← Product Code NLQ · K233471

# Reprocessed HARMONIC 1100 Shears, 5mm Diameter, 20cm length (HAR1120); Reprocessed HARMONIC 1100 Shears, 5mm Diameter, 36cm length (HAR1136) (K233471)

_Stryker Sustainability Solutions · NLQ · Feb 20, 2024 · SU · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K233471

## Device Facts

- **Applicant:** Stryker Sustainability Solutions
- **Product Code:** NLQ
- **Decision Date:** Feb 20, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** SU
- **Attributes:** Therapeutic, Pediatric

## Intended Use

The Reprocessed HARMONIC 1100 Shears are indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instruments can be used as an adjunct to or substitute for electrosurgery, lasers and steel scalpels in general, pediatric, gynecologic, urologic, thoracic, sealing and transection of lymphatic vessels, and other open and endoscopic procedures. The instruments allow for the coagulation of veins up to and including 5 mm in diameter and arteries up to and including 7 mm in diameter, using the Energy button with Advanced Hemostasis.

## Device Story

Sterile, single-patient-use ultrasonic surgical instrument; used for dissection, grasping, coagulation, and cutting. Input: mechanical tissue interaction via clamp arm and blade. Operation: utilizes Adaptive Tissue Technology (ATT) to monitor instrument status; generator modulates power output based on tissue feedback. Output: audible/tactile feedback; thermal coagulation and mechanical transection. Used in OR/endoscopic settings by surgeons. Benefits: precise hemostasis, reduced thermal injury, and vessel sealing up to 7mm. Requires Generator 11 (GEN11) for operation.

## Clinical Evidence

No clinical data. Evidence includes bench testing (biocompatibility, sterilization, electrical safety per IEC 60601-1/1-2/2-2, functional performance) and preclinical animal studies (acute/chronic survival) evaluating thermal spread and hemostasis efficacy.

## Technological Characteristics

Ultrasonic surgical instrument; 5mm diameter; 20cm or 36cm shaft lengths. Features ergonomic handle, integrated handpiece, and two energy buttons. Utilizes Adaptive Tissue Technology for generator power modulation. Electrical safety per IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-2. Sterile, single-patient use.

## Regulatory Identification

Single-use reprocessed versions of the ultrasonic devices used in surgical procedures either for fragmentation, emulsification and aspiration of soft tissue and hard tissue or for ligation of vessels (under product code LFL). Reprocessing validation data for this device type must be included in a 510(k) submission.. (70 FR 56911, available at https://www.govinfo.gov/content/pkg/FR-2005-09-29/pdf/05-19510.pdf).

## Predicate Devices

- HARMONIC 1100 Shears 20cm Length; HARMONIC 1100 Shears 36cm Length ([K200841](/device/K200841.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 20, 2024

Stryker Sustainability Solutions Aphrodeja Crutch Staff Specialist, Regulatory Affairs 1810 Drake Dr. Tempe, Arizona 85283

Re: K233471

Trade/Device Name: Reprocessed HARMONIC 1100 Shears, 5mm Diameter, 20cm length (HAR1120); Reprocessed HARMONIC 1100 Shears, 5mm Diameter, 36cm length (HAR1136) Regulation Number: 21 CFR Unclassified Regulation Name: Regulatory Class: Class II Product Code: NLQ Dated: October 23, 2023 Received: October 25, 2023

Dear Aphrodeja Crutch:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

> Sincerely. Digitally signed by Mark Mark Trumbore -S Trumbore -S Date: 2024.02.20 14:09:02 -05'00' Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Enclosure:

# Models For This Submission

Description: Reprocessed HARMONIC 1100 Shears Model Numbers: HAR1120 and HAR1136

#### Physical Description/Specifications

The Reprocessed HARMONIC 1100 Shears sterile instruments are used for dissection, grasping, coaqulation, and cutting between the blade and clamp arm. Each device model consists of an ergonomic handle, an integrated hand piece, and two energy delivery buttons:

- Energy Button - User can adjust power levels from 1-5
- . Energy button with Advanced Hemostasis – For larger vessel sealing; user cannot adjust.

