Disposable Hemostatic Clips

K232969 · Ningbo Xinwell Medical Technology Co., Ltd. · PKL · Jun 7, 2024 · Gastroenterology, Urology

Device Facts

Record IDK232969
Device NameDisposable Hemostatic Clips
ApplicantNingbo Xinwell Medical Technology Co., Ltd.
Product CodePKL · Gastroenterology, Urology
Decision DateJun 7, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Disposable Hemostatic Clips is indicated for Endoscopic clip placement within the Gastrointestinal tract in adult populations only via a straight or side viewing flexible endoscope for the purpose of : 1) Endoscopic marking; 2) Hemostasis for: A) Mucosal/sub-mucosal defects < 3cm. B) Bleeding ulcers. C) Polyps < 1.5cm in diameter. D) Diverticula in the colon. E) Arteries < 2 mm. F) Prophylactic clipping to reduce the risk of delayed bleeding post lesion resection. 3) As a supplementary method, closure of GI tract luminal perforations <20mm that can be treated conservatively 4) Anchoring to affix jejunal feeding tubes to the wall of the small bowel

Device Story

Disposable Hemostatic Clips consist of a delivery system (slide bar, rotation head, spring tube) and a stainless steel clip assembly. Operated by a physician via a flexible endoscope; the device allows for clip opening/closing (up to 5 times) for lesion repositioning before final deployment. Used for endoscopic marking, hemostasis of GI defects/ulcers/polyps/arteries, closure of small perforations, and anchoring feeding tubes. The device is provided sterile for single use. Clinical benefit includes effective management of GI bleeding and tissue defects through minimally invasive endoscopic intervention.

Clinical Evidence

No clinical data included. Substantial equivalence supported by bench testing, including appearance, rotation performance, release force, clamping strength, relocation, tensile strength, mechanical integrity, scope compatibility, and three-ring handle strength. Biocompatibility testing performed per ISO 10993-1.

Technological Characteristics

Materials: Stainless steel (SUS316L, SUS631, SUS316F, SUS304), ABS, TPE, LDPE. Dimensions: 1250-2700mm length; 9-16mm opening width. Principle: Mechanical clip deployment via delivery system. Connectivity: None. Sterilization: Ethylene Oxide (ISO 11135). Software: None.

Indications for Use

Indicated for adult patients requiring endoscopic clip placement in the GI tract via flexible endoscope for marking, hemostasis (mucosal/sub-mucosal defects <3cm, bleeding ulcers, polyps <1.5cm, colonic diverticula, arteries <2mm, post-resection prophylaxis), closure of GI perforations <20mm, or anchoring jejunal feeding tubes.

Regulatory Classification

Identification

A hemorrhoidal ligator is a device used to cut off the blood flow to hemorrhoidal tissue by means of a ligature or band placed around the hemorrhoid.

