The Cure Trio MC-310A is electrical stimulator designed for pain management and neuro muscular stimulation. a. When using Cure Trio to deliver Transcutaneous Electrical Nerve Stimulation (TENS Mode), it is intended to provide: [1] Symptomatic relief and management of chronic, intractable pain [2] Relief of pain associated with arthritis [3] Temporarily relieves pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities (arms) and lower extremities (legs) due to strain from exercise or normal household duties [4] Relief of post-surgical pain and post traumatic acute pain b. When using Electrical Muscle Stimulation (EMS Mode), Cure Trio is intended: [1] To temporarily increase local blood circulation in healthy leg muscles [2] To stimulate healthy muscles in order to improve and facilitate muscle performance [3] To relax muscle spasm [4] To increase blood flow circulation [5] For prevention of retardation of disuse atrophy [6] For muscle re-education [7] For maintaining or increasing range of motion (Note: The document provides similar variations of this statement for models MC-310B through MC-310U, specifying TENS-only or TENS/EMS combinations for each).
Device Story
Cure Trio (Models MC-310A-U) are electrical stimulators for pain management (TENS) and neuromuscular stimulation (EMS). Devices deliver electrical pulses via electrodes to skin/muscles; intended for clinical or home use by patients or clinicians. TENS mode provides symptomatic relief for chronic, acute, or post-surgical pain and arthritis. EMS mode stimulates healthy muscles to improve performance, increase circulation, relax spasms, prevent atrophy, and assist in re-education/range of motion. Output intensity and parameters are controlled by the user to achieve therapeutic effect. Benefits include non-pharmacological pain management and muscle rehabilitation.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Transcutaneous electrical nerve stimulator and powered muscle stimulator. Device operates via electrical pulse generation. Models vary by TENS-only or TENS/EMS functionality. Connectivity and specific material standards not disclosed in the provided text.
Indications for Use
Indicated for patients requiring pain management (TENS) or neuromuscular stimulation (EMS). TENS indications: chronic intractable pain, arthritis pain, sore/aching muscles (shoulder, waist, back, extremities) due to strain, post-surgical/post-traumatic acute pain. EMS indications: increase local blood circulation in healthy leg muscles, improve/facilitate muscle performance, relax muscle spasms, increase blood flow, prevent/retard disuse atrophy, muscle re-education, maintain/increase range of motion.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".
April 19, 2024
Oriental Inspiration Limited Francis Ko Director Unit D (01), 14/F.Block 1.Tai Ping Industrial Centre. No. 57 Ting Kok Road, Tai Po, New Territories Hong Kong, Hong Kong China
Re: K233046
Trade/Device Name: Electrical Neuromuscular Stimulator, Cure Trio Models MC-310A, MC-310B, MC-310C, MC-310D,MC-310E, MC-310F, MC-310G MC-310H, MC-310I, MC-310O, MC-310P, MC-310R, MC-310R, MC-310S, MC-310T, MC-310U Regulation Number: 21 CFR 882.5890, 21 CFR 890.5850 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief, Powered muscle stimulator Regulatory Class: Class II Product Code: NUH,GZJ, IPF,NGX, NYN, GZI Dated: March 11, 2024 Received: March 20, 2024
Dear Francis Ko:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{1}------------------------------------------------
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
# Doe W. Kumsa -S
for Pamela Scott, MS Assistant Director DHT5B: Division of Neuromodulation and Rehabilitation Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K233046
#### Device Name
Cure TrioModel: MC-310A, Model: MC-310B, Model: MC-310D, Model: MC-310E, Model: MC-310F, Model: MC-310G.Model: MC-310H, Model: MC-310I, Model: MC-3100, Model: MC-3100, Model: MC-310R,Model: MC-310S, Model: MC-310T, Model: MC-310U
Indications for Use (Describe)
#### Cure Trio (Model: MC-310A), Rx
The Cure Trio MC-310A is electrical stimulator designed for pain management and neuro muscular stimulation.
