CellFX Percutaneous Electrode System (SYS3000)

K233705 · Pulse Biosciences, Inc. · GEI · Mar 8, 2024 · General, Plastic Surgery

Device Facts

Record IDK233705
Device NameCellFX Percutaneous Electrode System (SYS3000)
ApplicantPulse Biosciences, Inc.
Product CodeGEI · General, Plastic Surgery
Decision DateMar 8, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The CellFX Percutaneous Electrode System is intended for ablation of soft tissue in percutaneous and intraoperative surgical procedures. The CellFX Percutaneous Electrode System (Percutaneous Electrode) is not intended for use in cardiac procedures.

Device Story

CellFX Percutaneous Electrode System uses nanosecond Pulsed Field Ablation (nsPFA) to ablate soft tissue. System includes console, percutaneous electrode, adapter, and pneumatic footswitch. Device delivers nanosecond-duration, high-amplitude electrical pulses via bipolar electrodes to create nanopores in lipid membranes, disrupting cellular homeostasis and causing regulated cell death (apoptosis). Used in percutaneous and intraoperative surgical procedures by clinicians. Output is nonthermal ablation of targeted soft tissue. Benefits include precise, nonthermal tissue destruction without thermal damage to surrounding structures. Clinical decision-making supported by ultrasound guidance for electrode placement.

Clinical Evidence

Prospective single-arm clinical study of 10 subjects with benign thyroid nodules. Ablation zones monitored via ultrasound at 30 and 90 days. Primary endpoint: safety and effectiveness. Results showed 95% average reduction in ablation zones by 30 days. No serious adverse events reported. Supported by two in vivo animal safety studies.

Technological Characteristics

Energy-based platform for nanosecond Pulsed Field Ablation (nsPFA). Bipolar electrode (Stainless Steel 304, 95mm length, 2.4mm diameter). Pulse parameters: 800V-15kV amplitude, 100-500ns width, 1-10Hz frequency. Max power 15W. EO sterilization. Single-use. Complies with IEC 60601-1, 60601-1-2, 60601-2-2, 60601-1-6, 62366, 62304.

