TD 200® Automated TEE Probe Disinfector with TD-12® High-Level Disinfectant with QwikCheck™ Chemical Indicator
Applicant
Cs Medical, LLC
Product Code
PSW · Radiology
Decision Date
Nov 21, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The TD 200® Automated TEE Probe Disinfector with TD-12® High-Level Disinfectant with QwikCheck™ Chemical Indicators is designed to provide high-level disinfection of Transesophageal (TEE) ultrasound probes. The system can use TD-12® disinfectant, which is designed to be used only with the TD 200® disinfector. The disinfectant bottles cannot be reused in the system. TD-12® disinfectant is intended for use as single use high-level disinfectant to be used exclusively in the TD 200® disinfector for high-level disinfection of TEE ultrasound probes. TD-12® high level disinfectant and TD 200® disinfector is intended for use by qualified individuals trained in its use. TD-12® disinfectant should be used with the following contact conditions in TD 200® disinfector: High-Level Disinfectant: TD-12®; Time: 3 minutes; Temperature: 38°C; Minimum Recommended Concentration: 1750 ppm peracetic acid. QwikCheck™ Chemical Indicators are used prior to the disinfection cycle to ensure that the disinfectant used is above the minimum recommended concentration (MRC) of 1750ppm peracetic acid.
Device Story
Automated system for high-level disinfection of TEE ultrasound probes; utilizes TD-12 peracetic acid-based disinfectant. Process: bedside/manual pre-cleaning of probe; loading of fresh TD-12 bottle; automated cycle heating disinfectant to 38°C; 3-minute soak; rinse; printout of verification report (time, temperature, success/failure). QwikCheck chemical indicator used prior to cycle to verify peracetic acid concentration >1750 ppm. Used in clinical settings by trained personnel. Vapor management system uses air circulation and activated carbon filtration to control user exposure. Benefits: standardized, automated disinfection; reduced user exposure to disinfectant; verification of disinfection parameters.
Clinical Evidence
No clinical trials performed. Evidence consists of bench testing, including automated simulated-use studies demonstrating complete microbial kill, vapor management capacity testing, and material compatibility testing with GE, Philips, and Siemens TEE probes.
Technological Characteristics
Automated disinfector; 110 VAC; microcontroller-controlled. Uses peracetic acid (TD-12) at 38°C for 3 minutes. Vapor management via activated carbon filtration. QwikCheck indicator uses starch/iodide test paper. Electrical safety per UL 61010-1. Connectivity includes built-in printer for cycle verification.
Indications for Use
Indicated for high-level disinfection of Transesophageal (TEE) ultrasound probes by qualified, trained individuals.
Regulatory Classification
Identification
A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.
Predicate Devices
TD 200/TD-12 disinfector/disinfectant system (K192228)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
November 21, 2023
CS Medical LLC Kendall Ashe Vice President, General Manager 2179 East Lyon Station Rd Creedmoor, North Carolina 27522
Re: K233049
Trade/Device Name: TD 200® Automated TEE Probe Disinfector with TD-12® High-Level Disinfectant with QwikCheck™ Chemical Indicator Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic Ultrasonic Transducer Regulatory Class: Class II Product Code: PSW, MED, JOJ Dated: September 25, 2023 Received: September 25, 2023
Dear Kendall Ashe:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Bifeng Qian -S
Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K233049
Device Name
TD 200® Automated TEE Probe Disinfector with TD-12® High Level Disinfectant with OwikCheck™ Chemical Indicators
Indications for Use (Describe)
The TD 2008 Automated TEE Probe Disinfector with TD-12® High Level Disinfectant with QwikCheck™ Chemical Indicators is designed to provide high-level disinfection of Transesophageal (TEE) ultrasound probes. The system can use TD-12® disinfectant, which is designed to be used only with the TD 2008 disinfector. The disinfectant be reused in the system.
TD-12® disinfectant is intended for use as single use high-level disinfectant to be used exclusively in the TD 200® disinfector for high-level disinfection of TEE ultrasound probes.
TD-12® high level disinfectant and TD 200® disinfector is intended for use by qualified individuals trained in its use.
