K233056 · Guidestar Medical Devices · FMF · Nov 2, 2023 · General Hospital
Device Facts
Record ID
K233056
Device Name
EpiZact
Applicant
Guidestar Medical Devices
Product Code
FMF · General Hospital
Decision Date
Nov 2, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
EpiZact is intended for use with an epidural needle for detecting a loss of resistance, which aids a clinician in verifying needle tip placement in the epidural space.
Device Story
EpiZact is a single-use, manually powered, self-pressurizing piston syringe; connects to standard 16-18 gauge Tuohy needles via Luer connector. Used in hospital settings by clinicians performing epidural procedures. Device utilizes loss of resistance (LOR) technique to detect epidural space entry. Provides both visual and tactile feedback upon detection; tactile feedback assists in reducing forward needle motion. Output allows clinician to confirm needle tip placement. Benefits include clear endpoint for needle advancement and enhanced safety through tactile feedback. Constructed from copolyester, ABS, stainless steel springs, and silicone components.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including ISO 7886-1 (syringes), ISO 80369-7 (connectors), USP 788 (particulates), and biocompatibility testing (ISO 10993-4, -5, -10, -11, -23). Sterile barrier integrity validated per ASTM F1980, F2096, F88, and F1886. Sterilization validated per ISO 11137.
Technological Characteristics
Self-pressurizing piston syringe. Materials: Copolyester, ABS, stainless steel springs, silicone O-rings/valves/lubricant. Luer slip connector (ISO 80369-7). Sterilization: E-beam radiation (ISO 11137). Manual operation; provides visual and tactile feedback.
Indications for Use
Indicated for use with an epidural needle to aid clinicians in verifying needle tip placement in the epidural space. No specific patient population, age, or disease state defined.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 2, 2023
GuideStar Medical Devices % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114
Re: K233056
Trade/Device Name: EpiZact Regulation Number: 21 CFR 880.5860 Regulation Name: Piston syringe Regulatory Class: Class II Product Code: FMF Dated: October 23, 2023 Received: October 23, 2023
Dear Prithul Bom:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.
{2}------------------------------------------------
See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Bradley Q. Quinn -S
Bradley Quinn Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known)
## Device Name EpiZact
Indications for Use (Describe)
EpiZact is intended for use with an epidural needing a loss of resistance, which ads a clinician in verifying needle tip placement in the epidural space.
Type of Use (Select one or both, as applicable)
Z Prescription Use (Part 21 CFR 801 Subpart D) [ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displaysa currently valid OMB number."
{4}------------------------------------------------
# 510(k) SUMMARY
#### SUBMITTER 1
| Submitter: | GuideStar Medical Devices Inc. |
|------------|--------------------------------|
| | 201-2067 Cadboro Bay Rd. |
| | Victoria, BC, Canada, V8R 5G4 |
| | 604-970-2714 |
Contact Person: Michael Dolphin, CEO 604-970-2714 mdolphin@guidestarmd.com
Date of Preparation: July 24, 2023
# 2 DEVICE INFORMATION
| Trade Name: | EpiZact |
|--------------------|-----------------|
| Regulation Number: | 21 CFR 880.5860 |
| Regulation Name: | Piston Syringe |
| Regulatory Class: | Class II |
| Product Code: | FMF |
#### 3 PREDICATE DEVICE
| Predicate Trade Name: | EpiFaith Syringe |
|-------------------------|-----------------------|
| Premarket Notification: | K192421 |
| Manufacturer: | Flat Medical Co., Ltd |
| Regulation Number: | 21 CFR 880.5860 |
| Regulation Name: | Piston Syringe |
| Regulatory Class: | Class II |
| Product Code: | FMF |
#### 4 DEVICE DESCRIPTION
EpiZact is a self-pressurizing piston syringe that uses the same loss of resistance technique used by physicians to detect the epidural space. The device is single use, sterile, hanually powered, and designed to connect to standard epidural Tuohy needles (16-18 gauge) with a standard Luer connector.
EpiZact provides tactile feedback, in addition to visual feedback, when a loss of resistance is detected. The tactile feedback is also designed to help reduce the forward motion of the needle upon detection of loss of resistance. The device is comprised of copolyester plastic, stainless steel springs, silicone Orings, silicone and ABS valve, and polydimethylsiloxane (silicone) lubricant.
{5}------------------------------------------------
The device is intended to be used in a hospital setting, with sterile medical supplies. The intended user of this device is a clinician who is performing an epidural procedure.