The instruments are available in two shaft lengths, 20 cm and 36 cm. An integrated audible and tactile mechanism in the handle indicates full trigger closure. The instrument has a clamp arm and coated curved blade that are deigned to work through a 5 mm trocar, through a 5mm reducer cap in a larger diameter trocar, or through an incision without the use of a trocar. The instrument shafts can be rotated continuously to facilitate visualization and access to targeted tissue. The two dashes on the instrument are intended to represent relative vessel size. The energy button is indicated for vessels up to 5 mm in diameter. When the energy button is used, cutting speed is the fastest. The energy button with Advanced Hemostasis is designed for larger vessels and is indicated for vessels up to 7 mm in diameter. When the Energy button with Advanced Hemostasis in used, cutting speed in reduced and hemostasis is maximized. The instrument utilizes Adaptive Tissue Technology. This provides the generator with the ability to identify and monitor the instrument during use, which enables the generator to modulate and adjust its power output as well as provide audible feedback to the user as appropriate.

{3}------------------------------------------------

## Indications for Use

510(k) Number (if known) K233471

#### Device Name

Reprocessed HARMONIC 1100 Shears, 5mm Diameter, 20cm Length (HAR1120); Reprocessed HARMONIC 1100 Shears, 5mm Diameter, 36cm Length (HAR1136)

#### Indications for Use (Describe)

The Reprocessed HARMONIC 1100 Shears are indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instruments can be used as an adjunct to or substitute for electrosurgery, lasers and steel scalpels in general, pediatric, gynecologic, thoracic, sealing and transection of lymphatic vessels, and other open and endoscopic procedures. The instruments allow for the coagulation of veins up to and including 5 mm in diameter and arteries up to and including 7 mm in diameter, using the Energy button with Advanced Hemostasis.

| Type of Use (Select one or both, as applicable)                            |                                                                           |
|----------------------------------------------------------------------------|---------------------------------------------------------------------------|
| <div> <span>☑</span> Prescription Use (Part 21 CFR 801 Subpart D)   </div> | <div> <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)   </div> |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

