← Product Code [QAS](/productcode/QAS) · K233247

# Heuron ICH (K233247)

_Heuron Co., Ltd. · QAS · May 15, 2024 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K233247

## Device Facts

- **Applicant:** Heuron Co., Ltd.
- **Product Code:** [QAS](/productcode/QAS.md)
- **Decision Date:** May 15, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2080
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** AI/ML, Software as a Medical Device

## Indications for Use

Heuron ICH is radiological computer-aided triage and notification software designed for the analysis of non-contrast head CT images in adults or transitional adolescents aged 18 and older. This device is intended to aid appropriately trained medical specialists and hospital networks in streamlining workflow by identifying and communicating suspected positive findings of Intracranial hemorrhage (ICH). Heuron ICH employs an artificial intelligence algorithm to analyze non-contrast CT images, flagging cases with identified findings through a dedicated application that operates in parallel with the standard of care image interpretation process. Users receive notifications for cases with suspected findings, which include compressed preview images provided for informational purposes only and are not intended for diagnostic use beyond notification. Importantly, Heuron ICH does not modify the original medical images and is not intended to serve as a diagnostic device. The results generated by Heuron ICH are intended to complement other patient information and assist medical specialists in prioritizing and triaging medical images. Notified medical specialists are responsible for viewing the full non-contrast CT images in accordance with established standard of care practices.

## Device Story

Heuron ICH is an AI-based triage software for non-contrast head CT images; operates in parallel with standard-of-care (SOC) image interpretation. Device connects to hospital PACS via IP/Port/TLS; automatically receives DICOM images. CNN-based deep learning algorithm analyzes images for suspected ICH; if detected, sends notifications (email, SMS, mobile app) to medical specialists. Notifications include compressed, non-diagnostic preview images. Device does not perform segmentation, mark images, or alter original DICOM data; does not remove or deprioritize cases from SOC reading queues. Used in hospital networks by radiologists/specialists to prioritize review of time-critical cases. Benefits include reduced time-to-notification compared to SOC, facilitating faster clinical decision-making for potential ICH patients.

## Clinical Evidence

Retrospective, multi-center study of 600 NCCT images (46.2% positive for ICH) from three US hospitals, independent of training data. Primary endpoints: sensitivity 86.3% (95% CI: 81.9-90.3) and specificity 87.6% (95% CI: 83.9-91.0). AUC was 0.945. Mean time-to-notification was 60.3 ± 39.3 seconds. Subgroup analyses performed for slice thickness, gender, age, manufacturer, and co-existing findings. Study met all acceptance criteria (lower bound of 95% CI > 80% for sensitivity and specificity).

## Technological Characteristics

Software-only device; utilizes deep learning (CNN) algorithm. Inputs: DICOM NCCT images. Connectivity: Networked via PACS integration (IP, Port, AE title, TLS authentication). Output: Notifications (email, SMS, push) with compressed preview images. No hardware components. No image segmentation or annotation. Siemens CT images outside 2.5-5mm slice thickness are automatically excluded.

## Regulatory Identification

Radiological computer aided triage and notification software is an image processing prescription device intended to aid in prioritization and triage of radiological medical images. The device notifies a designated list of clinicians of the availability of time sensitive radiological medical images for review based on computer aided image analysis of those images performed by the device. The device does not mark, highlight, or direct users' attention to a specific location in the original image. The device does not remove cases from a reading queue. The device operates in parallel with the standard of care, which remains the default option for all cases.

## Special Controls

Radiological computer aided triage and notification software must comply with the following special controls: 1. Design verification and validation must include: i. A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations. ii. A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (e.g., improved time to review of prioritized images for pre-specified clinicians). iii. Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (e.g., subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment. iv. Standalone performance testing protocols and results of the device. v. Appropriate software documentation (e.g., device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results). 2. Labeling must include the following: i. A detailed description of the patient population for which the device is indicated for use. ii. A detailed description of the intended user and user training that addresses appropriate use protocols for the device. iii. Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (e.g., poor image quality for certain subpopulations), as applicable. iv. A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images. v. Device operating instructions. vi. A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (e.g., improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (e.g., confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.

