The Disposable Trocars have application in a variety of endoscopic procedures to provide a port of entry for endoscopic instruments and/or endoscope.
Device Story
Disposable trocars provide access ports for endoscopic instruments and endoscopes during minimally invasive surgery. Device consists of cannula and obturator; variants include normal, balloon-stabilized, threaded, and single-port designs. Normal trocars feature various obturator types (ingot, optical, auto-shielded, blunt) and optional accessories like filters or cleaning rings. Balloon trocars use inflatable balloons for stabilization; threaded trocars enhance stability during instrument exchange; single-port trocars allow multiple instruments through one incision. CO2 insufflation is supported via stopcock valves. Used in clinical settings by surgeons; provides stable access to abdominal/thoracic cavities while maintaining pneumoperitoneum. Single-use, EO-sterilized. Benefits include reduced surgical trauma, improved instrument handling, and maintained abdominal access.
Clinical Evidence
No clinical data. Evidence is based on bench testing and non-clinical performance evaluations. Testing included appearance, obturator compatibility, insertion/cannula stability, air leakage, balloon rigidity, and fixation retention. Biocompatibility testing followed ISO 10993 standards (cytotoxicity, sensitization, irritation, acute systemic toxicity, pyrogenicity). Sterilization and shelf-life validated per ISO/EN standards.
Technological Characteristics
Materials: Polycarbonate, silicone. Design: Cannula and obturator assembly; diameters 3-20mm, lengths 55-160mm. Energy: Mechanical (manual insertion). Connectivity: None. Sterilization: Ethylene Oxide (EO). Standards: ISO 10993 (biocompatibility), ISO 11135 (sterilization), ISO 11607 (packaging), ASTM D4169 (transportation).
Indications for Use
Indicated for use in endoscopic procedures to provide a port of entry for endoscopic instruments and/or endoscopes in patients undergoing minimally invasive surgery.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
April 9, 2024
Hangzhou Kangji Medical Instrument Co., Ltd. % Esther Zhang Shanghai Ling Fu Technology Co., Ltd. 4F No. 585-2, Wanyuan Rd. Minhang District Shanghai, Shanghai 201102 China
Re: K233263
Trade/Device Name: Disposable Trocars Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: GCJ Dated: September 28, 2023 Received: September 29, 2023
Dear Esther Zhang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{2}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely. Mark Trumbore -S Date: 2024.04.09 12:02:16 -04'00 Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{3}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
K233263
Device Name
Disposable Trocars
Indications for Use (Describe)
The Disposable Trocars have application in a variety of endoscopic procedures to provide a port of entry for endoscopic instruments and/or endoscope.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) summary K233263
## | Submitter
Device submitter: Hangzhou Kanqji Medical Instrument Co., Ltd.
No. 1668 Chuniiang East Road. Economic Development Zone. Tongly. Hangzhou, 311501, China.
Contact person: Martin Sun
Manager of Regulatory Affairs Phone: +86-0571-69901712 Fax: +86-0571-69901712 Email: martin.sun@kanqji.com
Date: April 9, 2024
## II Device
Trade Name of Device: Disposable Trocars Common Name: Trocars Regulation Number: 21 CFR 876.1500 Regulation Name: Laparoscope, General & Plastic Surgery Requlatory Class: II Product code: GCJ Review Panel: General and Plastic Surgery
## III Correspondent
Shanghai Ling Fu Technology Co., Ltd. 4F No. 585-2, Wanyuan Rd. Minhang District, Shanghai, P.R.China Contact: Esther ZHANG Email: Esther.zhang@llins-tech.com
## IV Device description
The Disposable Trocars includes Normal trocars (model A), Balloon trocars (model B), Thread trocars (model C) and Single-port trocars (model E). They are generally composed of cannula and obturator, with or without filter/cleaning ring/protector.
The Disposable Trocars have application in a variety of endoscopic procedures to provide a port of entry for endoscopic instruments and/or endoscope. If needed, the CO2 can be infused through the valve. The product is sterilized by Ethylene oxide and is a single-use product.
Normal trocars have 3 different cannulas (normal cannula, normal cannula with filter, normal cannula with cleaning ring) and 5 different obturators (Ingot-type, Optical obturator A. Optical obturator B. Auto-shielded obturator, Blunt obturator). Normal cannula consists of Cannula Body, Cannula Tube Seat, Choke Valve, Sheath Cap and Stopcock. It may be equipped with a filter or cleaning ring as an accessory. Cleaning ring is equipped inside the cannula and used for removing blood and foreign matter from the laparoscope. Filter is detachable, installed in stopcock, and used for filtering out any
{5}------------------------------------------------
foreign matter that spills out of the body during surgery. Stopcock can be used for insufflation if needed. Ingot-type obturator is easy to insertion which consists of Needle Tip, Needle Tube and Needle Seat. Optical obturator A has a transparent needle tip while Optical obturator B has a transparent needle tip and needle tube. Needle tube of these two optical obturators is hollow which allows laparoscope go inside to see if it has passed the abdominal wall. Auto-shielded obturator has a hided blade which only showing up in puncturing process. Blunt obturator has a blunt needle tip which can protect internal organ during the procedure. Normal trocars are available from 3-15mm of diameter and 65-160mm of length.