Special Controls

*Classification.* Class II (special controls). Except for a hemostatic metal clip intended for use in the gastrointestinal tract, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. June 7, 2024 Ningbo Xinwell Medical Technology Co., Ltd % Diana Hong General Manager Mid-Link Consulting Co, Ltd. P.O. Box 120-119 Shanghai. 200120 CHINA Re: K232969 > Trade/Device Name: Disposable Hemostatic Clips Regulation Number: 21 CFR 876.4400 Regulation Name: Hemorrhoidal Ligator Regulatory Class: Class II Product Code: PKL Dated: September 21, 2023 Received: May 9, 2024 Dear Diana Hong: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" {2}------------------------------------------------ (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). {3}------------------------------------------------ Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. # Sivakami Venkatachalam -S for Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use Submission Number (if known) K232969 Device Name ## Disposable Hemostatic Clips Indications for Use (Describe) The Disposable Hemostatic Clips is indicated for Endoscopic clip placement within the Gastrointestinal tract in adult populations only via a straight or side viewing flexible endoscope for the purpose of : 1) Endoscopic marking; 2) Hemostasis for: A) Mucosal/sub-mucosal defects < 3cm. B) Bleeding ulcers. C) Polyps < 1.5cm in diameter. D) Diverticula in the colon. E) Arteries < 2 mm. F) Prophylactic clipping to reduce the risk of delayed bleeding post lesion resection. 3) As a supplementary method, closure of GI tract luminal perforations <20mm that can be treated conservatively 4) Anchoring to affix jejunal feeding tubes to the wall of the small bowel Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {5}------------------------------------------------ # 510(k) Summary This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92. The assigned 510(k) Number: K232969 - 1. Date of Preparation: 06/03/2024 - 2. Sponsor Identification Ningbo Xinwell Medical Technology Co., Ltd. No.188 Binjiang Road, Cixi High-tech Industrial Development Zone, Cixi City, Zhejiang,P.R. China Establishment Registration Number: 3013526170 Contact Person: Lei Xi Position: RA manager Tel: +86-021-34693382 Email: xilei@xinhaigroup.com - 3. Designated Submission Correspondent Ms. Diana Hong (Primary Contact Person) Ms. Tingting Su (Alternative Contact Person) Mid-Link Consulting Co., Ltd P.O. Box 120-119, Shanghai, 200120, China Tel: +86-21-22815850, Fax: 360-925-3199 Email: info@mid-link.net - 4. Identification of Proposed Device Trade Name: Disposable Hemostatic Clips Common Name: Hemostasis Clipping Device Regulatory Information Classification Name: Hemorrhoidal Ligator; {6}------------------------------------------------ Classification: II, Product Code: PKL, Regulation Number: 21CFR 876.4400 Review Panel: Gastroenterology/Urology, #### Indications for Use: The Disposable Hemostatic Clips is indicated for Endoscopic clip placement within the Gastrointestinal tract in adult populations only via a straight or side viewing flexible endoscope for the purpose of: - 1) Endoscopic marking; - 2) Hemostasis for: - A) Mucosal/sub-mucosal defects < 3cm. - B) Bleeding ulcers. - C) Polyps < 1.5cm in diameter. - D) Diverticula in the colon. E) Arteries < 2 mm. F) Prophylactic clipping to reduce the risk of delayed bleeding post lesion resection. 3) As a supplementary method, closure of GI tract luminal perforations <20mm that can be treated conservatively 4) Anchoring to affix jejunal feeding tubes to the wall of the small bowel. - 5. Device Description The proposed device is consisted of a clip section. The delivery section. The delivery section include: slide bar, slip ring, protective sleeve, rotation head, coated spring tube, slip ring holder, traction assembly, connecting tube, decoupled piece, decoupled shrapnel, rotation sleeve. And the clip section include: clip, sliding block, shaft and rocker arm. The clip is made of stainless steel. The clip can be opened and closed more than five times prior to deployment, aiding in repositioning of the lesion site. The proposed device is available in single ring and three ring models, and each case is divided into rotatable and non-rotatable clip. The effective length of the proposed device includes 1250mm, 1650mm, 2350mm and the opening width is 9mm, 11mm. 13mm and 16mm. The proposed devices are provided in sterile and single use. - Identification of Predicate Device 6. Predicate Device 510(k) Number: K202333 Product Name: Lockado™ Repositionable Hemostasis Clip {7}------------------------------------------------ #### Non-Clinical Test Conclusion 7. Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards: - ♪ ISO 10993-3:2014 Biological evaluation of medical devices Part 3: Tests for genotoxicity, carcinogenicity and reproductive toxicity - > ISO 10993-5:2009 Biological evaluation of medical devices- Part 5: Test for in vitro cytotoxicity - > ISO 10993-6:2016 Biological evaluation of medical devices -- Part 6: Tests for local effects after implantation - > ISO 10993-7:2008 Biological evaluation of medical devices- Part 7: Ethylene Oxide Sterilization Residuals - > ISO 10993-10:2021 Biological evaluation of medical devices Part 10: Tests for skin sensitization - > ISO 10993-11: 2017 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity - > ISO 10993-23:2021 Biological evaluation of medical devices Part 23: Tests for irritation - > ASTM F1929-15 Standard test method for detecting seal leaks in porous medical packaging by dye penetration. - > ASTM F88/F88M-15 standard method for seal strength of flexible barrier materials - > ASTM F1886 / F1886M-16, Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection - > USP <85> Bacterial Endotoxins Test - A ASTM D4169-22 Standard Practice for Performance Testing of Shipping Containers and Systems #### Performance testing The following bench tests were performed on the Disposable Hemostatic Clips: - . Appearance - Rotation performance • - Release force - . Clamping strength - . Relocation - . Tensile strength - . Dimension - . Mechanical integrity of clip assembly - Scope compatibility/usability - . Endoscope damage - . Torque - Clip approach . - Three ring handle strength Above all tests were passed and demonstrated the result can meet the product requirements. {8}------------------------------------------------ ### Biocompatibility testing The biocompatibility evaluation for the proposed device was conducted in accordance with FDA Guidance, Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a risk management process" issued on September 4, 2020. The following tests were conducted for the clip component of Disposable Hemostatic Clips: - . Cytotoxicity - Irritation . - Skin Sensitization . - Acute Systemic Toxicity - . Sub-acute Systemic Toxicity - . Genotoxicity - . Pyrogenicity - Implantation Test The following tests were conducted for the delivery component of Disposable Hemostatic Clips: - . Cytotoxicity - Irritation ● - . Skin Sensitization - . Acute Systemic Toxicity - . Pyrogenicity #### Sterility, Shipping, and Shelf -life The proposed device sterilization process using Ethylene Oxide (EO) has been validated in accordance with ISO 11135:2014 to achieve a sterility assurance level (SAL) of 106. EO and Ethylene Chlorohydrin (ECH) residuals were below the limits specified in ISO 10993-7:2008. Bacterial Endotoxin Levels were below the level of 2.15 EU/device in accordance with USP <85>. Both baseline and accelerated shelf life testing were conducted demonstrating the device will perform as intended to support the proposed 3 year shelf-life. - . Package integrity test - after environmental conditioning, simulated transportation testing in accordance to ASTM D4169-22 on final, packaged, and sterile device. - Sterile Barrier Packaging performed on the proposed device: . - Seal Strength ASTM F88/F88-15 o - Dye penetration ASTM F1929-15 O - Visual Inspection ASTM F1886/F1886M-16 O - . Shelf-life of 3-years is validated using FDA recognized standard ASTM F1980 -16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices. {9}------------------------------------------------ #### Clinical Test Conclusion 8. No clinical study is included in this submission. - 9. Substantially Equivalent (SE) Comparison | Table 1. General Comparison | | | | |--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------| | Item | Proposed Device | Predicate Device | Remark | | Product Code | PKL | PKL | Same | | Regulation Number | 21CFR 876.4400 | 21CFR 876.4400 | Same | | Indications for Use | The Disposable<br>Hemostatic Clips is indicated<br>for Endoscopic clip placement<br>within the Gastrointestinal tract<br>in adult populations only via a<br>straight or side viewing flexible<br>endoscope for the purpose of :<br>1) Endoscopic marking;<br>2) Hemostasis for:<br>A) Mucosal/sub-mucosal<br>defects < 3cm.<br>B) Bleeding ulcers.<br>C) Polyps < 1.5cm in<br>diameter.<br>D) Diverticula in the<br>colon.<br>E) Arteries < 2 mm.