8MP Color LCD Displays C811W, C811WT, PA27 and PA27T
Applicant
Shenzhen Beacon Display Technology Co., Ltd.
Product Code
PZZ · Pathology
Decision Date
Dec 28, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.3700
Device Class
Class 2
Indications for Use
8MP Color LCD Displays C811W, C811WT, PA27 and PA27T are intended for in vitro diagnostic use to display digital images of histopathology slides acquired from IVD-labeled whole-slide imaging scanners and viewed using IVD-labeled digital pathology image viewing software that have been validated for use with this device. They are an aid to the pathologist to review and interpret digital images of histopathology slides for primary diagnosis. It is the responsibility of the pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images using 8MP Color LCD Displays C811W, C811WT, PA27 and PA27T. The displays are not intended for use with digital images from frozen section, cytology, or non-formalin-fixed, paraffin embedded (non-FFPE) hematopathology specimens.
Device Story
Device consists of 27-inch 8MP color LCD monitors (models C811W, C811WT, PA27, PA27T) used in clinical pathology settings. Monitors receive digital image data from IVD-labeled whole-slide imaging scanners via DVI-D, HDMI, or DisplayPort interfaces. Displays utilize LED backlighting and built-in brightness stabilization control circuits to maintain consistent luminance. Models C811WT and PA27T include capacitive touch screens. Pathologists use these displays to view and interpret histopathology slides for primary diagnosis. Performance is maintained through front sensor calibration and feedback systems. Benefits include high-resolution visualization of digital pathology images, aiding diagnostic accuracy in clinical workflows.
Clinical Evidence
Bench testing only. Performance evaluated against IDMS 1.03 standards, including spatial resolution (MTF), pixel defects, temporal response, luminance uniformity, and color gamut volume. System-level testing performed with Leica GT450DX scanner using test targets and tissue sample images; results were acceptable.
Technological Characteristics
27-inch TFT Color LCD (AHVA) panel; LED backlight; 3840 x 2160 resolution; 500 cd/m² calibrated luminance; 1000:1 contrast ratio. Interfaces: DVI-D, HDMI, DisplayPort. Features: brightness stabilization control circuits, front sensor calibration, capacitive touch (select models). Standards: ANSI AAMI ES60601-1, IEC 60601-1-2, AAPM TG18, IEC 62304, IDMS 1.03.
Indications for Use
Indicated for use by pathologists to review and interpret digital histopathology images for primary diagnosis. Not for use with frozen section, cytology, or non-FFPE hematopathology specimens.
Regulatory Classification
Identification
The whole slide imaging system is an automated digital slide creation, viewing, and management system intended as an aid to the pathologist to review and interpret digital images of surgical pathology slides. The system generates digital images that would otherwise be appropriate for manual visualization by conventional light microscopy.
Special Controls
A whole slide imaging system must comply with the following special controls: (1) Premarket notification submissions must include the following information: (i) The indications for use must specify the tissue specimen that is intended to be used with the whole slide imaging system and the components of the system. (ii) A detailed description of the device and bench testing results at the component level, including for the following, as appropriate: (A) Slide feeder; (B) Light source; (C) Imaging optics: (D)Mechanical scanner movement; (E) Digital imaging sensor; (F) Image processing software; (G)Image composition techniques; (H)Image file formats; (I) Image review manipulation software; (J) Computer environment; (K)Display system. (iii)Detailed bench testing and results at the system level, including for the following, as appropriate: (A)Color reproducibility; (B) Spatial resolution; (C) Focusing test; (D) Whole slide tissue coverage; (E) Stitching error: (F) Turnaround time. (iv) Detailed information demonstrating the performance characteristics of the device, including, as appropriate: (A)Precision to evaluate intra-system and inter-system precision using a comprehensive set of clinical specimens with defined, clinically relevant histologic features from various organ systems and diseases. Multiple whole slide imaging systems, multiple sites, and multiple readers must be included. (B) Reproducibility data to evaluate inter-site variability using a comprehensive set of clinical specimens with defined, clinically relevant histologic features from various organ systems and diseases. Multiple whole slide imaging systems, multiple sites, and multiple readers must be included. (C) Data from a clinical study to demonstrate that viewing, reviewing, and diagnosing digital images of surgical pathology slides prepared from tissue slides using the whole slide imaging system is non-inferior to using an optical microscope. The study should evaluate the difference in major discordance rates between manual digital (MD) and manual optical (MO) modalities when compared to the reference (e.g., main sign-out diagnosis). (D) A detailed human factors engineering process must be used to evaluate the whole slide imaging system user interface(s). (2) Labeling compliant with 21 CFR 809.10(b) must include the following: The intended use statement must include the information described in paragraph (i) (1)(i) of this section, as applicable, and a statement that reads, "It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images obtained using this device." (ii) A description of the technical studies and the summary of results, including those that relate to paragraph (1)(ii) and (1)(iii) of this section, as appropriate. (iii) A description of the performance studies and the summary of results, including those that relate to paragraph (1)(iv) of this section, as appropriate. (iv) A limiting statement that specifies that pathologists should exercise professional judgment in each clinical situation and examine the glass slides by conventional microscopy if there is doubt about the ability to accurately render an interpretation using this device alone.
