K233566 · Orthocon, Inc. · MQV · Aug 21, 2024 · Orthopedic
Device Facts
Record ID
K233566
Device Name
Montage-XT Settable, Resorbable Bone Putty
Applicant
Orthocon, Inc.
Product Code
MQV · Orthopedic
Decision Date
Aug 21, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3045
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Orthocon Montage-XT Settable, Resorbable Bone Putty is indicated to fill bony voids or gaps in the skeletal system (i.e., extremities and pelvis). These defects may be surgically created, or osseous defects created as the result of traumatic injury to the bone. Montage-XT is indicated only for filling bony voids or gaps that are not intrinsic to the integrity of the bony structure. When hardened in situ, Montage-XT may be used to augment provisional hardware (e.g., k-wires, plates and screws) and to help support bone fragments during the surgical procedure. The hardened putty acts only as a temporary support medium and is not intended to provide structural support during the healing process. Montage-XT can be drilled and tapped, and hardware can be placed through it at any time during the setting process.
Device Story
Sterile, resorbable, two-part putty-like material; composed of granular calcium phosphate, calcium stearate, vitamin E acetate, triglyceride, polyalcohol, and lactide-diester/polyester-based polymers. Mixed manually at time of surgery to form cohesive, settable material. Applied by surgeon to bone defects in extremities/pelvis. Hardens in situ to act as temporary support medium for provisional hardware (k-wires, plates, screws) and bone fragments. Can be drilled/tapped during setting. Resorbs over time via dissolution, hydrolysis, and cellular removal. Benefits patient by filling voids and providing temporary support during surgical procedures.
Clinical Evidence
No clinical data. Bench testing verified handling properties (pH, dissolution, reaction temperature, stiffness, spreadability, stickiness, electrocautery compatibility, working time). Biocompatibility testing performed per ISO 10993 (irritation, sensitization, toxicity, genotoxicity, implantation, hemolysis, pyrogenicity). Animal study (rabbit critical-sized femoral defect) showed new bone formation (20.4% for Montage-XT vs 20.8% for predicate) and similar resorption profiles at 12 weeks.
Technological Characteristics
Two-part putty; components: granular calcium phosphate (hydroxyapatite/beta-tricalcium phosphate), calcium stearate, vitamin E acetate, triglyceride, polyalcohol, lactide-diester/polyester-based polymers. Sterile (gamma irradiation). Resorbable. Sets in situ without significant exothermic reaction. 4-minute working time. Biocompatible per ISO 10993.
Indications for Use
Indicated for filling bony voids or gaps in the skeletal system (extremities and pelvis) resulting from surgery or traumatic injury. Not for use in defects intrinsic to bony structure integrity. Used to augment provisional hardware and support bone fragments during surgery; not for structural support during healing.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
MONTAGE-XT Settable, Resorbable Hemostatic Bone Putty (K220315)
Submission Summary (Full Text)
{0}------------------------------------------------
August 21, 2024
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Orthocon, Inc. Howard Schrayer Regulatory Consultant 700 Fairfield Ave. - Suite 1 Stamford, Connecticut 06902
Re: K233566
Trade/Device Name: Montage-XT Settable, Resorbable Bone Putty Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable Calcium Salt Bone Void Filler Device Regulatory Class: Class II Product Code: MQV, OIS Dated: July 3, 2024 Received: July 3, 2024
Dear Howard Schrayer:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, JESSE MUIR Digitally signed by JESSE
> MUIR-S Date: 2024.08.21 -2 10:28:43 -04'00' Jesse Muir, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# K233566 - Howard Schrayer
Enclosure
{3}------------------------------------------------
## Indications for Use
Submission Number (if known)
K233566
Device Name
Montage-XT Settable, Resorbable Bone Putty
Indications for Use (Describe)
Orthocon Montage-XT Settable, Resorbable Bone Putty is indicated to fill bony voids or gaps in the skeletal system (i.e., extremities and pelvis). These defects may be surgically created, or osseous defects created as the result of traumatic injury to the bone. Montage-XT is indicated only for filling bony voids or gaps that are not intrinsic to the integrity of the bony structure.