a. When using Cure Trio to deliver Transcutaneous Electrical Nerve Stimulation (TENS Mode), it is intended to provide: [1] Symptomatic relief and management of chronic, intractable pain
- [2] Relief of pain associated with arthritis
- [3] Temporarily relieves pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities
- (arms) and lower extremities (legs) due to strain from exercise or normal household duties
- [4] Relief of post-surgical pain and post traumatic acute pain
- b. When using Electrical Muscle Stimulation (EMS Mode), Cure Trio is intended:
- [1] To temporarily increase local blood circulation in healthy leg muscles
- [2] To stimulate healthy muscles in order to improve and facilitate muscle performance
- [3] To relax muscle spasm
- [4] To increase blood flow circulation
- [5] For prevention of retardation of disuse atrophy
- [6] For muscle re-education
- [7] For maintaining or increasing range of motion
Cure TrioModel: MC-310B), Rx
The Cure Trio MC-310B is electrical stimulator designed for pain management and neuro muscular stimulation.
a. When using Cure Trio to deliver Transcutaneous Electrical Nerve Stimulation (TENS Mode), it is intended to provide:
- [1] Symptomatic relief and management of chronic, intractable pain
- [2] Relief of pain associated with arthritis
- [3] Temporarily relieves pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities
- (arms) and lower extremities (legs) due to strain from exercise or normal household duties
[4] Relief of post-surgical pain and post traumatic acute pain
- b. When using Electrical Muscle Stimulation (EMS Mode), Cure Trio is intended:
- [1] To temporarily increase local blood circulation in healthy leg muscles
- [2] To stimulate healthy muscles in order to improve and facilitate muscle performance
- [3] To relax muscle spasm
- [4] To increase blood flow circulation
- [5] For prevention of retardation of disuse atrophy
Cure Trio (Model: MC-310C), Rx and OTC
The Cure Trio MC-310C is electrical stimulator designed for pain management and neuro muscular stimulation.
a. When using Cure Trio to deliver Transcutaneous Electrical Nerve Stimulation (TENS Mode), it is intended to provide:
- [1] Symptomatic relief and management of chronic, intractable pain
[2] Relief of pain associated with arthritis
[3] Temporarily relieves pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities
(arms) and lower extremities (legs) due to strain from exercise or normal household duties
[4] Relief of post-surgical pain and post traumatic acute pain
b. When using Electrical Muscle Stimulation (EMS Mode), Cure Trio is intended:
[1] To temporarily increase local blood circulation in healthy leg muscles
{4}------------------------------------------------
- [2] To stimulate healthy muscles in order to improve and facilitate muscle performance
- [3] To relax muscle spasm
- [4] To increase blood flow circulation
- [5] For prevention of retardation of disuse atrophy
Cure TrioModel: MC-310D), Rx and OTC
The Cure Trio MC-310D is Transcutaneous Electrical Nerve Stimulation device(Only TENS) for pain management. It is intended to provide:
[1] Symptomatic relief and management of chronic, intractable pain
- [2] Relief of pain associated with arthritis
- [3] Temporarily relieves pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities
(arms) and lower extremities (legs) due to strain from exercise or normal household duties
- [4] Relief of post-surgical pain and post traumatic acute pain
Cure TrioModel: MC-310E, Rx and OTC
The Cure Trio MC-310E is Transcutaneous Electrical Nerve Stimulation device(Only TENS) for pain management. It is intended to provide:
- [1] Symptomatic relief and management of chronic, intractable pain
- [2] Relief of pain associated with arthritis
[3] Temporarily relieves pain associated with sore and aching muscles in the shoulder, waist, back, upper extremittes
(arms) and lower extremities (legs) due to strain from exercise or normal household duties
[4] Relief of post-surgical pain and post traumatic acute pain
Cure TrioModel: MC-310FRx and OTC
The Cure Trio MC-310F is electrical stimulator designed for pain management and neuro muscular stimulation.
