GENIOVA Aligners are indicated for use in the alignment of permanent teeth (i.e., all second molars) through orthodontic treatment of misalignment and malocclusion. The aligners guide teeth to their final position by way of continuous gentle forces.
Device Story
GENIOVA Aligners are thermoformed plastic trays used for orthodontic tooth movement. Input consists of traditional mold impressions or digital intraoral scans of patient dentition. Dental professionals use 3Shape Ortho System software to create sequential 3D dental models representing incremental tooth realignment. These models are 3D printed, and thermoplastic sheets are thermoformed over them to create the aligners. The device is prescribed by dentists or orthodontists and worn by the patient 20-22 hours daily, removed for eating and cleaning. The aligners apply continuous gentle forces to guide teeth to the target position. Treatment is monitored by the prescribing professional throughout the sequence. The device benefits patients by correcting malocclusion through a removable, patient-specific orthodontic appliance.
Clinical Evidence
No clinical data. Bench testing only. Validation included dimensional accuracy testing of the manufacturing process (intraoral scans, 3D printed molds, and final aligners) using digital calipers, confirming translational measurements within 0.150 mm of target values. Biocompatibility testing performed per ISO 10993-1 and ISO 10993-5:2009.
Technological Characteristics
Thermoformed plastic aligners; material: Essix thermoplastic (K062828). Patient-specific dimensions. Non-sterile. Software-driven design using 3Shape Ortho System. Mechanical force application via sequential tray geometry.
Indications for Use
Indicated for alignment of permanent teeth (including second molars) in patients requiring orthodontic treatment for misalignment and malocclusion.
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 12, 2024
Geniova Technologies Sociedad Limitada % Juan Tezak Consultant Compliance4Devices 118 W Prive Cr. Delray Beach, Florida 33445
Re: K233152
Trade/Device Name: GENIOVA aligners Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: NXC Dated: September 27, 2023 Received: September 27, 2023
Dear Juan Tezak:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{3}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# MICHAEL E. ADJODHA -S
Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
Submission Number (if known)
K233152
Device Name
GENIOVA Aligners
#### Indications for Use (Describe)
GENIOVA Aligners are indicated for use in the alignment of permanent teeth (i.e., all second molars) through orthodontic treatment of misalignment and malocclusion. The aligners guide teeth to their final position by way of continuous gentle forces.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for GENIOVA, a company that specializes in fast aligners. The logo is composed of the company name in bold, sans-serif font, with a horizontal line running through the letters "OVA". A small, pink starburst is located in the center of the "O". Below the company name, the tagline "THE FAST ALIGNER" is written in a smaller font.
# 510(k) Summary
# GENIOVA Aligners K233152 March, 2024
#### ADMINISTRATIVE INFORMATION
Applicant
GENIOVA TECHNOLOGIES SOCIEDAD LIMITADA Calle Valentín Beato 21, planta-1. 28037 Madrid - Spain. Phone: +34629266297
# Establishment Registration Number
Official Contact
Antonio Baselga de la Vega CEO Calle Valentín Beato 21, planta-1. 28037 Madrid - Spain. Phone: +34629266297 E-mail: abaselga@geniova.com
#### Representative/Consultant
Juan Tezak Carlos Marín Compliance4Devices 118 W Prive Cr. Delray Beach Fl, 33445 Phone: +1 561-789-2411 E-mail: compliance4devices@gmail.com
#### DEVICE AND CLASSIFICATION NAME
Device Trade Name: Common Name: Device Classification: Classification Panel: Product Code:
GENIOVA Aligners Orthodontic plastic bracket. Class II, 21 CFR 872.5470 Dental NXC — Orthodontic plastic bracket
#### Predicate Device Information
Predicate Device: Reference Device: K202792 - BRIUS Clear Aligners K180941 - Ortho System K062828 – Essix ACE Plastic
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for "GENIOVA THE FAST ALIGNER". The word "GENIOVA" is in large, bold, sans-serif font, with a horizontal line running through the middle of the "O", creating a visual effect. Below "GENIOVA", in a smaller font size, are the words "THE FAST ALIGNER", also in a sans-serif typeface.
# Intended Use
GENIOVA Aligners are indicated for use in the alignment teeth (i.e., all second molars) through orthodontic treatment of misalignment and malocclusion. The aligners guide teeth to their final position by way of continuous gentle forces.
# Device Description
GENIOVA aligners are thermoformed plastic aligners designed to be used in sequence to facilitate the movement of teeth to the desired final position. Sequential aligners introduce incremental movements that move teeth by a gentle continuous force. Aligners should be worn 20 to 22 hours a day and should be removed for eating and cleaning.
GENIOVA aligners are designed based on traditional mold impression or digital scans of the patient's dentition submitted by a dental health professional (e.g., a dentist or orthodontist). From the scan, specialized orthodontic CAD/CAM software is used to develop the treatment plans, which consist of sequential dental models in which the teeth are gradually realigned at each step. For these 510(k) applications, the Ortho system from 3Shape A/S (K180941) will be used.
(K180941) from 3Shape A/S. Ortho System is approved for use in orthodontic model management, systematic inspection, detailed analysis, treatment simulation and virtual appliance design options. The specialized orthodontic treatment planning software has 510k clearance for intended use under FDA Classification Product Code PNN, regulation 872.5470.
Once the treatment plan is reviewed and approved by a dental health professional, each 3D model from the treatment plan is manufactured. The aligner trays are then manufactured by thermoforming a dental thermoplastic sheet over each model. The Essix thermoplastic material used for BRIUS Clear Aligners has 510(k) clearance (K062828) for use as an aligner material; the 510(k) holder (of the material) conducted the physical properties testing for the material.