*Classification.* Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include the following information:
(i) The indications for use must specify the tissue specimen that is intended to be used with the whole slide imaging system and the components of the system.
(ii) A detailed description of the device and bench testing results at the component level, including for the following, as appropriate:
(A) Slide feeder;
(B) Light source;
(C) Imaging optics;
(D) Mechanical scanner movement;
(E) Digital imaging sensor;
(F) Image processing software;
(G) Image composition techniques;
(H) Image file formats;
(I) Image review manipulation software;
(J) Computer environment; and
(K) Display system.
(iii) Detailed bench testing and results at the system level, including for the following, as appropriate:
(A) Color reproducibility;
(B) Spatial resolution;
(C) Focusing test;
(D) Whole slide tissue coverage;
(E) Stitching error; and
(F) Turnaround time.
(iv) Detailed information demonstrating the performance characteristics of the device, including, as appropriate:
(A) Precision to evaluate intra-system and inter-system precision using a comprehensive set of clinical specimens with defined, clinically relevant histologic features from various organ systems and diseases. Multiple whole slide imaging systems, multiple sites, and multiple readers must be included.
(B) Reproducibility data to evaluate inter-site variability using a comprehensive set of clinical specimens with defined, clinically relevant histologic features from various organ systems and diseases. Multiple whole slide imaging systems, multiple sites, and multiple readers must be included.
(C) Data from a clinical study to demonstrate that viewing, reviewing, and diagnosing digital images of surgical pathology slides prepared from tissue slides using the whole slide imaging system is non-inferior to using an optical microscope. The study should evaluate the difference in major discordance rates between manual digital (MD) and manual optical (MO) modalities when compared to the reference (
*e.g.,* main sign-out diagnosis).(D) A detailed human factor engineering process must be used to evaluate the whole slide imaging system user interface(s).
(2) Labeling compliant with 21 CFR 809.10(b) must include the following:
(i) The intended use statement must include the information described in paragraph (b)(1)(i) of this section, as applicable, and a statement that reads, “It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images obtained using this device.”
(ii) A description of the technical studies and the summary of results, including those that relate to paragraphs (b)(1)(ii) and (iii) of this section, as appropriate.
(iii) A description of the performance studies and the summary of results, including those that relate to paragraph (b)(1)(iv) of this section, as appropriate.
(iv) A limiting statement that specifies that pathologists should exercise professional judgment in each clinical situation and examine the glass slides by conventional microscopy if there is doubt about the ability to accurately render an interpretation using this device alone.
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
## I Background Information:
A 510(k) Number
K233119
B Applicant
Shenzhen Beacon Display Technology Co., Ltd.
C Proprietary and Established Names
8MP Color LCD Displays C811W, C811WT, PA27 and PA27T
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| PZZ | Class II | 21 CFR 864.3700 - Whole Slide Imaging System | PA - Pathology |
## II Review Summary:
A Purpose for Submission:
New digital pathology displays as follows:
8MP Color LCD Display C811W
8MP Color LCD Display C811WT
8MP Color LCD Display PA27
8MP Color LCD Display PA27T
B Type of Test:
Digital pathology display monitors.
