AI/ML, Software as a Medical Device, PCCP, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K233030 · Mar 1, 2024
BoneMRI
Mriguidance B.V.
Retrospective clinical imaging data (MRI and CT scans); Previously conducted clinical investigations
Retrospective clinical data from 76 pelvic and 117 spine patients were used to validate the quantitative accuracy of the BoneMRI software by comparing its output against co-registered CT scans.
Patients requiring pelvic or spine imaging (including adolescents and adults); Sample Size: 193 total (76 pelvic, 117 spine); Number of Sites: Multiple medical sites
Co-registered CT scans
Cortical delineation error (mm), Mean deviation in tissue/bone (HU), Correlation coefficient in bone
BoneMRI is an image processing software that can be used for image enhancement in MRI images. It can be used to visualize the bone structures in MRI images with increased contrast with respect to the surrounding soft tissue. It is to be used in the pelvic region, which includes the bony anatomy of the sacrum, hip bones and femoral heads; and the spine, which includes the bony anatomy of the cervical, thoracic, lumbar, and S1 vertebrae. BoneMRI is indicated for use in patients 12 years and older. BoneMRI is not to be used for diagnosis or monitoring of (primary or metastatic) tumors. BoneMRI images are not intended to replace CT images in general but can be used to visualize 3D bone morphology, tissue radiodensity and tissue radiodensity contrast.
Device Story
Standalone image processing software; analyzes 3D gradient echo MRI scans to construct 3D tomographic radiodensity contrast images (BoneMRI). Used by radiologists/orthopedic surgeons in clinics/hospitals. Operates on hospital networks (on-premise or managed cloud). Inputs DICOM MRI files; outputs enhanced DICOM images for viewing on standard PACS workstations. Enhances bone-to-soft-tissue contrast; assists in visualizing 3D bone morphology and radiodensity. Benefits clinicians by providing CT-like bone visualization from MRI, potentially reducing need for additional CT scans.
Clinical Evidence
Bench testing only. Quantitative voxel-by-voxel validation performed on 76 pelvic and 117 spine patient datasets comparing BoneMRI to co-registered CT. Primary endpoints: cortical delineation error (<1.0 mm), mean radiodensity deviation (<25 HU overall, <55 HU bone), and HU correlation coefficient (>0.75). Results confirmed clinically acceptable accuracy across diverse subgroups (age, BMI, scanner vendor, field strength).
Technological Characteristics
Standalone software package; Linux-based. Inputs/outputs DICOM MRI files. Uses convolutional neural network (CNN) for image enhancement via cascade of filter banks, thresholding, and scaling. Assigns Hounsfield Unit (HU) values to volume elements based on intensity and context. Deployed on-premise or via managed cloud; requires GPU hardware.
Indications for Use
Indicated for patients 12 years and older requiring visualization of bone structures in the pelvic region (sacrum, hip bones, femoral heads) and spine (cervical, thoracic, lumbar, S1 vertebrae) via MRI image enhancement. Contraindicated for diagnosis or monitoring of primary or metastatic tumors.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 1, 2024
MRIguidance B.V % Sujith Shetty Executive Vice President Maxis Medical 3031 Tisch Way Suite 1010 San Jose, California 95128
Re: K233030
Trade/Device Name: BoneMRI Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: OIH Dated: January 17, 2024 Received: January 22, 2024
Dear Sujith Shetty:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
FDA's substantial equivalence determination also included the review and clearance of your Predetermined Change Control Plan (PCCP). Under section 515C(b)(1) of the Act, a new premarket notification is not required for a change to a device cleared under section 510(k) of the Act, if such change is consistent with an established PCCP granted pursuant to section 515C(b)(2) of the Act. Under 21 CFR 807.81(a)(3), a new premarket notification is required if there is a major change or modification in the intended use of a device. or if there is a change or modification in a device that could significantly affect the safety or effectiveness of the device, e.g., a significant change or modification in design, material, chemical composition, energy source, or manufacturing process. Accordingly, if deviations from the established PCCP result in a major change or modification in the intended use of the device, or result in a change or modification in the device that could significantly affect the safety or effectiveness of the a new premarket notification would be required consistent with section 515C(b)(1) of the Act and 21 CFR 807.81(a)(3). Failure to submit such a premarket submission would constitute adulteration and misbranding under sections 501(f)(1)(B) and 502(o) of the Act, respectively.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100. Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn
{3}------------------------------------------------
(https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
D.G.K.