TD-12® disinfectant should be used with the following contact conditions in TD 200® disinfector:
| High-Level Disinfectant | Time | Temperature | Minimum Recommended Concentration |
|-------------------------|-----------|-------------|-----------------------------------|
| TD-12® | 3 minutes | 38°C | 1750 ppm peracetic acid |
OwikCheck™ Chemical Indicators are used prior to the disinfection cycle to ensure that the disinfectant used is above the minimum recommended concentration (MRC) of 1750ppm peracetic acid.
Type of Use (*Select one or both, as applicable*)
| | Prescription Use (Part 21 CFR 801 Subpart D) |
|--|----------------------------------------------|
| | Over-The-Counter Use (21 CFR 801 Subpart C) |
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{4}------------------------------------------------
## 510(k) Summary
## K233049
510(k) Owner CS Medical L.L.C. 2179 E. Lyon Station Road Creedmoor, North Carolina 27522 Phone 919-255-9472 Fax 919-528-3400
Contact Name Kendall Ashe Vice President and General Manager, CS Medical, LLC 2179 E. Lyon Station Road Creedmoor. North Carolina 27522 Email: kendallashe@csmedicalllc.com
Submission Prepared 25-Sept-2023
CS Medical Trade Name TD 200® Automated TEE Probe Disinfector with TD-12® High Level Disinfectant with QwikCheck™ Chemical Indicators
Common Name TD 200 disinfector TD-12 disinfectant and QwikCheck chemical indicator
Classification Name Diagnostic ultrasonic transducer (21 CFR 892.1570, Primary Product Code. PSW Secondary Product Code, MED, JOJ
Legally Marketed Predicate Devices
1. TD 200/TD-12 disinfector/disinfectant system (K192228) - Primary Predicate
2. IBT Scientific Peracid Test (K964264)
## Description of the CS Medical TD 200® Disinfector with TD-12® High Level Disinfectant and QwikCheck™ Indicator
The TD 200 disinfector provides high-level disinfection of transesophageal (TEE) ultrasound probes when used according to the operating instructions and when used with TD-12 disinfectant. The TD 200 disinfector is for use only with TD-12 disinfectant. The TD-12 is for use only in the TD 200 disinfector. Thus, the TD 200 disinfector and TD-12 disinfectant represent a dedicated system. Each soiled TEE probe is bedside cleaned and manually cleaned according to the TEE probe manufacturer's instructions before insertion into the TD 200 disinfector. A fresh, unopened bottle of TD-12 disinfectant is loaded into the TD 200 disinfector. The TD 200 disinfector heats the TD-12 disinfectant to at least 38°C, soaks the TEE probe at least three (3) minutes, and then thoroughly rinses the disinfectant off the TEE probe before the cycle is complete. The TD 200 disinfector prints a verification report indicating a successful disinfection cycle, the time and the average temperature during the disinfection. The TEE probe is then removed from the TD 200 disinfector and dried according to the TEE probe manufacturer's instructions. The TD 200 disinfector is ready for a new cycle immediately after the preceding cvcle is completed. A chemical indicator is used to ensure that the solution is above the MRC.
{5}------------------------------------------------
## TD 200, TD-12, and QwikCheck Indications for Use Statement:
The TD 200® disinfector is designed to provide high-level disinfection of Transesophageal (TEE) ultrasound probes. The system can use TD-12® disinfectant, which is designed to be used only with the TD 200® disinfector. The disinfectant bottles cannot be reused in the system.
TD-12® disinfectant is intended for use as single use high-level disinfectant to be used exclusively in the TD 200® disinfector for high-level disinfection of TEE ultrasound probes.
TD-12® high level disinfectant and TD 200® disinfector is intended for use by qualified individuals trained in its use.
TD-12® disinfectant should be used with the following contact conditions in TD 200® disinfector:
| High-level Disinfectant | Time | Temperature | Minimum Recommended Concentration |
|-------------------------|-----------|-------------|-----------------------------------|
| TD-12® | 3 minutes | 38°C | 1750 ppm peracetic acid |
QwikCheck™ Chemical Indicators are used prior to the disinfection cycle to ensure that the disinfectant used is above the minimum recommended concentration (MRC) of 1750 ppm peracetic acid.
