K233398 · Oertli Instrumente AG · HQC · Jun 27, 2024 · Ophthalmic
Device Facts
Record ID
K233398
Device Name
Faros Surgical System
Applicant
Oertli Instrumente AG
Product Code
HQC · Ophthalmic
Decision Date
Jun 27, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.4670
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Faros is a surgery system used by ophthalmologists during cataract surgery. It is designed for use in anterior segment procedures that require simultaneous lens fragmentation, irrigation, and aspiration, as well as ancillary functions such as vitreous cutting along with bipolar diathermy. The Faros is used for these surgical interventions in the anterior eye segment: Irrigation and aspiration (I/A function) Ultrasound phaco (PHACO function) Bipolar diathermy for hemostasis and coaptation of the conjunctiva (DIA function) Bipolar diathermic capsulotomy (CAPS function) Operation of a vitrectomy instrument (VIT function) The device may only be used with the instruments recommended and supplied by Oertli.
Device Story
Faros Surgical System is an AC-powered, tower-like unit for anterior segment ophthalmic surgery; primarily cataract removal. System inputs include user control via foot pedal and control panel; fluidic system management; and surgical handpieces. Principle of operation: high-frequency ultrasound (28 kHz) for lens emulsification; peristaltic pump for irrigation/aspiration; pneumatic guillotine for vitrectomy; 500 kHz bipolar diathermy for hemostasis/capsulotomy. Used in clinical settings by ophthalmologists. Output: emulsified lens material removal; hemostasis; tissue cutting. Healthcare providers monitor system state and control surgical parameters via the interface. Benefits include efficient lens removal and ancillary surgical capabilities during cataract procedures.
Clinical Evidence
Bench testing only. No clinical data provided. Compliance demonstrated via electrical safety (IEC 60601-1, IEC 60601-2-2, IEC 80601-2-58), EMC (IEC 60601-1-2), software validation (IEC 62304), cybersecurity (AAMI SW96, IEC 81001-5-1), usability (IEC 62366-1), and biocompatibility (ISO 10993-1, ISO 10993-5).
Technological Characteristics
AC-powered tower unit; peristaltic pump; 28 kHz longitudinal ultrasonic phaco; 500 kHz bipolar diathermy; pneumatic guillotine vitrectomy. Connectivity: foot pedal, handpiece ports. Software: continuous state monitoring and control signal generation. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2, IEC 80601-2-58, IEC 62304, ISO 10993.
A phacofragmentation system is an AC-powered device with a fragmenting needle intended for use in cataract surgery to disrupt a cataract with ultrasound and extract the cataract.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged in a circular pattern. To the right of the seal, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on top and "ADMINISTRATION" below. The logo is simple and professional, reflecting the FDA's role in regulating food and drugs.
June 27, 2024
Oertli Instrumente AG % Meagan Fagan Principal Consultant, CardioMed Device Consultants LLC CardioMed Device Consultants LLC 1783 Forest Drive Suite 254 Annapolis, Maryland 21401
Re: K233398
Trade/Device Name: Faros Surgical System Regulation Number: 21 CFR 886.4670 Regulation Name: Phacofragmentation System Regulatory Class: Class II Product Code: HQC, HQE, GEI Dated: May 28, 2024 Received: May 28, 2024
Dear Meagan Fagan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{1}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product: and 21 CFR 820.100. Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
3
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Claudine H. Krawczyk -S
Claudine Krawczyk Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K233398
Device Name Faros Surgical System
#### Indications for Use (Describe)
The Faros is a surgery system used by ophthalmologists during cataract surgery. It is designed for use in anterior segment procedures that require simultaneous lens fragmentation, irrigation, as well as ancillary functions such as vitreous cutting along with bipolar diathermy.
Faros is used for these surgical interventions in the anterior eye segment:
- Irrigation and aspiration (I/Afunction)
- Ultrasound phaco (PHACO function)
- Biopolar diathermy for hemostasis and coaptation of the conjunctiva (DIA function)
- Bipolar diathermic capsulotomy (CAPS function)
- Operation of a vitrectomy instrument (VIT function)
The device may only be used with the instruments recommended and supplied by Oertli.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span style="font-size:10pt">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------------------------------------------|
| <span style="font-size:10pt">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Oerli Switzerland. The logo is in a dark brown color, except for a red square above the "i" in Oerli. Below the logo, the word "SWITZERLAND" is written in a smaller font.
# 510(k) Summary
(as required by 21 CFR 807.92)
#### Faros Surgical System
#### 510(k) K233398
| Date Prepared | June 20th, 2024 |
|--------------------------|------------------------------------------------------------------------------|
| Applicant | Oertli Instrumente AG<br>Hafnerwisenstrasse 4<br>9442 Berneck<br>Switzerland |
| Contact Name: | Ms. Marianne Kesseli |
| Title: | Regulatory Affairs Manager |
| Email: | marianne.kesseli@oertli-instruments.com |
| Telephone: | +41 71 747 42 74 |
| Name of the Device | Faros Surgical System |
| Common Name | Phacofragementation Unit |
| Classification Name | Phacofragmentation System (21 CFR 886.4670) |
| Regulatory Class | II |
| Product Code | HQC |
| Subsequent Product Codes | HQE, GEI |
| Predicate Device | CataRhex 3 Cataract Surgery System (K133562) |
#### Intended Use/Indications for Use:
The Faros is a surgery system used by ophthalmologists during cataract surgery. It is designed for use in anterior segment procedures that require simultaneous lens fragmentation, irrigation and aspiration, as well as ancillary functions such as vitreous cutting along with bipolar diathermy.
