The Guided Bone Regeneration (GBR) System is used to stabilize and fix bone filling materials and/or barrier membranes used for bone regeneration in the oral cavity.
Device Story
GBR System consists of titanium screws used to stabilize bone grafts and barrier membranes during oral bone regeneration procedures. Device is used in dental clinics by clinicians. Screws are manually implanted into bone to provide mechanical fixation of graft materials or membranes. System includes various diameters and lengths to accommodate different anatomical sites. Device is provided non-sterile and requires autoclave sterilization by the user before application. Output is mechanical stabilization of the graft site, which facilitates bone regeneration and prepares the site for subsequent dental implant placement. Device is MR conditional.
Clinical Evidence
Bench testing only. Mechanical performance verified via insertion torque, fracture torque, and pull-out tests per TS EN ISO 19023 and ASTM F543-23. Biocompatibility assessment leveraged from K231100. Sterilization validated per EN 17665-1.
Technological Characteristics
Material: Pure Titanium Grade 4 (ASTM F67-13, ISO 5832-2). Form factor: Bone fixation screws (diameters 1.0-2.0mm). Energy source: Manual. Sterilization: Autoclave (EN 17665-1). Connectivity: None.
Indications for Use
Indicated for stabilization and fixation of bone filling materials and/or barrier membranes for bone regeneration in the oral cavity.
Regulatory Classification
Identification
An intraosseous fixation screw or wire is a metal device intended to be inserted into fractured jaw bone segments to prevent their movement.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 03, 2025
Proimtech Saglik Urunleri Anonim Sirketi Hakan Cevik Official Correspondent Dudullu OSB Mahallesi Imes Sanayi Sitesi E Blok 501. Sokak N o:25 Dudullu OSB Mahallesi Imes Sanayi Sitesi C Blok 305. Istanbul, UMRANIYE 34775 TURKEY
Re: K233419
Trade/Device Name: GBR System Regulation Number: 21 CFR 872.4880 Regulation Name: Intraosseous Fixation Screw Or Wire Regulatory Class: Class II Product Code: DZL Dated: December 9, 2024 Received: December 9, 2024
Dear Hakan Cevik:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Sherrill Lathrop Blitzer
for Andrew Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K233419
Device Name GBR System
Indications for Use (Describe)
The Guided Bone Regeneration (GBR) System is used to stabilize and fix bone filling materials and/or barrier membranes used for bone regeneration in the oral cavity.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| <div> <span> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary 510k number: K233419 GBR System Proimtech Saglik Urunleri Anonim Sirketi
| Company Name | : | Proimtech Saglik Urunleri A.S. |
|---------------------------|---|-----------------------------------------------------------------------------------|
| Authorised Person / Title | : | Hakan Çevik / General Manager |
| Address-1 | : | Dudullu OSB Mahallesi Imes 305 Sokak C Blok No:3 PK:34775<br>Umraniye / İstanbul |
| Address-2 | : | Dudullu OSB Mahallesi Imes 501 Sokak E Blok No:25 PK:34775<br>Umraniye / İstanbul |
| Phone | : | +90 (216) 999 5041-44 |
| Web/Mail | : | www.bilimplant.com / info@bilimplant.com |
| Official Contact | : Hakan Cevik, General Manager of Proimtech Saglik Urunleri Anonim Sirket |
|------------------|---------------------------------------------------------------------------|
| Phone Number | : +90 535 355 51 00 |
| E-Mail | : hakancevik@bilimplant.com |
#### DEVICE INFORMATION
Preparation Date: December 31, 2024
Device Trade Name: GBR System
Common Name: GBR Kit
Classification Name: Intraosseous fixation screw or wire (21 CFR 872.4880)
Device Class: II
Product Code: DZL
Regulation Number: 872.4880
Review Panel: Dental
{5}------------------------------------------------
### INDICATION FOR USE
The Guided Bone Regeneration (GBR) System is used to stabilize and fix bone filling materials and/or barrier membranes used for bone regeneration in the oral cavity.
