The NextAR Spine platform is intended as an aid for precisely locating anatomical structures in either open or percutaneous spine procedures. It is indicated for any medical condition in which the use of stereotaxic surgery may be appropriate, when reference to a rigid anatomical structure, such as vertebrae or pelvis, can be identified relative to images of the anatomy. This can include posterior approach spinal procedures, such as: - · Pedicle Screw Placement (Thoracic and Lumbosacral spine) - · Iliosacral Screw Placement NextAR Spine is also intended to provide planning tools for measuring and selecting the fixation rod for the thoracic and lumbosacral spine. The NextAR Spine platform is intended to be used in combination with NextARTM Stereotaxic instruments and / or Medacta preoperative planning. In the case of pre-operative planning software is used pre-operatively to plan the surgical placement of pedicle screws based upon radiological images of the patient. As an optional display, the NextAR Smart Glasses can be used auxiliary to the NextAR Spine Platform to view stereotaxic information as presented by the NextAR Spine Platform. The NextAR Smart Glasses should not be relied upon solely and should always be used in conjunction with the primary computer display.
Device Story
NextAR Spine Platform is a CT-based computer-assisted surgical navigation system for open or percutaneous spine surgery. Inputs include pre-operative CT scans and/or intra-operative 3D scans. System uses active infrared camera and active trackers (targets) mounted on bones to monitor spatial location of surgical instruments in real-time. Software registers patient anatomy to navigation elements via Direct 3D or 3D-3D registration. System provides intra-operative screw planning and rod planning (length/geometry) tools. Output is real-time visualization of instrument position/orientation relative to anatomy on a screen monitor; optional auxiliary display via AR smart glasses. Used by surgeons in clinical settings to execute surgical plans, improve accuracy of screw/rod placement, and potentially reduce surgical complications. System includes PC hardware, optical tracking, fiducial blocks, and reusable surgical instruments.
Clinical Evidence
No clinical data on human subjects. Evidence consists of bench testing, software validation, accuracy testing, and a cadaver study to verify performance and registration accuracy.
Technological Characteristics
CT-based computer-assisted navigation platform. Components: PC hardware, active infrared optical tracking system, fiducial block, AR smart glasses, and reusable surgical instruments. Registration methods: Direct 3D (intra-operative scan) or 3D-3D (pre-operative CT + intra-operative 3D scan). Connectivity: Networked/integrated system components. Sterilization: Tracking sensors, fiducial block, pins, and drills provided sterile.
Indications for Use
Indicated for patients requiring stereotaxic surgery for spinal procedures, including pedicle screw placement (thoracic/lumbosacral) and iliosacral screw placement, where rigid anatomical structures (vertebrae/pelvis) can be referenced to imaging.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{1}------------------------------------------------
April 24, 2024
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Medacta International S.A. % Chris Lussier Senior Director, Quality, Regulatory and Clinical Research Medacta USA 3973 Delp Street Memphis, Tennessee 38118
### Re: K233172
Trade/Device Name: NextAR™ Spine Platform Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO, JWH, PBF Dated: September 27, 2023 Received: September 28, 2023
### Dear Chris Lussier:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{2}------------------------------------------------
2
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Tejen D. Soni -S
For
Shumaya Ali, MPH Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K233172
Device Name NextAR™ Spine Platform
### Indications for Use (Describe)
The NextAR Spine platform is intended as an aid for precisely locating anatomical structures in either open or percutaneous spine procedures. It is indicated for any medical condition in which the use of stereotaxic surgery may be appropriate, when reference to a rigid anatomical structure, such as vertebrae or pelvis, can be identified relative to images of the anatomy. This can include posterior approach spinal procedures, such as:
- · Pedicle Screw Placement (Thoracic and Lumbosacral spine)
- · Iliosacral Screw Placement
NextAR Spine is also intended to provide planning tools for measuring and selecting the fixation rod for the thoracic and lumbosacral spine.
The NextAR Spine platform is intended to be used in combination with NextARTM Stereotaxic instruments and / or Medacta preoperative planning. In the case of pre-operative planning software is used pre-operatively to plan the surgical placement of pedicle screws based upon radiological images of the patient. As an optional display, the NextAR Smart Glasses can be used auxiliary to the NextAR Spine Platform to view stereotaxic information as presented by the NextAR Spine Platform. The NextAR Smart Glasses should not be relied upon solely and should always be used in conjunction with the primary computer display.
Type of Use (Select one or both, as applicable)
| <span style="text-decoration: overline;"></span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-----------------------------------------------------------------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# K233172 510(k) Summary
#### I. Submitter
Medacta International SA Strada Regina 6874 Castel San Pietro (CH) Switzerland Phone (+41) 91 696 60 60 Fax (+41) 91 696 60 66
Contact Person: Stefano Baj, Regulatory Affairs Director, Medacta International SA Applicant Correspondent: Christopher Lussier, Sr. Director of Quality and Regulatory, Medacta USA Date Prepared: September 25, 2023 Date Modified: April 24, 2024
#### II. Device
| Device Proprietary Name: | NextAR™ Spine Platform |
|--------------------------|---------------------------------------------------|
| Common or Usual Name: | Navigation System<br>Total Joint Replacement |
| Classification Name: | Stereotaxic Instrument |
| Primary Product Code: | OLO<br>JWH<br>PBF |
| Regulation Number: | 21 CFR 882.4560, 21 CFR 888.3560, 21 CFR 888.3030 |
| Device Classification | II |
#### III. Predicate Device
Substantial equivalence is claimed to the following devices:
Primary predicate device:
- NextARTM Spine Platform K223769 Medacta International SA .
