Hardy Diagnostics' Viral Transport Medium (VTM) is intended for the collection and transport of clinical specimens for the preservation of viral agents including, Influenza A, Influenza B, Adenovirus, and Echovirus from the collection site to the testing laboratory. Hardy Diagnostics' VTM is a culture-based media that is intended to be used for standard laboratory procedures of virus culture and diagnostic assays which utilize stable recoverable infectious viral particles.
Device Story
Hardy Diagnostics' VTM is a non-propagating, culture-based transport medium used to collect and transport clinical specimens suspected of containing specific viruses (Influenza A/B, Adenovirus, Echovirus). The device consists of a screw-cap polypropylene tube containing 3mL of medium (Hank's Balanced Salt Solution, fetal bovine serum, sucrose, Amphotericin B, Gentamicin Sulfate, and phenol red). It is used at the point of collection by healthcare personnel and transported to a laboratory. The medium preserves viral viability for up to 48 hours at 2-25°C. Laboratory staff use the recovered viral particles for standard culture or diagnostic assays. The device is single-use and supplied either as tubes alone or in a kit with a mini-tip flocked swab.
Clinical Evidence
Bench testing only. Viral recovery performance was evaluated by spiking known concentrations of Influenza A, Influenza B, Adenovirus, and Echovirus into negative clinical matrix. Samples were stored at 2-8°C and 20-25°C for 0, 24, and 48 hours. Viral titers were determined using the Reed-Muench method (TCID50/mL). Results showed viral recovery changes within ±90% of baseline, meeting acceptance criteria. Shelf-life stability was validated over 12 months at 2-25°C.
Technological Characteristics
Polypropylene screw-cap tube containing 3mL of transport medium. Formulation: Hank's Balanced Salt Solution, bovine serum albumin, Amphotericin B, Gentamicin Sulfate, L-Glutamic acid, sucrose, and phenol red (pH indicator). pH 7.3 ± 0.2 at 25°C. Shelf-life: 12 months at 2-25°C. Single-use device.
Indications for Use
Indicated for the collection and transport of clinical specimens for the preservation of viral agents (Influenza A, Influenza B, Adenovirus, Echovirus) from collection site to laboratory for culture-based diagnostic assays. For prescription use only.
Regulatory Classification
Identification
A transport culture medium is a device that consists of a semisolid, usually non-nutrient, medium that maintains the viability of suspected pathogens contained in patient specimens while in transit from the specimen collection area to the laboratory. The device aids in the diagnosis of disease caused by pathogenic microorganisms and also provides epidemiological information on these diseases.
Predicate Devices
Copan Universal Transport Medium (utm-rt) System (K042970)
Submission Summary (Full Text)
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K233534
B Applicant
Hardy Diagnostics
C Proprietary and Established Names
Viral Transport Medium
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| JSM | Class I, reserved | 21 CFR 866.2390 - Transport Culture Medium | MI - Microbiology |
## II Submission/Device Overview:
A Purpose for Submission:
To obtain substantial equivalence determination for the Hardy Diagnostics’ Viral Transport Medium for collection and transport of viral organisms for standard laboratory culture and testing.
B Measurand:
Not applicable.
C Type of Test:
Non-propagating transport culture medium.
## III Intended Use/Indications for Use:
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{2}
K233534 - Page 2 of 7
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
Hardy Diagnostics' Viral Transport Medium (VTM) is intended for the collection and transport of clinical specimens for the preservation of viral agents including, Influenza A, Influenza B, Adenovirus, and Echovirus from the collection site to the testing laboratory. Hardy Diagnostics' VTM is a culture-based media that is intended to be used for standard laboratory procedures of virus culture and diagnostic assays which utilize stable recoverable infectious viral particles.
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
D Special Instrument Requirements:
None.
IV Device/System Characteristics:
A Device Description:
The Hardy Diagnostics Viral Transport Medium is a non-propagating transport media used for the collection and transport of clinical specimens suspected of containing viruses for viral culture and detection. The VTM consists of Hank's Balanced Salt Solution, a buffer to maintain a neutral pH, and fetal bovine serum, sucrose for specimen stabilization, along with Amphotericin B and Gentamicin Sulfate to inhibit bacterial and fungal growth. VTM also contains a pH indicator (phenol red) to provide a visual check on medium pH. The VTM appears translucent and light peach in color.
The Hardy Diagnostics Viral Transport Medium includes a polypropylene screw-cap tube with skirted conical bottom containing 3mL of transport medium. The product is supplied in multiple configurations described in table 1 below: tubes alone, or in a kit format with a swab.