### DEVICE DESCRIPTION
Image /page/5/Figure/3 description: The image shows three different types of screws and their functions. The first screw is silver and is used for block bone graft fixation. The second screw is purple and stabilizes the graft particles by creating a tent effect under the membrane. The third screw is blue and fixes the membrane.
#### MATERIALS:
Only materials approved for medical purposes are used in the manufacture of products: Pure Titanium Grade 4 according to standard ISO 5832-2 and ASTM F67-13.
#### Non-Clinical Tests Conducted
Substantial equivalence to the predicate devices was determined by the following tests:
- > The product is manufactured from pure titanium (Grade 1641) raw material in accordance with EN 1642, TS EN ISO 19023, ISO 5832-2 and ASTM F67-13 Standard.
- > Our biocompatibility assessment is leveraged from our FDA approved application K231100. The devices in K231100 are made of the same materials and underwent the same manufacturing processes compared to the subject device.
- > Our Insertion Torque, Fracture Torque, Pull-out tests, which we had performed in neutral laboratories for the mechanical strength of our products, were carried out in accordance with TS EN ISO 19023 and ASTM F543-23 Standards.
- > Our non-sterile products should be sterilized with Autoclave before use by the users in accordance with the parameters specified in our user manuals.
- > The autoclave sterilization method declared by us has been validated in accordance with EN 17665-1 Standard. Our device has been verified by sterility tests.
{6}------------------------------------------------
### MRI Compatibility:
#### Rationale for relying on the Woods et al. Article
Non-clinical worst-case MRI review was performed to evaluate the Proimtech devices in the MRI environment using scientific rationale and published literature (e.g., Woods, Terry O., Jana G. Delfino, and Sunder Rajan. "Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices." Journal of Testing and Evaluation 49.2 (2019): 783-795), based on the entire system including all variations (all compatible implant bodies, dental abutments, and fixation screws) and material composition. The rationale addressed parameters per the FDA guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment," including magnetically induced displacement force and torque. The GBR System is labeled as MR conditional.
### Legally Marketed Device to Which
#### Claim Substantial Equivalence:
For primary device GBR System: K011698; Straumann GBR System Straumann USA
For reference device GBR System: K182881; Bone Tack, OSSTEM Implant Co., Ltd.
For reference device GBR System: K182881; Bone Screw, OSSTEM Implant Co., Ltd
For reference device GBR System: K161857; Tenting Screw, Salvin Dental
{7}------------------------------------------------
# SUMMARIES OF TECHNOLOGICAL CHARACTERISTICS & SUBSTANTIAL EQUIVALENCE DISCUSSION
{8}------------------------------------------------
| Technological<br>Characteristics | Subject Device<br>GBR System | Primary Predicate Device<br>Straumann GBR System | Reference Device<br>Bone Tack | Reference Device<br>Tenting Screw | Reference Device<br>Bone Screw |
|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K233419 | K011698 | K182881 | K161857 | K182881 |
| Manufacturer | Proimtech Saglik Urunleri A.S | Straumann USA | Osstem Implant Co.,<br>Ltd. | Salvin Dental | Osstem Implant Co.,<br>Ltd. |
| Product Code | DZL | JEY and DZL | DZL | DZL | DZL |
