Clevaligner Software (V1.0.0)

K233616 · Clevaligner, Ltd. · PNN · Jan 11, 2024 · Dental

Device Facts

Record IDK233616
Device NameClevaligner Software (V1.0.0)
ApplicantClevaligner, Ltd.
Product CodePNN · Dental
Decision DateJan 11, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.5470
Device ClassClass 2
AttributesSoftware as a Medical Device

Intended Use

The Clevaligner software is intended for use as a medical device standalone software, providing tools for management of orthodontic models, systematic inspection, detailed analysis, treatment simulation and virtual design of a series of dental casts, which may be used for sequential aligner trays or retainers, based on 3D models of the patient's dentition before the start of an orthodontic treatment. It can also be applied during the treatment to inspect and analyze the progress of the treatment. It can be used at the end of the treatment to evaluate if the outcome is consistent with the planned/desired treatment objectives. The use of Clevaligner Software requires the user to have the necessary training and domain knowledge in the practice of orthodontics, as well to have received a dedicated training in the use of the software.

Device Story

Clevaligner Software is a standalone orthodontic planning tool used by dental professionals in clinical settings. It imports 3D digital scans (STL, OBJ, PLY) and 2D overlays (PNG, JPG) of patient dentition. The software performs automatic segmentation of 3D scans and automatic design of ideal arch forms, which are reviewed and approved by orthodontists via a 2D/3D interface. It provides tools for arch shape analysis, tooth width, space analysis, overjet/overbite, and occlusion mapping. The software generates a treatment path from initial to final stage, which can be exported for fabrication of orthodontic appliances (aligners/retainers). By enabling virtual treatment simulation and design, it assists clinicians in planning and monitoring orthodontic progress, potentially improving treatment outcomes through precise, customized appliance design.

Clinical Evidence

No clinical data. Bench testing only. Software verification and validation were performed per IEC 62304 and FDA guidance. Testing confirmed that all user needs and performance requirements were met, including cybersecurity analysis. An aligner manufacturing validation was conducted to ensure that treatment plans generated by the software match the accuracy of 3D-printed/thermoformed aligner trays.

Technological Characteristics

Standalone software; Windows 10/Mac OS compatible. Supports STL, OBJ, JPG, PNG, PLY file formats. Features automatic segmentation and arch form design algorithms. Operates via 2D/3D interface for orthodontic analysis and CAD design. Cybersecurity controls implemented per FDA guidance. Software documentation level: Basic.

Indications for Use

Indicated for patients with malocclusion requiring orthodontic treatment planning, management, inspection, measurement, analysis, and virtual design of dental casts for sequential aligner trays or retainers.

