AXIOS Stent and Electrocautery-Enhanced Delivery System (6mmx8mm); AXIOS Stent and Electrocautery-Enhanced Delivery System (8mmx8mm); AXIOS Stent and Electrocautery-Enhanced Delivery System (10mmx10mm); AXIOS Stent and Electrocautery-Enhanced Delivery System (15mmx10mm); AXIOS Stent and Electrocautery-Enhanced Delivery System (20mmx10mm); AXIOS Stent and Electrocautery-Enhanced Delivery System (15mmx15mm)
K233318 · Boston Scientific Corporation · KNS · Oct 27, 2023 · Gastroenterology, Urology
Device Facts
Record ID
K233318
Device Name
AXIOS Stent and Electrocautery-Enhanced Delivery System (6mmx8mm); AXIOS Stent and Electrocautery-Enhanced Delivery System (8mmx8mm); AXIOS Stent and Electrocautery-Enhanced Delivery System (10mmx10mm); AXIOS Stent and Electrocautery-Enhanced Delivery System (15mmx10mm); AXIOS Stent and Electrocautery-Enhanced Delivery System (20mmx10mm); AXIOS Stent and Electrocautery-Enhanced Delivery System (15mmx15mm)
Applicant
Boston Scientific Corporation
Product Code
KNS · Gastroenterology, Urology
Decision Date
Oct 27, 2023
Decision
SESE
Submission Type
Special
Regulation
21 CFR 876.4300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
6mmx8mm and 8mmx8mm stents: The AXIOS Stent and Electrocautery-Enhanced Delivery System is indicated for use to facilitate transgastric or transduodenal endoscopic drainage of symptomatic pancreatic pseudocysts ≥ 6 cm in size, that are adherent to the gastric or bowel wall and are free of solid debris. The stent is intended for implantation up to 60 days and should be removed upon confirmation of pseudocyst resolution. 10mmx10mm and 15mmx10mm stents: The AXIOS Electrocautery-Enhanced-Stent and Delivery System is indicated for use to facilitate transgastric or transduodenal endoscopic drainage of symptomatic pancreatic pseudocysts ≥ 6 cm in size and walled-off necrosis ≥ 6 cm in size that are adherent to the gastric or bowel wall. Once placed, the AXIOS Stent functions as an access port allowing passage of standard and therapeutic endoscopes to facilitate debridement, irrigation and cystoscopy. The stent is intended for implantation up to 60 days and should be removed upon confirmation of pseudocyst or walled-off necrosis resolution. The AXIOS Stent and Electrocautery-Enhanced Delivery System is indicated for use to facilitate transgastric or transduodenal endoscopic drainage of the gallbladder in patients with acute cholecystitis who are at high risk for surgery. 20mmx10mm and 15mmx15mm stents: The AXIOS Electrocautery-Enhanced Stent and Delivery System is indicated for use to facilitate transgastric or transduodenal endoscopic drainage of symptomatic pancreatic pseudocysts ≥ 6 cm in size and walled-off necrosis ≥ 6 cm in size that are adherent to the gastric or bowel wall. Once placed, the AXIOS Stent functions as an access port allowing passage of standard and therapeutic endoscopes to facilitate debridement, irrigation and cystoscopy. The stent is intended for implantation up to 60 days and should be removed upon confirmation of pseudocyst or walled-off necrosis resolution.
Device Story
AXIOS Stent and Electrocautery-Enhanced Delivery System facilitates endoscopic drainage of pancreatic pseudocysts, walled-off necrosis, and gallbladder; device comprises flexible, fully-covered, self-expanding nitinol stent with dual flanges for tissue apposition; preloaded in delivery catheter with integrated handle; used with echoendoscopes (≥ 3.7 mm channel) and 0.035-inch guidewires; connects to external electrosurgical unit (ESU) for cautery-assisted access; physician operates device under endoscopic visualization; stent acts as access port for endoscope passage, debridement, and irrigation; intended for temporary implantation (up to 60 days) until resolution; sterile, single-use.
Clinical Evidence
Bench testing only. Performance testing demonstrated equivalent settings and output between the ERBE VIO 3 ESU and the baseline 300D ESU through range-finding and confirmation testing.
Technological Characteristics
Fully-covered, self-expanding braided nitinol stent; dual-flange design; MR conditional; delivery system includes catheter and manual handle; compatible with 3.7 mm+ working channel echoendoscopes and 0.035-inch guidewires; connects to IEC 60601-1-2 and IEC 60601-2-2 compliant electrosurgical units; sterile, single-use.
Indications for Use
Indicated for transgastric/transduodenal endoscopic drainage of symptomatic pancreatic pseudocysts (≥ 6 cm) or walled-off necrosis (≥ 6 cm) adherent to gastric/bowel wall; gallbladder drainage in acute cholecystitis patients at high surgical risk. Contraindicated for pseudocysts with solid debris (for 6mm/8mm stents).
