Mochida Nerve Cuff is indicated for the repair of peripheral nerve discontinuities where gap closure can be achieved by flexion of the extremity, or the management of peripheral nerve injuries in which there has been no substantial loss of nerve tissue.
Device Story
Mochida Nerve Cuff is a biodegradable, resorbable, suturable sheet composed of polyglycolic acid (PGA) embedded in covalently cross-linked alginate gel. Device is used by surgeons to repair peripheral nerve injuries; it is cut to size and wrapped around the damaged, severed, or excised nerve section. It acts as a physical scaffold to support the growth of axons and Schwann cells while protecting the regeneration site from scar tissue ingrowth. The device remains in the body, degrading over time as the nerve repairs. It is supplied sterile for single-use. Clinical benefit includes facilitating nerve regeneration in patients with peripheral nerve discontinuities or injuries.
Clinical Evidence
No human clinical data. Evidence consists of comparative bench testing (tensile strength, suture retention, stiffness) and two in vivo animal studies using a rat sciatic nerve model. Studies evaluated nerve repair, degradation, and local tissue response at 1, 8, 13, and 26 weeks compared to predicate devices and sham controls. Results showed similar safety and performance profiles. Biocompatibility testing (ISO 10993) confirmed non-cytotoxic, non-sensitizing, non-irritant, non-pyrogenic, and non-genotoxic status.
Technological Characteristics
Materials: Sodium Alginate and Polyglycolic acid (PGA). Form: Freeze-dried, pressed sponge sheet (5.5 cm x 5.5 cm). Principle: Biodegradable scaffold for nerve axon regeneration. Energy: None. Connectivity: None. Sterilization: eBeam. Single-use. Suturable. Resorbable.
Indications for Use
Indicated for repair of peripheral nerve discontinuities where gap closure is achievable via extremity flexion, or management of peripheral nerve injuries without substantial loss of nerve tissue.
Regulatory Classification
Identification
A nerve cuff is a tubular silicone rubber sheath used to encase a nerve for aid in repairing the nerve (e.g., to prevent ingrowth of scar tissue) and for capping the end of the nerve to prevent the formation of neuroma (tumors).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the date June 21, 2024. The month is June, the day is the 21st, and the year is 2024. The date is written in a clear, legible font. The text is black on a white background.
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Mochida Pharmaceutical Co., Ltd. % Kelliann Payne Partner Hogan Lovells U.S. LLP 1735 Market Street, Suite 2300 Philadelphia, Pennsylvania 19103
Re: K233322
Trade/Device Name: Mochida Nerve Cuff Regulation Number: 21 CFR 882.5275 Regulation Name: Nerve Cuff Regulatory Class: Class II Product Code: JXI Dated: May 24, 2024 Received: May 24, 2024
Dear Kelliann Payne:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Image /page/1/Picture/8 description: The image shows a digital signature. The signature is for Adam D. Pierce -S. The date of the signature is 2024.06.21. The time of the signature is 17:44:13 -04'00'.
Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices
{2}------------------------------------------------
OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/20/2023 See PRA Statement on last page
510(k) Number (if known)
K233322
Device Name
Mochida Nerve Cuff
Indications for Use (Describe)
Mochida Nerve Cuff is indicated for the repair of peripheral nerve discontinuities where gap closure can be achieved by flexion of the extremity, or the management of peripheral nerve injuries in which there has been no substantial loss of nerve tissue.
