The APERTO Lucent System is an imaging device, and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The MR system produces transverse, coronal, sagittal, oblique, and curved crosssectional images that display the internal structure of the head, body, or extremities. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination.
Device Story
The APERTO Lucent is a 0.4T permanent magnet MRI system used for non-invasive diagnostic imaging. It captures proton signals via RF coils, applying magnetic field gradients for spatial localization. The system reconstructs images based on T1, T2, proton density, and flow properties. Operated by trained clinicians in a clinical setting, the device provides cross-sectional images (transverse, coronal, sagittal, oblique, curved) to assist in diagnosis. It includes various pulse sequences and processing tasks (e.g., volume rendering, angiography, cardiac gating). The system supports DICOM networking for image storage and management. Clinical benefits include high-quality anatomical visualization without ionizing radiation. The device is a modification of the AIRIS Elite, featuring increased field strength (0.4T), updated gradient strength (22mT/m), and a Windows 10 IoT-based software platform.
Clinical Evidence
No formal clinical trial; clinical evidence consists of collected clinical image examples for new features and RF coils. A radiologist validated that the images provided have acceptable quality for clinical use.
Technological Characteristics
0.4T permanent magnet MRI system; 22mT/m gradient strength; 55 T/m/sec slew rate. Complies with NEMA MS 1-5, 8, 14 and IEC 60601-1, 60601-1-2, 60601-2-33, 62304, 62464-1. Connectivity via DICOM (Storage commitment, Modality Performed Procedure Step). Software runs on Windows 10 IoT. Includes various RF coils (QD Head, QD Flexible Body, Joint, QD Knee, QD Wrist, C-Spine, MA Shoulder).
Indications for Use
Indicated for diagnostic imaging of the head, body, spine, and extremities using proton excitation to provide T1, T2, proton density weighted imaging, diffusion weighted imaging, and MR Angiography.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
May 10, 2024
FUJIFILM Healthcare Corporation Kotei Aoki Manager, Regulatory Affairs 81 Hartwell Avenue, Suite 300 Lexington, Massachusetts 02421
Re: K233629
Trade/Device Name: APERTO Lucent MRI System Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: Class II Product Code: LNH Dated: April 2, 2024 Received: April 11, 2024
Dear Kotei Aoki:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Page 2
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
D.G.K.
Daniel M. Krainak, Ph.D Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
Submission Number (if known)
K233629
Device Name
#### APERTO Lucent MRI System
#### Indications for Use (Describe)
The APERTO Lucent System is an imaging device, and is intended to provide the physician with physiological and clinical information, obtained non invasively and without the use of ionizing radiation. The MR system produces transverse, coronal, sagittal, oblique, and curved cross sectional images that display the internal structure of the head, body, or extremities. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin lattice relaxation time (T1), spin spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination. Anatomical Region: Head, Body, Spine, Extremities Nucleus excited: Proton
Diagnostic uses:
- · T1, T2, proton density weighted imaging
- · Diffusion weighted imaging
- · MR Angiography
- · Image processing
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### Submitter Information
| Submitter: | FUJIFILM Healthcare Corporation |
|-------------------|------------------------------------|
| | 2-1, Shintoyofuta |
| | Kashiwa-Shi, Chiba, 277-0804 Japan |
| | Contact: |
| Telephone number: | 765-246-2931 |
| E-mail: | kotei.aoki@fujifilm.com |
| Date: | November 10, 2023 |
## Subject Device Name
| Trade/Proprietary Name: | APERTO Lucent MRI system |
|-------------------------|--------------------------------------------|
| Regulation Number: | 21 CFR 892.1000 |
| Regulation Name: | System, Nuclear Magnetic Resonance Imaging |
| Product Code: | LNH |
| Class: | 2 |
| Panel: | Radiology |
# Predicate Device Name
| Predicate Device(s): | AIRIS Elite V4.9 MRI system (K032232) |
|----------------------|--------------------------------------------|
| Regulation Number: | 21 CFR 892.1000 |
| Regulation Name: | System, Nuclear Magnetic Resonance Imaging |
| Product Code | LNH |
| Class | 2 |
| Panel | Radiology |
## Device Intended Use
The APERTO Lucent System is an imaging device, and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The MR system produces transverse, coronal, sagittal, oblique, and curved crosssectional images that display the internal structure of the head, body, or extremities. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination.
