Folysil catheters are intended for use up to 30 days in adult and pediatric populations for: - Bladder drainage by urethral catheterization, - Bladder drainage by suprapubic catheter replacement only (for non-grooved, straight 2-way Folysil catheters with a maximum balloon volume of 15mL). - Bladder instillation of physiological saline solution.
Device Story
Folysil Silicone Catheter is a thin, hollow tube with a balloon, funnel, and valve; inserted into the bladder via urethral or suprapubic approach to drain urine into a collection bag. Balloon is inflated with sterile water or aqua-glycerin to secure device in bladder. Used in clinical settings for adult and pediatric patients. Device facilitates bladder drainage or saline instillation. Healthcare providers insert and manage the catheter. Benefits include effective urine drainage and bladder management over a 30-day indwell period.
Clinical Evidence
No clinical data. Evidence consists of biocompatibility testing per ISO 10993-1 and bench performance testing per ASTM F623-19 and ISO 20696:2018. Bench tests included visual inspection, catheter integrity, flow rate, balloon resistance/volume maintenance, deflation reliability, tensile strength, kink resistance, and radiopacity. All pre-determined acceptance criteria were met.
Technological Characteristics
Silicone construction; 2-way lumen; 6-24 Fr sizes; 25-41 cm lengths; balloon volume 1.5-30 mL. Features include radiopaque line, various tip shapes (Tiemann, straight, open), and stylet/introducer for pediatric sizes. Sterilized via ethylene oxide (ANSI/AAMI/ISO 11135:2014). Complies with ASTM F623-19 and ISO 20696:2018.
Indications for Use
Indicated for adult and pediatric patients requiring bladder drainage via urethral or suprapubic catheterization (replacement only) or bladder instillation of saline for up to 30 days.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left, there is a seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around a stylized image. To the right of the seal, there is a blue square with the letters "FDA" in white. Next to the square, the words "U.S. FOOD & DRUG" are written in blue, with the word "ADMINISTRATION" written in a smaller font size below.
April 15, 2024
Coloplast Corp % Kristen Swanson Regulatory Consultant Kompass Regulatory Consulting 1583 Northrop St. Falcon Heights, Minnesota 55108
Re: K233411
Trade/Device Name: Folysil® Silicone Catheter Regulation Number: 21 CFR 876.5130 Regulation Name: Urological Catheter and accessories Regulatory Class: II Product Code: EZL Dated: October 6, 2023 Received: October 6, 2023
Dear Kristen Swanson:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{3}------------------------------------------------
the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Sharon M. Andrews -S
for Jessica K. Nguyen, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K233411
Device Name Folysil® Silicone Catheter
Indications for Use (Describe)
Folysil catheters are intended for use up to 30 days in adult and pediatric populations for:
- Bladder drainage by urethral catheterization,
- Bladder drainage by suprapubic catheter replacement only (for non-grooved, straight 2-way Folysil catheters with a maximum balloon volume of 15mL).
- Bladder instillation of physiological saline solution.
Type of Use (Select one or both, as applicable)
| <span></span> | Prescription Use (Part 21 CFR 801 Subpart D) |
|---------------|----------------------------------------------|
| | Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
K233411 Page 1 of 5 Coloplast 1601 West River Road North Minneapolis, MN 55411
# TRADITIONAL 510(k) SUMMARY
#### SUBMITTER INFORMATION I.
| Applicant | Coloplast A/S |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| Legal Manufacturer Address: | Holtedam 1<br>3050 Humlebaek, Denmark |
| Official Correspondent: | Kristen Swanson<br>Regulatory Consultant<br>1583 Northrop St.<br>Falcon Heights, MN 55108<br>Phone: 651-245-2678<br>Email: uskrs@coloplast.com) |
| Date Prepared: | April 12, 2024 |
#### II. DEVICE NAME
| Trade or Proprietary Name: | Folysil® Silicone Catheter |
|----------------------------|-----------------------------------------|
| Common or Usual Name: | Foley Catheter |
| Classification Regulation: | 21 CFR 876.5130 |
| Device Class: | II |
| Product Code: | EZL (Catheter, Retention Type, Balloon) |
| Panel: | Gastroenterology/Urology |
#### III. PREDICATE DEVICES IDENTIFICATION
| Primary Predicate device: | Folysil® Silicone Catheter (K013174) |
|-----------------------------|------------------------------------------------------|
| Secondary Predicate device: | Modified Pediatric Silicone Foley Catheter (K900031) |
{6}------------------------------------------------
#### IV. DEVICE DESCRIPTION
Folysil catheters are thin, hollow tubes equipped with a balloon, funnel and valve, to be inserted into the bladder by the urethral or supra-pubic approach to drain urine into a urine collection bag. Folysil catheters are held in place within the bladder by the balloon, which is filled with sterile water or with aqua-glycerin solution.