## Comparative Table for Indications for Use
| | Proposed Device | Predicate Device | Comparison |
|-----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Element | TD 200 Disinfector with TD-<br>12 High-Level disinfectant<br>with QwikCheck Chemical<br>Indicator | TD 200 Disinfector with TD-<br>12 High-Level Disinfectant | |
| Classification<br>Name (CFR;<br>Product code) | Diagnostic ultrasonic<br>transducer (21 C.F.R. §<br>892.1570, product code<br>PSW) | Diagnostic ultrasonic<br>transducer (21 C.F.R. §<br>892.1570, product code<br>PSW) | Same |
| Indications for<br>Use | The TD 200 disinfector is<br>designed to provide high-<br>level disinfection of<br>Transesophageal (TEE)<br>ultrasound probes. The<br>system can use TD-12<br>disinfectant, which is<br>designed to be used only<br>with the TD 200 disinfector.<br>The disinfectant bottles<br>cannot be reused in the<br>system.<br>TD-12 disinfectant is<br>intended for use as single<br>use high-level disinfectant<br>to be used exclusively in the | The TD 200 disinfector is<br>designed to provide high-<br>level disinfection of<br>Transesophageal (TEE)<br>ultrasound probes. The<br>system can use TD-12<br>disinfectant, which is<br>designed to be used only<br>with the TD 200 disinfector.<br>The disinfectant bottles<br>cannot be reused in the<br>system.<br>TD-12 disinfectant is<br>intended for use as single<br>use high-level disinfectant<br>to be used exclusively in the | Same |
{6}------------------------------------------------
| TD 200 disinfector for high-<br>level disinfection of TEE<br>ultrasound probes.<br>TD-12 high level<br>disinfectant and TD 200<br>disinfector is intended for<br>use by qualified individuals<br>trained in its use.<br>TD-12 disinfectant should<br>be used with the following<br>contact conditions in TD<br>200 disinfector: | TD 200 disinfector for high-<br>level disinfection of TEE<br>ultrasound probes.<br>TD-12 high level<br>disinfectant and TD 200<br>disinfector is intended for<br>use by qualified individuals<br>trained in its use.<br>TD-12 disinfectant should<br>be used with the following<br>contact conditions in TD<br>200 disinfector: |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| High-level Disinfectant TD-<br>12<br>Time - 3 Minutes<br>Temperature - 38°C<br>Minimum Recommended<br>Concentration (MRC) 1750<br>ppm Peracetic Acid | High-level Disinfectant TD-<br>12<br>Time - 3 Minutes<br>Temperature - 38°C<br>Minimum Recommended<br>Concentration (MRC) 1.75%<br>Peracetic Acid |
| QwikCheck Chemical<br>Indicators are used prior to<br>the disinfection cycle to<br>ensure that the disinfectant<br>used is above the minimum<br>recommended<br>concentration (MRC) of<br>1750ppm peracetic acid. | |
## Comparison of Proposed TD 200 Device to Primary Predicate TD 200 Device
| Features | Proposed Device<br>TD 200® Automated<br>TEE Probe<br>Disinfector with TD-<br>12® High Level<br>Disinfectant with<br>QwikCheck Chemical<br>Indicator | Predicate Device<br>TD 200® Automated<br>TEE Probe<br>Disinfector with TD-<br>12® High Level<br>Disinfectant<br>(K192228) | Comparison |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|------------|
| The system shall utilize a<br>disinfectant that has at least a<br>one-year shelf life from the<br>date of manufacture. | Yes | Yes | Same |
| Single-use disinfectants only | Yes | Yes | Same |
| User hazard to disinfectant<br>contact reduced by bottle<br>loading system | Yes | Yes | Same |
| The system shall provide a<br>disinfection cycle that has a<br>disinfectant contact time, at<br>least 180 seconds and<br>disinfects at 38°C | Yes | Yes | Same |
| Disinfectant residues shall be<br>rinsed until the disinfectant<br>residue is below a<br>toxicologically significant<br>level. | Yes | Yes | Same |
| Reduced height of machine<br>for easy load of TEE probe | Yes | Yes | Same |
| Machine serial number, lot<br>number of disinfectant, probe<br>number, and operator number<br>on output ticket | Yes | Yes | Same |
| Minimal footprint, secured to<br>wall for stability | Yes | Yes | Same |
| Design | | | |
| Intended Use | Automated high-level<br>disinfection of TEE<br>probes | Automated high-level<br>disinfection of TEE<br>probes | Same |
| Manual Pre-cleaning | Yes | Yes | Same |
| Operating principles | Peracetic Acid high-<br>level disinfection | Peracetic Acid high-<br>level disinfection | Same |
| Process monitors | Digital display<br>screen, printout | Digital display<br>screen, printout | Same |
| Software/firmware control | Yes | Yes | Same |
| Performance claims | High-level<br>disinfection | High-level<br>disinfection | Same |
| Provides high-level<br>disinfection for heat-sensitive<br>TEE probes | Yes | Yes | Same |
