: November 21, 2023
Dear Glen Sabin:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Daniel M. Krainak, Ph.D. Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
K233728
Device Name
SIGNA™ Champion
#### Indications for Use (Describe)
The SIGNA™ Champion is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMI, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body.
Depending on the region of interest being imaged, contrast agents may be used. The images produced by SIGNA™ Champion reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the GE HealthCare logo. The logo consists of a purple circle with a white GE monogram inside, followed by the words "GE HealthCare" in purple. The font is sans-serif and the overall design is clean and modern.
## 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
| Date | January 19, 2024 | | |
|-----------------------------|---------------------------------------------------------------------------------------------------------------------------------|--|--|
| Submitter | GE Healthcare (Tianjin) Company Limited<br>No. 266 Jingsan Road, Tianjin Airport Economic Area<br>Tianjin, P.R. China 300308 | | |
| Primary<br>Contact Person | Xinyu Song<br>Lead Specialist, Regulatory Affairs, MR<br>GE HealthCare<br>Phone: +86 186 1188 4503<br>E-mail: Xinyu.Song@ge.com | | |
| Secondary<br>Contact Person | Glen Sabin<br>Director - Regulatory Affairs, MR Strategy<br>GE HealthCare<br>Phone: 262 894-4968<br>E-mail: glen.sabin@ge.com | | |
| Device Trade<br>Name | SIGNA™ Champion | | |
| Common/Usual<br>Name | Magnetic Resonance Diagnostic Device | | |
| Classification<br>Names | Magnetic Resonance Diagnostic Device per 21 CFR 892.1000 | | |
| Product Code | LNH, LNI | | |
| Predicate<br>Device | SIGNA™ Voyager (K192426) | | |
| Reference<br>Device | (1) SIGNA™ Victor (K223439)<br>(2) SIGNA™ Hero (K213668) | | |
## Reason for Submission:
This 510(k) is being submitted due to the introduction of SIGNA™ Champion, a new 1.5T MR system from GE HealthCare.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the GE HealthCare logo. The logo consists of a circular emblem on the left and the text "GE HealthCare" on the right. The emblem features a stylized "GE" monogram inside a circle with decorative flourishes. The text "GE HealthCare" is written in a clean, sans-serif font and is colored in the same purple as the emblem.
## Device Description:
SIGNA™ Champion is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan time. The system uses a combination of time-varying magnet fields (Gradients) and RF transmissions to obtain information regarding the density and position of elements exhibiting magnetic resonance. The system can image in the sagittal, coronal, axial, oblique, and double oblique planes, using various pulse sequences, imaging techniques and reconstruction algorithms. The system features a 1.5T superconducting magnet with 70cm bore size. The system is designed to conform to NEMA DICOM standards (Digital Imaging and Communications in Medicine).
A high-level summary of significant hardware and software modifications is provided below:
Hardware changes (compared with Predicate Device SIGNA™ Voyager):
- Magnet Enclosure -
- -Patient Table and related support components
- Host Computer -
- RF Transmit Chain -
- Wireless Gating -
Software Changes (compared with Reference Device 2 SIGNA™ Hero)
- Sonic DL -
- AIR Recon DL -
#### Indications for Use
The SIGNA™ Champion is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body.
Depending on the region of interest being imaged, contrast agents may be used. The images produced by SIGNA™ Champion reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance.
These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the GE HealthCare logo. On the left is the GE monogram in purple, which is a stylized version of the letters G and E intertwined. To the right of the monogram is the text "GE HealthCare" in purple. The text is in a simple, sans-serif font.
### Technology
The SIGNA™ Champion employs the same fundamental scientific technology as its predicate device.
SIGNA™ Champion is built with superconducting magnet, RF transmit architecture, RF receive chain and software application suite.
### Comparison of Indications for Use
The changes in technology do not impact the indications for use.
The indications for use have not been changed, other than to reflect the SIGNA™ Champion product name.
Therefore, the intended use is the same as the predicate device in accordance with the FDA's guidance document "The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)]", dated 28 July 2014.
### Comparison of Technological Characteristics
Overall, the SIGNA™ Champion employs the same fundamental scientific technology and operating principles as the predicate device.
