The KLS Martin Drill-Free MMF Screw is indicated for temporary stabilization of mandibular and maxillary fractures to maintain proper occlusion during surgery in adults and adolescents (age 12 and older) in whom permanent teeth have erupted.
Device Story
KLS Martin Drill-Free MMF Screw is a bone-borne, self-drilling screw used for maxillomandibular fixation (MMF). The device provides stabilization of mandibular and maxillary fractures and maintains proper occlusion during intraoperative bone fixation. The screw features a head with a hole for passing and securing ligature wire. It is manufactured from stainless steel (ASTM F138) and is available in threaded lengths of 2.0 x 8.0 mm to 2.0 x 12.0 mm. The device is applied by a surgeon during surgical procedures. It is available in both sterile and non-sterile configurations. The device benefits patients by providing stable fixation during fracture management.
Clinical Evidence
No clinical data was required. Substantial equivalence was determined based on non-clinical bench testing, including comparative static and dynamic testing against the predicate, and mechanical testing (torsional strength, drive torque, pullout strength) per ASTM F543-23. Biocompatibility was confirmed per ISO 10993-1:2018. Sterilization validation was performed per ISO 11137-1:2015 and ISO 11137-2:2015 (VDmax25 method). Packaging validation was conducted per ISO 11607-1 and ASTM D7386. Endotoxin testing met specifications per ANSI/AAMI ST72:2019.
Technological Characteristics
Material: Stainless steel (ASTM F138). Design: Self-drilling, bone-borne screw with head hole for ligature wire. Dimensions: 2.0 x 8.0 mm to 2.0 x 12.0 mm. Sterilization: Gamma radiation (ISO 11137-1:2015, ISO 11137-2:2015). Packaging: PETG blister pack with Tyvek cover (ISO 11607-1, ASTM D7386).
Indications for Use
Indicated for temporary stabilization of mandibular and maxillary fractures to maintain proper occlusion during surgery in adults and adolescents (age 12 and older) with erupted permanent teeth.
Regulatory Classification
Identification
An intraosseous fixation screw or wire is a metal device intended to be inserted into fractured jaw bone segments to prevent their movement.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 17, 2024
KLS-Martin L.P. Liza Gordillo Regulatory Affairs Project Manager 11201 Saint Johns Industrial Parkway South Jacksonville, Florida 32246
Re: K233721
Trade/Device Name: KLS Martin Drill-Free MMF Screw Regulation Number: 21 CFR 872.4880 Regulation Name: Intraosseous Fixation Screw Or Wire Regulatory Class: Class II Product Code: DZL Dated: August 19, 2024 Received: August 19, 2024
Dear Liza Gordillo:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
Sherrill Lathrop Blitzer
for Andrew Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
K233721
Device Name
KLS Martin Drill-Free MMF Screw
Indications for Use (Describe)
The KLS Martin Drill-Free MMF Screw is indicated for temporary stabilization of mandibular and maxillary fractures to maintain proper occlusion during surgery in adults and adolescents (age 12 and older) in whom permanent teeth have erupted.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
510(k) #: K233721
510(k) Summary
Prepared on: 2024-09-17
## Contact Details
21 CFR 807.92(a)(1)
| Applicant Name | KLS-Martin L.P. | |
|----------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------|--------------|
| Applicant Address | 11201 Saint Johns Industrial Pkwy S Jacksonville FL 32246 United States | |
| Applicant Contact Telephone | 800-625-1557 | |
| Applicant Contact | Ms. Melissa Bachorski | |
| Applicant Contact Email | rapm_na@klsmartin.com | |
| Correspondent Name | KLS-Martin L.P. | |
| Correspondent Address | 11201 Saint Johns Industrial Pkwy S Jacksonville FL 32246 United States | |
| Correspondent Contact Telephone | 800-625-1557 | |
| Correspondent Contact | Ms. Liza Gordillo | |
| Correspondent Contact Email | liza.gordillo@klsmartin.com | |
| Device Name<br>21 CFR 807.92(a)(2) | | |
| Device Trade Name | KLS Martin Drill-Free MMF Screw | |
| Common Name | Screw, Fixation, Intraosseous | |
| Classification Name | Intraosseous fixation screw or wire | |
| Regulation Number | 872.4880 | |
| Product Code | DZL | |
| Legally Marketed Predicate Devices<br>21 CFR 807.92(a)(3) | | |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K083432 | Drill Free MMF Screw | DZL |
| K231599 | Stryker MMF Screw | DZL |
| K153482 | KLS Martin Thoracic Plating System | HRS |
| Device Description Summary<br>21 CFR 807.92(a)(4) | | |
| The KLS Martin Drill-Free MMF Screw are bone-borne, self-drilling screws for maxillomandibular fixation (MMF). The screws are intended | | |
to provide stabilization of mandibular and maxillary fractures as well as maintain proper occlusion during intraoneralive bone fixation. These screws may be applied prior to or after exposure of the Drill Free MMF Screw is manufactured from stainless steel (ASTM F138) with a head designed with a hole to allow passing and securing ligature wire during the procedure and is available in threaded lengths of 2.0 x 8.0 mm - 2.0 x 12.0mm. Implants are available both sterile and non-sterile.
