Smart Scope® CX is intended for gynecological examination. It provides a portable means of magnified visualization of the tissues of the vagina, cervix, and external genitalia to aid in diagnosing abnormalities and selecting areas for biopsy. The Net4Medix® application is intended to provide documentation of the field of view of the scope. The Smart Scope® CX camera probe is intended to be inserted into the a speculum by trained medical personnel in hospitals, clinics, and private offices, and is not intended for home use.
Device Story
Handheld, reusable transvaginal digital camera probe; connects via USB to dedicated tablet running Net4Medix® software. Probe features 5MP CMOS sensor, 10X magnification, and integrated white/green LEDs for illumination. Operator (physician/clinician) inserts probe into speculum; captures live/still images via tablet interface. Software manages patient records, visit history, and image storage. Device provides magnified visualization to assist in identifying abnormalities and biopsy site selection. Benefits include portable, high-definition visualization for gynecological exams. Security managed via Enterprise-Device-Management (EDM) software, restricting tablet to Net4Medix® application only.
Clinical Evidence
Bench testing only. Performance evaluated against ISO 8600-3 (field of view), ISO 8600-5 (optical resolution), IEC 62471 (photobiological safety), and electrical safety/EMC standards (IEC 60601-1, IEC 60601-1-2). No clinical data provided.
Technological Characteristics
Handheld digital colposcope; ABS plastic housing, glass camera protector, thermoplastic polyurethane buttons. 5MP CMOS sensor; 4 white/4 green LEDs. USB 2.0 connectivity to tablet. Powered via USB from tablet. Non-sterile, reusable. Software: Net4Medix® application on Android 6+ tablet. Cybersecurity: EDM software restricts tablet to dedicated application.
Indications for Use
Indicated for gynecological examination in hospitals, clinics, and private offices by trained medical personnel. Used for magnified visualization of vagina, cervix, and external genitalia to aid in diagnosing abnormalities and selecting biopsy sites. Not for home use.
Regulatory Classification
Identification
A colposcope is a device designed to permit direct viewing of the tissues of the vagina and cervix by a telescopic system located outside the vagina. It is used to diagnose abnormalities and select areas for biopsy. This generic type of device may include a light source, cables, and component parts.
Special Controls
*Classification.* Class II (special controls). The device, when it is a standard colposcope (or colpomicroscope) that uses only a white light source, does not use filters other than a green filter, does not include image analysis software, and is not smartphone-based, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
June 14, 2024
Periwinkle Technologies Pvt. Ltd % Ankur Naik Managing Director IZiel Healthcare Pentagon P1, Office No. 601 and 604, Magarpatta City Pune. Mab 411028 INDIA
Re: K233043
Trade/Device Name: Smart Scope® (CX) Regulation Number: 21 CFR 884.1630 Regulation Name: Colposcope Regulatory Class: Class II Product Code: HEX Dated: May 9, 2024 Received: May 10, 2024
Dear Ankur Naik:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at
https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov
{2}------------------------------------------------
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-devicereporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).
{3}------------------------------------------------
Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumereducation-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Reginald K. Avery -S
for Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K233043
Device Name Smart Scope® CX
#### Indications for Use (Describe)
Smart Scope® CX is intended for gynecological examination. It provides a portable means of magnified visualization of the tissues of the vagina, cervix, and external genitalia to aid in diagnosing abnormalities and selecting areas for biopsy. The Net4Medix® application is intended to provide documentation of the field of view of the scope. The Smart Scope® CX camera probe is intended to be inserted into the a speculum by trained medical personnel in hospitals, clinics, and private offices, and is not intended for home use.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
#### 510(k) Summary
510(k) summary of substantial equivalence for Smart Scope® CX is provided in accordance with 21 CFR 807.92.
| Date Prepared: | 14 June 2024 |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter (Owner): | Veena Moktali<br>Founder & CEO<br>Periwinkle Technologies Pvt Ltd.<br>B1, Samrat Ashok CHS,<br>S. No. 88/2, Veerbhadra Nagar, Baner, Pune -<br>411045, Maharashtra, India.<br>P: +91 9021147173<br>Email: vrmoktali@periwinkletech.com |
| 510(k) Contact Person: | Ankur Naik<br>Managing Director<br>IZiel Healthcare<br>14, Hadapsar Industrial Estate,<br>Hadapsar, Pune – 411013, India.<br>P: +91 72762 2555 M: +91 7069553814<br>Email: ankur.naik@izielhealthcare.com |
| Device Trade Name: | Smart Scope CX |
| Device Common Name | Colposcope |
| Classification Number: | 21 CFR 884.1630 |
| Classification Name: | Colposcope |
| Review Panel: | Obstetrics/Gynecology |
| Device Class: | Class II |
| Product Code: | HEX |
| Predicate Device: | Pocket Colposcope System (K181034) |
| | The predicate device has not been subjected to a<br>design-related recall. |
### Device Description
Smart Scope® CX is a handheld, reusable transvaginal digital examination camera designed for use in a hospital or clinical setting. It is intended for close examination and magnified visualization of the external genitalia, vagina, and cervix. The Smart
{6}------------------------------------------------
Scope® CX was designed, developed, and manufactured at Periwinkle Technologies Pvt. Ltd. The model number is indicated below:
| Model<br>Number | Model Name | Model Description |
|-----------------|-----------------|----------------------------------------------|
| CX | Smart Scope® CX | Digital Cervical Examination (Trans-vaginal) |
The Smart Scope® CX packaging includes:
- Smart Scope® CX Probe ●
- . Lenovo Smart Tab M8 Tablet with Net4Medix® Software
- . Stand for tablet.