## III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
K233119 - Page 2 of 8
## B Indication(s) for Use:
8MP Color LCD Displays C811W, C811WT, PA27 and PA27T are intended for in vitro diagnostic use to display digital images of histopathology slides acquired from IVD-labeled whole-slide imaging scanners and viewed using IVD-labeled digital pathology image viewing software that have been validated for use with this device. They are an aid to the pathologist to review and interpret digital images of histopathology slides for primary diagnosis. It is the responsibility of the pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images using 8MP Color LCD Displays C811W, C811WT, PA27 and PA27T. The displays are not intended for use with digital images from frozen section, cytology, or non-formalin-fixed, paraffin embedded (non-FFPE) hematopathology specimens.
## C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
## IV Device/System Characteristics:
### A Device Description:
The C811WT, C811WT, PA27 and PA27T are 8MP Color LCD displays, specifically intended for review and interpretation of surgical pathology slides from IVD-labeled whole-slide imaging scanners that have been validated for use with the display. The displays are equipped with a 27-inch color LCD panel with a fine pixel pitch and LED backlight panels. The built-in brightness stabilization control circuits enable the brightness of these displays are stable throughout the product life. Displays models C811WT and PA27T have capacitive touch screen and models C811W, PA27 do not have a capacitive touch screen.
### V Substantial Equivalence Information:
#### A Predicate Device Name(s):
Leica Aperio GT 450 DX
#### B Predicate 510(k) Number(s):
K232202
#### C Comparison with Predicate:
{2}
K233119 - Page 3 of 8
| Device & Predicate
Device(s): | Device
K233119 | Predicate
K232202 |
| --- | --- | --- |
| Device Trade Name | Shenzhen Beacon 8MP Color LCD displays C811WT, PA27T, C811W and PA27 | Aperio GT 450 DX (Barco MDPC-8127 display component) |
| General Device
Characteristic
Similarities | | |
| Screen size | 684 mm (27") | 684 mm (27") |
| Backlight type | LED | LED |
| Color Calibration | sRGB | sRGB |
| Contrast Ratio (typical) | 1000:1 | 1000:1 |
| General Device
Characteristic
Differences | | |
| Display Technology | TFT Color LCD Panel (AHVA) | TFT Color LCD Panel (IPS) |
| Resolution | 3840 x 2160 @ 60 Hz (16:9 aspect ratio) | 3840 x 2160 @ 120 Hz (16:9 aspect ratio) |
| Pixel pitch | 0.15525 x 0.15525 mm | 0.155 x 0.155 mm |
| Calibrated luminance | 500 cd/m² | 450 cd/m² |
| Display Interface | DVI-D x 1
HDMI x 1
DisplayPort x 1 | DisplayPort x 2 |
| Ambient light sensor | N/A | Yes |
| Touch screen technology | Capacitive touch screen | N/A |
| Calibration tools | Front sensor;
Calibration Feedback system | Front sensor;
QAWeb |
{3}
General Technological Characteristics Comparison of C811WT and PA27T
| Attributes | Proposed Device
8MP Color LCD Monitor
(C811WT, PA27T) | Predicate Device
8MP Color LCD Monitor
Barco MDPC-8127 |
| --- | --- | --- |
| Display Technology | TFT Color LCD Panel
(AHVA) | TFT Color LCD Panel (IPS) |
| Screen size | 684 mm (27") | 684 mm (27") |
| Resolution | 3840 x 2160 @ 60 Hz (16:9 aspect ratio) | 3840 x 2160 @ 120 Hz (16:9 aspect ratio) |
| Pixel pitch | 0.15525 x 0.15525 mm | 0.155 x 0.155 mm |
| Backlight type | LED | LED |
| Color Calibration | sRGB | sRGB |