Faros is used for these surgical interventions in the anterior eye segment:
- . Irrigation and aspiration (I/A function)
- . Ultrasound phaco (PHACO function)
- . Bipolar diathermy for hemostasis and coaptation of the conjunctiva (DIA function)
- Bipolar diathermic capsulotomy (CAPS function) ●
- . Operation of a vitrectomy instrument (VIT function)
The device may only be used with the instruments recommended and supplied by Oertli.
#### Device Description
The Faros Surgery System is designed for use in anterior segment eye surgery, primarily for removal of cataractous lenses. The system generates high-frequency ultrasound to emulsify the eye's natural lens,
{5}------------------------------------------------
which is then aspirated using fluidics. The system also includes functions for bipolar diathermy (for hemostasis, conjunctival coaptation, and capsulotomy), as well as for anterior vitrectomy, a procedure associated with cataract surgery.
The Faros Surgery System consists of a main unit (an AC-powered tower-like device with a control panel governing surgical functions, and ports to connect surgical instruments), with a mounted infusion pole and a foot pedal used by the surgeon to control system functions. An irrigation tubing system is mounted onto the main unit. Surgical handpieces connected to the main unit are used to perform irrigation and aspiration, ultrasound phacoemulsification, bipolar diathermy, bipolar diathermic capsulotomy, and anterior vitrectomy functions.
#### Comparison with Predicate Device
Comparison of the Faros Surgery System and the predicate device shows that the technological characteristics of the subject device (Faros Surgery System) such as the software and operating principle are similar to the currently marketed predicate device. The intended use of the subject device and the predicate are the same.
| | Subject Device - Faros | Predicate Device - CataRhex 3 |
|------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Oertli Instrumente AG | Oertli Instrumente AG |
| 510(k) | K233398 | K133562 |
| Indication for<br>Use/Intended Use | The Faros Surgery System is used by<br>ophthalmologists during cataract surgery. It is<br>designed for use in anterior segment<br>procedures that require simultaneous lens<br>fragmentation, irrigation, and aspiration, as<br>well as ancillary functions such as vitreous<br>cutting along with bipolar diathermy.<br><br>The Faros Surgery System is used for these<br>surgical interventions in the anterior eye<br>segment:<br>Irrigation and aspiration (I/A function) Ultrasound phaco (PHACO function) Biopolar diathermy for hemostasis and<br>coaptation of the conjunctiva (DIA<br>function) Bipolar diathermic capsulotomy (CAPS<br>function) Operation of a vitrectomy instrument (VIT<br>function) The device may only be used with instruments<br>recommended and supplied by Oertli<br>Instrumente AG. | The CataRhex 3 is a surgery system used by<br>ophthalmologists during cataract surgery. It is<br>designed for use in anterior segment procedures<br>that require simultaneous lens fragmentation,<br>irrigation and aspiration, as well as ancillary<br>functions such as vitreous cutting along with bipolar<br>diathermy.<br><br>The CataRhex 3 is used for surgical interventions in<br>the anterior eye segment:<br>Irrigation and aspiration (I/A function) Ultrasound phaco (PHACO function) Biopolar diathermy for hemostasis and<br>coaptation of the conjunctiva (DIA function) Bipolar diathermic capsulotomy (CAPS<br>function) Operation of a vitrectomy instrument (VIT<br>function) The device may only be used with instruments<br>recommended and supplied by Oertli Instrumente<br>AG. |
| Hardware Design | Main unit, tower-like device<br>Transportable: mobile<br>Foot pedal with additional switches<br>Control panel with backlit keys, with<br>additional keys<br>Automated irrigation pole<br>Colour coded ports for hand pieces<br>Voltage indicator LED<br>Fluidic system<br>Optional instrument table | Main unit, compact console<br>Transportable: portable<br>Foot pedal<br>Control panel with backlit keys<br>Manual irrigation pole<br>Colour coded ports for hand pieces<br>Fluidic system |
| Software | Software features:<br>continuous state of device monitoring control signals generation interfacing the parts of the device | Software features:<br>continuous state of device monitoring control signals generation interfacing the parts of the device |
| Electrical data | Mains voltage: 100 - 240 VAC<br>Mains frequency: 50 - 60 Hz<br>Power consumption: 150 VA<br>Operating mode: continuous | Mains voltage: 100 - 240 VAC<br>Mains frequency: 50/60 Hz<br>Power consumption: 280 VA<br>Operating mode: continuous |
| Irrigation / Aspiration | Gravity feed irrigation | Gravity feed irrigation |
{6}------------------------------------------------
| | Subject Device - Faros | Predicate Device - CataRhex 3 |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|
| Function | I/A pump principle: peristaltic<br>I/A modes | I/A pump principle: peristaltic<br>I/A modes |