Balloon Trocars are equipped an inflatable balloon with a fixed (Balloon cannula A) or movable holder (Balloon cannula B) to provide stabilization during a laparoscopic procedure. The Cannula consists of Balloon, Cannula body, Holder A, One-way vale, Cannula Tube Seat, Stopcock, Sheath Cap and Choke Valve. Stopcock can be used for insufflation if needed. Movable holder can be slid down to adjust its height at a desired position for stabilization. There are five different obturators as same as normal trocars: lngot-type, Optical obturator A, Optical obturator B, Auto-shielded obturator, Blunt obturator. Balloon Trocars are available in from 3-15mm of diameter and 55-140mm of length.
Thread trocars consists of a thread obturator and a Thread cannula A (thoracic trocars) or B (Orthopedics trocars). Threaded cannula enhances stability when exchanging instruments. Thread cannula B consists of Cannula body, Cannula Tube Seat, Stopcock, Sheath Cap and Choke Valve while Thread cannula A does not have a stopcock. Stopcock can be used for insufflation if needed. Thoracic trocars are available in from 5-20mm of diameter and 65-85mm of length. Orthopedics trocars are designed for Orthopedics laparoscopic surgery and available in from 5-20mm of diameter and 65-120mm of length.
Single-port Trocars are designed to provide access for multiple instruments and/or endoscope to the abdominal cavity through a single incision during minimally invasive endoscopic surgery. It consists of cannula, universal seal system, a variable height protector or fixed height protector, with or without obturator. Single port cannula A and B use same materials and only different in shapes. Universal seal system A supports the free combination and arrangement of trocar with different numbers and sizes to meet different surgical needs. Obturator is used with Single port cannula A. Universal seal system B is fully transparent and visible for surgical operation. Besides, soft material makes it good for handling the endoscopic instruments and reducing surgical compression congestion to the patient skin. Universal seal system C has a larger handling space than system B. Protector, which has fixed height or variable height, is used for protect surgical incision.
Due to the different structure and function, diameter, length of the Cannula and Obturator, and package amount, there are many combinations for final product.
## V Indications for use
The Disposable Trocars have application in a variety of endoscopic procedures to provide a port of entry for endoscopic instruments and/or endoscope.
{6}------------------------------------------------
# VI Substantial Equivalence
| Model of<br>proposed<br>device | Predicate device | Manufacturer | 510(k) No. |
|--------------------------------|-------------------------------------------|--------------------------------|------------|
| K112358 | Unimax Trocar System | Unimax Medical<br>Systems Inc. | GCJ |
| K211577 | SoftFix™ Balloon Trocar | Unimax Medical<br>Systems Inc. | GCJ |
| K141715 | Glove Port | NELIS | OTJ |
| K093372 | SILS™ Port | COVIDIEN LP | GCJ |
| K073719 | ASC TriPort Laparoscopic<br>Access Device | Advanced Surgical<br>Concepts. | GCJ |
able Trocars are substantially equivalent to the predicate devices: osed Dien
# Comparison of technological characteristics with the predicate devices
{7}------------------------------------------------
| Device<br>feature | Subject Device<br>(Model A/B/C) | Predicate Device K112358<br>(Unimax Trocar System) | Predicate Device<br>K211577( <b>SoftFixTM</b><br>Balloon Trocar) | Comment |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| Class | II | II | II | Equivalent |
| Product code | GCJ | GCJ | GCJ | Equivalent |
| Regulation<br>number | 21 CFR 876.1500 | 21 CFR 876.1500 | 21 CFR 876.1500 | Equivalent |
| Intended use | The Disposable Trocars<br>have application in a variety<br>of endoscopic procedures to<br>provide a port of entry for<br>endoscopic instruments<br>and/or endoscope. | The Unimax Trocar System, Model:<br>Auto-Locking Trocar, Bladeless<br>Trocar, Visible Trocar, Hasson<br>Trocar, Dilating Trocar Secondary<br>Trocar, and Thoracic Trocar has<br>application in a variety of endoscopic<br>procedures to provide a port of<br>entry for endoscopic instruments. | The <b>SoftFixTM</b> Balloon<br>Trocar have application in a<br>variety of endoscopic<br>procedures to provide a port<br>of entry for endoscopic<br>instruments into the body<br>cavity of patients. | Equivalent. |
| Models | Normal trocars (Model A) | Bladeless, Visible and auto-<br>locking,<br>Secondary trocars | / | Different<br>Comment |
| | Balloon trocars (Model B) | Hasson trocars | Balloon trocar (bladed<br>obturator with safety lock for<br>the shield, bladeless<br>obturator, optical obturator,<br>hasson obturator, cannula<br>sleeve with a stopcock for<br>insufflation) | 1 |
