K233707 · Invent Medical Corporation · BTT · Jun 14, 2024 · Anesthesiology
Device Facts
Record ID
K233707
Device Name
HFT150
Applicant
Invent Medical Corporation
Product Code
BTT · Anesthesiology
Decision Date
Jun 14, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5450
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
HFT150 is for spontaneously breathing adult and pediatric patients (10 kg and up) who would benefit from receiving high flow warmed and humidified respiratory gases. The flow may be from 2 - 60 L/min depending on the patient interface. HFT150 is used in hospital and professional healthcare environments.
Device Story
HFT150 is a heated humidifier with integrated flow source and heated breathing tube; delivers warmed, humidified gas flow (2-60 L/min) with supplemental oxygen (21-100%) to patients. Comprised of motorized blower assembly and integrated humidification tub. Operated by nurses, respiratory therapists, or doctors in hospital/professional healthcare environments. Software monitors ambient temperature, humidity, and flow to optimize humidity delivery and minimize condensation. Includes optional integrated Nonin Xpod 3012LP pulse oximetry for SpO2 and pulse rate monitoring. Disposable nasal cannulas, heated tubes, and water tubs used for patient interface. Provides audible/visual alarms for temperature, flow, and oxygen fraction. Benefits patients by providing respiratory support via high-flow therapy; integrated monitoring reduces need for standalone pulse oximetry devices.
Clinical Evidence
Bench testing only. Verification and validation testing performed per IEC 60601-1, IEC 60601-1-2, ISO 8185, and ISO 80601-2-74. Biocompatibility testing conducted per ISO 18562 series and ISO 10993 series. Software verification and validation performed per 21 CFR 820.30 and FDA guidance. No clinical data presented.
Technological Characteristics
Integrated blower and humidifier system. Electrical energy source. Flow range 2-60 L/min. Oxygen range 21-100%. Temperature settings 31°C, 34°C, 37°C. User interface: LCD touch screen and rotary encoder. Connectivity: Optional integrated Nonin Xpod 3012LP pulse oximetry. Materials: Externally communicating, tissue, prolonged duration. Disposable patient-contacting components (nasal cannulas, tubes, tubs). Software-controlled feedback loops with hardware backups.
Indications for Use
Indicated for spontaneously breathing adult and pediatric patients (10 kg and up) requiring high flow warmed and humidified respiratory gases (2-60 L/min) in hospital and professional healthcare settings.
Regulatory Classification
Identification
A respiratory gas humidifier is a device that is intended to add moisture to, and sometimes to warm, the breathing gases for administration to a patient. Cascade, gas, heated, and prefilled humidifiers are included in this generic type of device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
June 14, 2024
Invent Medical Corporation David Gramse Sr. Director, Regulatory Affairs 2788 Loker Ave. W. Carlsbad, California 92010
Re: K233707
Trade/Device Name: Hft150 Regulation Number: 21 CFR 868.5450 Regulation Name: Respiratory gas humidifier Regulatory Class: Class II Product Code: BTT Dated: May 9, 2024 Received: May 9, 2024
Dear David Gramse:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Ethan L. Nyberg -S
Ethan Nyberg, Ph.D. Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and
{2}------------------------------------------------
Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K233707
Device Name
HFT150
Indications for Use (Describe)
HFT150 is for spontaneously breathing adult and pediatric patients (10 kg and up) who would benefit from receiving high flow warmed and humidified respiratory gases. The flow may be from 2 - 60 L/min depending on the patient interface. HFT150 is used in hospital and professional healthcare environments.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------|----------------------------------------------------------------------|
| <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Invent Medical Corporation. The word "invent" is written in a sans-serif font, with the "in" in gray and the "vent" in blue. Below the word "invent" is the phrase "MEDICAL CORPORATION" in smaller, gray, sans-serif letters. The logo is simple and modern, with a focus on the company's name.