*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians).(iii) Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (
*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment.(iv) Stand-alone performance testing protocols and results of the device.
(v) Appropriate software documentation (
*e.g.,* device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use;
(ii) A detailed description of the intended user and user training that addresses appropriate use protocols for the device;
(iii) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality for certain subpopulations), as applicable;(iv) A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images;
(v) Device operating instructions; and
(vi) A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (*e.g.,* confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.

## Predicate Devices

- BriefCase ([K221240](/device/K221240.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

Heuron Co., Ltd.
% John Smith
Partner
Hogan Lovells US LLP
Columbia Square 555 Thirteenth Street NW
Washington, District of Columbia 20004

Re: K233247
May 15, 2024

Trade/Device Name: Heuron ICH
Regulation Number: 21 CFR 892.2080
Regulation Name: Radiological Computer Aided Triage And Notification Software
Regulatory Class: Class II
Product Code: QAS
Dated: April 5, 2024
Received: April 5, 2024

Dear John Smith:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K233247 - John Smith
Page 2

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica Lamb
Assistant Director
Imaging Software Team
DHT8B: Division of Radiologic Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{2}

FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
046

|  DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.  |
| --- | --- |
|  510(k) Number (if known) K233247  |   |
|  Device Name Heuron ICH  |   |
|  Indications for Use (Describe) Heuron ICH is radiological computer-aided triage and notification software designed for the analysis of non-contrast head CT images in adults or transitional adolescents aged 18 and older. This device is intended to aid appropriately trained medical specialists and hospital networks in streamlining workflow by identifying and communicating suspected positive findings of Intracranial hemorrhage (ICH). Heuron ICH employs an artificial intelligence algorithm to analyze non-contrast CT images, flagging cases with identified findings through a dedicated application that operates in parallel with the standard of care image interpretation process. Users receive notifications for cases with suspected findings, which include compressed preview images provided for informational purposes only and are not intended for diagnostic use beyond notification. Importantly, Heuron ICH does not modify the original medical images and is not intended to serve as a diagnostic device. The results generated by Heuron ICH are intended to complement other patient information and assist medical specialists in prioritizing and triaging medical images. Notified medical specialists are responsible for viewing the full non-contrast CT images in accordance with established standard of care practices.  |   |
|  Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)  |   |
|  CONTINUE ON A SEPARATE PAGE IF NEEDED.  |   |
|  This section applies only to requirements of the Paperwork Reduction Act of 1995. "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."  |   |
|  The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov  |   |
|  "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."  |   |

{3}

K233247

# 510(k) SUMMARY
Heuron Co., Ltd.'s Heuron ICH

## Submitter information

Company's name: Heuron Co., Ltd.
Company's address: 10F, C, 150, Yeongdeungpo-ro, Yeongdeungpo-gu, Seoul, 07292, Republic of Korea
Owner/Operator number: 10080283

Contact Person: John J. Smith, M.D., J.D.
Phone: +1 202 637 3638
Fax: +1 202 637 5910
Date Prepared: April 5, 2024

Name of Device: Heuron ICH
Common or Usual Name: Medical Imaging Software
Classification Name: Radiological Computer-Assisted Triage And Notification Software
Regulatory Class: Class II
Regulation number: 21 CFR 892.2080
Product Code: QAS

## Predicate Device: BriefCase (K221240)

## Device Description

Heuron ICH registers with the hospital’s Picture Archiving and Communication System (PACS) using IP, Port, AE title, and TLS authentication details. It automatically receives Non-Contrast Computed Tomography (NCCT) images in DICOM format from PACS. Upon connection request from PACS to Heuron ICH, the system verifies the IP, Port, AE title, and TLS authentication information before accepting the image transmission. The product does not query PACS to retrieve images. Instead, it receives images automatically from PACS systems that are allowed access by registering a list (white list) of PACS systems capable of uploading images to the product.