a. When using Cure Trio to deliver Transcutaneous Electrical Nerve Stimulation (TENS Mode), it is intended to provide: [1] Symptomatic relief and management of chronic, intractable pain
[2]Relief of pain associated with arthritis
[3] Temporarily relieves pain associated with sore and aching muscles in the shoulder, waist, back, upper extremites
(arms) and lower extremities (legs) due to strain from exercise or normal household duties
[4] Relief of post-surgical pain and post traumatic acute pain
b. When using Electrical Muscle Stimulation (EMS Mode), Cure Trio is intended:
- [1] To temporarily increase local blood circulation in healthy leg muscles
- [2] To relax muscle spasm
- [3] To increase blood flow circulation
- [4] For prevention of retardation of disuse atrophy
Cure TrioModel: MC-310G) Rx and OTC
The Cure Trio MC-310G is electrical stimulator designed for pain management and neuro muscular stimulation.
a. When using Cure Trio to deliver Transcutaneous Electrical Nerve Stimulation (TENS Mode), it is intended to provide:
[1] Symptomatic relief and management of chronic, intractable pain
- [2] Relief of pain associated with arthritis
[3] Temporarily relieves pain associated with sore and aching muscles in the shoulder, waist, back, upper extremittes
(arms) and lower extremities (legs) due to strain from exercise or normal household duties
[4] Relief of post-surgical pain and post traumatic acute pain
- b. When using Electrical Muscle Stimulation (EMS Mode), Cure Trio is intended:
- [1] To temporarily increase local blood circulation in healthy leg muscles
- [2] For prevention of retardation of disuse atrophy
Cure TrioModel: MC-310HRx and OTC
The Cure Trio MC-310H is Transcutaneous Electrical Nerve Stimulation device(Only TENS) for pain management. It is intended to provide:
- [1] Symptomatic relief and management of chronic, intractable pain
- [2] Relief of pain associated with arthritis
[3] Temporarily relieves pain associated with sore and aching muscles in the shoulder, waist, back, upper extremittes
{5}------------------------------------------------
(arms) and lower extremities (legs) due to strain from exercise or normal household duties
Cure TrioModel: MC-310IRx and OTC
The Cure Trio MC-310I is electrical stimulator designed for pain management and neuro muscular stimulation.
a. When using Cure Trio to deliver Transcutaneous Electrical Nerve Stimulation (TENS Mode), it is intended to provide:
[1] Symptomatic relief and management of chronic, intractable pain
- [2] Relief of pain associated with arthritis
- [3] Temporarily relieves pain associated with sore and aching muscles in the shoulder, waist, back, upper extremittes
(arms) and lower extremities (legs) due to strain from exercise or normal household duties
- [4] Relief of post-surgical pain and post traumatic acute pain
- b. When using Electrical Muscle Stimulation (EMS Mode), Cure Trio is intended:
- [1] To temporarily increase local blood circulation in healthy leg muscles
- [2] To stimulate healthy muscles in order to improve and facilitate muscle performance
## Cure TrioModel: MC-310ORx and OTC
The Cure Trio MC-3100 is electrical stimulator designed for pain management and neuro muscular stimulation.
a. When using Cure Trio to deliver Transcutaneous Electrical Nerve Stimulation (TENS Mode), it is intended to provide:
- [1] Symptomatic relief and management of chronic, intractable pain
- [2] Relief of pain associated with arthritis
- [3] temporarily relieves pain associated with sore and aching muscles in the shoulder, wast, back, upper extremities
- (arms) and lower extremities (legs) due to strain from exercise or normal household duties
- [4] Relief of post-surgical pain and post traumatic acute pain
- b. When using Electrical Muscle Stimulation (EMS Mode), Cure Trio is intended:
- [1] To temporarily increase local blood circulation in healthy leg muscles
- [2] To stimulate healthy muscles in order to improve and facilitate muscle performance
- [3] To relax muscle spasm
- [4] To increase blood flow circulation
- [5] For prevention of retardation of disuse atrophy
- [6] For muscle re-education
- [7] For maintaining or increasing range of motion
Cure TrioModel: MC-310PRx and OTC
The Cure Trio MC-310P is Transcutaneous Electrical Nerve Stimulation device(Only TENS) for pain management. It is intended to provide:
- [1] Symptomatic relief and management of chronic, intractable pain
- [2] Relief of pain associated with arthritis
- [3] temporarily relieves pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities
- (arms) and lower extremities (legs) due to strain from exercise or normal household duties
- [4] Relief of post-surgical pain and post traumatic acute pain
## Cure TrioModel: MC-310QRx and OTC
The Cure Trio MC-310Q is Transcutaneous Electrical Nerve Stimulation device(Only TENS) for pain management. It is intended to provide:
- [1] Symptomatic relief and management of chronic, intractable pain
- [2] Relief of pain associated with arthritis
- [3] temporarily relieves pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities (arms) and lower extremities (legs) due to strain from exercise or normal household duties
- [4] Relief of post-surgical pain and post traumatic acute pain
## Cure TrioModel: MC-310R
The Cure Trio MC-310R is electrical stimulator designed for pain management and neuro muscular stimulation.