The aligner trays are then delivered to the patient by the prescribing dental health professional. This dental health professional then monitors the patient's treatment from the first aligner to the delivery of the final aligner and completion of treatment. The aligners are held in place by pressure and can be removed by the patient at any time.
## Equivalence to Marketed Device
GENIOVA Aligners are functionally equivalent to the following predicate device: BRIUS Clear Aligners (K202792). The following table demonstrates the functional specifications of GENIOVA Aligners are substantially equivalent to the predicate devices.
| Specification | Subject Device:<br>GENIOVA Aligners | Predicate Device:<br>BRIUS Clear Aligners (K202792) | Comparison<br>Result |
|-------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| Regulation<br>Number | 21 CFR 872.5470 | 21 CFR 872.5470 | Same |
| Classification<br>Name | Orthodontic Plastic Bracket | Orthodontic Plastic Bracket | Same |
| Product Code | NXC | NXC | Same |
| Classification | Class II | Class II | Same |
| Specification | Subject Device:<br>GENIOVA Aligners | Predicate Device:<br>BRIUS Clear Aligners (K202792) | Comparison<br>Result |
| OTC or Rx | Rx | Rx | Same |
| Material | Essix Thermoplastic<br>(K062828) | Essix Thermoplastic<br>(K062828) | Same |
| Material<br>Properties | Acceptable material properties<br>established for use as an aligner. | Acceptable material properties<br>established for use as an aligner. | Same |
| Biocompatible | Yes | Yes | Same |
| Sterile | Non-sterile | Non-sterile | Same |
| Device<br>Description | Sequential thermoformed plastic<br>aligners | Sequential thermoformed plastic<br>aligners | Same |
| Patient<br>Removable? | Yes | Yes | Same |
| Indication for<br>Use | GENIOVA Aligners are indicated for<br>use in the alignment of permanent<br>teeth (i.e., all second molars)<br>through orthodontic treatment of<br>misalignment and malocclusion. The<br>aligners guide teeth to their final<br>position by way of continuous<br>gentle forces. | BRIUS Clear Aligners are indicated<br>for use in the alignment of<br>permanent teeth (i.e., all second<br>molars) through orthodontic<br>treatment of misalignment and<br>malocclusion. The aligners guide<br>teeth to their final position by way<br>of continuous gentle forces. | Same |
| Mode of Action | Continuous gentle force applied to<br>teeth following the prescribed and<br>approved treatment plan to achieve<br>orthodontic movement | Continuous gentle force applied to<br>teeth following the prescribed and<br>approved treatment plan to achieve<br>orthodontic movement | Same |
| Anatomy<br>Location | Mouth; mucosal<br>membranes | Mouth; mucosal<br>membranes | Same |
| Size | Patient specific | Patient specific | Same |
| Manufacturing<br>Method | Thermoforming | Thermoforming | Same |
| Software used to<br>design aligners | 3Shape A/S's Ortho System<br>(K180941) | 3Shape A/S's Ortho System<br>(K180941) | Same |
Table 1. Comparison with predicate device for Summary
{7}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for GENIOVA, a company that specializes in fast aligners. The logo is in black and white, with the word "GENIOVA" in large, bold letters. A pink star is in the middle of the word. Below the word "GENIOVA" is the phrase "THE FAST ALIGNER" in smaller letters.
Based on the above comparison, the indications for use, the design, construction, and performance characteristics of the GENIOVA Aligners is similar to that of the BRIUS Clear Aligners (K202792). Thus, the GENIOVA Aligners can be considered substantially equivalent to its predicate device.
# Non-Clinical Testing Summary
The use of thermoplastic materials for sequential aligners intended to treat malocclusions has been well documented in scientific literature regarding incremental tooth moving forces. However, durability testing was conducted on the aligners. Real world use was simulated to ensure that the
{8}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for "GENIOVA THE FAST ALIGNER". The word "GENIOVA" is in large, bold, black letters. There is a pink star-like shape in the middle of the word. Below the word "GENIOVA" is the phrase "THE FAST ALIGNER" in smaller, black letters.
aligner material and manufacturing process produced aligners that were suitable for their prescribed period of use.
An internal manufacturing validation was performed to establish the dimensional accuracy of the manufacturing process for GENIOVA Aligners. The submitted intraoral scans, digital dentition models from treatment planning, 3D printed molds, and the final thermoformed aligners were all assessed quantitatively or qualitatively in the validation. Thus, each critical element in the manufacturing process was evaluated.
For the validation, digital calipers were used to perform point-to-point and critical displacement measurements, and visual inspections were performed to assess the aligner qualitatively.
Translational measurements were within 0.150 mm (150 microns) of the target input value, the predefined tolerance of the manufacturing process. There were no significant differences in the difference in the intended and measured values observed from any of the groups. This test has met the pre-established acceptance criteria to demonstrate dimensional accuracy.
The Essix thermoplastic material used for GENIOVA Aligners has 510(k) clearance (K062828) for use as an aligner material; the 510(k) holder (of the material) conducted the physical properties testing for the material. Biocompatibility testing for the aligner material, the only patient contacting material, was conducted by the 510(k) holder in accordance with International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process".
Additional cytotoxicity testing according to ISO 10993-5:2009 was performed on final manufactured GENIOVA Aligners.
## Clinical Testing Summary
Clinical performance testing was not conducted.
## Conclusion
Based on the information presented in these 510(k) premarket notifications the GENIOVA Aligners is considered substantially equivalent (as safe, as effective and performs as well as) to the currently marketed devices (BRIUS Clear Aligners K202792) citied in this submission. The differences noted between the GENIOVA Aligners and the predicate device do not impact safety or effectiveness based on the successfully conducted testing of the subject device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.