Folysil Silicone catheters are provided in a variety of configurations as shown in Table 1.
| | Pediatric Use | Adult Use |
|---------------------------|-----------------------------------|-----------------------------------|
| 2-way | 2-way | 2-way |
| Size, Fr | 6 Fr - 10 Fr | 12 Fr - 24 Fr |
| Tip Shapes | Straight round, Tiemann, Straight | Straight round, Tiemann, Straight |
| | Open | Open |
| Balloon Volume, mL | 1.5mL - 3.0mL | 10 mL - 30 mL |
| Drainage eyes | 2 | 1-6 |
| Catheter Total Length, cm | 26 cm - 32 cm | 25 cm - 41 cm |
| Tube Surface | Smooth | Smooth and grooved |
### Table 1: Folysil Silicone Catheter Range
#### INDICATIONS FOR USE V.
Folysil catheters are intended for use up to 30 days in adult and pediatric populations for: Bladder drainage by urethral catheterization, -
- Bladder drainage by suprapubic catheterization for catheter replacement only (for non-grooved, straight 2-way Folysil catheters with a maximum balloon volume of 15mL).
- -Bladder instillation of physiological saline solution.
#### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
| Comparison<br>Element | Device<br>Folysil® Silicone Catheters | Primary Predicate Device<br>Porges Folysil Silicone Foley<br>Catheters<br>K013174 | Secondary Predicate<br>Device<br>Modified Pediatric<br>Silicone Foley Catheters<br>K900031 |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | Folysil catheters are intended<br>for use up to 30 days in adult<br>and pediatric populations for:<br>- Bladder drainage by<br>urethral catheterization,<br>- Bladder drainage by<br>suprapubic catheterization<br>for catheter replacement<br>only (for non-grooved,<br>straight 2-way Folysil<br>catheters with a maximum<br>balloon volume of 15mL). | The 2-way PORGES Folysil<br>silicone Foley catheter is used<br>for urethral urinary<br>catheterization for short-term<br>drainage of vesical urines.<br>Only straight 2-way PORGES<br>Folysil silicone Foley catheter<br>with a maximum 15 ml<br>balloon volume may be used<br>in a supra-pubic way (except<br>the grooved catheters). | The PORGES paediatric<br>silicone Foley catheter is<br>used to pass fluids or from<br>the urinary tract for<br>children. |
## Table 2: Comparison between Subject Device and Predicate Device
{7}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Coloplast. The logo consists of a blue circle with horizontal lines on the left side and the word "Coloplast" in blue on the right side. The font of the word "Coloplast" is bold and sans-serif.