Daniel M. Krainak, Ph.D. Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### Indications for Use
510(k) Number (if known) K233030
Device Name BoneMRI
#### Indications for Use (Describe)
BoneMRI is an image processing software that can be used for image enhancement in MRI images. It can be used to visualize the bone structures in MRI images with increased contrast with respect to the surrounding soft tissue. It is to be used in the pelvic region, which includes the bony anatomy of the sacrum, hip bones and the spine, which includes the bony anatomy of the cervical, thoracic, lumbar, and S1 vertebrae. BoneMRI is indicated for use in patients 12 years and older.
BoneMRI is not to be used for diagnosis or monitoring of (primary or metastatic) tumors. BoneMRI images are not intended to replace CT images in general but can be used to visualize 3D bone morphology, tissue radiodensity and tissue radiodensity contrast.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for MRI Guidance. The logo consists of a circle with a right-angle arrow inside it, followed by the text "MRI guidance". The circle and the "MRI" text are in a gradient of light blue and green, while the "guidance" text is in gray.
# 510(k) Summary
# 1. 510(k) Information
| 510(k) Number: | K233030 |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared: | February 26, 2024 |
| 510(k) Submitter: | MRIguidance B.V.<br>Maliesingel 23, 3581 BG<br>Utrecht, the Netherlands |
| 510(k) Submitter Contact Person: | David Sparks<br>Head of Regulatory Affairs<br>MRIguidance B.V.<br>Email: david.sparks@mriguidance.com<br>Tel: +31 681741711 |
| Correspondent Contact Person: | Dr. Sujith Shetty<br>EVP Maxis Medical<br>3031 Tische Way, Suite 1010<br>San Jose, CA 95128 USA<br>Email: sjshetty@maxismedical.com |
## 2. Device Information
| Device Trade Name: | BoneMRI |
|-----------------------------|---------------------------------------------------------------------|
| Device Common Name: | MRI image enhancement software |
| Device Classification Name: | Medical image management and processing system (21<br>CFR 892.2050) |
| Device Classification: | Class II |
| Product Code: | QIH |
# 3. Predicate Device
| Device Trade Name: | BoneMRI |
|-----------------------------|---------------------------------------------------------------------|
| Manufacturer: | MRIguidance B.V. |
| Device 510(k) Clearance: | K230197 |
| Device Classification Name: | Medical image management and processing system (21<br>CFR 892.2050) |
| Device Classification: | Class II |
| Product Code: | QIH |
{6}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for MRI Guidance. The logo consists of a circle with a right angle symbol inside of it on the left, and the text "MRI guidance" on the right. The word "MRI" is in a blue color, and the word "guidance" is in a gray color.
# 4. Device Description
The BoneMRI application is a standalone image processing software application that analyses 3D gradient echo MRI scans acquired with a dedicated MRI scan protocol. From the analysis of the gradient echo MRI scan, 3D tomographic radiodensity contrast images, called BoneMRI images, are constructed.
The BoneMRI images can be used to visualize the bone structures in MR images with enhanced contrast with respect to the surrounding soft tissue. The application is designed to be used by imaging experts, such as radiologists or orthopedic surgeons, typically in a physician's office.