| Calibrated luminance | 500 cd/m² | 450 cd/m² |
| Contrast Ratio
(typical) | 1000:1 | 1000:1 |
| Display Interface | DVI-D x 1,
HDMI x 1,
DisplayPort x 1 | DisplayPort x 2 |
| Ambient light sensor | N/A | Yes |
| Touch screen
technology | Capacitive touch screen | N/A |
| Calibration tools | Front sensor;
Calibration Feedback System | Front sensor
QAWeb |
General Technological Characteristics Comparison of C811W and PA27
| Attributes | Proposed Device
8MP Color LCD Monitor
(C811W, PA27) | Predicate Device
8MP Color LCD Monitor
Barco MDPC-8127 |
| --- | --- | --- |
| Display Technology | TFT Color LCD Panel
(AHVA) | TFT Color LCD Panel (IPS) |
| Screen size | 684 mm (27") | 684 mm (27") |
| Resolution | 3840 x 2160 @ 60 Hz (16:9 aspect ratio) | 3840 x 2160 @ 120 Hz (16:9 aspect ratio) |
| Pixel pitch | 0.15525 x 0.15525 mm | 0.155 x 0.155 mm |
K233119 - Page 4 of 8
{4}
| Backlight type | LED | LED |
| --- | --- | --- |
| Color Calibration | sRGB | sRGB |
| Calibrated luminance | 500 cd/m² | 450 cd/m² |
| Contrast Ratio (typical) | 1000:1 | 1000:1 |
| Display Interface | DVI-D x 1,
HDMI x 1,
DisplayPort x 1 | DisplayPort x 2 |
| Ambient light sensor | N/A | Yes |
| Touch screen technology | N/A | N/A |
| Calibration tools | Front sensor;
Calibration Feedback System | Front sensor
QAWeb |
VI Standards/Guidance Documents Referenced:
- ANSI AAMI, ES60601-1:2005/R2012&A1:2012 C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021]
- IEC, 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION
- AAPM, TG18 Assessment of Display Performance for Medical Imaging System, 2005
- IEC, TR 60601-4-2 Edition 1.0 2016-05
- ISO, 14971 Third Edition 2019-12
- IEC, 62563-1 Edition 1.2 2021-07
- ICDM, IDMS v1.03
- IEC, 62304 Edition 1.1 2015-06 CONSOLIDATED VERSION
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
1. Precision/Reproducibility: Not applicable
2. Linearity: Not applicable
3. Analytical Specificity/Interference: Not applicable
4. Accuracy (Instrument): Not applicable
K233119 - Page 5 of 8
{5}
5. Carry-Over:
Not applicable
B Other Supportive Instrument Performance Characteristics Data:
Technical performance testing for the displays was performed as specified in the Tables below:
Table: C811W, C811WT, PA27 and PA27T Test Results
| Test | Test Method/Applicable Standards | C811W | C811WT | PA27 | PA27T |
| --- | --- | --- | --- | --- | --- |
| 1.User controls | Modified-Performance Modes, IDMS 1.03, Section 2.1 | Luminance target, Maximum: 500 cd/m2 Display function: sRGB White point: D65 (6500K) Color space: sRGB 30 minutes of warm-up time | Luminance target, Maximum: 500 cd/m2 Display function: sRGB White point: D65 (500K) Color space: sRGB 30 minutes of warm-up time | Luminance target, Maximum: 500 cd/m2 Display function: sRGB White point: D65 (6500K) Color space: sRGB 30 minutes of warm-up time | Luminance target, Maximum: 500 cd/m2 Display function: sRGB White point: D65 (6500K) Color space: sRGB 30 minutes of warm-up time |
| 2.Spatial resolution | Effective Resolution, IDMS 1.03, Section 7.7 | Vertical and Horizontal MTFs are 0.861 and 0.862 at Nyquist frequency | Vertical and Horizontal MTFs are 0.877 and 0.885 at Nyquist frequency | Vertical and Horizontal MTFs are 0.861 and 0.862 at Nyquist frequency | Vertical and Horizontal MTFs are 0.877 and 0.885 at Nyquist frequency |