### Anatomical Region: Head, Body, Spine, Extremities
Nucleus excited: Proton
#### Diagnostic uses:
- T1, T2, proton density weighted imaging
- . Diffusion weighted imaging
- MR Angiography ●
- Image processing
{4}------------------------------------------------
## Device Description
### Function
The APERTO Lucent is a modification of the AIRIS Elite MRI System. The APERTO Lucent has been revised to increase the clinical utility as compared to the AIRIS Elite Magnetic Resonance imaging system.
### .Scientific Concepts
Magnetic Resonance imaging (MRI) is based on the fact that certain atomic nuclei have electromagnetic properties that cause them to act as small spinning bar magnets. The most ubiquitous of these nuclei is hydrogen, which makes it the primary nuclei currently used in magnetic resonance imaging. When placed in a static magnetic field, these nuclei assume a net orientation or alignment with the magnetic field, referred to as a net magnetization vector. The introduction of a short burst of radiofrequency (RF) excitation of a wavelength specific to the magnetic field strength and to the atomic nuclei under consideration can cause a re-orientation of the net magnetization vector. When the RF excitation is removed, the protons relax and return to their original vector. The rate of relaxation is exponential and varies with the character of the proton and its adjacent molecular environment. This re-orientation process is characterized by two exponential relaxation times, called T1 and T2. A RF emission or echo that can be measured accompanies these relaxation events.
The emissions are used to develop a representation of the relaxation events in a three dimensional matrix. Spatial localization is encoded into the echoes by varying the RF excitation, applying appropriate magnetic field gradients in the x, y, and z directions, and changing the direction and strength of these gradients. Images depicting the spatial distribution of the NMR characteristics can be reconstructed by using image processing techniques similar to those used in computed tomography.
### Physical and Performance Characteristics
MRI is capable of producing high quality anatomical images without the associated risks of ionizing radiation. The biological properties that contribute to MR image contrast are different from those responsible for x-ray image contrast. In MR imaging, difference in proton density, blood flow, and T1 and T2 relaxation times can all contribute to image contrast. By varying the pulse sequence characteristics, the resulting images can emphasize T1, T2, proton density, or the molecular diffusion of water or other proton containing molecules.
### Performance Evaluation
The APERTO Lucent MRI System is equivalent to the AIRIS Elite V4.9 (K032232) with following exceptions:
- Field strength is changed from 0.3T to 0.4T. ●
- . Gradient field strength is change from 21mT/m to 22mT/m.
- . Patient table
- Software operating system is changed to Windows 10 IoT from UNIX. .
A rationale analysis was then conducted, and the results are contained in Table 1.
| Testing Type | Rationale Analysis |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Performance Testing - Bench | Performance bench testing was conducted on the applicable new feature.<br>Test data confirmed that new feature perform as intended for diagnostic use. |
| Performance Testing -<br>Clinical | Clinical image examples are provided for applicable new feature and that we judged to be sufficient to evaluate clinical<br>usability. In addition, a radiologist validated that the clinical images have acceptable image quality for clinical use. |
#### Table 1 Performance Analysis
{5}------------------------------------------------
# Device Technological Characteristics
The control and image processing hardware and the base elements of the system software are identical to the predicate device. The APERTO Lucent MRI system software is substantially equivalent to the AIRIS Elite V4.9A (K0323232). See tables below. The technological characteristics in regard to hardware of the APERTO Lucent MRI system and the predicate are listed in Table 2.