The BoneMRI application is a server application running on the clinic or hospital networks. It is available as fully on-premise software with specific GPU hardware requirements, or partly running as a managed cloud service, for which the environment in which the managed modules run is controlled by MRIguidance. The on-premise software is fully controlled by the clinic or hospital, and as such, no protected health information (PHI) will leave the clinic or hospital network. All data sent to the managed cloud server will be de-identified before it leaves the clinic or hospital network, and as such, the managed cloud service will not receive PHI.
Within the hospital network, the application communicates with a DICOM compatible imaging archive (e.g., a PACS) to receive input MRI and to return BoneMRI images. Reading of the resulting BoneMRI images is performed using regular DICOM compatible medical image viewing software.
The BoneMRI application uses an algorithm to detect bone images from MRIs obtained using a specific gradient echo acquisition sequence. The algorithm training sets included images from multiple clinical sites, multiple anatomies, and multiple scanners to ensure that the trained algorithm was robust with respect to the approved indications for use. None of the data used in the training dataset was used subsequently in the validation dataset.
# 5. Indications for Use
BoneMRI is an image processing software that can be used for image enhancement in MRI images. It can be used to visualize the bone structures in MRI images with increased contrast with respect to the surrounding soft tissue. It is to be used in the pelvic region, which includes the bony anatomy of the sacrum, hip bones and femoral heads; and the spine, which includes the bony anatomy of the cervical, thoracic, lumbar, and S1 vertebrae. BoneMRI is indicated for use in patients 12 years and older.
BoneMRI is not to be used for diagnosis or monitoring of (primary or metastatic) tumors. BoneMRI images are not intended to replace CT images in general but can be used to visualize 3D bone morphology, tissue radiodensity and tissue radiodensity contrast.
## 6. Comparison of Technological Characteristics with the Predicate Device
A comparison of the intended use, indication for use, and technological characteristics of the subject BoneMRI application to the predicate device (BoneMRI v1.6, K230197) is presented below. We have included the attributes suggested in the July 2018 Guidance "The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications" for this comparison.
{7}------------------------------------------------
Image /page/6/Picture/1 description: The image contains the logo for MRI Guidance. The logo consists of a circular icon with a stylized arrow pointing upwards and to the right, colored in a gradient of light blue and green. To the right of the icon, the text "MRI" is displayed in a similar gradient of light blue and green, while the word "guidance" is written in gray.
#### Table 1 Predicate device comparison
| | Predicate Device<br>(BoneMRI K230197) | Subject Device<br>(BoneMRI) | Comment |
|-------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Intended Use | BoneMRI is an image<br>processing software that can<br>be used for image<br>enhancement in MRI images.<br>It can be used to visualize the<br>bone structures in MRI<br>images with enhanced<br>contrast with respect to the<br>surrounding soft tissue. | BoneMRI is an image<br>processing software that can<br>be used for image<br>enhancement in MRI images.<br>It can be used to visualize the<br>bone structures in MRI<br>images with enhanced<br>contrast with respect to the<br>surrounding soft tissue. | The same |
| 21CFR Section | 829.2050 | 829.2050 | The same |
| Product Code | QIH | QIH | The same |
| Target<br>Population | Adults | Adolescents and Adults | Different |