| 3.Pixel defects (count and map) | Defective Pixels, IDMS 1.03, Section 7.6 | Total number of bright and dark pixels <= 5 | Total number of bright and dark pixels <= 5 | Total number of bright and dark pixels <= 5 | Total number of bright and dark pixels <= 5 |
| 4. Artifacts | Artifacts and Irregularities, IDMS 1.03, Section 10.4 | Deviation <0.54% | Deviation <0.62% | Deviation <0.34% | Deviation <0.68% |
| 5. Temporal response | Gray-to-Gray Response Time, IDMS 1.03, Section 10.2.3 | Tmax: 20.87ms Tmin: 9.88ms | Tmax: 20.09ms Tmin: 3.84ms | Tmax: 22.87ms Tmin: 6.11ms | Tmax: 22.09ms Tmin: 3.56ms |
| 6. Maximum, minimum luminance and calibrated luminance | Vantage-Point Suite of Measurement, IDMS 1.03, Section 2.4 | Maximum: 523.6cd/m² Minimum: 0.501cd/m² | Maximum: 502.9cd/m² Minimum: 0.4805cd/m² | Maximum: 510cd/m² Minimum: 0.499cd/m² | Maximum: 502.6cd/m² Minimum: 0.515cd/m² |
| 7. Grayscale | Grayscale, IDMS 1.03, Section 6.1 | Maximum deviation:5.94 % | Maximum deviation:5.33 % | Maximum deviation:5.52% | Maximum deviation:6.07 % |
K233119 - Page 6 of 8
{6}
| | | | | | |
| --- | --- | --- | --- | --- | --- |
| 8. Luminance uniformity and Mura test | Sampled Vantage-Point Uniformity, IDMS 1.03, Section 8.1.2 | 11.16% non-uniformity on 100% video level | 6.89% non-uniformity on 100% video level | 9% non-uniformity on 100% video level | 9.26% non-uniformity on 100% video level |
| 9. Stability of luminance and chromaticity response with temperature and lifetime | Warm-up time, IDMS 1.03, Section 10.1 | Deviation from target luminance (500 cd/m²): 0.48% Variations for chromaticity: < 0.55% | Deviation from target luminance (500 cd/m²): 0.66% Variations for luminance and chromaticity: < 0.64% deviation | Deviation from target luminance (500 cd/m²): 0.66%, Variations for chromaticity: < 0.3% | Deviation from target luminance (500 cd/m²): 0.63% Variations for luminance and chromaticity: < 0.7% deviation |
| 10. Bidirectional reflection distribution function | Diagnostic: Characterizing Hemisphere Uniformity, IDMS 1.03, Section 11.12 | Specular reflection coefficient: 4.4%, Diffuse reflection coefficient: 0.92% | Specular reflection coefficient: 4.19%, Diffuse reflection coefficient: 0.8% | Specular reflection coefficient: 4.4%, Diffuse reflection coefficient: 0.92% | Specular reflection coefficient: 4.19%, Diffuse reflection coefficient: 0.8% |
| 11. Gray Tracking | Gray-scale Color Changes IDMS v1.03, Section 6.15 | 0.0006 Δu'v' | 0.0016Δu'v': | 0.0006 Δu'v' | 0.0014Δu'v': |
| 12. Color scale | Gray- and Color-Scale Measurement and 5.4 Color-Signal White, IDMS 1.03, Section 6 | Maximum color error:<2ΔE Average color error:<1ΔE | Maximum color error:<2ΔE Average color error:<1ΔE | Maximum color error:<3ΔE Average color error:<2ΔE | Maximum color error:<3ΔE Average color error:<2ΔE |
| 13. Color gamut volume | Volume-Color-Reproduction Capability, IDMS 1.03, Section 5.18 | 2D color gamut overlapped with sRGB:99.5% | 2D color gamut overlapped with sRGB:99.6% | 2D color gamut overlapped with sRGB:98.7% | 2D color gamut overlapped with sRGB:99.6% |
System-level test was also performed to test the displays with the Leica GT450DX system by displaying a test target and multiple tissue sample images. The results were acceptable.
## VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
K233119 - Page 7 of 8
{7}
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K233119 - Page 8 of 8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.