Regulatory Classification

Identification

An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath in a smaller font. January 11, 2024 Clevaligner Ltd % Ora Msika QA/RA Manager Rs Ness Hamelacha street 45 Netanya, 4250574 Israel Re: K233616 Trade/Device Name: Clevaligner Software (V1.0.0) Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: PNN. LLZ Dated: November 10, 2023 Received: November 13, 2023 Dear Ora Msika: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" {1}------------------------------------------------ (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Bobak Shirmohammadi -S For Michael E. Adjodha, M. ChE., RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices {2}------------------------------------------------ OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use Submission Number (if known) K233616 Device Name Clevaligner Software (V1.0.0) #### Indications for Use (Describe) The Clevaligner software is intended for use as a medical device standalone software, providing tools for management of orthodontic models, systematic inspection, detailed analysis, treatment simulation and virtual design of a series of dental casts, which may be used for sequential aligner trays or retainers, based on 3D models of the patient's dentition before the start of an orthodontic treatment. It can also be applied during the treatment to inspect and analyze the progress of the treatment. It can be used at the end of the treatment to evaluate if the outcome is consistent with the planned/desired treatment objectives. The use of Clevaligner Software requires the user to have the necessary training and domain knowledge in the practice of orthodontics, as well to have received a dedicated training in the use of the software. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ #### K233616 ### 510(k) SUMMARY In accordance with 21 C.F.R. § 807.92 the following summary of information is provided: DATE: November 10, 2023 #### SUBMITTER - Clevaligner LTD King George St 20 9426208 Jerusalem Israel ## CONTACT PERSON Jonathan Tabet Chief Technology Officer (СТО) Email: jonathan@clevaligner.com King George St 20 9426208 Jerusalem Israel T: +972-534719048 #### DEVICE PROPRIETARY NAME: Clevaligner Software V1.0.0 COMMON NAME: Orthodontic Software CLASSIFICATION REGULATION NAMES: Orthodontic plastic bracket CLASSIFICATION REGULATION: 21 C.F.R. § 872.5470 DEVICE CLASS: II PRIMARY PRODUCT CODE: PNN: Orthodontic Software (Under 21 C.F.R. 872.5470) ADDITIONAL PRODUCT CODE: LLZ: System, Image Processing, Radiological (under 21 C.F.R. 892.2050) Page 1 of 7 {5}------------------------------------------------ For the planned and proposed 510(k), the Clevaligner Software will have a primary predicate device as SoftSmile, Inc. Vision (K212770). ## DEVICE DESCRIPTION Clevaligner Software imports patient 3-D digital scans and provides the orthodontic treatment planning of the patient under dental professional supervision. The Clevaligner Software performs the automatic segmentation of the 3-D digital scans, achieves an automatic design of an ideal arch form, which is approved by orthodontists using a 2D software interface. The Clevaligner Software provides an initial to ideal final stage treatment plan, with each step of the treatment plan path presented in the 3D software interface. Every 3D digital model from the path treatment plan, generated by the Clevaligner Software, can be exported for fabrication of orthodontic appliances, either to an orthodontic laboratory or directly to orthodontic appliance manufacturers for further use in orthodontic treatment. ### INDICATIONS FOR USE Clevaligner Software is intended for use as a medical device standalone software, providing tools for management of orthodontic models, systematic inspection, detailed analysis, treatment simulation and virtual design of a series of dental casts, which may be used for sequential aligner trays or retainers, based on 3D models of the patient's dentition before the start of an orthodontic treatment. It can also be applied during the treatment to inspect and analyze the progress of the treatment. It can be used at the end of the treatment to evaluate if the outcome is consistent with the planned/desired treatment objectives. The use of Clevaligner Software requires the user to have the necessary training and domain knowledge in the practice of orthodontics, as well to have received a dedicated training in the use of the software. ## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE The proposed Clevaligner Software has the same indication for use as the legally marketed predicate device to which substantial equivalency is claimed, the SoftSmile, {6}------------------------------------------------ Inc. Vision (K212770). The proposed Clevaligner Software has identical intended use, operational and functional features as the cleared software, SoftSmile, Inc. Vision (K212770). Clevaligner Software uses the same fundamental technology as the legally marketed predicate software, to which substantial equivalency is claimed, the SoftSmile, Inc. Vision. The subject device supports in addition PLY file format whereas the predicate does not. Table 1 Substantial Equivalence Justification: comparison of technological characteristics with the predicate device | | Clevaligner Software | SoftSmile, Inc. Vision (K212770)<br>(Primary Predicate) | |--------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indication for use | The Clevaligner Software is intended<br>for use as a medical device<br>standalone, providing tools for<br>management of orthodontic models,<br>systematic inspection, detailed<br>analysis, treatment simulation and<br>virtual design of a series of dental<br>casts, which may be used for<br>sequential aligner trays or retainers,<br>based on 3D models of the patient's<br>dentition before the start of an<br>orthodontic treatment. It can also be<br>applied during the treatment to inspect<br>and analyze the progress of the<br>treatment. It can be used at the end of<br>the treatment to evaluate if the<br>outcome is consistent with the<br>planned/desired treatment objectives.<br>The use of Clevaligner Software<br>requires the user to have the<br>necessary training and domain<br>knowledge in the practice of<br>orthodontics, as well to have received<br>a dedicated training in the use of the<br>software. | The SoftSmile Vision is intended for<br>use as a medical front-end device<br>providing tools for management of<br>orthodontic models, systematic<br>inspection, detailed analysis,<br>treatment simulation and virtual<br>design of a series of dental casts,<br>which may be used for sequential<br>aligner trays or retainers, based on<br>3D models of the patient's dentition<br>before the start of an orthodontic<br>treatment. It can also be applied<br>during the treatment to inspect and<br>analyze the progress of the<br>treatment. It can be used at the end<br>of the treatment to evaluate if the<br>outcome is consistent with the<br>planned/desired treatment<br>objectives. The use of SoftSmile<br>Vision requires the user to have the<br>necessary training and domain<br>knowledge in the practice of<br>orthodontics, as well to have<br>received a dedicated training in the<br>use of the software. | | Intended