Regulatory Classification
Identification
An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.
Predicate Devices
AXIOS Stent and Electrocautery-Enhanced Delivery System (DEN230019, K22011)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 27, 2023
Boston Scientific Corporation Mr. Ian Broome Senior Regulatory Affairs Specialist 100 Boston Scientific Way Marlborough, Massachusetts 01752
### Re: K233318
Trade/Device Name: AXIOS Stent and Electrocautery-Enhanced Delivery System (6 mm x8 mm); AXIOS Stent and Electrocautery-Enhanced Delivery System (8 mm x8 mm); AXIOS Stent and Electrocautery-Enhanced Delivery System (10 mm x10 mm); AXIOS Stent and Electrocautery-Enhanced Delivery System (15 mm x10 mm); AXIOS Stent and Electrocautery-Enhanced Delivery System (20 mm x10 mm); AXIOS Stent and Electrocautery-Enhanced Delivery System (15 mm x15 mm) Regulation Number: 21 CFR 876.4300 Regulation Name: Endoscopic Electrosurgical Unit and Accessories Regulatory Class: Class II Product Code: KNS Dated: September 29, 2023 Received: September 29, 2023
Dear Mr. Ian Broome:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{2}------------------------------------------------
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{3}------------------------------------------------
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
## Glenn B. Bell -S
Glenn B. Bell, Ph.D Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
| Submission Number (if known) | |
|------------------------------|--|
|------------------------------|--|
K233318
Device Name
AXIOS Stent and Electrocautery-Enhanced Delivery System (6mmx8mm) (M00553680): AXIOS Stent and Electrocautery-Enhanced Delivery System (8mmx8mm) (M00553690); AXIOS Stent and Electrocautery-Enhanced Delivery System (10mmx10mm) (M00553640); AXIOS Stent and Electrocautery-Enhanced Delivery System (15mmx10mm) (M00553650); AXIOS Stent and Electrocautery-Enhanced Delivery System (20mmx10mm) (M00553660); AXIOS Stent and Electrocautery-Enhanced Delivery System (15mmx15mm) (M00553670) Indications for Use (Describe)
6mmx8mm and 8mmx8mm stents: The AXIOS Stent and Electrocautery-Enhanced Delivery System is indicated for use to facilitate transgastric or transduodenal endoscopic drainage of symptomatic pancreatic pseudocysts ≥ 6 cm in size, that are adherent to the gastric or bowel wall and are free of solid debris. The stent is intended for implantation up to 60 days and should be removed upon confirmation of pseudocyst resolution.
10mmx10mm and 15mmx10mm stents: The AXIOS Electrocautery-Enhanced-Stent and Delivery System is indicated for use to facilitate transgastric or transduodenal endoscopic drainage of symptomatic pancreatic pseudocysts ≥ 6 cm in size and walled-off necrosis ≥ 6 cm in size that are adherent to the gastric or bowel wall. Once placed, the AXIOS Stent functions as an access port allowing passage of standard and therapeutic endoscopes to facilitate debridement, irrigation and cystoscopy. The stent is intended for implantation up to 60 days and should be removed upon confirmation of pseudocyst or walled-off necrosis resolution. The AXIOS Stent and Electrocautery-Enhanced Delivery System is indicated for use to facilitate transgastric or transduodenal endoscopic drainage of the gallbladder in patients with acute cholecystitis who are at high risk for surgery.
20mmx10mm and 15mmx15mm stents: The AXIOS Electrocautery-Enhanced Stent and Delivery System is indicated for use to facilitate transgastric or transduodenal endoscopic drainage of symptomatic pancreatic pseudocysts ≥ 6 cm in size and walled-off necrosis ≥ 6 cm in size that are adherent to the gastric or bowel wall. Once placed, the AXIOS Stent functions as an access port allowing passage of standard and therapeutic endoscopes to facilitate debridement, irrigation and cystoscopy. The stent is intended for implantation up to 60 days and should be removed upon confirmation of pseudocyst or walled-off necrosis resolution.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
{5}------------------------------------------------
| K233318 - Page 1 of 2 | | |
|------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|
| 510(k) Summary | | Prepared on: 2023-10-26 |
| Contact Details | | 21 CFR 807.92(a)(1) |
| Applicant Name | Boston Scientific Corporation | |
| Applicant Address | 100 Boston Scientific Way Marlborough MA 01752 United States | |
| Applicant Contact Telephone | +1 (508) 683-4000 | |
| Applicant Contact | Mr. Ian Broome, M.S., RAC | |
| Applicant Contact Email | ian.broome@bsci.com | |
| Device Name | | 21 CFR 807.92(a)(2) |
| Device Trade Name | AXIOS™ Stent and Electrocautery-Enhanced Delivery System (6mmx8mm) (M00553680); AXIOS™ Stent and Electrocautery-Enhanced Delivery System (8mmx8mm) (M00553690); AXIOS™ Stent and Electrocautery-Enhanced Delivery System (10mmx10mm) (M00553640); AXIOS™ Stent and Electrocautery-Enhanced Delivery System (15mmx10mm) (M00553650); AXIOS™ Stent and Electrocautery-Enhanced Delivery System (20mmx10mm) (M00553660); AXIOS™ Stent and Electrocautery-Enhanced Delivery System (15mmx15mm) (M00553670) | |
| Common Name | Endoscopic electrosurgical unit and accessories | |
| Classification Name | Unit, Electrosurgical, Endoscopic (With Or Without Accessories) | |
| Regulation Number | 876.4300 | |
| Product Code | KNS | |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| DEN230019, K22011 | AXIOS Stent and Electrocautery-Enhanced Delivery System | KNS |
| Device Description Summary | | 21 CFR 807.92(a)(4) |
AXIOS™ Stent:
The AXIOS™ Stent is a flexible, MR conditional, fully-covered, self-expanding braided nitinol stent, which comes preloaded into a delivery system. The AXIOS™ stent is designed with two flanges on each end to enable tissue plane apposition and a "saddle" in between the flanges to span the tissue implant distance.