1601 West River Road North
| Minneapolis, MN 55411 | | | |
|------------------------------------|---------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|
| Comparison<br>Element | Device<br>Folysil® Silicone Catheters | Primary Predicate Device<br>Porges Folysil Silicone Foley<br>Catheters<br>K013174 | Secondary Predicate<br>Device<br>Modified Pediatric<br>Silicone Foley Catheters<br>K900031 |
| | Bladder instillation of<br>physiological saline<br>solution. | The 3-way PORGES Folysil<br>silicone Foley catheter is used<br>for urethral urinary<br>catheterization to allow short<br>term drainage of the vesical<br>urines, irrigation/injection<br>following surgery. | |
| Target<br>Population | Pediatric and Adult, male and<br>female | Pediatric and Adult, male and<br>female | Pediatric patients. |
| Duration<br>of<br>Use/Indwell time | Up to 30 days. | Up to 30 days. | Up to 30 days. |
| Common Name | Catheter, Retention Type,<br>Balloon | Catheter, Retention Type,<br>Balloon | Catheter, Retention Type,<br>Balloon |
| French Size | 6 - 24 Fr | 12 - 24 Fr | 6-10 Fr |
| Length | 25 - 41 cm | 25 - 41 cm | 30 — 35 cm |
| Lumen | 2-way | 2-way or 3-way | 2-way |
| Balloon | Yes | Yes | Yes |
| Balloon Size | 1.5mL-30mL | 10mL-30mL | 1.5 - 3mL |
| Single Use | Yes | Yes | Yes |
| Sterile | Yes, via EO | Yes, via EO | Yes, via EO |
| Main<br>Shaft<br>Material | Silicone | Silicone | Silicone |
| Shelf Life | 5 years | 5 years | 5 years |
| Radiopaque line | Yes | Yes, (only for neobladder<br>catheters) | - |
| Eyes | Up to 6 | Up to 6 | up to 6 |
| Tip Shapes | Tiemann (coude tip), straight<br>cylindrical," Over the guide"<br>(straight open), straight round | Tiemann (coude tip), straight<br>cylindrical, "Over the guide"<br>(straight open), straight round | Tiemann (coude tip),<br>straight cylindrical," Over<br>the guide" (straight open),<br>straight round |
| Stylet/Introducer | Yes (for pediatric sizes),<br>Looped stylet | Not for adult sizes | Yes (for pediatric sizes),<br>Straight stylet |
As evidenced by the above table, both the subject and the predicate devices have same intended use, but the subject and predicate devices have different technological characteristics. The differences in technological characteristics between the subject and the predicate does not raise different questions of safety or effectiveness.
#### VII. PERFORMANCE DATA
Device performance data was provided in support of the substantial equivalence determination.
{8}------------------------------------------------
### Biocompatibility Testing
Biocompatibility testing was conducted based upon ISO 10993-1 (2018): Biological evaluation of medical devices - Part 1: "Evaluation and testing within a risk management process" and FDA Guidance document (2023). "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" | FDA". The subject device is an externally communicating device in contact with mucosal tissue for 30 days or more and following tests were conducted in accordance with the relevant ISO 10993 standards:
- Cytotoxicity
- Irritation ●
- Sensitization ●
- Acute Systemic Toxicity ●
- Material-mediated Pyrogenicity ●
- Genotoxicity Ames and Mouse Lymphoma ●
- Implantation 4 and 13 weeks ●
- Systemic toxicity 13 weeks ●
The results of these tests demonstrate that the subject device is biocompatible.
### Bench Testing
The following tests were performed per FDA recognized standard ASTM F623-19, ISO 20696:2018 and FDA guidance document (2020), "Conventional Foley Catheters - Performance Criteria for Safety and Performance Based Pathway | FDA".
- Visual tests ●
- Catheter integrity test
- Balloon size and shaft size ●
- Flow rate through drainage lumen ●
- Balloon integrity (resistance to rupture)
- Inflated balloon response to traction
- . Balloon volume maintenance
- Deflation reliability test (failure to deflate) ●
- Funnel security of fit test
- Tensile test on funnel junction ●
- Tensile test on tip
- Simulation of use test
- Kink resistance test
- Radiopacity test
- Tensile test on valve junction .
All pre-determined acceptance criteria were met for 14-24 Fr. catheters. The performance testing on the catheter sizes 6-12 Fr. was conducted according to the sponsor's internal test protocols as these catheter sizes are not covered under the above referenced standards. The protocol and acceptance criteria for these tests were scientifically adequate and clinically relevant, and the acceptance criteria were met for these tests as well.
{9}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the Coloplast logo and address. The logo is a blue circle with white lines inside. The text above the logo says "K233411 Page 5 of 5". The address is "1601 West River Road North Minneapolis, MN 55411".
### Sterility, Packaging and Distribution
The Folysil® Silicone Catheters were validated to be sterilized using ethylene oxide in accordance with ANSVAAMI/ISO 11135:2014. To support the claimed shelf life, the Folysil Silicone Catheters were subjected to bench performance testing, transportation testing and package integrity testing to demonstrate that the product and package would be undamaged throughout the product life and maintain device sterility.
### VIII. CONCLUSIONS
Based on the information presented in this submission, it can be concluded that the subject device is substantially equivalent to the predicates.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.