| Indications for Use | The Guided Bone Regeneration (GBR)<br>System is used to stabilize and fix<br>bone grafts, bone filling materials<br>and/or barrier membranes used for<br>bone regeneration in the oral cavity. | The Straumann Guided Bone<br>Regeneration (GBR) System is<br>used to stabilize and fixate<br>bone grafts, bone filling<br>materials, and/or barrier<br>membranes used for<br>regeneration of bone in the<br>oral cavity. The GBR<br>technique can make it<br>possible for the placement of<br>dental implants in previously<br>unsuitable sites. | It is indicated for use<br>to stabilize and<br>support bone graft<br>and/or fractured<br>bone segments with<br>or without bone<br>plates or titanium<br>mesh in oral and<br>maxillofacial site<br>defects. | The Salvin Tenting Scr<br>ew is used to stabilize<br>, fixate, and/or suppo<br>rt bone grafts, bone fi<br>lling materials and /o<br>r barrier membranes<br>used for regeneration<br>of bone in the oral ca<br>vity. | Bone Screw is used<br>to stabilize and fixate<br>bone grafts, bone<br>filling material,<br>and/or barrier<br>membranes used for<br>regeneration of bone<br>in the oral cavity |
| Design | Image: Blue screw | Image: Purple screw | Image: Yellow screw | Image: Silver screw | Image: Two silver screws |
| Material | Titanium (Ti Grade 4, ASTM F67) | Titanium (Ti Grade 4, ASTM<br>F67) | Titanium Alloy (Ti-<br>6Al-4V, ASTM F136) | Ti-6Al-<br>4V ELI (ASTM F136) | Titanium Alloy (Ti-<br>6Al-4V, ASTM F136) |
| Diameter | 1<br>1.2<br>1.5<br>2 | 1.2 mm<br>1.5 mm | 2.5 mm | 1.5 | 1.4 mm<br>1.95 mm |
| Length | Ø1*3.7 mm<br>Ø1.2*6 mm<br>Ø1.2*8 mm<br>Ø1.2*10 mm<br>Ø1.2*12 mm<br>Ø1.2*14 mm<br>Ø1.5*6 mm<br>Ø1.5*7 mm<br>Ø1.5*8 mm<br>Ø1.5*9 mm<br>Ø1.5*10 mm<br>Ø1.5*11 mm<br>Ø1.5*12 mm<br>Ø2*5 mm<br>Ø2*6 mm<br>Ø2*7 mm | 3.0 mm<br>4.0 mm<br>5.0 mm<br>7.0 mm<br>8.0 mm<br>9.0 mm<br>10.0 mm<br>12.0 mm<br>14.0 mm | 3.0 mm | 7 mm<br>8 mm<br>9 mm | Ø1.4:<br>4.0 mm<br>6.0 mm<br>8.0 mm<br>Ø1.95:<br>8.0 mm<br>10.0 mm<br>12.0 mm<br>14.0 mm<br>16.0 mm |
| Single Use | Yes | Yes | Yes | Yes | Yes |
| Sterilization | Non-sterile | Non-sterile | Sterile | Non-sterile | Sterile |
| Principle of Operation | The function of proposed device is<br>used in application for fixating or<br>stabilizing barrier membranes in GBR | The function of predicate<br>devices is used in application<br>for maintaining the relative<br>position of and/or bone grafts<br>in reconstruction of maxillary<br>and/or mandibular areas | The function of<br>proposed device is<br>used in application<br>for fixating or<br>stabilizing barrier<br>membranes in GBR | The Salvin Tenting Scr<br>ew System is a set of<br>bone screws and inst<br>rumentation. The sys<br>tem consists of a set<br>of screws (one diame<br>ter and three lengths<br>) and the driver used<br>to implant these scre<br>ws. The devices are d<br>elivered non-<br>sterile. The screws ar<br>e made from titaniu<br>m alloy (Ti-<br>6Al4V ELI), as describ<br>ed by ASTM F136. | The function of<br>proposed devices is<br>used in application<br>for maintaining the<br>relative position of<br>and/or bone grafts in<br>reconstruction of<br>maxillary and/or<br>mandibular areas |
{9}------------------------------------------------
{10}------------------------------------------------
## Similarities
The device has similar characteristics compared to the predicate and reference devices in the following respects. Indication for use, design, sterilization, disposable and operating principle.
## Difference
The material, diameter and length dimensions of the device in question are slightly different from the the predicate and reference devices. However, as a result of the insertion torque and pull-out tests, it was determined that there was no significant difference in product performance and that this difference did not affect the substantial equivalence.
## Discussion
On the basis of the above discussion, it is concluded that the device in question is substantially equivalent to the predicate and reference devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.