Indications for Use

Indicated for ablation of soft tissue in percutaneous and intraoperative surgical procedures. Not for use in cardiac procedures.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged in a circular pattern. To the right of the seal, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on top and "ADMINISTRATION" below. The logo is simple and professional, reflecting the FDA's role in regulating food and drugs. March 8. 2024 Pulse Biosciences, Inc. Uyen Mai Director, Regulatory Affairs, Quality Assurance, and Quality Control 3957 Point Eden Way Hayward. California 94545 #### Re: K233705 Trade/Device Name: CellFX Percutaneous Electrode System (SYS3000) Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: February 2, 2024 Received: February 5, 2024 Dear Uyen Mai: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). {1}------------------------------------------------ Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reporting-mdr-howreport-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Digitally signed by Mark Mark Trumbore -S Trumbore -S Date: 2024.03.08 12:47:48 -05'00' Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K233705 Device Name CellFX Percutaneous Electrode System (SYS3000) Indications for Use (Describe) The CellFX Percutaneous Electrode System is intended for ablation of soft tissue in percutaneous and intraoperative surgical procedures. The CellFX Percutaneous Electrode System (Percutaneous Electrode) is not intended for use in cardiac procedures. Type of Use (Select one or both, as applicable) | <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> | |------------------------------------------------------------|---------------| | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) | <span></span> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary #### I. Submitter | Applicant: | Pulse Biosciences, Inc.<br>3957 Point Eden Way<br>Hayward, CA 94545<br>Phone: (833) 257-3393 | |-----------------|-------------------------------------------------------------------------------------------------------------| | Contact Person: | Uyen Mai, MS<br>Director, Regulatory Affairs, QA and QC<br>Pulse Biosciences, Inc.<br>Phone: (510) 906-4691 | | Date Prepared: | November 16, 2023 | #### DEVICE INFORMATION II. CellFX® Percutaneous Electrode System Trade Name: | Regulation Number | 21 CFR § 878.4400 | |-----------------------|----------------------------------------------------------------| | Regulation Name: | Electrosurgical Cutting and Coagulation Device and Accessories | | Regulation Class: | Class II | | Product Code: | GEI | | Classification Panel: | General and Plastic Surgery | #### III. PREDICATE DEVICE K190052 - Surgnova Healthcare Technologies (Zhejiang) Co., Ltd, Radio Frequency Ablation System Reference Devices K052796 - Valleylab Cool-tip RF Ablation System and K203299 – CellFX System #### IV. DEVICE DESCRIPTION The CellFX® Percutaneous Electrode System includes the CellFX Console, CellFX Percutaneous Electrode and Accessories: Percutaneous Adapter, pneumatic footswitch, and system software. The CellFX® Console is a proprietary energy-based platform for nanosecond Pulsed Field Ablation™ (nsPFA™) technology. nsPFA is a cell-specific, nonthermal ablation technology that delivers nanosecond-duration pulses of highamplitude electrical energy to tissue via bipolar electrodes. The pulses create nanopores in lipid membranes, which disrupt the ability of the cell and internal organelles to maintain cellular homeostasis - ultimately leading to regulated cell death (e.g., apoptosis). The electrode is provided sterile, for single use only. {4}------------------------------------------------ ## V. INDICATIONS FOR USE STATEMENT The CellFX® Percutaneous Electrode System is intended for ablation of soft tissue in percutaneous and intraoperative surgical procedures. It is not intended for use in cardiac procedures. ### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE To support a determination of substantial equivalence, Pulse Biosciences performed verification and validation testing demonstrating the subject device performs as intended; therefore, even though the CellFX Percutaneous Electrode System has different technological characteristics than the predicate device, based on the data provided, the CellFX Percutaneous Electrode System does not raise any new or different questions of safety and effectiveness. | | Subject Device | Predicate<br>Device<br>(K190052) | Ref Device<br>(K052796) | Ref Device<br>(K203299) | | |----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------|---------| | Trade Name | CellFX® Percutaneous<br>Electrode System | Dolphi RF<br>R150E<br>Ablation<br>System | Valleylab Cool-<br>tip RF Ablation<br>System | CellFX<br>System | | | Prescription Only | Yes | Yes | Yes | Yes | | | Regulation | 878.4400 | 878.4400 | 878.4400 | 878.4400 | | | Product Code | GEI | GEI | GEI | GEI | | | | | Indications for Use | | | | | Indications for<br>Use Statement | Intended for ablation of<br>soft tissue in<br>percutaneous and<br>intraoperative surgical<br>procedures. It is not<br>intended for use in<br>cardiac procedures. | Intended for<br>coagulation and<br>ablation of soft<br>tissue. It is not<br>intended for<br>use in cardiac<br>procedures. | Intended for the<br>use in<br>percutaneous,<br>laparoscopic, and<br>intraoperative<br>coagulation and<br>ablation of tissue. | Intended for<br>ablation and<br>resurfacing<br>of the skin. | | | | | Technological Characteristics | | | | | Components | RF Generator, Active<br>Electrode, Adapter,<br>Footswitch | Radio<br>Frequency<br>Generator,<br>Radio<br>Frequency<br>Electrode Kits,<br>Temperature<br>Probe, and Foot<br>Switch | RF Generator,<br>peristaltic pump,<br>electrodes,<br>inflow/outflow<br>tubing, and return<br>pads | RF<br>Generator,<br>Handpiece,<br>Treatment<br>tips | | | Mechanism of<br>Action | Nanosecond Pulse<br>Field Ablation<br>(nsPFA): ultrafast<br>nanosecond electrical | Radiofrequenc<br>y current for<br>ablation of soft<br>target tissue | Radiofrequency<br>current for<br>ablation of soft<br>target tissue | Same as<br>Subject<br>Device | | | | Subject Device | Predicate<br>Device<br>(K190052) | Ref Device<br>(K052796) | Ref Device<br>(K203299) | K233705 | | | pulses to the targeted<br>tissue | | | | | | Output Frequency | 1-10 Pulses per second | 470kHz | 480kHz±10% | Same