| | Thread trocars (Model C) | Thoracic trocars | / | |
| | | Dilating trocars | | |
| Device structure | Cannula with or without<br>cleaning ring/filter, balloon<br>cannula, cannula with<br>stopcock for insufflation,<br>obturator | Cannula with stopcock for insufflation,<br>obturator | Cannula with a stopcock for<br>insufflation, obturator | |
| Specification | Diameter: 3-15mm<br>Length: 60-160mm | Diameter: 3-15mm<br>Length: 65-150mm | Diameter: 5-12mm<br>Length: 70-100mm | Similar<br>Comment 2 |
| Reuse durability | Single use | Single use | Single use | Equivalent |
| Sterilization | EO | EO | EO | Equivalent |
| Single use | Yes | Yes | Yes | Equivalent |
| Performance | Obturator Compatibility;<br>Insertion & Cannula Stability;<br>Air Leakage; balloon rigidity;<br>fixation device retention | Obturator Compatibility;<br>Insertion&Cannula Stability;<br>Air Leakage | Obturator Compatibility;<br>Insertion & Cannula Stability;<br>AirLeakage; balloon rigidity;<br>fixation device retention | Equivalent |
| Biocompatibility | ISO 10993-1, ISO 10993-5,<br>ISO 10993-10, ISO 10993-<br>7, ISO 10993-12 | ISO 10993-1, ISO 10993-5, ISO<br>10993-10, ISO 10993-7, ISO 10993-<br>12 | ISO 10993-1, ISO 10993-5,<br>ISO 10993-10, ISO 10993-<br>7, ISO 10993-12 | Equivalent |
# Table 1 Substantial equivalence discussion- Model A/B/C
{8}------------------------------------------------
{9}------------------------------------------------
## Discussion:
## Comment 1
The subject device and the predicate device have intended use, to applicate in a variety of endoscopic procedures to provide a port of entry for endoscopic instruments. The obturator and cannula have miner difference on needle tip and cannula accessories (e.g. cleaning, filter), while these differences have been validated by performance test and do not influence use. This difference does not affect the clinical safety of the subject device.
## Comment 2
The Disposable Trocars are available in a range of Length is only to allow entry of other endoscopic instruments and does not have any effect on device performance. Additionally, length information is printed on the Obturator. Choice of Trocar specification depends on user's preference and clinical need. The differences on length do not raise new and effectiveness.
## Conclusion:
The Disposable Trocars (Model A/B/C) have the same intended use, technology, design and performance specifications are either identical or substantially equivalent to existing legally marketed predicate devices. The differences between the subject device and predicate devices do not alter suitability of the proposed device for its intended use.
{10}------------------------------------------------
| Device feature | Subject Device<br>(Model E Single-<br>port<br>Trocars) | Predicate<br>DeviceK141715<br>(Glove Port) | Predicate<br>DeviceK093372<br>(SILS™<br>Port) | Predicate<br>DeviceK073719<br>(ASC TriPort<br>Laparoscopic<br>Access<br>Device) | Comment |
|-------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Class | II | II | II | II | Equivalent |
| Regulation number | 21 CFR 876.1500 | 21 CFR 876.1500 | 21 CFR 876.1500 | 21 CFR 876.1500 | Equivalent |
| Intended use | The Disposable Trocars have<br>application in a variety of<br>endoscopic procedures to<br>provide a port of entry for<br>endoscopic instruments<br>and/orendoscope.<br>Single-port Trocars are<br>designed to provide access for<br>multiple instruments and/or<br>endoscope to the abdominal<br>cavity through a single incision<br>during minimally invasive<br>endoscopic surgery. | The Glove Port is<br>intended to provide<br>access for multiple<br>instruments and/or<br>endoscope to the<br>abdominal cavity<br>through a single<br>incision during<br>minimally invasive<br>laparoscopic<br>surgery. | The SILS™ Port is<br>indicated for<br>multiple instrument or<br>camera access to the<br>abdominal cavity<br>through a singleincision<br>for performingminimally<br>invasive laparoscopic<br>procedures. | The ASC TriPort<br>Laparoscopic Access<br>Device is intended<br>foruse as a multiple<br>instrument and/or<br>camera port during<br>minimally invasive<br>abdominal<br>laparoscopic surgery. | Equivalent |
## Table 2 Substantial equivalence discussion-Single-port trocars (Model E)
{11}------------------------------------------------
| Model | E01, E02, E03, E04, E05, E06, E07, E08 | Glove Port Original; Glove Port H; Glove Port A; Glove Port AT | SILS™ Port | ASC<br>Laparoscopic Access<br>Device | TriPort Similar Comment 1 |
|------------------------------------------------------|---------------------------------------------------------------------------------|--------------------------------------------------------------------------------|---------------------------------------------------------|--------------------------------------|---------------------------|