- . Tablet charger.
The Smart Scope® CX camera probe is equipped with an integrated green and white LED light, which serves to illuminate the object under observation. Additionally, it features a 10X magnification camera, facilitating the capture of color images. Users can switch between the white and green light illumination and zoom in or out to capture images at different working distances by pressing respective buttons in the smart scope probe. By establishing a USB connection between the Smart Scope and a tablet, users can access live images and capture necessary stills using the Net4Medix® application, installed on the tablet. These captured images are then stored directly on the tablet. The Net4Medix® software further enables the creation and management of patient records, visit histories, and seamless integration with the Smart Scope® CX for functions such as image streaming, capture, storage, and assignment to the respective patient records.
The Smart Scope® CX is designed and intended to be used only with speculum and it is not intended to come into direct contact with the body. Smart Scope® CX is not a substitute for histopathology and is not a diagnostic test.
The Smart Scope® CX should be connected to the tablet via USB connections. The Smart Scope® camera probe is powered from the tablet through a connected USB interface. During the tablet's charging process, the device cannot be utilized. The Smart Scope® CX is intended to be used only with a Smart tablet that is customized and supplied with the device.
To enhance security and manageability, Periwinkle Technologies Pvt Ltd has preloaded the tablet with Enterprise-Device-Management (EDM) software. This software serves as a tool for the secure distribution of the Net4Medix® application and facilitates policy control for data security. This security framework restricts the usage of any software other than Net4Medix®, ensuring that the tablet is dedicated exclusively to this critical application. Consequently, users are unable to connect to external networks or access other internet facilities.
{7}------------------------------------------------
#### Specification:
| Parameters | Specification |
|--------------------------|----------------------------------------------------------------------------------------------|
| Operating System | Android 6+ operating system |
| Dimensions | 211 x 124 x 8.2 mm or a similar |
| Display Screen | Touchscreen with 7 to 8-inch display |
| | Min. Resolution 1280X800, Gamut 60% NTSC,<br>84.6% sRGB, Brightness 350+ Nits, Contrast 552+ |
| Battery capacity | 4000+ mAh |
| Minimum storage capacity | 16 GB |
| Minimum RAM. | 2 GB |
## Smart Scope Features:
The Smart scope device is used like other colposcopes, and has the following features:
- Camera & Lens: 0° lens, 5 Mega Pixel CMOS sensor ●
- Light Source: Four (4) white and four (4) green LEDs arranged in a circular at . the probe tip.
- Probe (Nozzle): 74 mm in length, 22.4 mm of Outer Diameter. .
- Probe (Handle): 134.3 mm in length. ●
- Zoom, light controls on the probe handle. ●
- Image capture and pdf report generated through the Net4Medix® application. ●
- USB 2.0 cable attached to scope handle to connect to a tablet. .
- Packaged non-sterile and designed for reuse after cleaning and Disinfection. ●
- . High-definition output.
#### Intended Use / Indications for Use
Smart Scope® CX is intended for gynecological examination. It provides a portable means of magnified visualization of the tissues of the vagina, cervix, and external genitalia to aid in diagnosing abnormalities and selecting areas for biopsy. The Net4Medix® application is intended to provide documentation of the image in the field of view of the scope. The Smart Scope® CX camera probe is intended to be inserted into the vagina via a speculum by trained medical personnel in hospitals, clinics, and private offices, and is not intended for home use.