a. When using Cure Trio to deliver Transcutaneous Electrical Nerve Stimulation (TENS Mode), it is intended to provide: [1] Symptomatic relief and management of chronic, intractable pain
{6}------------------------------------------------
[3] Temporarily relieves pain associated with sore and aching muscles in the shoulder, waist, back, upper extremntes
(arms) and lower extremities (legs) due to strain from exercise or normal household duties
[4] Relief of post-surgical pain and post traumatic acute pain
b. When using Electrical Muscle Stimulation (EMS Mode), Cure Trio is intended:
[1] To temporarily increase local blood circulation in healthy leg muscles
- [2] To relax muscle spasm
- [3] To increase blood flow circulation
- [4] For prevention of retardation of disuse atrophy
- [5] For muscle re-education
- [6] For maintaining or increasing range of motion
Cure TrioModel: MC-310SRx and OTC
The Cure Trio MC-310S is electrical stimulator designed for pain management and neuro muscular stimulation.
a. When using Cure Trio to deliver Transcutaneous Electrical Nerve Stimulation (TENS Mode), it is intended to provide:
- [1] Symptomatic relief and management of chronic, intractable pain
- [2] Relief of pain associated with arthritis
- [3] Temporarily relieves pain associated with sore and aching muscles in the shoulder, waist, back, upper extremites
- (arms) and lower extremities (legs) due to strain from exercise or normal household duties
[4] Relief of post-surgical pain and post traumatic acute pain
- b. When using Electrical Muscle Stimulation (EMS Mode), Cure Trio is intended:
- [1] To temporarily increase local blood circulation in healthy leg muscles
- [2] To relax muscle spasm
- [3] To increase blood flow circulation
- [4] For prevention of retardation of disuse atrophy
- [5] For muscle re-education
- [6] For maintaining or increasing range of motion
Cure TrioModel: MC-310TRx and OTC
The Cure Trio MC-310T is Transcutaneous Electrical Nerve Stimulation device(Only TENS) for pain management. It is intended to provide:
[1] Symptomatic relief and management of chronic, intractable pain
[2] Relief of pain associated with arthritis
[3] temporarily relieves pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities (arms) and lower extremities (legs) due to strain from exercise or normal household duties
Cure TrioModel: MC-310URx and OTC
The Cure Trio MC-310U is electrical stimulator designed for pain management and neuro muscular stimulation.
a. When using Cure Trio to deliver Transcutaneous Electrical Nerve Stimulation (TENS Mode), it is intended to provide:
- [1] Symptomatic relief and management of chronic, intractable pain
- [2] Relief of pain associated with arthritis
[3] Temporarily relieves pain associated with sore and aching muscles in the shoulder, waist, back, upper extremittes (arms) and lower extremities (legs) due to strain from exercise or normal household duties
[4] Relief of post-surgical pain and post traumatic acute pain
b. When using Electrical Muscle Stimulation (EMS Mode), Cure Trio is intended:
[1] To temporarily increase local blood circulation in healthy leg muscles
[2] To stimulate healthy muscles in order to improve and facilitate muscle performance
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
|X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{7}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.