#### HFT150 510(k) Summary
This 510(k) summary of safety and effectiveness is submitted in accordance with the requirements of 21 CFR 807.92.
| SUBMITTER: | Invent Medical Corporation |
|-----------------------|---------------------------------------------------------|
| | 2788 Loker Ave. W. |
| | Carlsbad, CA 92010 |
| Contact Person: | David Gramse |
| | Sr. Director Regulatory Affairs |
| | P: (442) 888-7673 |
| | E: dave@invent-medical.com |
| Preparation Date: | May 08, 2024 |
| Name of Device: | HFT150 |
| Common or Usual Name: | Humidifier, Respiratory Gas, (Direct Patient Interface) |
| Regulation Name: | Respiratory gas humidifier |
| Regulation Number: | 21 CFR 868.5450 |
| Product Code: | BTT |
| Device Class: | Class 2 |
| | |
| Primary Predicate: | |
| Trade Name: | AIRVO 2 Series Humidifier |
| 510(k) Number: | K131895 |
| | |
| Reference Device: | |
| Trade Name: | ApneaLink Air |
| 510(k) Number: | K143272 |
| | |
| Reference Device: | |
| Trade Name: | HVT 2.0 |
| 510(k) Number: | K203357 |
#### DEVICE DESCRIPTION:
The HFT150 is a heated humidifier with integrated flow source and a heated breathing tube to deliver warmed and humidified gas flow to a patient. The HFT150 includes an integrated oxygen module, delivering an oxygen range of 21% to 100% to the patients.
The HFT150 humidifier is comprised of two connected functional units. One is a motorized blower assembly that provides air flow and the delivery of supplemental oxygen to patients via an integrated oxygen module. The blower speed is directly related to total delivered flow and is controlled by software with a flow sensor. The blower assembly output connects directly to an integrated humidification tub.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Invent Medical Corporation. The word "invent" is written in a sans-serif font, with the "in" in light gray and the "vent" in blue. Below the word "invent" is the phrase "MEDICAL CORPORATION" in smaller, gray, sans-serif letters. The logo is simple and modern, and the use of color helps to make it stand out.
The second functional unit of the HFT150 is a heated humidifier. The tub is integral to the HFT150. Software monitors ambient temperature, humidity, and flow to optimize the humidity delivery to the patient and minimize condensation in the patient circuit. The HFT150 is protected from contaminants and pathogens by an integral anti-bacterial filter. The nasal cannulas, heated tubes, and water tubs are disposable and are for single use only. The HFT150 may be operated by nurses, respiratory therapists, or doctors and water tubs are disposable and are for single use only. The HFT150 may be operated by nurses, respiratory therapists, or doctors.
# Indications for Use
HFT150 is for spontaneously breathing adult and pediatric patients (10 kg and up) who would benefit from receiving high flow warmed and humidified respiratory gases. The flow may be from 2 - 60 L/min depending on the patient interface. HFT150 is used in hospital and professional healthcare environments.
| HFT150 Indications for Use<br>(Subject Device – K233707) | AIRVO 2 Series Humidifier<br>(Primary Predicate Device – K131895) | Assessment of Differences |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| HFT150 is for spontaneously breathing<br>adult and pediatric patients (10 kg and<br>up) who would benefit from receiving<br>high flow warmed and humidified<br>respiratory gases. The flow may be from<br>2- 60 L/min depending on the patient<br>interface. HFT150 is used in hospital and<br>professional healthcare environments. | The AIRVO 2 humidifier is for the<br>treatment of spontaneously breathing<br>patients who would benefit from<br>receiving high flow warmed and<br>humidified respiratory gases. This<br>includes patients who have had upper<br>airways bypassed. The flow may be from<br>2 - 60 L/min depending on the patient<br>interface. The AIRVO 2 is for patients in<br>hospitals and long-term care facilities. | The indications for use are<br>substantially equivalent. The<br>devices are indicated to<br>deliver humidified<br>respiratory gases to<br>spontaneously breathing<br>patients. |
## SUBJECT AND PREDICATE DEVICE INDICATIONS FOR USE COMPARISON:
# TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE:
A technological comparison table is provided below that compares the subject device and predicate device:
| Functionality | HFT150<br>(Subject Device – K233707) | AIRVO 2 Series Humidifier<br>(Primary Predicate Device –<br>K131895) | Assessment of Differences | |
