The Uresta bladder support is indicated for use in adult women over 18 years of age who experience stress urinary incontinence (involuntary urine loss, such as a pee leak that occurs when you cough, laugh, sneeze or with some physical tasks).
Device Story
Uresta is a bell-shaped vaginal pessary designed to support the urethra and reduce stress urinary incontinence. The device features a tapered tip for vaginal insertion and a handle for placement and removal. It is intended for self-use by adult women, either daily or during specific physical activities. The device is non-sterile and may be used with water-based lubricant. It is available in five sizes (34 mm to 52 mm). The device functions mechanically to provide bladder support; it does not involve electronic components, software, or automated analysis. Users select and fit the device based on provided instructions and labeling.
Clinical Evidence
No clinical efficacy trials were conducted for this submission. The manufacturer leveraged non-clinical and clinical data from the predicate device (K081385). A prospective labeling usability and self-selection study was performed with a diverse cohort of naïve users. Results demonstrated that participants could accurately self-select based on criteria and successfully insert/fit the device using only the provided instructions, confirming suitability for OTC use.
Technological Characteristics
Bell-shaped vaginal pessary; material: non-latex thermoplastic rubber; dimensions: 34 mm to 52 mm (five sizes); non-sterile; mechanical support principle; no energy source; no software; no connectivity.
Indications for Use
Indicated for adult women over 18 years of age experiencing stress urinary incontinence (involuntary urine loss during physical tasks like coughing, laughing, or sneezing).
Regulatory Classification
Identification
A vaginal pessary is a removable structure placed in the vagina to support the pelvic organs and is used to treat conditions such as uterine prolapse (falling down of uterus), uterine retroposition (backward displacement), or gynecologic hernia.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 3, 2024
Resilia Inc. % Roshana Ahmed President Ouaras. LLC 2101 Camino Rey Fullerton, California 92833
Re: K233548
> Trade/Device Name: Uresta® Regulation Number: 21 CFR 884.3575 Regulation Name: Vaginal Pessary Regulatory Class: II Product Code: HHW Dated: February 26, 2024 Received: February 26, 2024
Dear Roshana Ahmed:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}------------------------------------------------
the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Sharon M. Andrews -S
Sharon M. Andrews Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Submission Number (if known)
K233548
Device Name
Uresta®
Indications for Use (Describe)
The Uresta bladder support is indicated for use in adult women over 18 years of age who experience stress urinary incontinence (involuntary urine loss, such as a pee leak that occurs when you cough, laugh, sneeze or with some physical tasks).
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
K233548 Page 1 of 3
Resilia Inc.
# 510(k) Summary
I. Submitter
Resilia Inc. 644 Main Street Suite S400, P.O. Box 1368 Moncton, NB E1C 1E2 Canada
Contact Person: Sam Imbeault, Chief Operating Officer Phone: 506-874-8584 Date Prepared: February 23, 2024
#### II. Device
| Device Proprietary Name: | Uresta® |
|--------------------------|------------------------|
| Common or Usual Name: | Vaginal pessary |
| Classification Name: | Vaginal pessary |
| Regulation Number: | 21 CFR 884.3575 |
| Product Code: | HHW (pessary, vaginal) |
| Device Classification | 2 |
#### III. Predicate Device
Uresta®, K081385, EastMed Inc. The predicate device has not been subject to a design-related recall.
#### IV. Device Description
Uresta® is a bell-shaped vaginal bladder support that is intended to support the urethra to help reduce leaks caused by stress urinary incontinence. It has a tapered tip for insertion into the vaginal introitus and a handle end to enable placement and removal. It may also be used with a water-based lubricant for easier insertion.
Uresta® is available in five (5) sizes, ranging from 34 mm (size 2) to 52 mm (size 6). It can be inserted in the morning and removed at bedtime every day or used just for specific activities such as exercise and removed after the activity.
{5}------------------------------------------------
Uresta® is molded from non-latex thermoplastic rubber and is provided non-sterile for single patient use.
#### V. Indications for Use
The Uresta bladder support is indicated for use in adult women over 18 years of age who experience stress urinary incontinence (involuntary urine loss, such as a pee leak that occurs when you cough, laugh, sneeze or with some physical tasks).
#### VI. Comparison of Technological Characteristics
Uresta® has the same intended use as the predicate device and is intended for use in the same user population; however, the subject device is intended for both Rx and OTC use, while the predicate device is indicated for Rx use only. With the exception of available product sizes, the subject device is identical to the predicate device with respect to material of construction and design. There have been no changes to the product since clearance under K081385.
| | Uresta®<br>Subject Device | Uresta®<br>(K081385) | Analysis |
|-------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use/Indications<br>for Use | The Uresta bladder support<br>is indicated for use in adult<br>women over 18 years of<br>age who experience stress<br>urinary incontinence<br>(involuntary urine loss,<br>such as a pee leak that<br>occurs when you cough,<br>laugh, sneeze or with some<br>physical tasks). | This device is for the use in<br>adult women, over 18<br>years of age who<br>experience involuntary<br>urine loss with physical<br>activity (stress urinary<br>incontinence). | Different; slight<br>differences in wording<br>do not alter the intended<br>use of the subject device<br>when compared to the<br>predicate device. |
| User Population | Women, ages 18 and older | Women, ages 18 and older | Identical |
| Use | Rx and OTC | Rx | Different; performance<br>testing demonstrates<br>patients can self-select<br>and use the device<br>appropriately in an OTC<br>setting. |
| Design | Bell-shaped pessary | Bell-shaped pessary | Identical |
| Number of models | Sizes 2 – 6 | Sizes 1 – 7 | Different, sizes 1 and 7<br>have been removed from<br>the product offering. |
| Reuse | Yes | Yes | Identical |
| Sterile | No | No | Identical |
A comparison of the subject and predicate devices is provided in the table below.
{6}------------------------------------------------
#### VII. Performance Data
Non-clinical and clinical data submitted within K083185 is being leverage for the subject device.
In addition, a self-selection and labeling usability study was performed and demonstrates that the subject device can be appropriately self-selected and fit by the intended lay user population. Women who were naïve to the device reviewed the external product packaging and determined if they met the criteria to be an intended user. The self-selection was confirmed via clinical diagnosis. Intended users then participated in a labeling usability simulated study in which they reviewed the Instructions for Use and independently inserted and fit the device without guidance. Participant demographics were diverse with respect to race, ethnicity, age, educational level, and health literacy and representative of the US patient population. The results of the self-selection and usability study demonstrate that patients can self-select and use the device appropriately.
#### VIII. Conclusion
The information provided above supports that Uresta® is substantially equivalent to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.