as<br>Subject<br>Device | | | Pulse<br>Amplitude/Voltage | 800V to 15kV | Up to 200 watts<br>@50 ohm | Up to 200 watts<br>@50 ohm | Same as<br>Subject<br>Device | | | Pulse Width | 100 ns – 500 ns | N/A | N/A | Same as<br>Subject<br>Device | | | Power Input | 100-240 VAC, 50/60 Hz,<br>2A | 100-240 VAC,<br>50/60 Hz | 100-240 VAC,<br>50/60 Hz | Same as<br>Subject<br>Device | | | Max Power<br>Output | 15 Watts<br>(when pulsing at 15kV pulse voltage with<br>500 ns pulse width and 10 Hz pulse<br>frequency into a 75 Ohm load) | Up to 200<br>Watts@50<br>ohms | 0-200 watts max<br>output @ 50 ohm<br>load | Same as<br>Subject<br>Device | | | | | Active Electrodes | | | | | Energy Output | Bipolar RF | Monopolar RF | Monopolar RF | Same as<br>Subject<br>Device | | | Electrode Length<br>(mm) | 95 | 70, 150, 200,<br>250 | 100, 144, 150,<br>200, 250 | Same as<br>Subject<br>Device | | | Exposure Length<br>(mm) | 14 | 5, 7, 10, 15, 20,<br>30, 40 | 7, 10, 20, 30 | Same as<br>Subject<br>Device | | | Outer Diameter<br>(mm) | 2.4 | 1.2, 1.47 | 1.2 | Same as<br>Subject<br>Device | | | Tip Type | Trocar | Trocar | Trocar | Microneedle<br>Array | | | Material | Stainless Steel 304 | Stainless Steel<br>304 | Stainless Steel<br>304 | Same as<br>Subject<br>Device | | | Disposable/Singe -<br>Use Device | Single-Use | Single-Use | Single-Use | Same as<br>Subject<br>Device | | | | Subject Device | Predicate<br>Device<br>(K190052) | Ref Device<br>(K052796) | Ref Device<br>(K203299) | | | Biocompatibility | Patient-contacting<br>materials are<br>biocompatible | Patient-<br>contacting<br>materials are<br>biocompatible | Patient-<br>contacting<br>materials are<br>biocompatible | Same as<br>Subject<br>Device | | | Sterility | EO Sterilization | EO<br>Sterilization | EO sterilization | Same as<br>Subject<br>Device | | | System Physical Characteristics | | | | | | | Voltage Supply | 100-240 VAC, 50/60<br>Hz, 2A | 100-240 VAC, 50/60 Hz, 2A | 100-240 VAC, 50/60 Hz, 2A | Same as<br>Subject<br>Device | | {5}------------------------------------------------ {6}------------------------------------------------ ### VII. PERFORMANCE DATA The following performance data was referenced or provided in support of the substantial equivalence determination. #### Biocompatibility Testing The testing was conducted in accordance with FDA Guidance, issued September 4, 2020: Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process." All biocompatibility tests met their respective acceptance criteria, for each of the following tests: - · Pyrogenicity, Material Mediated - · Systemic Toxicity - · Sensitization - Irritation - · Cytotoxicity - · Hemolysis The biocompatibility test results successfully demonstrate biocompatibility of the CellFX Percutaneous Electrode. #### Electrical Safety and Electromagnetic Compatibility (EMC) Testing The CellFX System complies with IEC 60601-1, IEC 60601-1-2, 60601-2-2, IEC 60601-1-6, IEC 62366, and IEC 62304. All test conditions were performed as outlined within the IEC standards. All electrical safety and EMC tests passed. #### Bench Testing The performance bench testing was conducted to ensure that the CellFX Percutaneous Electrode System performs as intended per its specifications. #### Software Verification and Validation Testing {7}------------------------------------------------ Software verification and validation testing were conducted and enhanced documentation was provided as recommended by FDA's Guidance, "Content of Premarket Submissions for Device Software Functions." issued on June 14, 2023. The intended use of this device is for general surgical procedures and in accordance with FDA's Guidance, "Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery," issued on March 9, 2020, an enhanced documentation level of software testing was included in this 510(k) that supports the safety and effectiveness of the device. Cybersecurity controls have been implemented to mitigate the risk of malware being introduced into the CellFX Percutaneous Electrode System as recommended by FDA's Guidance, "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices," issued on September 27, 2023 #### Pre-Clinical Animal Safety and Performance Studies In accordance with FDA's Guidance, "Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery," two in vivo animal studies were conducted to support the safety and performance of the CellFX Percutaneous Electrode System. Based on the in vivo animal safety studies, the data provided in this submission demonstrates and supports that the CellFX Percutaneous Electrode System is safe and effective for ablation of soft tissue and is substantially equivalent to the predicate device. #### Human Safety Study A prospective single-arm clinical study of ten (10) subjects was completed to demonstrate safety and effectiveness of the CellFX Percutaneous Electrode System. Two to four ablation zones were completed in benign thyroid nodules under ultrasound guidance and subjects were followed at 30 and 90 days post-treatment. Subjects showed an average of 95% reduction in ablation zones by 30 days. No serious adverse events were observed. Overall, the findings of this study provided strong evidence of the safety and effectiveness of the system in a surgical application in soft tissue. #### Summary The performance testing in pre-clinical and clinical safety studies was conducted to ensure that the CellFX Percutaneous Electrode System performs as intended per its specifications. CellFX Percutaneous Electrode System treatments are deemed nonthermal and did not result in any safety concerns during its intended use. Therefore, the performance testing verified that the device performs as intended and that the tested energy settings and treatment profile in soft tissue does not raise any different questions of safety or effectiveness. ### VIII. CONCLUSION The CellFX® Percutaneous Electrode System has the same indication for use, technological characteristics and principles of operation as the predicate device. The non-clinical performance testing, in vivo animal data, and clinical performance study data provided in this submission demonstrate and support that the CellFX Percutaneous Electrode System performs as intended and is as safe and as effective as the predicate device. Therefore, the CellFX Percutaneous Electrode System is substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...