Segmental Plating System (SPS);IdentiTi SPS Interbody System;IdentiTi NanoTec SPS Interbody System; Transcend SPS Interbody System;Transcend NanoTec SPS Interbody System
K233640 · Alphatec Spine, Inc. · KWQ · Feb 22, 2024 · Orthopedic
Device Facts
Record ID
K233640
Device Name
Segmental Plating System (SPS);IdentiTi SPS Interbody System;IdentiTi NanoTec SPS Interbody System; Transcend SPS Interbody System;Transcend NanoTec SPS Interbody System
Applicant
Alphatec Spine, Inc.
Product Code
KWQ · Orthopedic
Decision Date
Feb 22, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3060
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Segmental Plating System (SPS) The Segmental Plating System (SPS) is intended for anterior screw fixation to the cervical spine (C2-T1) for the following indications: degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), trauma (including fractures), tumors, deformity (kyphosis or scoliosis), pseudarthrosis, failed previous fusions, spondylolisthesis, and spinal stenosis. IdentiTi SPS Interbody System The IdentiTi SPS Interbody System is an anterior cervical interbody fusion system intended for spinal fusion procedures in skeletally mature patients with cervical disc degeneration and/or cervical spinal instability, as confirmed by imaging studies (radiographs, CT, MRI), that results in radiculopathy, myelopathy, and/or pain at multiple contiguous levels from C2-T1. The IdentiTi SPS Interbody System is intended for use with supplemental fixation systems. The system is designed for use with autograft, allograft comprised of cortical, cancellous, and/or corticocancellous bone graft, demineralized allograft with bone marrow aspirate, or a combination thereof. IdentiTi NanoTec SPS Interbody System The IdentiTi SPS Interbody System with advanced NanoTec surface treatment is an anterior cervical interbody fusion system intended for spinal fusion procedures in skeletally mature patients with cervical disc degeneration and/or cervical spinal instability, as confirmed by imaging studies (radiographs, CT, MRI), that results in radiculopathy, and/or pain at multiple contiguous levels from C2-T1. The IdentiTi NanoTec SPS Interbody System is intended for use with supplemental fixation systems. The system is designed for use with autograft, allograft comprised of cortical, cancellous, and/or corticocancellous bone graft, demineralized allograft with bone marrow aspirate, or a combination thereof. Transcend SPS Interbody System The Transcend SPS Interbody System is an anterior cervical interbody fusion system intended for use in skeletally mature patients with cervical disc degeneration and/or cervical spinal instability, as confirmed by imaging studies (radiographs, CT, MRI), that results in radiculopathy, and/ or pain at multiple contiguous levels from C2-T1. The Transcend SPS Interbody System is intended for use with supplemental fixation systems. The system is designed for use with autograft comprised of cortical, cancellous and/or corticocancellous bone graft, demineralized allograft with bone marrow aspirate, or a combination thereof. Transcend NanoTec SPS Interbody System The Transcend SPS PEEK Interbody System with advanced NanoTec surface treatment is an anterior cervical interbody fusion system intended for use in skeletally mature patients with cervical disc degeneration and/or cervical spinal instability, as confirmed by imaging studies (radiographs. CT, MRI), that results in radiculopathy, and/or pain at multiple contiguous levels from C2-T1. The Transcend NanoTec SPS Interbody System is intended for use with supplemental fixation systems. The system is designed for use with autograft, allograft comprised of cortical, cancellous and/or corticocancellous bone graft, demineralized allograft with bone marrow aspirate, or a combination thereof.
Device Story
System comprises Segmental Plating System (SPS) and various interbody fusion spacers (IdentiTi/Transcend, with/without NanoTec surface treatment). SPS uses titanium alloy plates and screws for anterior cervical fixation. Interbody spacers (PEEK, titanium, tantalum) provide structural support for fusion; feature internal graft apertures and endplate teeth to mitigate expulsion. NanoTec implants feature 20-40nm hydroxyapatite surface treatment. Used in surgical procedures by surgeons to stabilize cervical spine segments; requires supplemental fixation. Implants facilitate bone graft containment to promote fusion. SPS provided non-sterile (steam sterilized by user); interbody spacers provided sterile. Benefits include spinal stabilization and fusion in patients with degenerative conditions or instability.