| ITEM | | PREDICATE DEVICE | SUBJECT DEVICE | DIFFERENCE |
|----------------------|----------------------------|-----------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|------------|
| | | AIRIS ELITE(K032232) | APERTO LUCENT | |
| System | Standards Met | NEMA: MS 1, MS 2, MS 3, MS 4, MS 5, MS 8,<br>IEC: 60601-1, 60601-1-2, 60601-2-33, 62304 | NEMA: MS 1, MS 2, MS 3, MS 4, MS 5, MS 8, MS 14,<br>IEC: 60601-1, 60601-1-2, 60601-2-33, 62304 | Yes |
| Magnet and<br>Gantry | Type and Field<br>Strength | 0.3T Permanent magnet | 0.4T Permanent magnet | Yes |
| | Resonant Frequency | 12.7 MHz | 16.18MHz | Yes |
| Gradient System | Gradient Strength | 21mT/m | 22mT/m | Yes |
| | Slew Rate | 55 T/m/sec | 55 T/m/sec | No |
| | Rise Time | 380µsec to 21mT/m | 400µsec to 22mT/m | Yes |
| | Audible Noise (MCAN) | | | |
| | Ambient | - | 47.4 dBA | Yes |
| | Lpeak | - | 104.4 dB | Yes |
| | Leq | < 104dBA | 91.4 dBA | Yes |
| RF System | Transmitter channels | 4 | 4 | No |
| | Peak Envelop Power | 5 kW | 5 kW | No |
| | Duty Cycle | 15% square wave | 10% square wave | Yes |
| | RF receiver channel | 2 | 2 | No |
Table 2 Comparison: Hardware
The hardware differences from the predicate device to the APERTO Lucent MRI System are analyzed in Table 3.
#### Table 3 Hardware Comparison Analysis
| FDA<br>Requirements | Analyze why any differences between the subject device and predicate(s) do not render the device NSE (e.g., does not constitute a new intended use; and any differences in technological characteristics are accompanied by information that demonstrates the device is as safe and effective as the predicate and do not raise different questions of safety and effectiveness than the predicate), affect safety or effectiveness, or raise different questions of safety and effectiveness (see section 513(i)(1)(A) of the FD&C Act and 21 CFR 807.87(f)). | | | | |
|------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|--------------|----------------------------------------------------------------------------------|--|
| Device<br>Modification<br>Summary | • Field strength is changed from 0.3T to 0.4T.<br>• Resonant Frequency is changed from 12.7MHz to 16.18MHz.<br>• Gradient field strength is changed from 21mT/m to 22mT/m.<br>• Audible Noise is re-measured.<br>• Duty Cycle is changed from 15% to 10%. | | | | |
| Significant<br>Changes | □ Manufacturing Process | □ Labeling | □ Technology | □ Performance | |
| | □ Engineering | □ Materials | □ Others | <span style="text-decoration: overline;">☑</span> None (See rationale statement) | |
| FUJIFILM<br>Rationale<br>Statement | Modified specification doesn't constitute a new intended use. There are no significant changes in technological characteristics. For safety, gradient system and RF system is controlled according to same regulation as AIRIS Elite. So, safety and effectively of the device are same as AIRIS Elite(K032232). | | | | |
{6}------------------------------------------------
The technological characteristics in regards to coils of the APERTO Lucent MRI System and the predicate are listed in Table 4.
| Table 4 Comparison: RF Coils | | | | |
|------------------------------|------------------------------------------|-------------------------------------|-----------------------------|-----|
| ITEM | PREDICATE DEVICE<br>AIRIS ELITE(K032232) | SUBJECT DEVICE<br>APERTO LUCENT | DIFFERENCE | |
| RF Coils | Receiver Coils | Open MA head coil | QD Head coil R | Yes |
| | | MA Flexible Body coil (Medium) | N/A | N/A |
| | | MA Flexible Body coil (Large) | QD Flexible Body coil N (L) | Yes |
| | | MA Flexible Body coil (Extra Large) | N/A | N/A |
| | | Medium extremity(neck) coil | Joint coil(M) | Yes |
| | | Large extremity(neck) coil | Joint coil(L) | Yes |
| | | MA Knee coil | QD Knee coil R | Yes |
| | | TMJ coil | N/A | N/A |
| | | MA Wrist coil | QD Wrist coil | Yes |
| | | N/A | C-Spine coil | N/A |
| | | N/A | QD Head coil (L) | N/A |
| | | N/A | MA Shoulder coil | N/A |
Table 4 Comparison: RF Coils
The coil differences from the predicate device to the APERTO Lucent MRI system are analyzed in Table 5.