#### 5 INTENDED USE
EpiZact is intended for use with an epidural needle for detecting a loss of resistance, which aids a clinician in verifying needle tip placement in the epidural space.
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH PREDICATE б
| Characteristics | EpiZact | EpiFaith | Differences |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | Intended for use with an<br>epidural needle for detecting a<br>loss of resistance, which aids a<br>clinician in verifying needle tip<br>placement in the epidural<br>space. | Intended for use with an<br>epidural needle for detecting a<br>loss of resistance, which aids a<br>clinician in verifying needle tip<br>placement in the epidural<br>space. | Same |
| Patient<br>Population | Not designed for the treatment<br>of any specific disease or<br>condition, nor does it have a<br>target population. | Not designed for the treatment<br>of any specific disease or<br>condition, nor does it have a<br>target population. | Same |
| Principle of<br>operation | Loss of resistance (LOR) | Loss of resistance (LOR) | Same |
| Signal<br>mechanism | A visual and tactile signal is<br>triggered when the needle tip<br>enters the epidural space,<br>indicating a clear endpoint for<br>the advancement of the<br>needle. | A visual signal is triggered<br>when the needle tip enters the<br>epidural space, indicating a<br>clear endpoint for the<br>advancement of the needle. | Similar. EpiZact has tactile<br>feedback in addition to visual<br>feedback. |
| Nozzle type | Compatible with 16–18 gauge<br>epidural needles. One model:<br>Luer slip (as per ISO 80369-7) | Compatible with 16-18-gauge<br>epidural needles. Two models<br>(two colors):<br>Luer slip (as per ISO80369-7)<br>NRFit™ (as per ISO80369-6) | Luer slip connector (ISO 80369-<br>7) is the same.<br>NRFit™ connector (ISO 80369-<br>6) is not part of this EpiZact<br>application. |
| Materials | Plastic Components:<br>Copolyester and ABS<br>O-Ring/Seal/Valve: Silicone<br>Springs: Stainless steel | Plastic Components:<br>Polypropylene<br>O-Ring/Seal: Synthetic and<br>silicone rubber<br>Spring: Stainless steel | Similar. The material<br>differences between the<br>subject and predicate devices<br>do not alter suitability of the<br>proposed device for its<br>intended use. Materials were<br>subject to a biocompatibility<br>assessment per ISO 10993- 1. |
| Lubricant | Silicone oil | Silicone oil | Same |
| Biocompatibility | Biocompatibility Assessment<br>meets ISO 10993-1. | Biocompatibility Assessment<br>meets ISO 10993-1. | Same |
| Single use | Yes | Yes | Same |
| Characteristics | EpiZact | EpiFaith | Differences |
| Sterilization | E-beam radiation as per ISO<br>11137.<br>Sterile assurance level: 10-6 | E.O gas sterilization as per ISO<br>11135.<br>Sterile assurance level: 10-6 | Similar. Both sterilization<br>methods are recognized as<br>Established Category A<br>sterilization methods.<br>Sterilization methodology was<br>verified through sterilization<br>validation. |
| Performance<br>standards | ISO 7886-1<br>ISO 80369-7<br>USP 788 particulates<br>Operational Bench Tests | ISO 7886-1<br>ISO 80369-7<br>USP 788 particulates | Same, with additional bench<br>tests performed on EpiZact. |
| Packaging | Tyvek and PET pouch with SBS<br>backer card. | Tyvek and PET pouch. | Same, with additional SBS<br>cardboard backer card in<br>EpiZact packaging. |
{6}------------------------------------------------
The subject device (EpiZact) has the same indications for use, use environment, patient population, and principle of operation as the predicate device (EpiFaith). Both the subject and predicate use the same lubricant and are single use. The nozzle type (Luer) is identical to one of the two models of the predicate. The technical characteristics that are not identical are as follows:
Signal Mechanism - EpiZact provides tactile feedback in addition to visual feedback when a loss of resistance is detected. The tactile feedback is also designed to help reduce the forward motion of the needle. This adds no new concerns for safety or efficacy, and ultimately adds a margin of safety by providing additional feedback to the user and reduced needle motion.
Materials - The plastic components of EpiFaith are made from polypropylene, whereas the plastic components of EpiZact are made from copolyester and ABS. All EpiZact materials are medical grade and of similar properties to the materials of EpiFaith. The entire EpiZact assembly was validated through a biocompatibility assessment per ISO 10993.