| Uses PAA-based<br>disinfectants | Yes | Yes | Same |
| Automated disinfection cycle | Yes | Yes | Same |
| Disinfection process with<br>success/failure print out. | Yes | Yes | Same |
| User hazard to vapor<br>exposure controlled by vapor<br>management system utilizing<br>air circulation and activated<br>carbon filtration | Yes | Yes | Same |
| The system shall be<br>controlled by a<br>microcontroller processor. | Yes | Yes | Same |
| The system shall provide a<br>built-in printer. | Yes | Yes | Same |
| Specifications | | | |
| Dedicated Disinfectants for<br>use with the device | Peracetic acid<br>(TD-12) | Peracetic acid<br>(TD-12) | Same |
| Disinfectant Minimum<br>Recommended Concentration | 1750 ppm Peracetic<br>Acid | 1.75% Peracetic Acid | Same |
| 110 VAC | Yes | Yes | Same |
| Potable Water Inlet | Yes | Yes | Same |
| User hazard to vapor<br>exposure controlled by<br>vapor management<br>system utilizing air<br>circulation and filtration<br>with no room air<br>circulation required | Yes | Yes | Same |
| Use a chemical indicator | Yes | Yes | Same |
{7}------------------------------------------------
{8}------------------------------------------------
## Comparison of Proposed TD 12 Device to Primary Predicate TD 12 Device
| Physical<br>Properties | Proposed Device<br>TD-12 | Predicate Device<br>TD-12 (K192228) | Comparison |
|--------------------------------------------------|---------------------------------------|---------------------------------------|------------|
| Active<br>Ingredients | 0.30%<br>peracetic acid | 0.30%<br>peracetic acid | Same |
| Inert<br>Ingredients | 99.7% | 99.7% | Same |
| Water-Based<br>Liquid | No | No | Same |
| pH Value | 8.5 – 9.0 | 8.5 – 9.0 | Same |
| Buffer System | Sodium Carbonate and<br>Sulfamic Acid | Sodium Carbonate and<br>Sulfamic Acid | Same |
| Minimum<br>Recommended<br>Concentration<br>(MRC) | 1750 ppm | 1.75% | Same |
| Dilution<br>Required | Yes | Yes | Same |
| Activation<br>Required | Yes | Yes | Same |
| High-Level<br>Disinfection<br>Claim | Yes | Yes | Same |
| Sterilization<br>Claim | No | No | Same |
| Maximum Re-<br>Use Period | Single Use | Single Use | Same |
| PAA Test | Yes | Yes | Same |
{9}------------------------------------------------
| Strip Available | | | |
|----------------------------------------|-----------|-----------|------|
| Use in<br>Automated<br>Disinfector | Required | Required | Same |
| Manual Use | No | No | Same |
| Disinfectant<br>Temperature<br>Minimum | 38°C | 38°C | Same |
| Disinfection<br>Time | 3 Minutes | 3 Minutes | Same |
## Comparative Table between QwikCheck Chemical Indicators
| Element | QwikCheck Chemical<br>Indicator (proposed<br>device) | QwikCheck Chemical<br>Indicator (device used<br>in conjunction with<br>K192228) | Comparison |
|--------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Classification Name<br>(CFR; Product code) | Sterilization Process<br>Indicator (21 C.F.R. §<br>880.2800, product code<br>JOJ) | Sterilization Process<br>Indicator (21 C.F.R. §<br>880.2800, product code<br>JOJ) | Same |
| Indications for Use<br>for Chemical<br>Indicator | QwikCheck Chemical<br>Indicators are used prior<br>to the disinfection cycle<br>to ensure that the<br>disinfectant used is<br>above the minimum<br>recommended<br>concentration (MRC) of<br>1750ppm peracetic acid. | The QwikCheck™<br>Chemical Indicator is a<br>qualitative chemical<br>indicator dedicated for<br>use in determining<br>whether the<br>concentration of<br>peracetic acid, the active<br>ingredient in CS<br>Medical's TD-12 Solution,<br>is above or below the<br>established Minimum<br>Recommended<br>Concentration (MRC)<br>according to the solution<br>manufacturer. | Similar |
| Analyte | Peracetic acid in TD-12<br>Solution | Peracetic acid in TD-12<br>Solution | Same |
| Indicator Strip | 0.2" x 0.2" test paper<br>attached to a plastic<br>handle | 0.2" x 0.2" test paper<br>attached to a plastic<br>handle | Same |
| Indicator Agents | Starch and iodide | Starch and iodide | Same |
| Minimum Time of<br>Color Change | 30 seconds | 30 seconds | Same |
| Results | A test pad which is<br>similar in color to the<br>FAIL color block<br>indicates the solution is ≤<br>1750 ppm. If the color of<br>the test pad is similar to<br>the PASS block, the<br>solution is > 1750 ppm | A test pad which is<br>similar in color to the<br>FAIL color block<br>indicates the solution is ≤<br>1750 ppm. If the color of<br>the test pad is similar to<br>the PASS block, the<br>solution is > 1750 ppm | Same |
{10}------------------------------------------------
The only one difference between the subject and predicate devices is a change of supplier of the QwikCheck Chemical Indicator.