<br>F) Prophylactic clipping to<br>reduce the risk of delayed<br>bleeding post lesion resection.<br>3) As a supplementary<br>method, closure of GI tract<br>luminal perforations <20mm<br>that can be treated<br>conservatively<br>4) Anchoring to affix jejunal<br>feeding tubes to the wall of the<br>small bowel; | The Lockado™ Repositionable<br>Hemostasis Clip is indicated for<br>Endoscopic clip placement<br>within the Gastrointestinal tract<br>in adult patients only via a<br>straight or side viewing flexible<br>endoscope for the purpose of:<br>(1) Endoscopic marking,<br>(2) Hemostasis for<br>(a)Mucosal / sub-mucosal<br>defects < 3cm,<br>(b) Bleeding ulcers,<br>(c) Polyps < 1.5cm in diameter,<br>(d) Diverticula in the colon,<br>(e) Arteries < 2 mm<br>(f) Prophylactic clipping to<br>reduce the risk of delayed<br>bleeding post lesion resection;<br>(3) As a supplementary method,<br>closure of GI tract luminal<br>perforations <20mm that can be<br>treated conservatively<br>(4) Anchoring to affix jejunal<br>feeding tubes to the wall of the<br>small bowel; | Same | | Configuration | Delivery system and clip | Delivery system and clip | Same | | | assembly | assembly | | | Open width | 9mm, 11mm, 13mm and 16mm | 8mm, 11mm and 16mm, 22mm | Different | | Minimal working channel | 2.8mm | 2.8mm | Same | | Working length | 1250mm, 1650mm, 2350mm and 2700mm | 1650mm 1950mm, 2350mm and 2700mm | Different | | Ring number | Single ring and three rings | Single ring | Different | | Single Use | Single Use | Single Use | Same | | Labeling | Comply with 21 CFR Part 801 | Comply with 21 CFR Part 801 | Same | | Material | | | | | Clip assembly | Stainless Steel (SUS316L)<br>Stainless Steel (SUS631)<br>Stainless Steel (SUS316F) | | | | Traction assembly | Stainless Steel (SUS304) | | | | Outer tube assembly | Stainless Steel (SUS304)<br>Stainless Steel (SUS316F)<br>Polyethylene (PE) | | | | Slide bar | Acrylonitrile Butadiene Styrene (ABS) | Unknown | Different | | Slip ring | Thermoplastic Elastomer (TPE) | | | | Rotation head | Acrylonitrile Butadiene Styrene (ABS) | | | | | Thermoplastic Elastomer (TPE) | | | | Protective sleeve | High pressure-low density polyethylene (LDPE) | | | | Biocompatibility for Delivery system | | | | | Cytotoxicity | No cytotoxicity | | | | Skin Sensitization | No skin sensitization | | Same | | Irritation | No irritation | Comply with ISO 10993 standards | | | Acute Systemic Toxicity | No acute toxicity | | | | Sub-acute Systemic Toxicity | No sub-acute toxicity | | | | Biocompatibility for clip assembly | | | | | Cytotoxicity | No cytotoxicity | | | | Skin Sensitization | No skin sensitization | | Same | | Irritation | No irritation | Comply with ISO 10993 standards | | | Acute Systemic Toxicity | No acute toxicity | | | | Sub-acute Systemic Toxicity | No sub-acute toxicity | | | | Implantation | No abnormal reaction | | | | Genotoxicity | No genotoxicity | | | | Sterilization | | | | | Method | Ethylene oxide | Ethylene oxide | Same | | SAL | $10^{-6}$ | $10^{-6}$ | Same | | Endotoxin Limit | 20EU | 20EU | Same | Table 1. General Comparison {10}------------------------------------------------ {11}------------------------------------------------ #### Different - Open width The open width for proposed device is available in 9mm, 11mm, 13mm and 16mm four different specifications. The predicate device is available in 8mm, 11mm 16mm and 22mm four different specifications. The 9mm and 13mm specification is different from predicate device. However, these specifications can be covered in the open width range of predicate device. Different working length devices will be selected by physician per patient's condition. Therefore, this difference will not raise new questions on safety and effectiveness of the proposed device. ### Different - Working length The working length for proposed device is available in 1250mm, 2350mm and 2700mm. The working length of predicate device is available in 1650mm 1950mm and 2700mm. Different working length devices will be selected by physician per patient's condition. The length for proposed device of 1650mm, 2350mm and 2700mm can be covered by predicate device length range. Although 1250mm effective length device are not included in the effective length range of the predicate device, it is very close to the minimum length of predicate device. Therefore, this difference will not raise new questions on safety and effectiveness of the proposed device. #### Different- Ring number The proposed device has two models of sing…
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...