Sterilization - EpiZact is sterilized by E-beam. EpiFaith is sterilized by E.O. gas. Both methods of sterilization are "Established Category A" methods, with equivalent sterilization method for EpiZact was validated through sterilization testing per ISO 11137.
Performance Testing - EpiZact completed all the same performance tests as EpiFaith, with the addition of benchtop performance tests. These custom tests were added to assess operational performance of the device.
Packaging - EpiZact and EpiFaith both utilize Tyvek and PET (polyethylene terephthalate). EpiZact packaging contains an additional backer made from SBS (solid bleached sulfate). The packaging configuration for EpiZact was validated with distribution conditioning, environmental conditioning, and sterile package integrity. Biocompatibility testing conducted on packaged devices verify that this packaging presents no new concerns for safety.
{7}------------------------------------------------
#### 7 PERFORMANCE DATA
The following tests were performed on EpiZact in support of the substantial equivalence determination:
## Performance Tests
ISO 7886-1:2017 – Sterile hypodermic syringes for single use — Part 1: Syringes for manual use
- . Section 6.1 – Extraneous Matter (Visual Inspection)
- . Section 6.2, Annex A - Limits for Acidity
- . Section 6.3, Annex A – Limits for Extractable Metals
- Section 7 - Lubricant (Visual Inspection)
- . Section 12.2 – Position of Nozzle on End of Barrel (Visual Inspection)
- Section 12.3 – Nozzle Lumen (Visual Inspection)
- . Section 13.2, Annex B – Freedom from Air Leakage Past Plunger Stopper
- Section 13.4 Fit of Plunger/Stopper in Barrel .
ISO 80369-7:2021 – Small-bore connectors for liquids and gases in healthcare applications — Part 7: Connectors for intravascular or hypodermic applications
- . Section 5 - Dimensioning
- Section 6.2, Annex D Sub-Atmospheric Pressure Air Leakage .
- . Section 6.1.3, Annex C - Falling Drop Positive Pressure Liquid Leakage
- . Section 6.4, Annex F – Resistance to Separation from Axial Load
- . Section 6.5, Annex G - Resistance to Separation from Unscrewing
- Section 6.6, Annex H Resistance to Overriding ●
- Section 6.3, Annex E Stress Cracking ●
USP <788> – Particulate Matter in Injections
## Biocompatibility Tests
ISO 10993-4:2017 – Biological evaluation of medical devices — Part 4: Selection of tests for interactions with blood
- Hemolysis Test ●
ISO 10993-5:2009 – Biological evaluation of medical devices — Part 5: Tests for in vitro cytotoxicity
- Cytotoxicity Evaluation ●
ISO 10993-10:2021 – Biological evaluation of medical devices — Part 10: Tests for skin sensitization
- Dermal Sensitization
ISO 10993-11:2017 – Biological evaluation of medical devices — Part 11: Tests for systemic toxicity
- Acute Systemic Toxicity
ISO 10993-23:2021 – Biological evaluation of medical devices — Part 23: Tests for irritation
- . Intracutaneous Test
{8}------------------------------------------------
## Sterile Barrier Package Testing
ASTM F1980-21 – Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices
- ASTM D4332-14 Standard Practice for Conditioning Containers, Packaging Components for Testing
- ASTM D4169-22 Standard Practice for Performance Testing of Shipping Containers and Systems
- ASTM F2096-11 Standard Test Method for Detecting Gross Leaks in Porous Medical Packaging by Internal Pressurization (Bubble Leak)
- ASTM F88/F88M-15 Standard Test Method for Seal Strength of Flexible Barrier Materials
- ASTM F1886/F1886M-16 Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection
## Sterilization Testing
ISO 11137-1:2006/(R)2015 & A1:2013 & A2:2019 – Device Sterilization ISO 11137-2:2013/(R)2019 - Sterility Verification ISO 11137-3:2017 - Dose Mapping ISO 11737-1:2018 & A1:2021 — Bioburden USP <85> – Bacterial Endotoxins
Performance Benchtop Testing
Production Benchtop Testing Human Factors Studies
#### 8 CONCLUSION
Testing conducted under performance, biocompatibility, sterile barrier packaging, and sterilization demonstrate the subject device is substantially equivalent to the predicate device. The differences between the devices do not raise any new questions of safety or effectiveness.
The conclusions drawn from the nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well or better than the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.