## Summary of Non-Clinical Testing
## Software Validation
All qualification activities defined in the TD 200 Automated TEE Probe Disinfector with TD-12 High Level Disinfectant Verification and Validation Plan were successfully completed and provided documented evidence that the TD 200 consistently performs as intended. All data collected during test execution met the established acceptance criteria, was recorded on the data forms, and satisfactorily completed. The TD 200 Verification and Validation Summary Report and the associated qualification documents for the TD 200 detail the successful completion of all requirements of the TD 200 Master Verification and Validation Plan.
## Electrical Safety Testing
The TD 200 Automated TEE Probe Disinfector withTD-12 High LevelDisinfectant has been investigated and passed all testing in accordance with UL 61010-1, 3rd Edition; Standard for Safety, Electrical Equipment for Measurement, Control, and Laboratory Use; Part 1: General Requirements; Rev. April 29, 2016.
## Biocompatibility
Under conditions of the study, TD-12 is not a sensitizer, nor an irritant and is non-cytotoxic. Transesophageal probes processed in the TD 200 Automated TEE Probe Disinfector with TD-12 High Level Disinfectant were evaluated for disinfectant residue. The analysis indicates that the level of residue on the TEE probes are not likely to have toxic effects on humans.
#### Material Compatibility
The materials used to construct the TD 200 Automated TEE Probe Disinfector with TD-12 High Level Disinfectant were exposed to TD-12 disinfectant for a pre-determined period of time. There were no observable effects from exposure to TD-12 disinfectant on materials.
## Compatibility Testing with TEE Probes
The TD 200 Automated TEE Probe Disinfector withTD-12 High Level Disinfectant should be used only with TEE probes which have been tested and approved by TEE probe manufacturers. Material Compatibility has been performed with GE, Philips and Siemens TEE probes. These probes did not have cosmetic nor functional deterioration. TD-12 probe compatibility testing was
completed using a use-suspension of 3% w/v (approximately 3000ppm) PAA TD-12 at 38°C.
{11}------------------------------------------------
## Performance Testing
TD-12 High Level Disinfectant vapor exposure studies were performed in a room with zero air exchanges and showed the main vapor management filter has capacity for 12 months of use. Automated Simulated use studies with TD-12 were conducted in the TD 200 disinfector. In all cases there was a measurably complete kill of microorganisms after TEE probe processing.
#### Clinical In-Use Testing
After routine clinical use, soiled TEE probes were subjected to disinfection in the TD 200 Automated TEE Probe Disinfector with TD-12 High Level Disinfectant using standard operating parameters. In all cases, there was a complete kill of microorganisms after TEE probe processing.
No clinical in-use testing was performed with the QwikCheck chemical indicators as no clinical in-use testing was performed with the predicate chemical indicators so this testing is not applicable.
## Conclusion
Based on the intended use, technological characteristics, and non-clinical performance data the TD 200 Automated TEE Probe Disinfector with TD-12 High Level Disinfectant with QwikCheck Chemical Indicators is as safe, as effective, and performs as well as or better than the legally marketed predicate device CS Medical TD 200 Automated TEE Probe Disinfector with TD-12 High Level Disinfectant (K192228).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.