Type of Use (Select one or both, as applicable)
図 Prescription Use (Part 21 CFR 801 Subpart D)
□ Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
# FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### 510(k) SUMMARY Mochida's Nerve Cuff
### Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Mochida Pharmaceutical Co., Ltd. 1-7 Yotsuva. Shiniuku-ku Tokyo, Japan Phone: +81-3-3225-6306 Fax: +81-3-3225-5886
Contact: Akiko Koarai, Head of Medical Device Quality Management and Regulatory Affairs Department
Date Prepared: June 20, 2024
#### Name of Device
Trade name: Mochida Nerve Cuff Common name: Cuff. Nerve Classification name: Nerve cuff (21 CFR 882.5275) Regulatory class: II Product code: JXI
#### Predicate Devices
Collagen Matrix, Inc.'s Collagen Nerve Wrap (K060952) (Predicate device) Collagen Matrix, Inc.'s Collagen Nerve Cuff (K012814) (Predicate device)
#### Device Description
The Mochida Nerve Cuff is made from a polyglycolic acid (PGA) sheet embedded into covalently cross-linked alginate gel, which is produced into a sponge form by freeze-drying and pressing into sheets.
The device is designed to repair peripheral nerve function by supporting the extension of the axon by providing a scaffold for growth of axons and Schwann cells following the degradation/absorption process after nerve injury. The device is to be implanted to cover the damaged, severed or excised section of the nerve axons.
When used, the device is cut according to the size of the nerve axon injury site, and fixed in place to cover the site.
#### Indications for Use
Mochida Nerve Cuff is indicated for the repair of peripheral nerve discontinuities where gap closure can be achieved by flexion of the extremity, or the management of peripheral nerve injuries in which there has been no substantial loss of nerve tissue.
{5}------------------------------------------------
### Comparison of Technological Characteristics with the Predicate Device
The subject device has similar technological characteristics as the predicate devices. Both are smooth, flexible products that are made of biodegradable materials, and are supplied in sterilized packages and for single use. Both devices aim to achieve the function of nerve repair through supporting nerve axon regeneration by providing a scaffold for growth of axons following the degradation/absorption process after nerve injury.
Both the subject and the predicate devices are made of biodegradable material. Specifically, the subject device consists of covalently cross-linked alginate gel, which is biodegradable, similar to the collagen material used in the predicate device.
Both the subject and the predicate devices are supplied in sheet/wrap forms. Both devices have the same fundamental science, which is to serve as scaffolds for Schwann and nerve cells to grow and to protect the regrowth site from the environment, such as ingrowth of scar tissue.
Both the subject and predicate devices remain in the body for a period of time until the injured nerves are repaired. The comparative animal studies showed similar host reaction and nerve regeneration between the subject device and Collagen Nerve Cuff predicate. Both the subject and predicate devices are provided sterile and are intended for single use. Sterilization validation testing demonstrated that the sterilization method used for the subject device is adequate to achieve sterility of the device. In addition, biocompatibility testing showed the biological safety of the device, further supporting the adequacy of the sterilization.
| | Mochida Nerve Cuff<br>(Subject device) | Collagen Matrix, Inc.'s<br>Collagen Nerve Wrap<br>(K060952)<br>(Predicate device) | Collagen Matrix, Inc.'s<br>Collagen Nerve Cuff<br>(K012814)<br>(Predicate device) |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulation | 21 CFR 882.5275 | 21 CFR 882.5275 | 21 CFR 882.5275 |
| Product Code | JXI | JXI | JXI |
| Resorbable | Yes | Yes | Yes |
| Suturable | Yes | Yes | Yes |
| Material | Sodium Alginate<br>Polyglycolic acid | Collagen | Type I Collagen |