| Indications for<br>Use | BoneMRI is an image<br>processing software that can<br>be used for image<br>enhancement in MRI images.<br>It can be used to visualize the<br>bone structures in MRI<br>images with enhanced<br>contrast with respect to the<br>surrounding soft tissue. It is<br>to be used in the pelvic<br>region, which includes the<br>bony anatomy of the sacrum,<br>hip bones and femoral heads;<br>and the lumbar spine region,<br>which includes the bony<br>anatomy of the vertebrae<br>from L3 to S1. BoneMRI is<br>not to be used for diagnosis<br>or monitoring of (primary or<br>metastatic) tumors.<br>Warning: BoneMRI images<br>are not intended to replace<br>CT images. | BoneMRI is an image<br>processing software that can<br>be used for image<br>enhancement in MRI images.<br>It can be used to visualize the<br>bone structures in MRI<br>images with increased<br>contrast with respect to the<br>surrounding soft tissue. It is<br>to be used in the pelvic<br>region, which includes the<br>bony anatomy of the sacrum,<br>hip bones and femoral heads;<br>and the spine, which includes<br>the bony anatomy of the<br>cervical, thoracic, lumbar, and<br>S1 vertebrae. BoneMRI is<br>indicated for use in patients<br>12 years and older.<br>BoneMRI is not to be used<br>for diagnosis or monitoring of<br>(primary or metastatic)<br>tumors. BoneMRI images are<br>not intended to replace CT<br>images in general but can be<br>used to visualize 3D bone<br>morphology, tissue<br>radiodensity and tissue<br>radiodensity contrast. | Similar |
| Technological Characteristics | | | |
| Device Nature | Software package | Software package | The same |
| Operating System | Linux | Linux | The same |
| Data Input | MRI images in DICOM format | MRI images in DICOM format | The same |
| Data Output | MRI images in DICOM format | MRI images in DICOM format | The same |
| Processing Algorithms | MRIguidance software<br>implements an image<br>enhancement algorithm using<br>convolutional neural network.<br>Original images are enhanced<br>by running them through a<br>cascade of filter banks, where<br>thresholding and scaling<br>operations are applied.<br>Separate neural network-<br>based filters are obtained to<br>assign a Hounsfield Unit (HU)<br>value to a single volume<br>element, based on intensity<br>and contextual information.<br>The parameters of the model<br>were obtained through an<br>algorithm development<br>pipeline. | MRIguidance software<br>implements an image<br>enhancement algorithm using<br>convolutional neural network.<br>Original images are enhanced<br>by running them through a<br>cascade of filter banks, where<br>thresholding and scaling<br>operations are applied.<br>Separate neural network-<br>based filters are obtained to<br>assign a Hounsfield Unit (HU)<br>value to a single volume<br>element, based on intensity<br>and contextual information.<br>The parameters of the model<br>were obtained through an<br>algorithm development<br>pipeline. | The same |
| User Interface | None - enhanced images are<br>viewed on existing PACS<br>workstations. | None – enhanced images are<br>viewed on existing PACS<br>workstations. | The same |
| Workflow | The software operates on<br>DICOM files on the file<br>system, enhances the images,<br>and stores the enhanced<br>images on the file system.<br>The receipt of original<br>DICOM image files and<br>delivery of enhanced images<br>as DICOM files depends on<br>other software systems.<br>Enhanced images co-exist<br>with the original images. | The software operates on<br>DICOM files on the file<br>system, enhances the images,<br>and stores the enhanced<br>images on the file system.<br>The receipt of original<br>DICOM image files and<br>delivery of enhanced images<br>as DICOM files depends on<br>other software systems.<br>Enhanced images co-exist<br>with the original images. | The same |
{8}------------------------------------------------
Image /page/7/Picture/1 description: The image is a logo for MRI Guidance. The logo consists of a circle with a right angle symbol inside of it, followed by the text "MRI guidance". The text "MRI" is in a larger font than the word "guidance". The colors of the logo are a gradient of light blue and green.
## 7. Summary of Changes
The changes to the BoneMRI application from the predicate device (BoneMRI v1.6, K230197) to the subject device are detailed in the table below.
{9}------------------------------------------------
Image /page/8/Picture/1 description: The image contains the logo for MRI Guidance. The logo consists of a green circle with a white arrow pointing upwards and to the right. To the right of the circle, the text "MRI" is written in a blue gradient, and below that, the text "guidance" is written in gray.
