MYOTouch Muscle Stimulator

K233098 · Sunmed, LLC · IPF · Aug 2, 2024 · Physical Medicine

Device Facts

Record IDK233098
Device NameMYOTouch Muscle Stimulator
ApplicantSunmed, LLC
Product CodeIPF · Physical Medicine
Decision DateAug 2, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.5850
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The MYOTouch Muscle Stimulator is indicated for use as an adjunct in the treatment of high and low anorectic malformations by helping with the identification of striated muscles to be used in anal reconstructions. The MYOTouch device is intended for use by trained personnel who are competent to apply the appropriate stimuli.

Device Story

MYOTouch is a handheld, 9V battery-powered, single-channel muscle stimulator used by pediatric surgeons in clinical/neurosurgical operating environments. It assists in identifying striated muscles during anorectal reconstruction (e.g., PSARP). The device delivers electrical stimulation via a sterile bipolar stylus (MS200). The surgeon uses manual push-button controls to select between two stimulation modes (single or continuous pulse). The device transforms battery power into controlled electrical pulses; it features automatic no-load and overload trip mechanisms for safety. The surgeon observes the patient's muscle response to the stimulation to guide surgical decision-making. The device benefits patients by facilitating accurate muscle identification during complex reconstructive procedures.

Clinical Evidence

No clinical tests were submitted. Substantial equivalence is supported by bench testing, including electrical safety, electromagnetic compatibility, and risk management standards.

Technological Characteristics

Powered muscle stimulator; 9V battery-powered; microprocessor-controlled; mono-phasic rectangular waveform; 50Hz frequency; 200μs/500μs pulse widths; 1.44mm diameter bipolar probe; ABS plastic housing. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, and IEC 62304. Biocompatibility tested per ISO 10993-1.

Indications for Use

Indicated for use as an adjunct in the treatment of high and low anorectal malformations in neonates, infants, children, and adolescents (<18 years) to identify striated muscles for anal reconstruction. Intended for use by trained personnel.