Table 1. List of Configurations: Hardy Diagnostics' Viral Transport Medium
| Catalog Number | Description | Quantity |
| --- | --- | --- |
| R99 | VTM, 16 × 100mm polypropylene tube, 3mL fill | 20 tubes/box |
| R64BX | VTM, 13 × 80mm polypropylene tube, 3mL fill | 100 tubes/box |
| TPV50 | TransPRO VTM System, single 16×100mm polypropylene tube, 3mL fill with individually wrapped, sterile, mini-tip, flocked swab with 80mm breakpoint. | 50 each/box |
B Principle of Operation:
Clinical specimens are aseptically collected using swabs (polyester (Dacron), rayon, or flocked nylon) and shafts (plastic or flexible wire) that are non-toxic to viral agents. Swabs with a scored breakpoint at or less than 100 mm are recommended. The TransPRO Viral Transport Medium (VTM) System includes an individually wrapped, sterile, mini-tip flocked swab with an 80mm
{3}
breakpoint. After collection of specimens, cap is removed from the tube and swab is inserted into the media. The swab shaft is cut with scissors or broken off by bending the shaft at the scored breakpoint against the rim of the tube. The us swab shaft is required to be below the rim of the tube to facilitate proper cap closure of the tube. The tubes may then be stored or transported at the permissible temperature.
Hardy Diagnostics' VTM has the following ingredients.
- Hank's Balanced Salts Solution (HBSS)
- Bovine Serum Albumin
- Amphotericin B
- Gentamicin Sulfate
- L-Glutamic Acid
- Sucrose
- Phenol Red
HBBS, bovine albumin, and sucrose provide stabilization to the specimen. Gentamicin Sulfate and Amphotericin B inhibit bacterial and fungal contamination, respectively. Phenol red functions as a pH indicator.
## V Substantial Equivalence Information:
### A Predicate Device Name(s):
Copan Universal Transport Medium (utm-rt) System
### B Predicate 510(k) Number(s):
K042970
### C Comparison with Predicate(s):
| Device & Predicate Device(s): | Device: K233534 | Predicate: K042970 |
| --- | --- | --- |
| Device Trade Name | Hardy Diagnostics Viral Transport Medium | Copan Universal Transport Medium (UTM-RT) System |
| Device Product Code and Classification | JSM, Class 1 | JSM, Class 1 |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | Hardy Diagnostics' Viral Transport Medium (VTM) is intended for the collection and transport of clinical specimens | Copan Universal Transport Medium (UTM-RT) System is intended for the collection and transport of clinical specimens |
K233534 - Page 3 of 7
{4}
K233534 - Page 4 of 7
| | for the preservation of viral agents including, Influenza A, Influenza B, Adenovirus, and Echovirus from the collection site to the testing laboratory. Hardy Diagnostics’ VTM is a culture-based media that is intended to be used for standard laboratory procedures of virus culture and diagnostic assays that utilize stable recoverable infectious viral particles. | containing viruses, Chlamydiae, Mycoplasma or Ureaplasma from the collection site to the testing laboratory. UTM-RT can be processed using standard clinical laboratory operating procedures for viral, chlamydial, mycoplasma and ureaplasma culture. |
| --- | --- | --- |
| Tube Material | Plastic Screw-Cap Tube | Same |
| pH | 7.3 ± 0.2 at 25°C | Same |
| Shelf-life | 12 months or 365 days | Same |
| Storage Temperature | 2-25°C | Same |
| Single Use Device | Yes | Same |
| General Device Characteristic Differences | | |
| Media Formulation | • Hank’s Balanced Salts Solution (HBSS)
• Bovine Serum Albumin
• Amphotericin B
• Gentamicin Sulfate
• L-Glutamic Acid
• Phenol Red
• Sucrose | • Hank’s Balanced Salts solution (HBSS)
• Bovine Serum Albumin (BSA)
• Vancomycin
• Amphotericin B
• Colistin
• L-Glutamic Acid
• L-Cysteine
• HEPES Buffer
• Phenol Red
• Gelatin
• Sucrose |
| Supported Strains | • Adenovirus
• Echovirus
• Influenza A
• Influenza B | • Adenovirus
• Cytomegalovirus
• Echovirus Type 30
• Herpes Simplex Virus Type 1
• Herpes Simplex Virus Type 2
• Influenza A
• Parainfluenza 3 |
{5}
K233534 - Page 5 of 7
| | | • Respiratory Syncytial Virus
• Varicella Zoster Virus
• Chlamydia pneumoniae
• Chlamydia trachomatis
• Mycoplasma hominis
• Mycoplasma pneumoniae
• Ureaplasma urealyticum |
| --- | --- | --- |
## VI Standards/Guidance Documents Referenced:
Not applicable.