| Phaco Function | Ultrasonic emulsification<br>Oscillation: longitudinal<br>Operating frequency: 28 kHz<br>Phaco modes<br>Preset aspiration mode called SPEEP | Ultrasonic emulsification<br>Oscillation: longitudinal<br>Operating frequency: 28 kHz<br>Phaco modes<br>Preset aspiration mode called CORTEX |
| Diathermy Function | Bipolar diathermy<br>Capsulotomy<br>Diathermy modes<br>Operating frequency: 500 kHz | Bipolar diathermy<br>Capsulotomy<br>Diathermy modes<br>Operating frequency: 500 kHz |
| Vitrectomy Function | Cutting principle: pneumatic<br>Push and pull movement<br>Cutting design: guillotine | Cutting principle: pneumatic<br>Push and pull movement<br>Cutting design: guillotine |
#### Summary of Performance Data
The Faros Surgery System and the predicate device. CataRhex 3. have similar safety, effectiveness, and performance profiles. Technological characteristics affecting surgical performance are similar to those of the predicate CataRhex 3. Non-clinical testing has demonstrated that the Faros Surgery System complies with specifications and requirements established for the Faros Surgery System and the proposed subject device is substantially equivalent to the predicate device.
#### Electrical safety and electromagnetic compatibility (EMC)
The Faros Surgery System electrical safety and electromagnetic compatibility has been evaluated and found to be in compliance with:
- ANSI AAMI ES60601-1:2005/(R)2012 & A1:2012. C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021] Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) [Including Amendment 2 (2021)]
- . FDA Guidance Electromagnetic Compatibility (EMC) of Medical Devices, 2022-06
- . IEC 60601-1-2 Edition 4.1 2020-09 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- . IEC 60601-2-2 Edition 6.0 2017-03 Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories
- . IEC 80601-2-58 Edition 2.0 2014-09 Medical electrical equipment - Part 2-58: Particular requirements for basic safety and essential performance of lens removal devices and vitrectomy devices for ophthalmic surgery [Including AMENDMENT 1 (2016)]
#### Software
The Faros Surgery System software has been evaluated and found to be in compliance with:
- FDA Guidance, Content of Premarket Submissions for Device Software Functions, 2023-06 .
- . IEC 62304 Edition 1.1 2015-06 Medical device software - Software life cycle processes
{7}------------------------------------------------
### Cybersecurity
The Faros Surgery System cybersecurity has been evaluated and found to be in compliance with:
- . FDA Guidance Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions, 2023-09
- . FDA Guidance for Post-market Management of Cybersecurity in Medical Devices, 2016-12-28
- AAMI TIR57:2016 . Principles for medical device security - Risk management
- . AAMI TIR97:2019 Principles for medical device security - Postmarket risk management for device manufacturers
- . ANSI AAMI SW96:2023 Standard for medical device security - Security risk management for device manufacturers
- . IEC 81001-5-1 Edition 1.0 2021-12 Health software and health IT systems safety effectiveness and security - Part 5-1: Security -Activities in the product life cycle
- . ISO/IEC 29147:2018-10 Information technology - Security techniques - Vulnerability disclosure
- . ISO/IEC 30111:2019-10 Information technology - Security techniques - Vulnerability handling processes
- . FIRST CVSS v3.X Common Vulnerability Scoring System version 3.X
#### Usability
The Faros Surgery System usability has been evaluated and found to be in compliance with:
- FDA Guidance Applying Human Factors and Usability Engineering to Medical Devices, 2016-02
- . IEC 62366-1 Edition 1.1 2020-06 Medical devices Part 1: Application of usability engineering to medical devices, including Amendment 1
- . IEC 60601-1-6 Edition 3.2 2020-07 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
#### Biocompatibility
The biocompatibility of the Faros Surgery System and its components and accessories was found to be in compliance with:
- FDA Guidance Use of International Standard ISO 10993-1, "Biological evaluation of medical ● devices - Part 1: Evaluation and testing within a risk management process, 2023-09
- ISO 10993-1 Fifth edition 2018-08 ● Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
{8}------------------------------------------------
- . ISO 10993-5 Third edition 2009-06-01 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
# Processing of Reusable Accessories (Moist Heat)
The reusable accessories of Faros Surgery System were found to be in compliance with:
- . ISO 17664-1 Second edition 2021-07 Processing of health care products - Information to be provided by the medical device manufacturer for the processing of medical devices
### Sterility of Accessories
The sterile accessories of Faros Surgery System were found to be in compliance with:
- . FDA Guidance Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile, 2024-01
#### Conclusion
The Faros surgery system has the same intended use, similar design and functional features as the predicate device. Conclusions drawn from the non-clinical tests demonstrate that the Faros Surgery System is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.