{8}------------------------------------------------
#### Substantial Equivalence Discussion
The Pocket Colposcope (K181034) has been selected as the primary predicate device for the substantial equivalence discussion regarding the Smart Scope® CX. The details regarding the substantial equivalence between the subject device and the predicate devices are explained below:
{9}------------------------------------------------
## Comparison to predicate devices:
| Comparable<br>Properties | Subject Device | Predicate Device (K181034) | Comparison<br>Results | | |
|-----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| Product Name | Smart Scope® CX | Pocket Colposcope | Not Applicable | | |
| Manufacturer | Periwinkle Technologies Pvt Ltd | Hadleigh Health Technologies,<br>LLC | Not Applicable | | |
| Regulation<br>Number | 21 CFR 884.1630 | 21 CFR 884.1630 | Identical | | |
| Product Code | HEX | HEX | Identical | | |
| Product Class | Class II | Class II | Identical | | |
| Indication for Use | Smart Scope® CX is intended for<br>gynecological examination. It<br>provides a portable means of<br>magnified visualization of the<br>tissues of the vagina, cervix, and<br>external genitalia to aid in<br>diagnosing abnormalities and<br>selecting areas for biopsy. The<br>Net4Medix® application is intended<br>to provide documentation of the<br>image in the field of view of the<br>scope. The Smart Scope® CX<br>camera probe is intended to be<br>inserted into the vagina via a<br>speculum by trained medical<br>personnel in hospitals, clinics, and<br>private offices, and is not intended<br>for home use. | The Pocket Colposcope is a digital<br>video colposcope intended for<br>qynecological examination. It<br>provides a portable means of<br>magnified visualization of the<br>tissues of the vagina, cervix, and<br>external genitalia in order to aid in<br>diagnosing abnormalities and<br>selecting areas for biopsy. The<br>image system is intended to<br>provide documentation of the<br>image in the field of view of the<br>colposcope. The Pocket<br>Colposcope System is intended to<br>be inserted into the vagina via a<br>speculum by trained medical<br>personnel in hospitals, clinics, and<br>private offices, and is not intended<br>for home use. | Similar | | |
| Multiple Use | Yes | Yes | Identical | | |
| Prescription (Rx<br>only) | Yes | Yes | Identical | | |
| Comparable<br>Properties | Subject Device | Predicate Device (K181034) | Comparison<br>Results | | |
| Anatomical Site | External genitalia, vagina, and<br>cervix | Cervix, vagina, and external<br>genitalia | Identical | | |
| Intended User | Trained medical professionals | Trained medical professionals | Identical | | |
| Use<br>Environment | Hospitals, clinics, and private<br>offices | Hospitals, clinics, and private<br>offices | Identical | | |
| Device Design | Image: Subject Device | Image: Predicate Device | Not Applicable | | |
| Standard<br>Configuration | Digital Camera connected to Tablet<br>via USB cable and used without<br>stand. | Digital CMOS camera used without<br>stand. | Identical | | |
| External Power<br>Source | Voltage: 5V<br>Frequency: DC<br>Input power: USB 5V from a tablet<br>device | Voltage: 5V<br>Frequency: DC<br>Input power: USB 5V | Identical | | |
| Materials | Housing: Acrylonitrile-butadiene-<br>styrene (ABS) Plastic<br>Buttons: Thermoplastic<br>polyurethane<br>Camera protector: Glass | Housing: ABS Plastic<br>Lens: PMMA | Different<br>The housing of the<br>subject and<br>predicate device is<br>made up of ABS<br>plastic. The other<br>subject device<br>components are<br>made of other<br>plastics or glass<br>that differ from the<br>predicate device. | | |
| Comparable<br>Properties | Subject Device | Predicate Device (K181034) | Comparison<br>Results | | |
| Fundamental<br>Scientific<br>Technology | The camera has optics that relay<br>optical information (live streaming<br>of images) to a smart tablet<br>(visualization/ annotation module). | The camera has optics that relay<br>optical video information to a PC or<br>laptop (visualization/ annotation<br>module). | However, the<br>differences do not<br>raise different<br>questions of safety<br>and effectiveness.<br>Different<br>The Net4Medix<br>application is<br>installed on the<br>Smart Tablet and<br>aids in visualization<br>and annotation.<br>The predicate<br>device required a<br>PC or laptop to<br>relay and visualize<br>the optical<br>information. This<br>difference does not<br>raise different<br>questions of safety | | |
| | Device Specification: Light | | and effectiveness. | | |
| Light Module | LEDs in circular form. | Single loop group LED light | Identical | | |
| Light Source | White and green LED light | White and green LED light | Identical | | |
| Light Source<br>Lifetime | ≥ 10,000 hrs | ≥ 10,000 hrs | Identical | | |
| Maximum<br>Illumination | White-450 lux at working distance<br>of 35-40mm | 20,000 lux at a working distance of<br>5mm. | Different | | |
| | Green-550 lux at working distance<br>of 35-40mm | 462 lux at a working distance of<br>50mm. | The illumination of<br>the subject device<br>does not exceed | | |
| Comparable<br>Properties | Subject Device | Predicate Device (K181034) | Comparison<br>Results | | |
| Illumination<br>Range/Beam<br>Diameter | 75 mm at a working distance of 40<br>mm | 74.5 mm at working distance of 50<br>mm | the illumination for<br>the predicate<br>device. These<br>differences do not<br>raise different<br>questions of safety<br>and effectiveness.<br>Different<br>The subject and<br>predicate device<br>have a similar<br>illumination range<br>at different working<br>distances. These<br>differences do not<br>raise different<br>questions of safety<br>and effectiveness. | | |
| | Device Specification: Camera | | | | |
| Magnification | 29 or 46X on tablet display<br>depending on orientation | 3X-52X | Different<br>The magnification<br>of the subject<br>device is within the<br>range specified by<br>predicate device.<br>This does not raise<br>any different<br>questions of safety<br>and effectiveness. | | |
| Comparable<br>Properties | Subject Device | Predicate Device (K181034) | Comparison<br>Results | | |
| System Resolution | ≥ 1944 TVL…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.