#### Table 2 Summary of changes for the BoneMRI application
| Change | Change Description |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Software architecture | Refactor of the application workflow; introducing multi-tenancy capabilities; and improving scalability of cloud deployment. |
| Validation strategy | Revision of the validation strategy to support the validation of a multi-vendor, multi-field strength algorithm. |
| Intended patient population | The intended patient population of BoneMRI is extended to include adolescents. The algorithm for the application has not been changed. |
| Predetermined Change Control Plan | Addition of a Predetermined Change Control Plan to support an iterative development approach for the machine learning models in the BoneMRI application. |
| Algorithm for the Spine region | The algorithm for the Spine region has been re-trained. With additional data for training and testing, the anatomical region of the algorithm was extended to include the Cervical Spine and the Thoracic Spine in addition to the Lumbar Spine. |
| Algorithm for the Pelvic region | No changes are made to the algorithm for the Pelvic region. Updates have been made to improve statistical testing of the algorithm and to test the algorithm on additional subgroups. |
### 8. Predetermined Change Control Plan
The BoneMRI application uses an algorithm derived from machine learning (ML) to detect bone images from MRIs obtained using a specific gradient echo acquisition sequence. The algorithm training sets included images from multiple clinical sites, multiple anatomies, and multiple scanners to ensure that the trained algorithm was robust with respect to the approved indications for use. MRIguidance will make future algorithm improvements under a Predetermined Change Control Plan (PCCP). In that plan, a protocol is provided to mitigate the risks of the algorithm changes leading to changes in the device's technical specifications or negatively affecting performance specifications directly associated with the indications for use of the device. Changes made under this PCCP are detailed in the table below. In accordance with the PCCP, all algorithm modifications will be trained, tuned, and locked prior to release of the application.
| Modification | Rationale |
|------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1. Re-training to improve ML<br>model performance with<br>additional training data | Re-training of the ML model with additional data to increase the<br>safety and performance of the device in any of the following<br>categories:<br>Increased accuracy; Increased performance for challenging cases such as rare<br>pathologies or artifacts; Increased robustness and generalization of the model. |
| 2. Validation of additional<br>scanner support | Validation of the ML model (either with or without additional re-<br>training of the ML model) in order to validate an additional MRI<br>vendor or field strength. |
Table 3 Summary of changes under a Predetermined Change Control Plan
{10}------------------------------------------------
Image /page/9/Picture/1 description: The image contains the logo for MRI Guidance. The logo consists of a circle with an arrow pointing to the upper right, and the text "MRI guidance" to the right of the circle. The word "MRI" is in a larger font size and a gradient color, while the word "guidance" is in a smaller font size and a gray color.
## 9. Performance Data
The following performance testing has been performed on BoneMRI:
- 1. Software verification and validation testing
- Studies that utilized retrospective clinical data to demonstrate the software enhanced imaging 2. quality in MR images via an enhancement of bone validated with CT.
### Software Verification and Validation Testing
Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" dated June 14, 2023.
### Performance Validation
A quantitative voxel-by-voxel validation of BoneMRI was performed on imaging data from 76 patients (pelvic region) and 117 patients (spine region), consisting of the BoneMRI and CT of the same patient in the same anatomical region, acquired using standard of care bone imaging protocols during previously conducted clinical investigations. Test data are acquired at different medical sites, departments or within different clinical studies than training data, and test data is unseen data that was not used in any way during developments. The validation was conducted by MRIguidance based on an algorithm to detect bone images from MRIs obtained using a specific gradient echo sequence. The demographics and performance data, including subgroup analysis of the patient population are described in the table below.