Clinical Evidence
Bench testing only. No clinical data provided. Mechanical testing included static/dynamic compression, static/dynamic torsion (ASTM F2077), subsidence (ASTM F2267), push-out/expulsion (ASTM Draft), and vertebrectomy model testing (ASTM F1717).
Indicated for skeletally mature patients requiring anterior cervical spinal fusion (C2-T1) due to degenerative disc disease, trauma, tumors, deformity, pseudarthrosis, failed fusions, spondylolisthesis, or spinal stenosis. Interbody systems indicated for cervical disc degeneration/instability resulting in radiculopathy, myelopathy, or pain. Contraindications: none stated.
Regulatory Classification
Identification
A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.
Divergence Mini-Plate Anterior Cervical Fusion System (K140417)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 22, 2024
Alphatec Spine Inc. Andrew Zhang Regulatory Affairs Specialist 1950 Camino Vida Roble Carlsbad, California 92008
# Re: K233640
Trade/Device Name: Segmental Plating System (SPS);IdentiTi SPS Interbody System;IdentiTi NanoTec SPS Interbody System: Transcend SPS Interbody System:Transcend NanoTec SPS Interbody System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: Class II Product Code: KWQ, ODP Dated: November 13, 2023 Received: January 24, 2024
Dear Andrew Zhang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Brent Showalter -S
Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
Enclosure
{3}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
K233640
Device Name
Segmental Plating System (SPS); ldentiTi SPS Interbody System; IdentiTi NanoTec SPS Interbody System; Transcend SPS Interbody System; Transcend NanoTec SPS Interbody System Indications for Use (Describe)
Segmental Plating System (SPS)
The Segmental Plating System (SPS) is intended for anterior screw fixation to the cervical spine (C2-T1) for the following indications: degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), trauma (including fractures), tumors, deformity (kyphosis or scoliosis), pseudarthrosis, failed previous fusions, spondylolisthesis, and spinal stenosis.
# IdentiTi SPS Interbody System
The IdentiTi SPS Interbody System is an anterior cervical interbody fusion system intended for spinal fusion procedures in skeletally mature patients with cervical disc degeneration and/or cervical spinal instability, as confirmed by imaging studies (radiographs, CT, MRI), that results in radiculopathy, myelopathy, and/or pain at multiple contiguous levels from C2-T1. The IdentiTi SPS Interbody System is intended for use with supplemental fixation systems. The system is designed for use with autograft, allograft comprised of cortical, cancellous, and/or corticocancellous bone graft, demineralized allograft with bone marrow aspirate, or a combination thereof.
# IdentiTi NanoTec SPS Interbody System
The IdentiTi SPS Interbody System with advanced NanoTec surface treatment is an anterior cervical interbody fusion system intended for spinal fusion procedures in skeletally mature patients with cervical disc degeneration and/or cervical spinal instability, as confirmed by imaging studies (radiographs, CT, MRI), that results in radiculopathy, and/or pain at multiple contiguous levels from C2-T1. The IdentiTi NanoTec SPS Interbody System is intended for use with supplemental fixation systems. The system is designed for use with autograft, allograft comprised of cortical, cancellous, and/or corticocancellous bone graft, demineralized allograft with bone marrow aspirate, or a combination thereof.
## Transcend SPS Interbody System
The Transcend SPS Interbody System is an anterior cervical interbody fusion system intended for use in skeletally mature patients with cervical disc degeneration and/or cervical spinal instability, as confirmed by imaging studies (radiographs, CT, MRI), that results in radiculopathy, and/ or pain at multiple contiguous levels from C2-T1. The Transcend SPS Interbody System is intended for use with supplemental fixation systems. The system is designed for use with autograft comprised of cortical, cancellous and/or corticocancellous bone graft, demineralized allograft with bone marrow aspirate, or a combination thereof.