Table 5 Coil Comparison Analysis
| FDA<br>Requirements | Analyze why any differences between the subject device and predicate(s) do not render the device NSE (e.g., does not<br>constitute a new intended use; and any differences in technological characteristics are accompanied by information that<br>demonstrates the device is as safe and effective as the predicate and do not raise different questions of safety and<br>effectiveness than the predicate ), affect safety or effectiveness, or raise different questions of safety and effectiveness<br>(see section 513(i)(1)(A) of the FD&C Act and 21 CFR 807.87(f)). | | | |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|--------------|----------------------------------|
| Device<br>Modification<br>Summary | Head Coil is changed from Open MA head coil to QD Head coil R.<br>Flexible Body coil is changed from MA Flexible Body coil(Large) to QD Flexible Body coil N(L).<br>Joint coil is changed Medium extremity(neck) coil to Joint coil(M).<br>Joint coil is changed Large extremity(neck) coil to Joint coil(L).<br>Knee coil is changed MA Knee coil to QD Knee coil R.<br>Wrist coil is changed MA Wrist coil to QD Wrist coil.<br>Deleted the following coils.<br>- MA Flexible Body coil (Medium)<br>- MA Flexible Body coil (Extra Large)<br>- TMJ coil<br>Added the following coils.<br>- C-Spine coil<br>- QD Head coil (L)<br>- MA Shoulder coil | | | |
| Significant<br>Changes | ☐ Manufacturing Process | ☐ Labeling | ☐ Technology | ☐ Performance |
| | ☐ Engineering | ☐ Materials | ☐ Others | ☑ None (See rationale statement) |
| FUJIFILM<br>Rationale<br>Statement | Revised coils do not constitute a new intended use. There are no significant changes in technological characteristics.<br>During transmitter coil operation, RF Coils are de-resonated for the safety function by same scheme as AIRIS<br>Elite(K032232) | | | |
{7}------------------------------------------------
The technological characteristics in regard to changes in functionality of the APERTO Lucent MRI System as compared to the predicate are listed in Table 6.
| ITEM | DIFFERENCES | ANALYSIS |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Operating System | Going from UNIX to Windows 10 IoT | See Table 7 |
| CPU Platform | Xeon E3-1275 v6 3.8GHz | See Table 7 |
| Application Software | Going from V4.9 to V7.1E | See Table 7 |
| Scan Tasks | Following Scan Tasks are added.<br>- RAPID is not available.<br>- Interactive Scan Control<br>- Radial Stack<br>- Auto Pose (Brain) | See Table 7 |
| 2D Processing Tasks | Following 2D Processing Tasks are added.<br>- Uniformity is not available.<br>- Image division<br>- Image Stitching<br>Add the parameter R2, R1 in Parameter Analysis | See Table 7 |
| 3D Processing Tasks | Following 3D Processing Tasks are added.<br>- Volume Rendering | See Table 7 |
| Analysis Tasks | None | No |
| Maintenance Tasks | None | No |
| Viewport Tools | Following Viewport Tools are added.<br>- SIR Map | See Table 7 |
| Film, Archive Tools | Following Film Tools are added.<br>- Auto filming is not available. | See Table 7 |
| Network Tools | Following Network Tool are added<br>- DICOM Modality Performed Procedure Step<br>- DICOM Storage commitment | See Table 7 |
| Protocol Enhancements | Following protocol enhancement are added.<br>- Multi-Slice/Multi-phase Cardiac Gating Navigator Echo Functions<br>- Time Resolved Angiography<br>- Fluoro Triggered Angiography<br>- Black blood imaging<br>- Arrhythmia rejection<br>- 3DGEIR<br>- Blood Sensitive Imaging (BSI)<br>- RADAR (2D FSE, 3D FSE, 2D FIR, 3D FIR, 2D SE, 2D GE, 3D GE)<br>- IP-Recon<br>- IP-Scan<br>- VASC-ASL | See Table 7 |
| Pulse Sequences | Following Pulse sequence are added.<br>- 2D Inversion Recovery(2D IR) is not available. (*1)<br>- 2D Time Reversed SARGE (2D TRSG) is not available.<br>- 3D Time Reversed SARGE (3D TRSG) is not available.<br>- 2D Fast Spin Echo (2D primeFSE)<br>- 3D Fast Spin Echo (3D primeFSE)<br>- 2D Fast Inversion Recovery (2D primeFIR)<br>- 3D Fast Inversion Recovery (3D primeFIR)<br>- 2D RADAR DW Fast Spin Echo (2D DW FSE)<br>- 2D Fast Spin Echo Fat Water Separation (2D FATSEPFSE) | See Table 7 |
| Powered by Machine<br>Learning | None | No |
Table 6 Comparison: Functionality
The functionality differences from the predicate device from the APERTO Lucent MRI System are analyzed in Table 7.