## VII Performance Characteristics (if/when applicable):
### A Analytical Performance:
1. **Precision/Reproducibility:**
Not applicable.
2. **Linearity:**
Not applicable.
3. **Analytical Specificity/Interference:**
Not applicable.
4. **Assay Reportable Range:**
Not applicable.
5. **Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):**
**Shelf-life:**
To evaluate Shelf-life, real-time stability studies were conducted with Hardy Diagnostics’ VTM lots in a real-time aging study at 2-8°C and 20-25°C. Three VTM lots were tested qualitatively for appearance (e.g., visual inspection for turbidity, precipitation, color, etc.), pH (acceptable range: 7.3 ± 0.2), fill volume (measured by weight), virus viability at every month from 0 to 12 months. All results met the study acceptance criteria and support a shelf-life stability of Hardy Diagnostics’ VTM for up to 12 months at 2-25°C.
{6}
6. Detection Limit:
# Performance Testing - Recovery Studies:
The performance of Hardy Diagnostics Viral Transport Medium (VTM) was evaluated for culture-based recovery of Influenza A, Influenza B, Echovirus, Adenovirus, and Respiratory Syncytial Virus (RSV). Pooled negative clinical matrix was spiked with known concentration of virus and then the contrived viral samples were transferred with swabs into subject VTM and held at $2 - 8^{\circ}\mathrm{C}$ and $20 - 25^{\circ}\mathrm{C}$ for 0, 24, and 48 hours. The viral recovery study was conducted with three variously aged lots of $3\mathrm{ml}$ VTM covering the claimed shelf-life. At the end of each time point media stored under each temperature range, an aliquot of incubated samples were serially diluted, and inoculated in triplicate into the susceptible host cell line. After 24-72 hours, Viral quantification was determined by microscopic observation of virus induced CPE to determine the fifty-percent tissue culture infective dose $(\mathrm{TCID}_{50} / \mathrm{mL})$ using the Reed-Muench method. Results were considered acceptable if the average viral recovery for each time point and storage condition demonstrate any percent changes within $\pm 90\%$ (i.e., 1 log change) from baseline $(T = 0)$ .
Table 2 and Table 3 below exhibit the viral recovery performance of the Hardy Diagnostics Viral Transport Medium at ${20} - {25}^{ \circ }\mathrm{C}$ and at $2 - 8{}^{ \circ }\mathrm{C}$ respectively.
Table 2: Viral recovery performance of Hardy Diagnostics Viral Transport Medium at ${20} - {25}^{ \circ }\mathrm{C}$
| Test Strain | Viral recovery (TCID50/mL) at T=0 hr. | Percent changes in viral recovery from the baseline (T= 0 hr.) (-ve indicates reduction) | |
| --- | --- | --- | --- |
| | | 24 hrs. | 48 hrs. |
| Influenza A | 2.12x103 | -92.24* | -59.76 |
| Influenza B | 7.23x102 | -23.86 | -3.45 |
| Echovirus | 1.40x105 | -20.55 | -25.23 |
| Adenovirus | 1.92x103 | 18.11 | -7.71 |
* Considered acceptable because subsequent timepoints, i.e., 48 h time point showed < 90% increase.
Table 3: Viral recovery performance of Hardy Diagnostics Viral Transport Medium at ${2 - 8}^{ \circ }\mathrm{C}$
| Test Strain | Viral recovery (TCID50/mL) at T=0 hr. | Percent Changes in viral recovery from the baseline (T= 0 hr.) (-ve indicates reduction) | |
| --- | --- | --- | --- |
| | | 24 hrs. | 48 hrs. |
| Influenza A | 1.32x103 | -5.98 | -58.59 |
| Influenza B | 8.05x102 | -36.32 | -44.27 |
| Echovirus | 1.27x105 | 21.91 | -3.98 |
| Adenovirus | 1.47x103 | -20.19 | 23.11 |
K233534 - Page 6 of 7
{7}
The viral recovery study results support the recovery of the tested viruses for up to 48 hours from collection when samples are stored in the Hardy Diagnostics Viral Transport Medium at 2-8°C or at room temperature (20-25°C).
7. Assay Cut-Off:
Not applicable.
B Comparison Studies:
1. Method Comparison with Predicate Device:
Not applicable.
2. Matrix Comparison:
Not applicable.
C Clinical Studies:
1. Clinical Sensitivity:
Not applicable.
2. Clinical Specificity:
Not applicable.
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Not applicable.
D Clinical Cut-Off:
Not applicable.
E Expected Values/Reference Range:
Not applicable.
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K233534 - Page 7 of 7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.