| Indications<br>For Use | Mochida Nerve Cuff is<br>indicated for the repair<br>of peripheral nerve<br>discontinuities where<br>gap closure can be<br>achieved by flexion of<br>the extremity, or the<br>management of<br>peripheral nerve injuries<br>in which there has been<br>no substantial loss of<br>nerve tissue. | Collagen Nerve Wrap is<br>indicated for the<br>management of peripheral<br>nerve injuries in which<br>there has been no<br>substantial loss of nerve<br>tissue and where gap<br>closure can be achieved<br>by flexion of the extremity. | Collagen Nerve Cuff is<br>intended for use in<br>repair of peripheral<br>nerve discontinuities<br>where gap closure can<br>be achieved by flexion<br>of the extremity. |
{6}------------------------------------------------
| | Mochida Nerve Cuff<br>(Subject device) | Collagen Matrix, Inc.'s<br>Collagen Nerve Wrap<br>(K060952)<br>(Predicate device) | Collagen Matrix, Inc.'s<br>Collagen Nerve Cuff<br>(K012814)<br>(Predicate device) |
|-------------------------|----------------------------------------------------------------------|-----------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|
| Physical<br>structure | Sheet | Wraps (Pre-rolled membrane) | Tubular matrix |
| Size | 5.5 cm × 5.5 cm<br>(Long ×Wide)<br>Can be cut to<br>appropriate size | 4 x 25 mm | Diameter: 1.0, 1.5, 2.0,<br>2.5, 3.0, 3.5, 4.0, 5.0,<br>6.0, 10.0 mm;<br>Lengths of 1, 2 and 3<br>cm. |
| | | 4 x 50 mm | |
| | | 6 x 25 mm | |
| | | 6 x 50 mm | |
| | | 12 x 25 mm | |
| | | 12 x 50 mm | |
| | | Can be trimmed or<br>shaped to appropriate<br>size | |
| | | Biocompatibility | |
| Preparation for<br>use | Use dry or minimally<br>hydrated | Hydrated for 5 minutes in<br>saline | Hydrated for 5 minutes<br>in saline |
| Reusable | Single-use | Single-use | Single-use |
| Sterilization<br>Method | eBeam | Gamma irradiation | Gamma irradiation |
| Packaging | Double Pouch | Double blister | not publicly available |
| Shelf life | 4 months | 3 years | 3 years |
# Performance Data
The technical characteristics of the subject and predicate devices have been tested using comparative bench tests, which showed very similar results. In addition, performance testing, including biocompatibility testing and animal testing, demonstrates similar safety and performance of the subject device and its predicates. The following performance testing has been conducted:
### Stability testing of the product and the packaging
Stability of the Mochida Nerve Cuff was demonstrated by the real-time and accelerated aging methodologies for the product and the packaging materials.
### Transportation testing
The transportation testing was conducted to demonstrate the ability of the Mochida Nerve Cuff and its packaging to withstand the distribution environment. The testing was conducted according to ASTM D4169-22.
### Pyrogenicity testing
Mochida Nerve Cuff is labeled as non-pyrogenic. Endotoxin testing was conducted to demonstrate that the product is non-pyrogenic.
{7}------------------------------------------------
# Biocompatibility testing
Biocompatibility testing was conducted according to FDA Guidance "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process.""
| Test | Test Summary | Conclusions |
|---------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| Chemical<br>characterization | The chemical characterization was<br>implemented based on ISO 10993-18. The<br>extractable and leachable that could be<br>released from the test article were identified and<br>quantified. | The study did not identify<br>any analytes that represent<br>a significant toxicological<br>risk. |
| Cytotoxicity | The device was evaluated for potential cytotoxic<br>effects following ISO 10993-5 guidelines. | The cytotoxicity test results<br>do not raise any safety<br>concerns. |
| Sensitization | The device was evaluated for the potential to<br>cause sensitization in guinea pigs based on ISO<br>10993-10. The test animals showed no<br>evidence of inducting sensitization. | Non-sensitizing |
| Irritation | The device was evaluated for the potential to<br>cause irritation following intracutaneous<br>injection in rabbits based on ISO 10993-23. The<br>extracts of the test article show no evidence of<br>irritation. | Non-irritant |
| Acute systemic<br>toxicity | The device was evaluated for acute systemic<br>toxicity in mice based on ISO 10993-11. No<br>change was observed in general conditions or<br>in necropsy. | No acute systemic toxicity |
| Subacute<br>toxicity | The device was evaluated for acute systemic<br>toxicity in rats based on ISO 10993-6 and ISO<br>10993-11. No abnormality occurred in general<br>conditions or surgical sites after implantation. | No subacute systemic<br>toxicity |