### 510(k) Summary

| Applicant Name:                              | Stryker Sustainability Solutions                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:                              | Aphrodeja Crutch<br>Staff Specialist<br>1810 Drake Drive<br>Tempe, AZ 85283<br>Phone: 404-993-1981<br>Email: Aphrodeja.Crutch@stryker.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Device Trade Name:                           | Reprocessed HARMONIC 1100 Shears, 5mm Diameter, 20cm<br>Length (HAR1120); Reprocessed HARMONIC 1100 Shears,<br>5mm Diameter, 36cm Length (HAR1136)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Common Name:                                 | Instrument, Ultrasonic Surgical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Classification Name:                         | Instrument, Ultrasonic Surgical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Regulation Number:                           | Unclassified                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Product Code:                                | NLO                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Device Class:                                | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Date Prepared:                               | 02/20/2024                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Predicate Device:                            | HARMONIC 1100 Shears 20cm Length; HARMONIC 1100<br>Shears 36cm Length (K200841)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Predicate Product Code:                      | LFL                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Description                                  | The Reprocessed HARMONIC 1100 Shears instrument is a<br>sterile, single patient use instrument used for dissection,<br>grasping, coagulation, and cutting between the blade and<br>clamp arm. It consists of an ergonomic handle with an<br>integrated hand piece and two energy delivery buttons.<br><br>1) Energy button for power levels 1-5, and<br><br>2) Energy with Advanced Hemostasis button for large vessel<br>sealing.<br><br>The instrument is available in two shaft lengths (20 cm and 36<br>cm). An integrated audible and tactile mechanism in the grip<br>housing indicates full trigger closure. The instrument has a<br>clamp arm and coated curved blade that are designed to work                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|                                              | through a 5 mm trocar, through a 5 mm reducer cap in a<br>larger diameter trocar, or through an incision without the use<br>of a trocar. The instrument shafts can be rotated continuously<br>to facilitate visualization and access to targeted tissue. The<br>two dashes on the instrument are intended to represent<br>relative vessel size. The Energy button is indicated for vessels<br>up to 5mm in diameter. When the Energy button is used,<br>cutting speed is the fastest. The energy button with Advanced<br>Hemostasis is designed for larger vessels and is indicated for<br>veins up to and including 5 mm in diameter and arteries up to<br>and including 7 mm in diameter. When the Energy button<br>with Advanced Hemostasis is used, cutting speed is reduced<br>and hemostasis is maximized. The instrument utilizes<br>Adaptive Tissue Technology. This provides the generator with<br>the ability to identify and monitor the instrument during use,<br>which enables the generator to modulate and adjust its power<br>output as well as provide audible feedback to the user as<br>appropriate. The HARMONIC 1100 Shears instrument is<br>designed for use exclusively with the Generator 11 (GEN11)<br>software version 2018-1 or later. |
| Indications for Use:                         | The Reprocessed HARMONIC 1100 Shears are indicated for<br>soft tissue incisions when bleeding control and minimal<br>thermal injury are desired. The instruments can be used as an<br>adjunct to or substitute for electrosurgery, lasers and steel<br>scalpels in general, pediatric, gynecologic, urologic, thoracic,<br>sealing and transection of lymphatic vessels, and other open<br>and endoscopic procedures. The instruments allow for the<br>coagulation of veins up to and including 5 mm in diameter<br>and arteries up to and including 7 mm in diameter, using the<br>Energy button with Advanced Hemostasis.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Summary of Intended Use and<br>Technologies: | The design, materials, and intended use of the Reprocessed<br>HARMONIC 1100 Shears are equivalent to the predicate<br>device. The mechanism of action of the reprocessed device is<br>identical to the predicate device in that the same standard<br>mechanical design, materials, and size is utilized. There are no<br>changes to the claims, intended use, clinical application,<br>patient population, performance specifications, or method of<br>operation. In addition, Stryker Sustainability Solutions'<br>reprocessing of the device includes removal of adherent<br>visible soil and decontamination. Each individual device is<br>tested for appropriate function of its components prior to<br>packaging and labeling operations. The only differences<br>between the Reprocessed HARMONIC 1100 Shears and the<br>predicate HARMONIC 1100 Shears are that the device is<br>reprocessed, and some device components are replaced with<br>equivalent components during the reprocessing operation.                                                                                                                                                                                                                                                     |
| Non-Clinical Testing:                        | Bench and laboratory testing was conducted to demonstrate<br>performance (safety and effectiveness) of the Reprocessed<br>HARMONIC 1100 Shears. This included the following tests:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|                                              | • Biocompatibility<br>• Validation of Reprocessing<br>• Sterilization<br>• Electrical Safety and Electromagnetic Compatibility<br>• Functional Performance Tests<br>Blade to Clamp Arm Angle<br>Actuating Trigger Force<br>Energy Button Activation Force<br>Advanced Hemostasis (AH) Button Activation Force<br>Rotation Knob Force<br>Jaw Clamp Force<br>Tissue Retention Force<br>Leak Test<br>Bend Test<br>ATT Functionality and Transection Time<br>Burst Pressure<br>Maximum Jaw and Shaft Temperature<br>Reliability Cycling<br>Burst Pressure (Post-Reliability)<br>Maximum Jaw and Shaft Temperature (Post-Reliability)<br>Performance Attribute Testing                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                                              | The functional performance testing involved electrical safety<br>and electromagnetic compatibility testing in accordance with<br>IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-2, and<br>verification/comparative testing (to the predicate device).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|                                              | Additionally, preclinical laboratory evaluations in an animal<br>model were performed, which included acute and chronic<br>survival studies. The studies were done to evaluate thermal<br>spread and the ability to achieve hemostasis of vessels of the<br>reprocessed device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Clinical Testing:                            | Clinical testing was not required for this submission.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Conclusion:                                  | The subject device Reprocessed HARMONIC 1100 Shears,<br>5mm Diameter, 20cm Length (HAR1120); Reprocessed<br>HARMONIC 1100 Shears, 5mm Diameter, 36 cm Length<br>(HAR1136) is substantially equivalent to the previously<br>cleared predicate device HARMONIC 1100 Shears 20cm<br>Length; HARMONIC 1100 Shears 36cm Length (K200841).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |

{5}------------------------------------------------

{6}------------------------------------------------

---

**Source:** [https://fda.innolitics.com/device/K233471](https://fda.innolitics.com/device/K233471)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