The Heuron ICH is an artificial intelligence-based solution that analyzes non-contrast CT images and provides a notification of suspected positive cases of intracranial hemorrhage (ICH) for prioritization of review. Heuron ICH uses deep learning (DL) technique of a convolutional neural network (CNN). Dataset obtained from the RSNA (Radiological Society of North America) Brain CT Hemorrhage Challenge 2019 was used for training and development of the model. Once the DICOM images transmitted from PACS are uploaded to the Heuron ICH server, the images become accessible through the worklist. The worklist displays patient identification information (Patient ID, name, age, etc.) and analysis status for convenient reference. Images received by Heuron ICH server are analyzed in the order of reception.

During the analysis, if ICH is suspected, the server provides users with a notification. The notifications include compressed preview images, which are not to be used for diagnostic use, but only for informational purposes. It is important to note that the software does not provide segmentation, analysis, or intermediate outputs to users. These notifications can be sent to registered email addresses, mobile SMS, and through the mobile app push notification feature.

## Intended Use / Indications for Use

Heuron ICH is radiological computer-aided triage and notification software designed for the analysis of non-contrast head CT images in adults or transitional adolescents aged 18 and older. This device is intended to

{4}

aid appropriately trained medical specialists and hospital networks in streamlining workflow by identifying and communicating suspected positive findings of Intracranial hemorrhage (ICH).

Heuron ICH employs an artificial intelligence algorithm to analyze non-contrast CT images, flagging cases with identified findings through a dedicated application that operates in parallel with the standard of care image interpretation process. Users receive notifications for cases with suspected findings, which include compressed preview images provided for informational purposes only and are not intended for diagnostic use beyond notification. Importantly, Heuron ICH does not modify the original medical images and is not intended to serve as a diagnostic device.

The results generated by Heuron ICH are intended to complement other patient information and assist medical specialists in prioritizing and triaging medical images. Notified medical specialists are responsible for viewing the full non-contrast CT images in accordance with established standard of care practices.

# Summary of Technological Characteristics

Both subject and the primary predicate BriefCase device are radiological computer-assisted triage and notification software for ICH triage. Both devices import NCCT images and use artificial intelligence algorithms to identify suspected ICH to aid in prioritization and triage of radiological medical images. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. At a high level, the subject and predicate devices are based on the following same technological elements:

Data import
- Image processing (identification of suspected ICH)
- Notification is provided

Similar to the predicate device's algorithms, ICH triage does not externalize any internal segmentation, analysis, or intermediate outputs used in determining if a suspected ICH is present in the NCCT, nor does the algorithm mark the analyzed NCCT image. Like the predicate, the subject devices neither remove cases from the standard of care reading queue nor deprioritize cases. The following technological differences exist between the subject and predicate devices:

- There are differences in the learning method and learning data of artificial intelligence algorithms.

A table comparing the key features of the subject and predicate devices is provided below.

|  Device name | Subject device Heuron ICH | Primary predicate device BriefCase (K221240)  |
| --- | --- | --- |
|  Manufacture | Heuron Co., Ltd. | Aidoc Medical, Ltd  |
|  Product code | QAS | QAS  |
|  Indications for Use | Heuron ICH is radiological computer-aided triage and notification software designed for the analysis of non-contrast head CT images in adults or transitional adolescents aged 18 and older. This device is intended to aid appropriately trained medical specialists and hospital networks in streamlining workflow by identifying and communicating suspected positive findings of Intracranial hemorrhage (ICH). | BriefCase is a radiological computer aided triage and notification software indicated for use in the analysis of nonenhanced head CT images in adults or transitional adolescents aged 18 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communication of suspected positive findings of Intracranial hemorrhage (ICH) pathologies.  |