{5}------------------------------------------------
## Intended Use/Indications for Use
The KLS Martin Drill-Free MMF Screw is indication of mandibular and maxillary fractures to maintain proper occlusion during surgery in adults and adolescents (age 12 and older) in whom permanent teeth have erupted.
## Indications for Use Comparison
The indications for use of the subject device, KLS Martin Drill Free MMF Screw, is identical to the primary predicate device, KLS Martin Drill Free MMF Screw (K083432). The potential impact on substantial equivalence of each technological difference was addressed through risk analysis and verification and validation testing.
## Technological Comparison
#### Similarities to Predicates
The subject device, KLS Martin Drill Free MMF Screw is substantially equivalent to the predicate device, Drill Free MMF Screw (K083432) with respect to intended use, design, materials, and manufacturing. Anatomical sites used for the subject and predicate device are identical.
The KLS Martin Drill Free MMF Screw is identical to the predicate device (K083432) in material, stainless steel (ASTM F138),
#### Differences from Predicate
The KLS Martin Thoracic Plating System (K153482) has been included as a reference device to support the difference in sterilization method between the KLS Martin Drill Free MMF Screw and predicate devices with regard to the sterile screws. The sterile screws will use the same sterilization method, same packaging, and the same sterlif life as the reference device (K153482).
The Stryker MMF Screw (K231599) has been included as a reference device to support the change in labeling expanding the intended patient population to include skeletally mature patients.
#### Conclusion
The sterile KLS Martin Drill Free MMF Screw has the same principles of operation, and technological characteristics as the predicate devices. No new issues of safety or effectiveness have been raised.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
#### Non-Clinical Performance Data
Comparative head-to-head static and dynamic bench testing was conducted on the subject and predicate (KLS Martin Drill Free MMF Screw (K083432)) devices to determine that the subject device has equivalent performance to the primary predicate. Additionally, comparative screw testing was performed to evaluate torsional strength, drive torque, and pullout strength in accordance with ASTM F543-23. Mechanical test results demonstrate that KLS Martin MMF Screw's performance is substantially equivalent to the primary predicate device. Biological safety risk assessments in compliance with ISO 10993-1:2018 concluded the devices are biocompatible and appropriate for their intended use. LAL endotoxin testing was conducted to address the presence of bacterial endore they meet pyrogen limit specifications in accordance with ANSI/AAMI ST72:2019.
The gamma sterilization process for the sterile in accordance with ISO 11137-12015 and ISO 11137-2:2015 using the VDmax25 method. The validation was also performed in accordance with ISO 11737-1:2018 and ISO 11737-2:2019.
Packaging validations were performed for the PETG blister pack with 1073B Tyvek cover in accordance with ISO 11607-1 and ASTM D7386.
#### Clinical Performance Data
Clinical testing was not necessary for the determination of substantial equivalence.
#### Conclusions
The KLS Martin Drill-Free MMF Screw has the same intended use and similar technological characteristics as the predicate device and reference devices. Technological differences have been addressed through performance data from the predicate and reference devices, in addition to analysis of peer-reviewed clinical studies. All information provided show the safe and effective use of the subject device for the intended patient population.
According to the comparison based on the requirements of 21 CFR 807.87 and the information provided herein, it is concluded that the information included in this submission supports substantial equivalence.
## 21 CFR 807.92(a)(5)
# 21 CFR 807.92(a)(6)
21 CFR 807.92(a)(5)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.