{5}------------------------------------------------
#### IV. Device Description
The NextARTM Spine Platform is a CT based computer-assisted surgical navigation platform used in either open/mini open or percutaneous spine surgery procedure and includes the following components:
- navigation software which displays information to the surgeon; ●
- Augmented Reality glasses; ●
- optical tracking system; ●
- PC based hardware platform; ●
- Fiducial Block;
- Adaptor for sensor: ●
- . Spine attachment instruments
- Reusable surgical instruments for spine surgery procedures. .
The system operates on the common principle of stereotaxic technology in which markers are mounted on the bones and an infrared camera is used to monitor the spatial location of the instruments. Tracking sensors attached to the bones enable the surgeon to view the position and orientation of the instrumentation relative to the intra-operative data in real-time while performing the surgical procedure. The tracking sensors, the fiducial block and a group of pins and drills are provided sterile.
The NextAR™ Spine Platform aid the surgeon in executing the surgical plan by visualizing all the information in real time in a screen monitor.
The NextAR Spine system is a surgical navigation platform which uses the information of either an intra-operative scan or pre-operative CT in combination with an intra-operative 3D scan in order to register the spine to navigation elements.
The registration can be performed with one of the following approaches:
- Direct 3D: based on the use of an intra-operative 3D scan
- 3D-3D: based on the use of a pre-operative CT scan and an intra-operative 3D scan .
NextAR gives the possibility of planning the screw positioning on the intraoperative DICOM images of the patient, in case of the 3D direct approach, just before the system setup. The application allows for navigating the spine with a screw planning superimposed on the acquired scan.
The NextAR Spine Platform includes also the rod planning that is intended to provide information on the length and geometry of the pre-bend rod that would fit with each screw position.
The system's navigation technology is based on an active infrared camera coupled with an active tracker (Target). These elements allow, by means of the different registration approaches
{6}------------------------------------------------
and use of compatible instruments, to accurately prepare trajectories in the vertebrae and/or to implant screws while visualizing information in real time on a screen monitor.
#### V. Indications for Use
The NextAR Spine platform is intended as an aid for precisely locating anatomical structures in either open/mini-open or percutaneous spine procedures. It is indicated for any medical condition in which the use of stereotaxic surgery may be appropriate, when reference to a rigid anatomical structure, such as vertebrae or pelvis, can be identified relative to images of the anatomy. This can include posterior approach spinal procedures, such as:
- · Pedicle Screw Placement (Thoracic and Lumbosacral spine)
- Iliosacral Screw Placement
NextAR Spine is also intended to provide planning tools for measuring and selecting the fixation rod for the thoracic and lumbosacral spine.
The NextAR Spine platform is intended to be used in combination with NextAR™ Stereotaxic instruments and / or Medacta preoperative planning. In the case of pre-operative planning. surgical planning software is used pre-operatively to plan the surgical placement of pedicle screws based upon radiological images of the patient. As an optional display, the NextAR Smart Glasses can be used auxiliary to the NextAR Spine Platform to view stereotaxic information as presented by the NextAR Spine Platform. The NextAR Smart Glasses should not be relied upon solely and should always be used in conjunction with the primary computer display.
#### VI. Comparison of Technological Characteristics
The NextAR™ Spine System and the predicate NextAR™ Spine Platform, K223769, share the following characteristics:
- Intended use ●
- principle of operation; ●
- user interface: ●
- power source;
- optical tracking system ●
- platform; ●
- Displaying Technology
- Glasses communication;
- use of surgical instruments for navigation; ●
- main system components;
- use of surgical instruments for navigation; ●
- Tracking System accuracy;
- use of active optical tracking system
- . computer hardware;
- pre-operative patient anatomy data acquisition. ●
{7}------------------------------------------------
- intra-operative patient anatomy data acquisition ●
- . use of surgical instruments for navigation;
The subject NextAR™ Spine System differs from the predicate device with respect to:
- Rod planning feature ●
- Intra Operative Screw planning feature ●
- . Fiducial Registration Accuracy
### Discussion
The differences between the subject and the primary predicate device are concerning the rod planning and intra operative screw planning features.
They do not raise different questions of safety or effectiveness when compared to the predicate device. Both navigation systems utilize stereotaxic technologies.
Minor differences are addressed by performing cadaveric testing/ rational/ test.
#### VII. Performance Data
Testing was conducted according to written protocols with acceptance criteria that were based on standards. The following studies were performed in support of a substantial equivalence determination:
- software validation;
- Accuracy test ●
- performance testing
- cadaver study. ●
### VIII. Conclusion
The information provided above supports that the NextAR™ Spine Platform is substantially equivalent to the identified predicate devices.
Substantial equivalence has been demonstrated through a comparison of design and technological characteristics, as well as performance evaluations. The NextAR™ Spine Platform can be considered substantially equivalent to the identified predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.