|------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Flow range | 2 – 60 L/min | 2 – 60 L/min | Equivalent: The HFT150 is<br>within the range of the<br>predicate | |
| Oxygen input | < 60 L/min<br>(internally regulated via<br>software and a flow sensor) | < 60 L/min | Identical | |
| Functionality | HFT150<br>(Subject Device – K233707) | AIRVO 2 Series Humidifier<br>(Primary Predicate Device – K131895) | Assessment of Differences | |
| Oxygen Range | 21 to 100% O2 via Oxygen<br>module | 25 to 95% O2 | Equivalent: The HFT150 has a<br>slightly larger range, and this<br>difference does not impact<br>safety or performance<br>specifications | |
| Design | Integrated blower and<br>humidifier with chamber and<br>heated breathing tube | Integrated blower and<br>humidifier with Integrated<br>blower and humidifier with<br>chamber and heated breathing<br>tube | Similar | |
| Energy source | Electrical from mains to<br>power blower and heating | Electrical from mains to power<br>blower and heating | Identical | |
| Temperature<br>setting | 31°C – 37°C | 31°C, 34°C, 37°C | Identical | |
| Operating principle | Delivery of humidified Air/02<br>at constant flow to the patient | Delivery of humidified Air/02 at<br>constant flow to the patient | Identical | |
| Patient interface | Nasal cannula only | Nasal cannula, tracheostomy<br>direct connection, face mask | Similar | |
| Alarms | Audible and visual for<br>Temperature, Flow and<br>Oxygen Fraction | Audible and visual for<br>Temperature, Flow and Oxygen<br>Fraction | Identical | |
| User interface | User set point adjustment via<br>rotary encoder knob or<br>pressing color LCD touch<br>screen for temperature and<br>flow | User set point adjustment via<br>menu system on color display<br>for temperature and flow | Equivalent: The difference<br>with user adjustment<br>settings does not impact<br>safety or performance. | |
| Display | LCD display | LED display for Temperature,<br>Flow, Oxygen Fraction | Similar | |
| Control | Software control using<br>feedback sensors with<br>hardware back ups<br>Nonin Xpod 3012LP | Software control using<br>feedback sensors with<br>hardware backups | Identical | |
| Integrated Pulse<br>Oximetry | Optional feature provided for<br>user convenience so users do<br>not have to use a separate,<br>standalone pulse oximetry<br>device. | Not included | Different: Refer to the<br>Reference Device<br>comparison table below. | |
| Transport mode | Limits power to flow only | Limits power to flow only | Identical | |
| Functionality | HFT150<br>(Subject Device -<br>K233707) | ApneaLink Air<br>(Reference Device<br>K143272) | HVT 2.0<br>Reference Device<br>K203357 | Assessment of<br>Differences |
| Integrated<br>Pulse<br>Oximetry | Nonin Xpod 3012LP<br>Optional feature provided<br>for user convenience so<br>users do not have to use a<br>separate, standalone pulse<br>oximetry device. | Nonin Xpod 3012LP | N/A | Identical technology:<br>This additional<br>feature is the same<br>as the reference<br>device, ApneaLink Air<br>(cleared in K143272).<br>This additional<br>feature does not<br>impact safety or<br>performance. |
| Patient<br>Contacting<br>Materials | Externally communicating,<br>tissue, prolonged duration | N/A | Externally communicating,<br>tissue, prolonged duration | Identical: both<br>devices have the<br>same duration of<br>contact. |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Invent Medical Corporation. The word "invent" is written in a sans-serif font, with the "in" in gray and the "vent" in blue. Below the word "invent" is the phrase "MEDICAL CORPORATION" in smaller, gray, sans-serif letters. The logo is simple and modern.
# _K233707 - HFT150: Traditional 510(k)
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Invent Medical Corporation. The word "invent" is written in a sans-serif font, with the "in" in gray and the "vent" in blue. Below the word "invent" is the phrase "MEDICAL CORPORATION" in a smaller, sans-serif font, also in blue. The logo is simple and modern, and the use of color helps to create a sense of trust and reliability.
### K233707 - HFT150: Traditional 510(k)
A technological comparison is provided below between the subject device and reference device:
## SUBSTANTIAL EQUVALENCE DISCUSSION:
The subject device, HFT150, is substantially equivalent in technology, design and function to the predicate device, AIRVO 2 Series Humidifier cleared under K131895. The minor differences between predicate and subject devices are identified in the table above and are described further below.