When hardened in situ, Montage-XT may be used to augment provisional hardware (e.g., k-wires, plates and screws) and to help support bone fragments during the surgical procedure. The hardened putty acts only as a temporary support medium and is not intended to provide structural support during the healing process.
Montage-XT can be drilled and tapped, and hardware can be placed through it at any time during the setting process.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
ver-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### 510(k) Summary K233566
| Contact: | Howard Schrayer<br>Orthocon, Inc.<br>8 Lookout<br>Hilton Head Island, SC 29928<br>Telephone: 609-273-7350<br>hs.ss@lucidmedical.net |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared: | August 20, 2024 |
| Device Trade Name: | MONTAGE-XT™ Settable, Resorbable Bone Putty |
| Manufacturer: | Orthocon, Inc.<br>700 Fairfield Avenue, Suite 1<br>Stamford, CT 10533 |
| Telephone: | (855) 475 - 9175 |
| Common Name: | Resorbable calcium salt bone void filler device |
| Classification: | Class II |
| Product Code: | MQV, OIS |
| Primary Predicate Device: | |
- MONTAGE® Settable, Resorbable Bone Putty Orthocon, Inc. 510(k) K222063
### Reference Predicate Device:
- MONTAGE-XT Settable, Resorbable Hemostatic Bone Putty Orthocon, Inc. 510(k) K220315
### Indications for Use:
Orthocon Montage-XT Settable, Resorbable Bone Putty is indicated to fill bony voids or gaps in the skeletal system (i.e., extremities and pelvis). These defects may be surgically created, or osseous defects created as the result of traumatic injury to the bone. Montage-XT is indicated only for filling bony voids or gaps that are not intrinsic to the integrity of the bony structure.
When hardened in situ. Montage-XT may be used to augment provisional hardware (e.g., k-wires, plates and screws) and to help support bone fragments during the
{5}------------------------------------------------
surqical procedure. The hardened putty acts only as a temporary support medium and is not intended to provide structural support during the healing process.
Montage-XT can be drilled and tapped, and hardware can be placed through it at any time during the setting process.
## Device Description:
Montage-XT Settable, Resorbable Bone Putty is a sterile, biocompatible, resorbable material for use in filling bony voids or gaps in skeletal bones of the extremities. The single use Montage-XT device contains two separate components of putty-like consistency containing granular calcium phosphate, calcium stearate, vitamin E acetate, a triglyceride, a polyalcohol and a mixture of a lactide-diester and polyesterbased polymers. When mixed together, the components of Montage-XT form a cohesive putty-like material that adheres to the bone surface and remains in place following application. The resulting hardened material is primarily calcium phosphate. Montage-XT can be drilled and tapped, and hardware can be placed through it at any time during the setting process.
Montage-XT differs from Montage by allowing for an extended working time of up to 4 minutes. The performance of Montage-XT was compared to Montage in a rabbit critical sized femoral defect model. At the 12-week timepoint, animal study data demonstrated new bone formation averages of 20.4% in the Montage-XT group, 20.8% in the Montage predicate group, and less than 10% in the empty defect negative control group. Animal study data show that approximately 70% of implant material remained in both the Montage-XT group and the Montage group at 12 weeks following implantation.
### Purpose of this Submission:
Montage-XT was previously cleared for use in the control of bleeding from cut or damaged bone by acting as a mechanical barrier or tamponade. The purpose of this submission is to add an indication for Montage-XT for use as a bone void filler device.
### Substantial Equivalence and Predicate Devices:
The subject device is nearly identical to and is substantially equivalent to its previously cleared primary predicate, Montage. The only difference between the subject device and its reference predicate is the addition of a statement regarding the use of the device as a bone void filler.