# Transcend NanoTec SPS Interbody System
The Transcend SPS PEEK Interbody System with advanced NanoTec surface treatment is an anterior cervical interbody fusion system intended for use in skeletally mature patients with cervical
{4}------------------------------------------------
disc degeneration and/or cervical spinal instability, as confirmed by imaging studies (radiographs. CT, MRI), that results in radiculopathy, and/or pain at multiple contiguous levels from C2-T1. The Transcend NanoTec SPS Interbody System is intended for use with supplemental fixation systems. The system is designed for use with autograft, allograft comprised of cortical, cancellous and/or corticocancellous bone graft, demineralized allograft with bone marrow aspirate, or a combination thereof.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for "atec". The "a" is green and the "tec" is dark blue. There is a trademark symbol next to the "c". There is a black line underneath the logo.
This 510(k) summary of safety and effectiveness is being submitted in accordance with the requirements of 21 CFR 807.92.
| I. | SUBMITTER: | Alphatec Spine, Inc.<br>1950 Camino Vida Roble<br>Carlsbad, CA 92008<br>Phone: (760) 494-6860<br>Fax: (760) 431-0289 |
|----|------------------------|----------------------------------------------------------------------------------------------------------------------|
| | Contact Person: | Andrew Zhang<br>Regulatory Affairs Specialist<br>Contact Phone: (760) 494-6860 |
| | Date Summary Prepared: | January 24, 2024 |
#### II. DEVICE
Trade or Proprietary Name:
| | Segmental Plating System (SPS) |
|----------------------|-----------------------------------------------------------------------------------------------------------|
| | IdentiTi™ SPS Interbody System |
| | IdentiTi™ NanoTec™ SPS Interbody System |
| | Transcend™ SPS Interbody System |
| | Transcend™ NanoTec™ SPS Interbody System |
| Common Name: | Spinal Intervertebral Body Fixation Orthosis<br>Intervertebral body fusion device |
| Classification Name: | Appliance, Fixation, Spinal Intervertebral Body<br>Intervertebral fusion device with bone graft, Cervical |
| Regulation Number: | 21 CFR 888.3060<br>21 CFR 888.3080 |
| Classification: | Class II |
| Product Code: | KWQ, ODP |
#### III. LEGALLY MARKETED PREDICATE DEVICES
| 510(k) | Product Name | Product Code | Clearance Date |
|---------|-----------------------------------------|--------------|-------------------|
| K213443 | Insignia Anterior Cervical Plate System | KWQ | December 15, 2021 |
| 510(k) | Product Name | Product Code | Clearance Date |
|---------|----------------------------------------------------|--------------|-------------------|
| K211903 | Shoreline Threaded TruProfile Plate | KWQ | August 30, 2021 |
| K072703 | Pioneer SlimFuse Anterior Cervical Plate<br>System | KWQ | January 10, 2008 |
| K222973 | IdentiTi and Transcend Interbody Systems | ODP, OVE | November 17, 2022 |
# Additional Predicate Devices:
## CONFIDENTIAL
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "atec" in a stylized font. The "a" is green, while the "tec" is a dark blue color. There is a trademark symbol to the right of the "c". A black line is underneath the word.