Regulatory Classification

Identification

A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font below. August 2, 2024 SunMed LLC. Robert Yamashita VP of Regulatory Affairs 2710 Northridge Dr. NW. Suite A Grand Rapids, Michigan 49544 Re: K233098 Trade/Device Name: MYOTouch Muscle Stimulator Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: IPF Dated: June 14, 2024 Received: June 14, 2024 ### Dear Robert Yamashita: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). {1}------------------------------------------------ Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerelv. Xiaorui Tang For: CDR Jitendra Virani Assistant Director DHT5B: Division of Neuromodulation and Rehabilitation Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use Submission Number (if known) K233098 Device Name MYOTouch Muscle Stimulator #### Indications for Use (Describe) The MYOTouch Muscle Stimulator is indicated for use as an adjunct in the treatment of high and low anorectic malformations by helping with the identification of striated muscles to be used in anal reconstructions. The MYOTouch device is intended for use by trained personnel who are competent to apply the appropriate stimuli. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary K233098 Date of Submission Prepared: 11 November 2023 #### l Submitter: | Submitter: | Sunmed L.L.C. | |------------------------------------|--------------------------------------------------------------------| | Address: | 2710 Northridge Dr. NW, Suite A,<br>Grand Rapids, MI 49544, U.S.A. | | Telephone: | (616) 259-8419 | | Fax: | (800) 400-8820 | | E-Mail: | ryamashita@sun-med.com | | Official Contact: | Robert Yamashita / VP of Regulatory Affairs | | FDA Establishment Registration No: | 1314417 | SunMed L.L.C. #### Address of the manufacturing facility: EasyMed Instruments Co Ltd 3/F-6F, Block A, No. 4, Fengxin Road, Fengxiang Industrial District, Daliang, Shunde, Foshan, Guangdong, 528300 China FDA Establishment Registration No: 3004049909 #### ll Submitted Device: | Trade Name: | MYOTouch | |-----------------------|----------------------------------| | Common Name: | Powered Muscle Stimulator Device | | Regulation Name: | Powered Muscle Stimulator | | Regulation Number: | 21 CFR 890.5850 | | Product Code: | IPF | | Classification Panel: | Physical Medicine | | Device Class: | Class II | #### ��� Predicate Device: | Trade / Device Name: | Digitimer DS7AP Constant Current Stimulator | |-----------------------|---------------------------------------------| | Submitter: | Digitimer Ltd | | 510(k) Number: | K172381 | | Regulation Name: | Powered Muscle Stimulator | | Regulation Number: | 21 CFR 890.5850 | | Product Code | IPF | | Classification Panel: | Physical Medicine | | Device Class: | Class II | {4}------------------------------------------------ #### IV Device Description: MYOTouch is a hand-held device designed to be used for surgical correction of neonatal anorectal malformations. MYOTouch is intended for prescription use. The device is a 9V battery-powered, one-channel muscle stimulator. The device is supplied with a Sterile Stylus, which connects to the control unit by cable and plug. The MYOTouch muscle stimulator has been specially designed to meet the requirements of pediatric surgeons who need a reliable and easy to use muscle locating stimulator when carrying out the Pena, "Pull-through", or posterior sagittal anorectoplasty (PSARP) anorectal reconstruction technique. A trained professional or pediatric surgeon must use the MYOTouch device. The pediatric surgeon uses manual, push-button controls to control the electrical stimulation level. The unit is intended for prescription use and is designed with simplicity and ease of use. It has two preset modes available. #### Intended Patient Population The reconstruction surgeries for which the MYOTouch is intended are typically performed in newborns, infants, children, and adolescents (<18 years of age). Philip K. Frykman, M.D., Ph.D., M.B.A., FACS, FAAP confirm the intended patient population. The intended population for the MYOTouch device is as listed below: - Neonate/Newborn (birth through 28 days) с - Infant (from 29 days to 2 years of age) O - Child (from 2 years to 12 years of age) o - Adolescent (from 12 years to 18 years of age) o ### Intended Part of the Body to be Interacted With The MYOTouch is used in the treatment of anorectal malformations by helping identify striated muscles to be used in anal reconstructions. #### Intended User Profile The intended user is a specialist clinician trained in neurophysiological studies or a pediatric surgeon trained in the specialist reconstruction surgeries for which the MYOTouch is indicated for. #### Accessories: - - MS200 Sterile Stylus specifications: - o MYOTouch Frykman-Kimble Stimulator Stylus part number = MS200 - Replacement battery specifications: - o 9 Volt Alkaline battery ### ENVIRONMENT OF USE: Clinics, Hospital The MYOTouch Muscle Stimulator is intended for static tabletop use in a clinical environment, typically a neurosurgery operating theatre. {5}------------------------------------------------ #### > Indications for Use of the device: The MYOTouch Muscle Stimulator is indicated for use as an adjunct in the treatment of high and low anorectic malformations by helping with the identification of striated muscles to be used in anal reconstructions. The MYOTouch device is intended for use by trained personnel who are competent to apply the appropriate stimuli. #### VI Equivalence Comparison to the Predicate Devices: Electrical muscle stimulation is the technological principle for the MYOTouch and the predicate device Digitimer DS7AP (K172381). The intended Patient Population of MYOTouch is different from those of the predicate device Digitimer DS7AP (K172381), including