Electrocautery-Enhanced Delivery System:
The AXIOS™ Electrocautery-Enhanced Delivery System consists of a catheter and an integrated handle with manual controls for positioning and deploying the AXIOS™ Electrocautery-Enhanced Delivery System is designed to be used in the gastrointestinal tract with commercially available echoendoscopes with a 3.7 mm diameter or larger working channel and is compatible with commercially available 0.035-inch insulated endoscopic guidewires.
{6}------------------------------------------------
The Electrocautery-Enhanced Delivery System connects with an off-the-shelf electrosurgical unit or generator that is compliant to IEC 60601-1-2 and IEC 60601-2-2.
The AXIOS™ Stent and Electrocautery-Enhanced Delivery System is provided sterile, disposable and intended for single use.
# Intended Use/Indications for Use
## 21 CFR 807.92(a)(5)
6mmx8mm and 8mmx8mm stents: The AXIOS™ Stent and Electrocautery-Enhanced Delivery System is indicated for use to facilitate transgastric or transduodenal endoscopic drainage of symptomatic pancreatic pseudocysts ≥ 6 cm in size, that are adherent to the gastric or bowel wall and are free of solid debris. The stent is intended for implantation up to 60 days and should be removed upon confirmation of pseudocyst resolution.
10mmx10mm and 15mmx10mm stents: The AXIOS™ Electrocautery-Enhanced-Stent and Delivery System is indicated for use to facilitate transgastric or transduodenal endoscopic drainage of symptomatic pancreatic pseudocysts ≥ 6 cm in size and walled-off necrosis ≥ 6 cm in size that are adherent to the gastric or bowel wall. Once placed, the AXIOS™ Stent functions as an access port allowing passage of standard and therapeutic endoscopes to facilitate debridement, irrigation and cystoscopy. The stent is intended for implantation up to 60 days and should be removed upon confirmation of pseudocyst or walled-off necrosis resolution.
The AXIOS™ Stent and Electrocautery-Enhanced Delivery System is indicated for use to facilitate transguodenal endoscopic drainage of the gallbladder in patients with acute cholecystitis who are at high risk for surgery.
20mmx10mm and 15mmx15mm stents: The AXIOS™ Electrocautery-Enhanced Stent and Delivery System is indicated for use to facilitate transgastric or transduodenal endoscopic drainage of symptomatic pancreatic pseudocysts ≥ 6 cm in size and walled-off necrosis ≥ 6 cm in size that are adherent to the gastric or bowel wall. Once placed, the AXIOS™ Stent functions as an access port allowing passage of standard and therapeutic endoscopes to facilitate debridement, irrigation and cystoscopy. The stent is intended for implantation up to 60 days and should be removed upon confirmation of pseudocyst or walled-off necrosis resolution.
# Indications for Use Comparison
The indications for use of the proposed AXIOS™ Stent and Electrocautery-Enhanced Delivery System devices remain identical to the predicate device.
# Technological Comparison
The technological characteristics of proposed AXIOS™ Stent and Electrocautery-Enhanced Delivery System are identical to the predicate device.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
Performance testing (bench) was completed to demonstrate equivalent settings for AXIOS devices when used with the ERBE VIO 3 Electrosurgical Unit (ESU).
The testing conducted was based on original clearance testing for the AXIOS Stent and Electrocautery-Enhanced Delivery System, along with consultation of the ESU manufacturer.
Two phases of ESU testing were performed:
· Range Finding, to establish ideal effect range based on baseline created with 300D ESU;
· Confirmation Testing, to determine ideal settings output and perform statistical analysis to ensure equivalent performance between both ESUs.
Based on the results of bench testing, the substantially equivalent to the predicates. The updates to labeling do not affect safety, effectiveness, performance or function of the device.
# 21 CFR 807.92(a)(6)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.