Both the HFT150 and the predicate device support the delivery of oxygen flow up to 60 L/min. The subject device utilizes an internal mechanism for oxygen flow delivery. The predicate device utilizes an external mechanism for oxygen flow delivery.
Both the HFT150 and the predicate device support a nasal cannula patient interface. The predicate device also supports a tracheotomy and mask interface.
The HFT150 utilizes both a rotary control knob and LCD touch screen for selecting device settings. The predicate device utilizes hard keys for selecting device settings.
The HFT150 provides users the option of an integrated Nonin Xpod 3012LP for SpO2% and Pulse Rate display and alarms for low SpO2%, high SpO2%, High Pulse Rate, and Low Pulse Rate. Although both the subject device and predicate device provide supplemental oxygen, the predicate device does not provide an integrated Pulse Oximetry option. For either the subject device or predicate device, the user may choose to use an external, standalone pulse oximetry device. As such, this submission includes the reference device, ApneaLink Air (K143272), which also utilizes the same well-established Nonin Xpod 3012LP technology. Additionally, a second
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Invent Medical Corporation. The word "invent" is written in a sans-serif font, with the "in" in gray and the "vent" in blue. Below the word "invent" is the phrase "MEDICAL CORPORATION" in smaller, gray, sans-serif letters. The logo is simple and modern.
reference device, HVT 2.0 (K203357), was added to establish the duration of patient contacting materials as externally communicating, tissue, prolonged duration.
The above differences between the HFT150 subject device and the AIRVO 2 Series Humidifier do not impact substantial equivalency.
## TESTING SUMMARY:
Comprehensive testing was conducted to verify and validate that the HFT150 meets specified requirements and is substantially equivalent to the predicate device, AIRVO 2 Series Humidifier. The HFT150 was subjected to verification and validation testing according to the following standards to demonstrate that it performs as intended:
- Safety testing in accordance with IEC 60601-1:2012+AMD1:2012+AMD2:2020, Medical Electrical Equipment – part 1: General requirements for basic safety and essential performance
- EMC testing in accordance with IEC 60601-1-2:2014 ED. 4.0, Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: Electromagnetic disturbances- Requirements and tests
- Biocompatibility testing in accordance with:
- ISO 18562-1:2017, Biocompatibility Evaluation of Breathing Gas Pathways in O Healthcare Applications – Evaluation and testing within a risk management process
- ISO 18562-2:2017, Biocompatibility Evaluation of Breathing Gas Pathways in O Healthcare Applications – Tests for emissions of particulate matter
- O ISO 18562-3:2017, Biocompatibility Evaluation of Breathing Gas Pathways in Healthcare Applications - Test for emissions of volatile organic compounds (VOCs)
- O ISO 10993-1:2018, Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process
- ISO 10993-5:2009, Biological Evaluation of Medical Devices Part 5: Tests for In O Vitro Cytotoxicity
- ISO 10993-10:2021, Biological Evaluation of Medical Devices Part 10: Tests for O Irritation and Skin Sensitization
- O ISO 10993-11:2017, Biological evaluation of medical devices – Part 11: Tests for systemic toxicity
- ISO 10993-17:2002, Biological evaluation of medical devices Part 17: o Establishment of allowable limits for leachable substances
- ISO 10993-23:2021, Biological evaluation of medical devices Part 23: Tests for o irritation
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Invent Medical Corporation. The word "invent" is written in a sans-serif font, with the "i" in gray and the rest of the letters in blue. Below the word "invent" is the phrase "MEDICAL CORPORATION" in smaller, gray, sans-serif letters.
- Particular Requirements testing in accordance with ISO 8185:2007, Respiratory track ● humidifiers for medical user – Particular requirements for respiratory humidification systems; and 80601-2-74:2017, Particular requirements for basic safety and essential performance of respiratory humidifying equipment
- . Device life testing
- . Shelf life testing
- Software Verification and Validation Testing in accordance with 21 CFR 820.30, the FDA guidance document, General Principles of Software Validation, and the FDA guidance document, Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.
#### CONCLUSION:
The information provided in this premarket notification established that the subject device, HFT150, and the predicate device, AIRVO 2 Series Humidifier, are substantially equivalent in comparison to indications for use, design, technology and performance.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.