| Port number | 3~4 | 3~4 | 3~4 | 3 | |
| Absolute size | 10~25 | 10~25mm | 10~25mm | 25mm | |
| Material | Polycarbonate, silicone | Polyurethane | Silicone | Silicone | Different<br>Comment<br>2 |
| Installation and<br>abdominal wall<br>inner fixation | Inner ring of the protector<br>being fixed inside the<br>abdominal wall inside. | Wound retractor<br>ring being fixed<br>inside the<br>abdominal wall<br>inside. | Distal ring being fixed in<br>the abdominal wallinside. | Non-fixed | Equivalent |
| Sterilization | EO gas sterilization | EO gas sterilization | Gamma sterilization | Gamma sterilization | Different<br>Comment<br>3 |
{12}------------------------------------------------
#### Discussion:
### Comment 1
The subject device and the predicate devices are endoscopic instrument access ports that are used to perform the same function as a trocar. They are all retracts a small abdominal incision to allow endoscopic instruments to pass through to abdomen, and maintains pneumoperitoneum in the abdomen during the surgical procedure, whether or not endoscopic instruments are passing through the port. The difference in models, port number and size will not influence their intended use. Besides, surgeons will select port number and size based on actual needs during procedure. Furthermore, these differences have been addressed by performance test and do not affect the basic design principle, usage, effectiveness and safety of the subject device. And no question is raised regarding to effectiveness and safety.
## Comment 2
The materials of the subject device and the predicate devices are partially different. However, the materials have been tested according to ISO 10993 series standards and performance test. The results showed they're biocompatible with human body and no impact of air leakage, connection firmness, Cannula stabilization and other functions. This difference does not affect the effectiveness and safety.
## Comment 3
The sterilization method of the subject device and the predicate devices are partially different. However, the subject device was ensured sterility by sterilization validation. Therefore, the differences on sterilization do not raise new questions about safety and effectiveness.
## Conclusion:
The Disposable Trocars (Model E) have the same intended use, technology, design and performance specifications are either identical or substantially equivalent to existing legally marketed predicate devices. The differences between the subject device and predicate devices do not alter suitability of the proposed device for its intended use.
#### Comparison Summary:
The subject and predicate devices have the same intended use and application. The subject and predicate device designs are similar. They are single-use devices. The differences in technological characteristics between the subject and predicate devices (i.e., needle tip, different size, port number) do not raise different questions of safety and effectiveness.
## VII Summary of non-clinical testing Performance testing
{13}------------------------------------------------
The Disposable Trocars has been evaluated by our Design Engineer and through performance studies and bench testing, as attached in Appendix C. Testing encompassed appearance, Obturator Compatibility, Insertion & Cannula Stability, Air Leakage, balloon rigidity, fixation device retention, etc. All the test results were "PASS". The performance of Disposable Trocars meets the technical standard requirements of Hangzhou Kangji Medical Instrument Co., Ltd. as compared to the predicate.
## Biocompatibility testing
Biocompatibility of the Disposable Veress needles was evaluated in accordance with ISO 10993-1:2018 for the body contact category. The following tests were performed, as recommended:
| Cytotoxicity | ISO 10993-5:2009 |
|-------------------------|-------------------|
| Sensitization | ISO 10993-10:2021 |
| Irritation | ISO 10993-23:2021 |
| Acute Systemic toxicity | ISO 10993-11:2017 |
| Pyrogenicity | USP<151> |
#### Sterilization and shelf life testing
- EO sterilization validation per EN ISO 11135-1, 11737-1,11737-2 A
- A Transportation test per ASTM D4169
- A Packaging validation per ISO 11607-1/-2
- The 3 years shelf life of the device is determined based on stability study which A includes ageing test.
- > Bacterial Endotoxin Testing per USP-NF:2023 <85>
## VIII Conclusion
The Disposable Trocars is substantially equivalent to its predicate devices as listed in VI Substantial Equivalence. The differences between the predicate and subject device do not raise any new or different questions of safety or effectiveness. The non-clinical testing demonstrates that the device is as safe, as effective and performs as well as the legally marketed device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.