The following table shows comparisons of the several characteristics of Montage-XT and the primary predicate device. The differences noted in the table below do not impact substantial equivalence.
{6}------------------------------------------------
### SUBSTANTIAL EQUIVALENCE INFORMATION
### Orthocon, Inc. Montage-XT Settable, Resorbable Bone Putty
#### Orthocon, Inc. Montage Settable, Resorbable Bone Putty
### 510(k) - 233566
#### 510(k) - K222063
### Similarities and Differences
| Orthocon Montage-XT Settable, | Orthocon Montage Settable, |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Resorbable Bone Putty is indicated to | Resorbable Bone Putty is indicated to |
| fill bony voids or gaps in the skeletal | fill bony voids or gaps in the skeletal |
| system (i.e., extremities and pelvis). | system (i.e., extremities and pelvis). |
| These defects may be surgically | These defects may be surgically |
| created, or osseous defects created | created, or osseous defects created |
| as the result of traumatic injury to the | as the result of traumatic injury to the |
| bone. Montage-XT is indicated only | bone. Montage is indicated only for |
| for filling bony voids or gaps that are | filling bony voids or gaps that are not |
| not intrinsic to the integrity of the bony | intrinsic to the integrity of the bony |
| structure. | structure. |
| When hardened in situ, Montage-XT | When hardened in situ, Montage may |
| may be used to augment provisional | be used to augment provisional |
| hardware (e.g., k-wires, plates and | hardware (e.g., k-wires, plates and |
| screws) and to help support bone | screws) and to help support bone |
| fragments during the surgical | fragments during the surgical |
| procedure. The hardened putty acts | procedure. The hardened putty acts |
| only as a temporary support medium | only as a temporary support medium |
| and is not intended to provide | and is not intended to provide |
| structural support during the healing | structural support during the healing |
| process. | process. |
| Montage-XT can be drilled and<br>tapped, and hardware can be placed<br>through it at any time during the<br>setting process. | Montage can be drilled and tapped,<br>and hardware can be placed through it<br>at any time during the setting process. |
| <br> | <br> |
| At the time of application, device is in | At the time of application, device is in |
| the form of a putty-like material | the form of a putty-like material |
| <br> | <br> |
| Device is designed to be manually | Device is designed to be manually |
| applied and spread onto bone defects | applied and spread onto bone defects |
| <br> | <br> |
| Montage-XT Settable, Resorbable<br>Bone Putty is formulated as a two-<br>part putty/putty device that forms a<br>"settable" (hardening) material when<br>mixed at the time of surgery | Montage-XT Settable, Bone Putty is<br>formulated as a two-part putty/putty<br>device that forms a "settable"<br>(hardening) material when mixed at<br>the time of surgery |
| Sterile mixture of two separate | Sterile mixture of two separate |
| components of putty-like consistency<br>comprised of granular calcium<br>phosphate, (hydroxyapatite and ß-<br>tricalcium phosphate), calcium<br>stearate, vitamin E acetate, a<br>triglyceride, a polyalcohol and a<br>mixture of a lactide-diester and<br>polyester-based (lactide and<br>caprolactone) absorbable polymers.<br>Montage-XT is to be mixed<br>immediately prior to use. Resulting<br>settable material from the two putties<br>is primarily comprised of calcium<br>phosphate similar to the mineral | components of putty-like consistency<br>comprised of granular calcium<br>phosphate, (hydroxyapatite and ß-<br>tricalcium phosphate), calcium<br>stearate, vitamin E acetate, a<br>triglyceride, a polyalcohol and a<br>mixture of a lactide-diester and<br>polyester-based (lactide and<br>caprolactone) absorbable polymers.<br>Montage is to be mixed immediately<br>prior to use. Resulting settable<br>material from the two putties is<br>primarily comprised of calcium<br>phosphate similar to the mineral |
| phase of native bone tissue. | phase of native bone tissue. |
| Implanted device is resorbable in | Implanted device is resorbable in |
| greater than 30 days primarily due to<br>presence of calcium phosphate. | greater than 30 days primarily due to<br>presence of calcium phosphate. |
| | |