use | Used by dental professionals in<br>orthodontic treatment planning<br>(before, during, and after treatment)<br>• Management of patients and models<br>• Inspection, measurement, and<br>analysis of orthodontic models<br>• Treatment simulation | Used by dental professionals in<br>orthodontic treatment planning<br>(before, during, and after treatment)<br>• Management of patients and models<br>• Inspection, measurement,<br>and analysis of orthodontic models<br>• Treatment simulation | | | Clevaligner Software | SoftSmile, Inc. Vision (K212770)<br>(Primary Predicate) | | | Virtual preparation of dental casts handling and export Provides digital file | Virtual preparation of dental casts, handling and export Provides digital file | | Software Environment of Use | Dental office | Dental office | | Software Intended User | Dental professional | Dental professional | | Intended Patient Population | Patients with malocclusion | Patients with malocclusion | | Target Anatomic Area | Maxilla, Mandible | Maxilla, Mandible | | Type of Patient Contact | None | None | | Principle of Operation | Apply digital imaging tools for use in orthodontic case archiving, treatment planning and CAD design of customized appliances. The system supports the following types of digital data: STL, OBJ, JPG, PNG and PLY. The subject device supports PLY file format whereas the predicate does not. | Apply digital imaging tools for use in orthodontic case archiving, treatment planning and CAD design of customized appliances. The system supports the following types of digital data: STL, OBJ, JPG, BMP, PNG. | | Technical attributes | Minimum Requirements: Operating System (OS): Windows 10, 64-bit, Mac OS 10 Browser system: Chrome Version 110, Safari 16 RAM: 8 GB Monitor Resolution: 1920x1440 or similar Graphic Card Memory: Video Card Memory: 2 GB or more discrete graphics card Available HDD Space: 80 GB CPU: IntelCore i3, AMD FX4300 or higher Network:8MB/s | Minimum Requirements: Supported PC formats: Windows 10 64-bit RAM: 4 GB Monitor Resolution: 1280x800 or similar Video Card Memory: 2 GB or more discrete graphics card Available HDD Space: 120 GB or more CPU: IntelCore i3, AMD FX-4300 or higher | | Management of patient/case base Data | Allows creating, editing, deleting, copying patient/case data | Allows creating, editing, deleting, copying patient/case data | | | Clevaligner Software | SoftSmile, Inc. Vision (K212770)<br>(Primary Predicate) | | Collection of Input | • Surface scan for intra-oral scanner<br>• Surface scan from STL or OBJ file<br>• 2D overlay: PNG, JPG | • Surface scan for intra-oral scanner<br>• Surface scan from STL or OBJ file<br>• 2D overlay: PNG, JPG, BMP | | Alignment of Input | • Aligning surface scan image<br>• Alignment of 2D overlays (e.g., ideal arch) | • Aligning surface scan image<br>• Alignment of 2D overlays (e.g., ideal arch) | | Measurement of Input | 3D measurement toolbox | 3D measurement toolbox | | Analysis of Input | • Arch shape<br>• Tooth width<br>• Space analysis<br>• Overjet/overbite<br>• Occlusion map | • Arch shape<br>• Tooth width<br>• Bolton<br>• Space analysis<br>• Overjet/overbite<br>• Occlusion map | | Treatment simulation | 3D simulation | 3D simulation | | Virtual appliance design | • Orthodontic dental cast virtual preparation<br>• Orthodontic dental cast design<br>• Orthodontic dental cast export | • Orthodontic dental cast search<br>• Orthodontic dental cast virtual preparation<br>• Orthodontic dental cast design<br>• Orthodontic dental cast export | | Virtual appliance options | Dental casts | Dental casts | {7}------------------------------------------------ {8}------------------------------------------------ {9}------------------------------------------------ ### PERFORMANCE DATA ### NON-CLINICAL TESTS Software testing has been conducted in accordance with the software life cvcle processes, as defined in IEC 62304. Software and integration verification and validation testing were performed in accordance with the FDA Guidance Document "General Principles of Software Validation; Final Guidance for Industry (issued June 11, 2002). Software verification and validation testing was conducted to ensure that all user needs and performance requirements according to the design inputs are fulfilled. Software verification and validation was conducted to ensure the functionality of the software and to support the safety and effectiveness of the proposed software. The verification and validation of the software has been completed prior to release of the Clevaligner Software. The testing includes validation of implemented mitigations, related to device hazards identified in the risk management procedures. All test results met acceptance criteria, demonstrating Clevaligner Software performs as intended, raises no new or different questions of safety or effectiveness and is substantially equivalent to the predicate device. The described approach for testing also includes testing of cybersecurity requirements determined in the cybersecurity analysis. A cybersecurity analysis was performed in accordance with the FDA Guidance Document "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions", issued on September 27, 2023. Clevaligner Software monitors vulnerabilities of the software during lifetime and analyses them in the risk management. The documentation level for the software was assessed per the FDA Guidance for the Content of Premarket Submissions for Device Software Functions" issued June 14, 2023) and determined to be Basic Documentation level, based on the risks of the software functions in the context of the software intended use. The test data to verify the performance of Clevaligner Software have been provided. where the results of this testing, combined with the comparison to the predicate device, support substantial equivalence and do not raise any new issues of safety or effectiveness. All test results have been reviewed and approved, demonstrating Clevaligner is substantially equivalent in terms of safety and effectiveness to the predicate software. ### CLINICAL TESTING Clinical testing has not been performed and is not required since the proposed device, Clevaligner Software, is a stand-alone medical device software which is used without direct patient contact. However, an aligner manufacturing validation was performed, in order to demonstrate that the digital aligners treatment planned proposed by the Clevaligner Software, matches the fabrication {10}------------------------------------------------ of aligner trays or retainers 3D printed from subsequent thermoforming steps with an acceptable level of accuracy. # CONCLUSIONS The proposed Clevaligner Software is intended to provide virtual design of dental casts, which may be used for sequential aligner trays or retainers such as the predicate device. Both devices create customized dental appliances based on computer models of the patient's pre-treatment and post-treatment dentition.The minor technological differences between this device and its predicate device raise no new issues of safety or effectiveness. Thus, the Clevaligner Software is found to be substantially equivalent to in terms of safety and effectiveness to the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...