individuals 3 - 18 years of age. Both devices have included the highest risk population, i.e., Neonate/Newborn (birth through 28 days), Infant (from 29 days to 2 years of age), and children <3 Years. Therefore, the difference does not raise new questions of safety or effectiveness. The technical characteristics of MYOTouch are similar to those of the predicate device in design, intended use, and function. The new device MYOTouch and predicate device Digitimer DS7AP (K172381) apply an electrical current via a disposable bipolar probe to a patient's striated muscles. The stimulation parameters of MYOTouch are similar to those of the predicate device Digitimer DS7AP (K172381). MYOTouch has two programs, and the parameters of MYOTouch are all in the similar range as those of the predicate device DS7AP (K172381). Table 1 below summarizes the shared and unique technological elements between the MYOTouch and DS7AP (K172381). | Attribute | Subject Device | Predicate Device | Comparison | |------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Product Name | MYOTouch | Digitimer DS7AP | | | Submitter | SunMed L.L.C. | Digitimer Ltd | | | 510(K) number | K233098 | K172381 | | | Product Code | IPF | IPF | Identical | | Regulation No. | 21 CFR 890.5850 | 21 CFR 890.5850 | Identical | | Device<br>Classification | Class II | Class II | Identical | | Attribute | Subject Device | Predicate Device | Comparison | | Indications<br>for Use | The MYOTouch Muscle<br>Stimulator is indicated for<br>use as an adjunct in the<br>treatment of high and low<br>anorectal malformations<br>by helping with the<br>identification of striated<br>muscles to be used in<br>anal reconstructions. The<br>MYOTouch device is<br>intended for use by<br>trained, competent<br>personnel to apply the<br>appropriate stimuli. | The Digitimer DS7AP<br>Constant Current<br>Stimulator is indicated for<br>use as an adjunct in the<br>treatment of high and low<br>anorectal malformations<br>by helping in the<br>identification of the striated<br>muscles to be used in anal<br>reconstructions. | Similar. The subject<br>device includes the<br>intended user<br>population in the<br>indications for use<br>statement | | Intended Use | Used to identify striated<br>muscle in the treatment of<br>high and low anorectal<br>malformations | Used to identify striated<br>muscle in the treatment of<br>high and low anorectal<br>malformations | Identical | | Prescription or<br>Over-the-<br>Counter | Prescription Use | Prescription Use | Identical | | Number of<br>output channels | One | One | Identical | | Regulated<br>Current or<br>Regulated<br>Voltage? | Yes, Constant Current | Yes, Constant Current | Identical | | Software/Firmw<br>are/<br>Microprocessor<br>Control? | Yes | No | Different<br>See Note 1 | | Automatic<br>No-<br>Load Trip? | Yes | Yes | Identical | | Automatic<br>Overload Trip? | Yes | Yes | Identical | | Automatic Shut<br>Off? | No | No | Identical | | User Override<br>Control? | Yes | Yes | Identical | | Indicator<br>Display: | Yes | Yes<br>N/A | Similar | | -On/Off status | Yes | Yes | See Note 2 | | Attribute | Subject Device | Predicate Device | Comparison | | -Low battery<br>-Voltage<br>/Current level<br>-Time to cut off | No | No | | | Mode<br>of<br>Operation | 2 | 2 | Identical | | Frequency | 50Hz | 50Hz | Identical | | Pulse Width | 200μς / 500μς | 200μς (LO) / 500μς (HI) | Identical | | Waveform | Mono-phasic | Mono-phasic | Identical | | Shape | Rectangular | Rectangular | Identical | | Maximum<br>Output Voltage | 56V@ 500Ω | 56V@ 500Ω | Identical | | Maximum<br>Output Current | 112mA@ 500Ω | 112mA@ 500Ω | Identical | | Maximum<br>Phase Charge | 56μC@ 500Ω | 56μC@ 500Ω | Identical | | Probe Diameter | 1.44mm | 0.73mm | Different | | Contact<br>Surface Area | 0.033cm² | 0.0084cm² | See Note 3<br>Different | | Maximum<br>Current Density | 3.44A/cm² @ 500Ω | 13.52A/cm²@ 500Ω | See Note 3<br>Similar | | Maximum<br>Power Density | 4.96 W/cm²@ 500Ω | 18.93W/cm²@ 500Ω | See Note 3<br>Similar | | Power Source | One 9V alkaline battery | Supply Mains (115/230V;<br>50/60Hz) | See Note 2<br>Different | | Dimensions [W<br>x H x D] | 113 x 61 x 25mm | 255 x 100 x 225 mm | Similar<br>The differences in<br>device dimensions do<br>not raise new<br>questions about<br>safety or<br>effectiveness. | | Weight | 154g with battery(0.34lb) | 2400g(4.85lb) | Similar<br>The difference in<br>device weight does<br>not raise new<br>questions about | | Attribute | Subject Device | Predicate Device | Comparison | | | | | safety or effectiveness. | | | | | Different | | Housing Materials and Construction | A.B.S. plastics | Thermo-plastic with metal front and rear panels | The difference in the housing material does not raise any additional safety and effectiveness problems.<br>The subject device was tested per ISO 10993-1, with passing results. | | Electrode lead wires and patient cable compliance with 21 CFR 898 | Yes | Yes | Identical | | Footswitch | NA | Yes | The predicate device (DS7AP) uses a footswitch to control the delivery of stimulation, while MYOTouch uses buttons on the main unit to control stimulation delivery.<br>This difference does not raise new questions about safety or effectiveness. | | Compliance with Voluntary Standards | Yes<br>See section 1, 1.3 Comparisons of Applied Standards | Yes<br>See section 1, 1.3 Comparisons of Applied Standards | Identical | | Operator Interaction: Functions Controllable: An explanation of how the device interacts | The operator can control the electrical stimulation levels, the starting and stopping stimulation. | The operator can control the electrical stimulation levels and the starting and stopping stimulation. | Identical | | Attribute | Subject Device | Predicate Device | Comparison | | with the<br>operator. | | | | | Operator<br>Interaction:<br>Programming<br>Capability<br>Whether the<br>device can be<br>programmed<br>and to what<br>extent | The operator sets<br>electrical stimulation<br>levels. Operators can<br>choose between two<br>stimulation modes. | The operator sets<br>electrical stimulation<br>levels. Operators can<br>choose between two<br>stimulation modes. | Identical | | Operator<br>Interaction:<br>Operator<br>Requirements<br>Knowledge and<br>training<br>required of the<br>operator | Prescription device.