| Subgroup | N | Gender | Age | Data<br>origin | Cortical delineation<br>error (mm) | Mean deviation<br>in all tissue and bone<br>(HU) | Correlation coefficient<br>in bone |
|----------|-----|----------------|-------------|----------------|------------------------------------|--------------------------------------------------|------------------------------------|
| Pelvis | 76 | 75% M<br>25% F | $53 \pm 26$ | US,<br>EU | < 1 mm | < 25 HU<br>< 55 HU | > 0.75 |
| Spine | 117 | 49% M<br>51% F | $48 \pm 23$ | US,<br>EU | < 1 mm | < 25 HU<br>< 55 HU | > 0.75 |
| Pelvis | 57 | 93% M<br>7% F | $66 \pm 17$ | US,<br>EU | < 1 mm | < 25 HU<br>< 55 HU | > 0.75 |
| Spine | 94 | 51% M<br>49% F | $59 \pm 16$ | US,<br>EU | < 1 mm | < 25 HU<br>< 55 HU | > 0.75 |
| Pelvis | 19 | 21% M<br>79% F | $17 \pm 3$ | US | < 1 mm | < 25 HU<br>< 55 HU | > 0.75 |
| Spine | 23 | 22% M<br>78% F | $15 \pm 2$ | EU | < 1 mm | < 25 HU<br>< 55 HU | > 0.75 |
Table 4 Validation data demographics and performance testing was with a significance level of p <0.05.
{11}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the logo for MRI guidance. The logo consists of a circle with a right angle symbol inside, and the text "MRI guidance" to the right of the circle. The "MRI" part of the text is in a blue gradient, while the "guidance" part is in gray.
| US | Pelvis | 25 | 24% M<br>76% F | 18 ± 4 | US | < 1 mm | < 25 HU<br>< 55 HU | > 0.75 |
|-----|------------------|----|----------------|---------|-----------|--------|--------------------|--------|
| | Spine | 23 | 33% M<br>67% F | 64 ± 10 | US | < 1 mm | < 25 HU<br>< 55 HU | > 0.75 |
| EU | Pelvis | 51 | 100% M<br>0% F | 71 ± 80 | EU | < 1 mm | < 25 HU<br>< 55 HU | > 0.75 |
| | Spine | 91 | 46% M<br>54% F | 49 ± 23 | EU | < 1 mm | < 25 HU<br>< 55 HU | > 0.75 |
| BMI | Obese | 11 | 64% M<br>36% F | 51 ± 16 | US,<br>EU | < 1 mm | < 25 HU<br>< 55 HU | > 0.75 |
| | Over-<br>weight | 17 | 60% M<br>40% F | 52 ± 18 | US,<br>EU | < 1 mm | < 25 HU<br>< 55 HU | > 0.75 |
| | Healthy | 18 | 22% M<br>78% F | 29 ± 22 | US,<br>EU | < 1 mm | < 25 HU<br>< 55 HU | > 0.75 |
| | Under-<br>weight | 10 | 20% M<br>80% F | 17 ± 3 | EU | < 1 mm | < 25 HU<br>< 55 HU | > 0.75 |
The objective was to validate the quantitative accuracy of BoneMRI using rigorous, objective, and unbiased statistical tests comparing bone morphology, radiodensity, and radiodensity contrast in BoneMRI and CT images. The endpoints of the testing were the metrics that describe the accuracy of 3D bone morphology, radiodensity, and radiodensity contrast versus co-registered CT scans in terms of voxel-by-voxel HUs and standard deviations around these HU values. Subgroup analyses for different MRI vendors, field strengths, age ranges, geographical locations and BMI was performed as part of the testing. The results demonstrated clinically acceptable accuracy on each of these endpoints.
The data provided demonstrate that BoneMRI application can:
- accurately reconstruct the 3D bone morphology with a mean absolute cortical delineation ● error below 1.0 mm on average;
- accurately reconstructs the tissue radiodensity, with a mean deviation below 25 HU on average and a mean deviation below 55 HU specifically for bone;
- 0 accurately reconstructs the tissue radiodensity contrast, with a mean HU correlation coefficient above 0.75 specifically for bone.
The BoneMRI application demonstrates accurate bone morphology, radiodensity contrast to qualitatively and quantitatively assess the bony anatomy of the pelvic and spine regions.
#### 10. Substantial Equivalence Conclusion
The subject BoneMRI application has the same intended use and a similar indication as the identified predicate device, its predecessor (BoneMRI, K230197). The technological features of the BoneMRI application are the same as the identified predicate device. Therefore, we conclude that the BoneMRI application is substantially equivalent to the identified predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.