{5}

|   | Heuron ICH employs an artificial intelligence algorithm to analyze non-contrast CT images, flagging cases with identified findings through a dedicated application that operates in parallel with the standard of care image interpretation process. Users receive notifications for cases with suspected findings, which include compressed preview images provided for informational purposes only and are not intended for diagnostic use beyond notification. Importantly, Heuron ICH does not modify the original medical images and is not intended to serve as a diagnostic device.
The results generated by Heuron ICH are intended to complement other patient information and assist medical specialists in prioritizing and triaging medical images. Notified medical specialists are responsible for viewing the full non-contrast CT images in accordance with established standard of care practices. | BriefCase uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings on a standalone desktop application in parallel to the ongoing standard of care image interpretation. The user is presented with notification for cases with suspected findings. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device.
The results of BriefCase are intended to be used in conjunction with other patient information and based on their professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care.  |
| --- | --- | --- |
|  Anatomical region | Head | Head  |
|  User | Hospital networks and appropriately trained medical specialists | Hospital networks and appropriately trained medical specialists  |
|  Compatible input data format and modality | Non-enhanced CT | Non-enhanced CT  |
|  Image format | DICOM | DICOM  |
|  Technical Implementation | AI/ML/Neural Network | AI/ML/Neural Network  |
|  Directs user to finding | No, the device does not highlight or direct a user's attention to a specific location in the image file. | No, the device does not highlight or direct a user's attention to a specific location in the image file.  |
|  Alteration of original image data | No | No  |
|  Notification/Prioritization | Yes – PACS, Workstation, email, mobile (Push, SMS) | Yes – PACS, Workstation, email, mobile  |

# Performance Data

Heuron conducted a retrospective, multi-center study using Heuron ICH software to evaluate the software's performance in detecting suspected intracranial hemorrhage in NCCT images. A total of 600 NCCT images were obtained from three different hospitals located in US for the study, which were newly acquired and

{6}

confirmed independent from the training and testing dataset used for model development. There were approximately equal numbers of positive and negative cases (46.2% images with ICH and 53.8% without ICH, respectively) included in the analysis.

For primary endpoint, the performance of Heuron ICH was evaluated by calculating sensitivity and specificity through comparing the software analysis results with the ground truth. The ground truth was determined by the two US board-certified neuroradiologists (truthers) interpreting each NCCT images, and in case of disagreement between the two truthers, a third truther reviewed the case for generating the final ground truth.

The standalone performance study results exceeded the acceptance criteria which were  $80\%$  for the lower bound of  $95\%$  Confidence Interval for both sensitivity and specificity. Sensitivity of Heuron ICH was  $86.3\%$  (95% CI: 81.9-90.3) and specificity was  $87.6\%$  (95% CI: 83.9-91.0). The lower bounds of each confidence interval exceeded  $80\%$ , thus the study met its goals for both sensitivity and specificity.

NPV was  $88.1\%$  and PPV was  $85.6\%$ .

Demographic information

|  Demographic |   | Total (N = 600)  |
| --- | --- | --- |
|  Gender | Female | 288 (48.0%)  |
|   |  Male | 312 (52.0%)  |
|  Age | ≤65 years | 263 (43.8%)  |
|   |  >65 tears | 337 (56.2%)  |
|  Race | Asian | 26 (4.3%)  |
|   |  Black or African American | 35 (5.8%)  |
|   |  White | 496 (82.7%)  |
|   |  Other | 28 (4.7%)  |
|   |  2 or more races | 3 (0.5%)  |
|   |  Declined | 4 (0.7%)  |
|   |  Unavailable | 8 (1.3%)  |
|  Ethnicity | Hispanic | 58 (9.7%)  |
|   |  Not Hispanic | 526 (87.7%)  |
|   |  Declined | 2 (0.3%)  |
|   |  Unavailable | 14 (2.3%)  |