#### Table 7 Functionality Comparison Analysis
| FDA | Analyze why any differences between the subject device and predicate(s) do not render the device NSE (e.g., does not |
|--------------|-------------------------------------------------------------------------------------------------------------------------|
| Requirements | constitute a new intended use; and any differences in technological characteristics are accompanied by information that |
{8}------------------------------------------------
| | demonstrates the device is as safe and effective as the predicate and do not raise different questions of safety and<br>effectiveness than the predicate ), affect safety or effectiveness, or raise different questions of safety and effectiveness<br>(see section 513(i)(1)(A) of the FD&C Act and 21 CFR 807.87(f)). | | | |
|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------|------------|-------------|
| Device<br>Modification<br>Summary | Application software is changed in V7.1E.<br>Following Scan Tasks are added.<br>- RAPID is not available.<br>- Interactive Scan Control<br>- Radial Stack<br>- Auto Pose (Brain)<br>Following 2D Processing Tasks are added.<br>- Uniformity is not available.<br>- Image division<br>- Image Stitching<br>Add the parameter R2, R1 in Parameter Analysis as 2D Processing.<br><br>Following 3D Processing Tasks are added.<br>- Volume Rendering<br>Following Viewport Tools are added.<br>- SIR Map<br>Following Film Tools are added.<br>- Auto filming is not available.<br>Following Network Tools are added.<br>- DICOM Modality Performed Procedure Step<br>- DICOM Storage commitment<br>Following protocol enhancement are added.<br>- Multi-Slice/Multi-phase Cardiac Gating Navigator Echo Functions<br>- Time Resolved Angiography<br>- Fluoro Triggered Angiography<br>- Black blood imaging<br>- Arrhythmia rejection<br>- 3DGEIR<br>- Blood Sensitive Imaging (BSI)<br>- RADAR (2D FSE, 3D FSE, 2D FIR, 3D FIR, 2D SE, 2D GE, 3D GE)<br>- IP-Recon<br>- IP-Scan<br>- VASC-ASL<br>Following Pulse sequence are added.<br>- 2D Inversion Recovery(2D IR) is not available. (*1)<br>- 2D Time Reversed SARGE (2D TRSG) is not available.<br>- 3D Time Reversed SARGE (3D TRSG) is not available.<br>- 2D Fast Spin Echo (2D primeFSE)<br>- 3D Fast Spin Echo (3D primeFSE)<br>- 2D Fast Inversion Recovery (2D primeFIR)<br>- 3D Fast Inversion Recovery (3D primeFIR)<br>- 2D RADAR DW Fast Spin Echo (2D DW FSE)<br>- 2D Fast Spin Echo Fat Water Separation (2D FATSEPFSE)<br>*1 2D IR sequence is integrated to SE using with IR-pulse. | | | |
| Significant<br>Changes | Manufacturing Process | Labeling | Technology | Performance |
{9}------------------------------------------------
| FUJIFILM<br>Rationale<br>Statement | Modified specification doesn't constitute a new intended use. There are no significant changes in technological<br>characteristics. So, safety and effectiveness of the device are same as AIRIS Elite (K032232). |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
{10}------------------------------------------------
# Substantial Equivalence
A summary decision was based on analysis of Table 8.