| Reference Devices: | | | |
|--------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|-------------------|
| 510(k) | Product Name | Product<br>Code | Clearance Date |
| K211805 | IdentiTiTM Porous Ti Interbody System,<br>TranscendTM PEEK Interbody System, IdentiTiTM<br>NanoTecTM Interbody System, TranscendTM<br>NanoTecTM Interbody System | MAX, OVD,<br>ODP, PHM | August 26, 2021 |
| K192938 | InvictusTM Spinal Fixation System | NKB, KWP | December 12, 2019 |
| K222028 | IdentiTiTM Porous Ti Interbody System, IdentiTiTM<br>NanoTecTM Interbody System, TranscendTM PEEK<br>Interbody System, TranscendTM NanoTecTM<br>Interbody System, IdentiTiTM ALIF Standalone<br>Interbody System, IdentiTiTM NanoTecTM ALIF<br>Standalone Interbody System | MAX, OVD,<br>PHM | October 07, 2022 |
| K140417 | Divergence Mini-Plate Anterior Cervical Fusion<br>System | KWQ, ODP | July 09, 2014 |
#### IV. DEVICE DESCRIPTION
The Segmental Plating System (SPS) is intended for anterior fixation to the cervical spine. The Segmental Plating System (SPS) consists of a variety of sizes of 2 - 4 holes plates and 3.5 mm and 4.0 mm screws that are manufactured from titanium alloy conforming to ASTM F136 and are offered non-sterile. The plate includes a screw anti-backout mechanism. The system will offer instrumentation for the delivery of the plate and screw construct. The instruments in this system are intended for use in surgical procedures. The plate system implants are provided non-sterile to be steam sterilized by the end user.
The IdentiTi and Transcend SPS Interbody Systems are cervical intervertebral body fusion systems designed to be inserted through anterior surgical approaches. The interbody spacers are manufactured from PEEK (polyetheretherketone) Optima LT1 per ASTM F2026, tantalum per ASTM F560, commercially pure titanium (CP Ti Grade 2) per ASTM F67, and an optional hydroxyapatite nano (HAMM) surface treatment. The subject system implants consist of various lengths, widths, heights and lordotic options to accommodate individual patient anatomy. To mitigate risk of expulsion, the interbody endplates feature teeth. All interbody spacers feature an internal graft aperture for placement of graft material to promote fusion through the cage. Additionally, the IdentiTi implants are offered with a microstructure due to the layering of material that forms the porous architecture. This porous geometry extends to the superior and inferior surfaces of the device for implant fixation. The subject IdentiTi and Transcend NanoTec SPS Interbody Systems interbody implant surfaces have been treated with a 20-40 nanometer thin hydroxyapatite (HA) surface treatment. The surface treatment presents a nano-scale topography on the entirety of the implant surface. in addition to macro-/micro-scale topography existing from prior to HA man treatment. The interbody spacers are provided individually packaged and sterile.
#### V. INDICATIONS FOR USE
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for "atec". The "a" is green, and the "tec" is dark blue. There is a horizontal line below the logo.
# Segmental Plating System (SPS)
The Segmental Plating System (SPS) is intended for anterior screw fixation to the cervical spine (C2-T1) for the following indications: degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), trauma (including fractures), tumors, deformity (kyphosis, lordosis or scoliosis), pseudarthrosis, failed previous fusions, spondylolisthesis, and spinal stenosis.
## IdentiTi SPS Interbody System
The IdentiTi SPS Interbody System is an anterior cervical interbody fusion system intended for spinal fusion procedures in skeletally mature patients with cervical disc degeneration and/or cervical spinal instability, as confirmed by imaging studies (radiographs, CT, MRI), that results in radiculopathy, myelopathy, and/or pain at multiple contiguous levels from C2-T1. The IdentiTi SPS Interbody System is intended for use with supplemental fixation systems. The system is designed for use with autograft, allograft comprised of cortical, cancellous, and/or corticocancellous bone graft, demineralized allograft with bone marrow aspirate, or a combination thereof.
## IdentiTi NanoTec SPS Interbody System
The IdentiTi SPS Interbody System with advanced NanoTec surface treatment is an anterior cervical interbody fusion system intended for spinal fusion procedures in skeletally mature patients with cervical disc degeneration and/or cervical spinal instability, as confirmed by imaging studies (radiographs, CT, MRI), that results in radiculopathy, myelopathy, and/or pain at multiple contiguous levels from C2-T1. The IdentiTi NanoTec SPS Interbody System is intended for use with supplemental fixation systems. The system is designed for use with autograft, allograft comprised of cortical, cancellous, and/or corticocancellous bone graft, demineralized allograft with bone marrow aspirate, or a combination thereof.