<br>Special knowledge and<br>training are required; a<br>detailed instruction<br>manual is provided | Prescription device.<br>Special knowledge and<br>training are required; a<br>detailed instruction manual<br>is provided | Identical | ### Table 1 Substantial Equivalence Comparison Table {6}------------------------------------------------ {7}------------------------------------------------ {8}------------------------------------------------ {9}------------------------------------------------ ### Different Technological Characteristics: 1. Software Control: MYOTouch uses a microprocessor to control the generation of desired outputs, while Digitimer DS7AP utilizes logical and analog circuits to generate outputs. The subject device is a complete digital device and gives the operator more accurate control and outputs. 2. Power Supply: Subject device MYOTouch uses a 9V battery as its power supply, while predicate device Digitimer DS7AP is A.C. main powered. The subject device provides the similar effective output characteristics of electrical pulses, compared to the predicate device. The subject device is much smaller in size, weight, and portability. Both devices are in compliance with IEC 60601-1 requirements. The differences in power source do not raise new questions about safety or effectiveness. 3. Stimulation Parameters: The stimulation parameters of the new device MYOTouch are the similar as those of the predicate device Digitimer DS7AP (K172381). MYOTouch has two modes, single and continuous pulse outputs, while Digitimer DS7AP (K172381) also has two modes, Low (200µs) and High (500µs), both being continuous output pulses. The new device, MYOTouch, offers the operator a single pulse output, which reduces risk to the patient because it uses minimum output energy application to achieve the similar muscle stimulation outcomes. The subject device's probe is larger in diameter compared to the predicate device, offering a greater contact surface area. The current and power densities are lower for the subject device when the same current intensity level is chosen. {10}------------------------------------------------ The power densities of both the predicate and subject device are greater than the recommended 0.25 W/cm² limit (per the FDA Guidance Document on Powered Muscle Stimulators) at certain stimulation level settings. Warnings have been added to the subject device's labelling to provide similar risk mitigations provided in the predicate device. 4. Mode of Operation: MYOTouch has two modes, single and continuous pulses outputs, while Digitimer DS7AP also has two modes, Low (200μs) and High (500μs), both being continuous output pulses. The new device, MYOTouch, offers the operator a single pulse output, which is safer for the patient because it uses minimum output energy application to achieve the similar muscle stimulation outcomes. 5. Housing Materials and Construction: MYOTouch has a complete A.B.S. plastic housing, while the predicate device uses Thermoplastic with metal front and rear panels. The housing material does not contact the patient. The subject device was tested per ISO 10993-1, with passing results. These differences in housing material and construction does not raise new concerns about safety and effectiveness in their intended use. {11}------------------------------------------------ #### Performance Tests: VII A series of safety and performance tests were conducted on the subject device MYOTouch. See below: | FDA recognition No. | Standard Title | |---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 19-49 | IEC 60601-1 Edition 3.2 2020-08 CONSOLIDATED VERSION<br>Medical electrical equipment - Part 1: General requirements for<br>basic safety and essential performance | | 19-36 | I.E.C. 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION<br>Medical electrical equipment - Part 1-2: General requirements for<br>basic safety and essential performance - Collateral Standard:<br>Electromagnetic disturbances - Requirements and tests | | 17-16 | I.E.C. 60601-2-10 Edition 2.1 2016-04<br>Medical electrical equipment - Part 2-10: Particular requirements for<br>the basic safety and essential performance of nerve and muscle<br>stimulators | | 5-125 | ISO 14971 Third Edition 2019-12<br>Medical devices - Application of risk management to medical<br>devices | | 13-79 | IEC 62304 Edition 1.1 2015-06 CONSOLIDATED VERSION<br>Medical device software - Software life cycle processes | All the test results demonstrate that MYOTouch meets the requirements of its predefined acceptance criteria and intended use. ## VIII Clinical Tests No clinical tests were submitted to support this 510(k) ## X Conclusion: - � The MYOTouch has similar technological characteristics and intended uses as the predicate Digitimer DS7AP (K172381), and - ◆ The labeling of the MYOTouch is concordant with the predicate device and FDA requirements, and - � The technological differences between subject device MYOTouch and predicate device DS7AP does not raise new questions about safety or effectiveness. It demonstrates with reasonable assurance based on established controls that the device is at least as safe and effective as the legally marketed predicate device. Therefore, the MYOTouch is substantially equivalent to the predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...