| The non-calcium salt and non-<br>polymeric components degrade via<br>dissolution; the polymer degrades via<br>hydrolysis and calcium salts degrade<br>via chemical dissolution and/or<br>cellular removal | The non-calcium salt and non-<br>polymeric components degrade via<br>dissolution; the polymer degrades via<br>hydrolysis and calcium salts degrade<br>via chemical dissolution and/or cellular<br>removal |
| | |
| Single-patient-use device is provided<br>sterile by gamma irradiation | Single-patient-use device is provided<br>sterile by gamma irradiation |
| The two putties are provided | The two putties are provided |
| separately within a single outer foil<br>pouch. The outer foil pouch contains<br>a desiccant. | separately within a single outer foil<br>pouch. The outer foil pouch contains<br>a desiccant. |
| Mixing for homogeneity takes 45 sec. | Mixing for homogeneity takes 45 sec. |
| Working time is 4 minutes | |
| | Working time is 2 minutes |
| Device cures with no appreciable | Device cures with no appreciable |
| exothermic reaction. | exothermic reaction |
| Device can be drilled and tapped, and | Device can be drilled and tapped, and |
| hardware can be placed through it at | hardware can be placed through it at |
| any time during the setting process. | any time during the setting process. |
| When hardened in situ, Montage-XT | When hardened in situ, Montage-XT may |
| may be used to augment provisional | be used to augment provisional |
| hardware (e.g., k-wires, plates and | hardware (e.g., k-wires, plates and |
| screws) and to help support bone | screws) and to help support bone |
| fragments during the surgical | fragments during the surgical |
| procedure. The hardened putty acts | procedure. The hardened putty acts |
| only as a temporary support medium | only as a temporary support medium |
| and is not intended to provide | and is not intended to provide |
| structural support during the healing | structural support during the healing |
| process. | process. |
{7}------------------------------------------------
{8}------------------------------------------------
### Technoloqical Characteristics:
The technological characteristics of the devices are unchanged. The only change is the addition of a statement regarding the use of Montage-XT in as a bone void filler.
### Performance Testing:
The only clinically meaningful difference between Montage and Montaqe-XT is the extension of working time from 2 to 4 minutes. Evaluations were previously conducted to compare MONTAGE-XT Settable, Resorbable Bone Putty with the predicate device. The evaluations demonstrated that Montage-XT is substantially equivalent to the Montage predicate in intended use, technological characteristics, and performance. This testing included the following:
Bench Testing was conducted to verify the device's handling properties, to characterize the device's performance. The following bench studies were completed: pH characterization, dissolution/solubility, reaction temperature, relative stiffness, spreadability, stickiness, temperature sensitivity, electrocautery compatibility and working time. Both devices are sterilized with the same validated gamma irradiation process and both have the same established shelf-life / stability characteristics.
Biocompatibility Testing was conducted to evaluate biocompatibility in accordance with the recommendations of ISO 10993. The following biocompatibility studies were conducted in accordance with the GLP requirements: irritation, sensitization, acute systemic toxicity, genotoxicity, implantation, systemic toxicity, hemolysis, endotoxicity and pyrogenicity.
{9}------------------------------------------------
Animal Testing included animal studies to quantify device resorption and new bone formation of Montage and Montage-XT in a rabbit critical sized femoral defect model. Because the devices have nearly identical formulations and have been found to be substantially equivalent in all other respects, the data supporting use as a bone void filler from studies of Montage is relevant to Montage-XT and was confirmed through testing of Montage-XT. Clinical performance has not been evaluated.
### Conclusion
MONTAGE-XT is substantially equivalent to Montage Settable, Resorbable Bone Putty with respect to intended use, general technological characteristics, and performance.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.