| Pyrogenicity | The device was evaluated for the potential to<br>induce a pyrogenic response following<br>intravenous injection in rabbits based on ISO<br>10993-11. The total rise of temperature was<br>within acceptable range. | Non-pyrogenic |
| Genotoxicity -<br>Bacterial<br>Reverse<br>Mutation | The device was evaluated for the potential to<br>induce reverse mutations in bacteria per ISO<br>10993-3.<br>The result indicated that the extracts of the<br>device were not genotoxic. | Non-genotoxic |
| Genotoxicity -<br>Chromosomal<br>Abnormality | The device was evaluated for the potential of<br>clastogenicity per ISO 10993-3. It showed that<br>this device is clastogenicity negative in cultured<br>mammalian cells. | Non-genotoxic |
| Test | Test Summary | Conclusions |
| Hemolysis<br>testing | The device was evaluated for hemolytic<br>potential when in contact with blood based on<br>ISO 10993-4 and ASTM F756 standard. Under<br>the conditions of this study, data obtained were<br>considered not analyzable. | No conclusion was drawn<br>for hemolytic potential of the<br>treated test article extract. |
| Subchronic<br>toxicity, Chronic<br>toxicity, and<br>Carcinogenicity | These endpoints were addressed with a<br>chemical characterization/toxicological risk<br>assessment approach. | Justified by chemical<br>characterization/toxicological<br>risk assessment |
{8}------------------------------------------------
# Bench testing
Comparative bench testing of Mochida Nerve Cuff and the predicate devices, including tensile strength, suture retention strength, pH, stiffness, degradation, swelling rate, and porosity testing, was conducted.
| Test | Test Method Summary | Results |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|
| Tensile strength | The device is cut into a piece, and then pull both ends of<br>the test piece and measure the strength when it breaks. | Pass |
| Suture retention<br>strength | After passing a suture with a needle though the wetted<br>specimen, fix the suture to blocks. Pull the specimen,<br>and measure the strength at which the device breaks. | Pass |
| pH | Measure the pH of the sample extract of the device. | N/A<br>(For<br>reference) |
| Stiffness | Wrap sample around the smooth shaft and record any<br>observation of cracking or breaking. | Pass |
| Degradation | Immerse the devise in a solution and measure the<br>weight of the devise after picking up from the solution<br>after a period of time. | N/A<br>(For<br>reference) |
| Swelling rate | Measure the absorbed amount of water per device<br>weight. | N/A<br>(For<br>reference) |
| Porosity | Porosimetry | N/A<br>(For<br>reference) |
# Animal study
In addition to the bench testing, the subject device has also been tested in two animal studies using the rat sciatic nerve model. The first study evaluated the nerve repair process, article degradation and local tissue response to Mochida Nerve Cuff when applied to a sciatic nerve transection (nerve gap) in the rat for 1, 8, and 26 weeks, as compared to the Collagen Nerve Cuff predicate. The second study evaluated the article degradation and local tissue response to Mochida Nerve Cuff when applied to a sciatic nerve without transection (no substantial loss of nerve tissue) in the rat for 1, 8, and 13 weeks, as compared to the Collagen Nerve Wrap predicate. In addition to the test article and control article (predicate device), the studies also
{9}------------------------------------------------
included a sham arm to evaluate the response that could be elicited from the surgical procedures.
For all evaluated endpoints in both studies, including neurological findings and histology findings, the results demonstrated that the Mochida Nerve Cuff has the similar safety profile and performance as the predicate devices. Therefore, the results of the in vivo animal study support the substantial equivalence.
### Conclusions
The Mochida Nerve Cuff and its predicate devices have intended use and similar indications, technological characteristics and principles of operation. The minor technological differences do not present any new issues of safety or effectiveness. Performance testing demonstrated that the subject device performs similarly as the predicate devices. Thus, the Mochida Nerve Cuff is substantially equivalent to the predicate devices.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.