# Subgroup analysis

Slice thickness

|  Subgroup | AUC (95% CI) | Sensitivity (95% CI) | Specificity (95% CI)  |
| --- | --- | --- | --- |
|  < 2.5 mm | 0.936 (0.910,0.958) | 87.7 (82.9,92.0) | 82.9 (77.6,87.6)  |
|  2.5 to 5 mm (inclusive) | 0.974 (0.961,0.985) | 89.1 (85.5,92.7) | 95.0 (92.5,97.2)  |

Gender

|  Subgroup | AUC (95% CI) | Sensitivity (95% CI) | Specificity (95% CI)  |
| --- | --- | --- | --- |
|  Female | 0.915 (0.878,0.947) | 79.8 (72.3,86.6) | 87.0 (81.7,91.7)  |
|  Male | 0.971 (0.954,0.984) | 91.1 (86.7,94.9) | 88.3 (83.1,92.9)  |

{7}

Age

|  Subgroup | AUC (95% CI) | Sensitivity (95% CI) | Specificity (95% CI)  |
| --- | --- | --- | --- |
|  ≤ 65 years | 0.955 (0.930,0.974) | 84.5 (77.7,91.3) | 90.0 (85.0,94.4)  |
|  > 65 years | 0.940 (0.912,0.963) | 87.4 (82.2,92.0) | 85.3 (79.8,90.8)  |

Manufacturer

|  Subgroup | AUC (95% CI) | Sensitivity (95% CI) | Specificity (95% CI)  |
| --- | --- | --- | --- |
|  GE Healthcare | 0.956 (0.920,0.982) | 85.7 (78.0,92.3) | 94.5 (89.9,98.2)  |
|  Siemens (slice thickness 2.5-5mm)* | 0.982 (0.963,0.994) | 93.7 (88.4, 97.9) | 92.1 (87.1, 97.0)  |
|  Toshiba | 0.978 (0.955,0.993) | 82.4 (73.6,90.1) | 97.3 (93.8,100.0)  |

*Siemens scanner CT images outside the slice thickness of 2.5-5mm will be automatically excluded from the analysis and such information on the non-processed data can be found on the device worklist with an icon with explanation pop-up.

Co-existing findings or abnormalities

|  Subgroup | AUC (95% CI) | Sensitivity (95% CI) | Specificity (95% CI)  |
| --- | --- | --- | --- |
|  No additional findings | 0.962 (0.935,0.982) | 89.1 (83.6,94.5) | 90.9 (86.4,94.9)  |
|  Any finding | 0.926 (0.897,0.951) | 84.4 (79.0,89.8) | 83.7 (77.6,89.1)  |

The secondary endpoints evaluated area under the receiver operating curve (AUC) and time-to-notification of this software. The AUC was 0.945 when compared with the consensus diagnosis from radiologists.

The time-to-notification, 60.3 seconds  $\pm 39.3$  seconds (range 9-161 seconds), was similar to other cleared devices. The time-to-notification was significantly shorter than the average time to notification as seen in the Standard of Care.

# Conclusions

The Heuron ICH is a standalone software that provides intracranial hemorrhage triage and notification to medical specialists. Both the subject device and the predicate device have the same intended use and similar indications for use and technological characteristics. Both devices send notifications of suspected time critical cases and display unannotated, compressed preview image which is not intended for diagnostic use. Both devices are not intended for diagnostic purposes, and users are instructed to review the original images from the PACS.

The Original images remain unaltered, with no markings provided, and no reordering of prioritization is performed. Therefore, the software does not interfere with the standard of care medical specialists. Both the subject and predicate devices contribute to faster workflow times by triaging and notifying time critical cases of suspected ICH.

The performance of this software has been verified through standalone performance testing, demonstrating shorter time-to-notification compared to the standard of care. Therefore, Heuron ICH device is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K233247](https://fda.innolitics.com/device/K233247)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