| Table 8 Rationale Analysis: APERTO Lucent vs. Predicate | |
|---------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ITEM | Overall Rationale Analysis |
| Hardware | Modified specification doesn't constitute a new intended use. There are no significant changes in technological<br>characteristics. For safety, gradient system and RF system is controlled according to same regulation as AIRIS Elite<br>V4.9A(K032232). So, safety and effectively of the device are same as AIRIS Elite V4.9A(K032232). |
| Coils | Revised coils do not constitute a new intended use. There are no significant changes in technological characteristics. |
| Functionality | Additional functions do not constitute a new intended use. There are no significant changes in technological<br>characteristics. So, safety and effectivity of the device are equivalent to AIRIS Elite (K032232). |
Therefore, based on a thorough analysis and comparison of the functions, scientific concepts, physical and performance characteristics, performance comparison and technological characteristics, the proposed APERTO Lucent is considered substantially equivalent to the currently marketed predicate device (AIRIS Elite V4.9A MRI System (K032232)) in terms of design features, fundamental scientific technology, indications for use, and safety and effectiveness.
{11}------------------------------------------------
# Summary of Non-Clinical Testing
The APERTO Lucent MRI System was subjected to the following laboratory testing.
- IEC60601-1:2005 + CORR. 1:2006 + CORR. 2:2007 + AM:2012. Medical electrical equipment -. part 1: general requirements for basic safety and essential performance
- IEC 60601-2-33 Edition 3.2 b:2015. medical electrical equipment part 2-33: particular . requirements for the basic safety and essential performance of magnetic resonance equipment for medical diagnostic.
- IEC 62304 Edition 1.1 2015-06, CONSOLIDATED VERSION medical device software software . life cycle processes.
- NEMA MS 1-2008, Determination of Signal-to-noise Ratio (SNR) in Diagnostic Magnetic . Resonance Images
- . NEMA MS 2-2008, Determination of Two-Dimensional Geometric Distortion in Diagnostic Magnetic Resonance Images
- . NEMA MS 3-2008, Determination of Image Uniformity in Diagnostic Magnetic Resonance Images
- . NEMA MS 4-2010, Acoustic Noise Measurement Procedure for Diagnostic Magnetic Resonance Imaging Devices
- NEMA MS 5-2018, Determination of Slice Thickness in Diagnostic Resonance Imaging
- NEMA MS 8-2016, Characterization of the Specific Absorption Rate for Magnetic Resonance . Imaging Systems
- . NEMA MS 14-2019, Characterization of Radiofrequency (RF) Coil Heating in Magnetic Resonance Imaging Systems
- . IEC 60601-1-2 Edition 4.0:2014, medical electrical equipment - part 1-2: general requirements for basic safety and essential performance - collateral standard: electromagnetic disturbances requirements and tests.
- . IEC 62464-1 Edition 2.0 2018-12 Magnetic resonance equipment for medical imaging Part1: Determination of many essential image quality
# Summary of Clinical Testing
Clinical images were collected and analyzed, to ensure that images from the new feature and three new RF coils meet user needs.
As a result of the analysis:
| Testing Type | Rationale Analysis |
|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Performance Testing -<br>Clinical | Clinical image examples are provided for applicable new feature and three new RF coils, and that we judged to be sufficient<br>to evaluate clinical usability. In addition, a radiologist validated that the clinical images have acceptable image quality for<br>clinical use. |
# Conclusions
It is the opinion of FUJIFILM, the APERTO Lucent, MRI system is substantially equivalent with respect to hardware, base elements of the software, safety, effectiveness, and functionality to the AIRIS Elite V4.9A MRI System (K032332).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.