## Transcend SPS Interbody System
The Transcend SPS Interbody System is an anterior cervical interbody fusion system intended for use in skeletally mature patients with cervical disc degeneration and/or cervical spinal instability, as confirmed by imaging studies (radiographs, CT, MRI), that results in radiculopathy, myelopathy, and/or pain at multiple contiguous levels from C2-T1. The Transcend SPS Interbody System is intended for use with supplemental fixation systems. The system is designed for use with autograft, allograft comprised of cortical, cancellous and/or corticocancellous bone graft, demineralized allograft with bone marrow aspirate, or a combination thereof.
# Transcend NanoTec SPS Interbody System
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for "atec". The "a" is green, while the "tec" is dark blue. There is a trademark symbol to the right of the "c". There is a black line underneath the logo.
The Transcend SPS PEEK Interbody System with advanced NanoTec surface treatment is an anterior cervical interbody fusion system intended for use in skeletally mature patients with cervical disc degeneration and/or cervical spinal instability, as confirmed by imaging studies (radiographs, CT, MRI), that results in radiculopathy, myelopathy, and/or pain at multiple contiguous levels from C2-T1. The Transcend NanoTec SPS Interbody System is intended for use with supplemental fixation systems. The system is designed for use with autograft, allograft comprised of cortical, cancellous and/or corticocancellous bone graft, demineralized allograft with bone marrow aspirate, or a combination thereof.
#### TECHNOLOGICAL COMPARISON TO PREDICATES VI.
The subject Segmental Plating System(SPS) is made with similar design, materials, and technological characteristics as those cleared in primary predicate Insignia Anterior Cervical Plate System (K213443) and additional predicates: Shoreline Threaded TruProfile Plate (K211903), and Pioneer SlimFuse Anterior Cervical Plate System (K072703).
The subject IdentiTi and Transcend SPS and IdentiTi and Transcend NanoTec SPS Interbody Systems' implants are a line extension to predicate IdentiTi and Transcend Interbody Systems (K222973). The subject interbody implants are similar in design and made of identical materials as the predicate IdentiTi and Transcend Cervical Interbody System (K222973) except the subject implants have scallop cuts on the anterior wall of the endplates.
The technological design features of the subject implants were compared to the predicates in intended use, indications for use, design, function, and technology and it was demonstrated that they are substantially equivalent.
#### VII. PERFORMANCE DATA
The following non-clinical testing was performed and included, where appropriate for the design, or referenced in predicate 510(k) submissions to support clearance of Segmental Plating and Interbody Systems:
ASTM F2077:2018 Test Methods for Intervertebral Body Fusion Devices
- Static Compression ●
- Dynamic Compression ●
- Static Torsion ●
- Dynamic Torsion ●
ASTM F2267:2004/(R)2018 Standard Test Method for Measuring Load Induced Subsidence of Intervertebral Body Fusion Device Under Static Axial Compression
- Subsidence ●
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for "atec". The "a" is green and the "tec" is dark blue. There is a trademark symbol to the right of the "c". There is a black line underneath the logo.
ASTM Draft Standard F 04.25.02.02 Static Push-out Test Method for Intervertebral Body Fusion Device
- . Expulsion
ASTM F1717:2021 Test Methods for Spinal Implant Constructs in a Vertebrectomy Model
- Static Compression
- Dynamic Compression
- Static Torsion ●
The results demonstrate that the subject Segmental Plating and Interbody Systems are substantially equivalent to other predicate devices for nonclinical testing.
## VIII. CONCLUSION
Based upon the information provided in this 510(k) submission, it has been determined that the subject device is substantially equivalent to legally marketed devices in regard to indications for use, intended use, design, technology, and performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.