There are some differences in characteristics between the proposed device and the predicate device, as summarized below:
| Subsystem or<br>Component | Predicate Device<br>SIGNA™ Voyager<br>(K192426) | Proposed Device<br>SIGNA™ Champion<br>(K233728) | Comments |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Magnet | 1.5T Superconducting Magnet with active<br>shielding | 1.5T Superconducting Magnet with active<br>shielding | SIGNA™ Champion uses the<br>same magnet as the<br>predicate SIGNA™ Voyager,<br>with a modified enclosure<br>design. |
| Gradient<br>Subsystem | Water cooled gradient coil with active<br>shielding. | Water cooled gradient coil with active<br>shielding. | SIGNA™ Champion uses the<br>same gradient subsystem as<br>the predicate SIGNA™<br>Voyager. |
| Subsystem or<br>Component | Predicate Device<br>SIGNA™ Voyager<br>(K192426) | Proposed Device<br>SIGNA™ Champion<br>(K233728) | Comments |
| RF Transmit<br>Subsystem | Transmit with integrated body coil and local<br>T/R coil. | Transmit with integrated body coil and local<br>T/R coil. | SIGNA™ Champion uses the<br>same integrated body coil as<br>the predicate SIGNA™<br>Voyager, but other<br>components in the transmit<br>chain are different. The<br>SIGNA™ Champion uses the<br>In Scan Room transmit<br>architecture found in<br>Reference Device 1 (SIGNA™<br>Victor). |
| RF Receive<br>Subsystem | Digitize-Per-Pin (DPP) receive chain<br>architecture. | Digitize-Per-Pin (DPP) receive chain<br>architecture. | SIGNA™ Champion uses the<br>same DPP architecture as<br>the predicate SIGNA™<br>Voyager. |
| RF Coils | Comprehensive suite of detachable coils for<br>imaging all anatomies | Comprehensive suite of detachable coils for<br>imaging all anatomies | Coils used by SIGNA™<br>Champion are also used by<br>the predicate SIGNA™<br>Voyager and/or Reference<br>Device 1 (SIGNA™ Victor). |
| Software<br>Features | Comprehensive suite of software features,<br>pulse sequences, and image processing<br>applications to support MR imaging of all<br>anatomies. | Comprehensive suite of software features,<br>pulse sequences, and image processing<br>applications to support MR imaging of all<br>anatomies. | Both SIGNA™ Champion and<br>the predicate SIGNA™<br>Voyager are fully capable MR<br>systems with a wide range of<br>software features. SIGNA™<br>Champion includes some<br>new and enhanced features<br>such as AIR Recon DL and<br>Sonic DL that were not<br>included in the predicate<br>K192426 submission. The<br>SIGNA™ Champion software<br>features are similar to those<br>found on Reference Device 2<br>(SIGNA™ Hero). |
| Gating<br>Accessories | Respiratory,<br>peripheral, and<br>cardiac gating. | Respiratory<br>peripheral and<br>cardiac gating with<br>wireless connection<br>option. | In addition to SIGNA™<br>Voyager's gating solution,<br>SIGNA™ Champion offers an<br>additional option to use<br>wireless connection. |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for GE HealthCare. The logo consists of the GE monogram in a purple circle on the left, followed by the words "GE HealthCare" in purple. The text is in a sans-serif font and is slightly larger than the monogram.
# SIGNA Champion
510(k) Premarket Notification
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the GE HealthCare logo. The logo consists of a circular emblem on the left and the text "GE HealthCare" on the right. The emblem is purple and features a stylized "GE" monogram. The text "GE HealthCare" is also purple.
These differences do not raise any different questions regarding safety and effectiveness. Both devices must address questions of whether they provide an adequate level of image quality appropriate for diagnostic use. The performance data described in this submission include results of both bench testing and clinical testing that show the image quality performance of SIGNA™ Champion compared to the predicate device.
## Determination of Substantial Equivalence
Summary of Non-Clinical Tests:
The SIGNA™ Champion and the predicate device were subject to similar risk management testing to demonstrate substantial equivalence of safety and performance.
Testing to the following voluntary standards included:
- ANSI AAMI ES60601-1 ●
- IEC 60601-1-2
- IEC 60601-2-33
- IEC 62304
- IEC 60601-1-6 ●
- IEC 62366-1
- . ISO 10993-1
In addition, the SIGNA™ Champion complies with applicable NEMA MS standards for MRI and NEMA PS3 standard for DICOM, as does the predicate device.
Both the SIGNA™ Champion and the predicate device have a successful biocompatibility track record, as demonstrated by ISO 10993 testing and by their history of use in previously cleared devices.
The following quality assurance measures were applied to the development of the subject device, as they were for the predicate device:
- Risk Analysis ●
- Requirements Reviews
- . Design Reviews
- Testing on unit level (Module verification)
- Integration testing (System verification)
- . Performance testing (Verification)
- . Simulated use testing (Validation)
#### Summary of Clinical Tests:
The subject of this premarket submission, the SIGNA™ Champion, did not require clinical studies to support substantial equivalence. Sample clinical images have been included in this submission.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the GE HealthCare logo. The logo consists of a purple circle with a white GE monogram inside. To the right of the circle is the text "GE HealthCare" in purple.
## SIGNA Champion
510(k) Premarket Notification
The sample clinical images demonstrate acceptable diagnostic image performance of the SIGNA™ Champion in accordance with the FDA Guidance "Submission of Premarket Notifications for Magnetic Resonance Diagnostic Devices" issued on October 10, 2023. The image quality of the SIGNA™ Champion is substantially equivalent to that of the predicate device.
Substantial Equivalence Conclusion:
The indications for use of the proposed device are comparable to the claimed predicate device. The SIGNA™ Champion employs equivalent technology to the claimed predicate device. Additionally, the results from the above non-clinical tests demonstrate that the device performs as intended. Therefore, the SIGNA™ Champion is substantially equivalent to the predicate device to which it has been compared.
#### Conclusion
In conclusion, GE HealthCare considers the SIGNA™ Champion to be as safe, as effective, with performance that is substantially equivalent to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.