K232925 · Alimetry , Ltd. · MYE · Nov 17, 2023 · Gastroenterology, Urology
Device Facts
Record ID
K232925
Device Name
Gastric Alimetry System
Applicant
Alimetry , Ltd.
Product Code
MYE · Gastroenterology, Urology
Decision Date
Nov 17, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1735
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The Gastric Alimetry System is intended to record, store, view and process gastric myoelectrical activity as an aid in the diagnosis of various gastric disorders.
Device Story
Gastric Alimetry System is an electrogastrography (EGG) device for non-invasive measurement of gastric myoelectrical activity. Input: 64-channel surface electrode array on adhesive patch; accelerometer for motion artifacts. Operation: Array records signals from skin surface; data digitized and transmitted to tablet App; App captures patient-reported symptoms. Processing: Fourier-transform spectral analysis; automated noise detection; correlation analysis between symptom severity and gastric amplitude. Output: Clinical report with myoelectrical waveforms, spectral maps, symptom summary panels, and gut-brain wellbeing questionnaire results. Used in clinical settings by GI physicians to aid diagnosis of gastric disorders. Benefits: Provides objective data on gastric rhythm and amplitude to correlate with patient symptoms, assisting in clinical decision-making.
Clinical Evidence
No clinical data. Bench testing only. Previous bench testing remains applicable; array electrical performance testing was repeated to confirm consistency with prior design iterations.
Technological Characteristics
Electrogastrography system; 64-channel electrode array (ANSI/AAMI EC12 compliant); accelerometer; 250 Hz raw sampling; 4 Hz logging; 0-2 Hz frequency range. Battery-powered; tablet-based App interface; cloud-server data storage. Software includes data acquisition, spectral analysis, and reporting. Reprocessed reader/dock; disposable electrodes. Complies with IEC 60601-1 (electrical safety) and IEC 60601-1-2 (EMC).
Indications for Use
Indicated for patients 18 years and older to record, store, view, and process gastric myoelectrical activity as an aid in the diagnosis of various gastric disorders. Used by medical professionals and specialist GI physicians.
Regulatory Classification
Identification
An electrogastrography system (EGG) is a device used to measure gastric myoelectrical activity as an aid in the diagnosis of gastric motility disorders. The device system includes the external recorder, amplifier, skin electrodes, strip chart, cables, analytical software, and other accessories.
Special Controls
*Classification.* Class II (Special Controls). The special controls are as follows:(1) The sale, distribution and use of this device are restricted to prescription use in accordance with § 801.109 of this chapter.
(2) The labeling must include specific instructions:
(i) To describe proper patient set-up prior to the start of the test, including the proper placement of electrodes;
(ii) To describe how background data should be gathered and used to eliminate artifact in the data signal;
(iii) To describe the test protocol (including the measurement of baseline data) that may be followed to obtain the EGG signal; and
(iv) To explain how data results may be interpreted.
(3) The device design should ensure that the EGG signal is distinguishable from background noise that may interfere with the true gastric myoelectric signal.
(4) Data should be collected to demonstrate that the device has adequate precision and the EGG signal is reproducible and is interpretable.
In addition to the general controls of the act, the 3CPM EGG Machine is subject to the following special controls: (1) The sale, distribution and use of this device are restricted to prescription use in accordance with 21 CFR 801.109. (2) The labeling must include specific instructions (a) to describe proper patient set-up prior to the start of the test, including the proper placement of electrodes; (b) to describe how background data should be gathered and used to eliminate artifact in the data signal; (c) to describe the test protocol (including the measurement of baseline data) which may be followed to obtain the EGG signal; and (d) to explain how data results may be interpreted. (3) The device design should ensure that the EGG signal is distinguishable from background noise that may interfere with the true gastric myoelectric signal. (4) Data should be collected to demonstrate that the device has adequate precision and the EGG signal is reproducible and is interpretable.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 17, 2023
Alimetry Ltd. % Janice Hogan Partner Hogan Lovells US LLP 1735 Market Street. Floor 23 Philadelphia, Pennsylvania 19115
Re: K232925
Trade/Device Name: Gastric Alimetry System Regulation Number: 21 CFR 876.1735 Regulation Name: Electrogastrography System Regulatory Class: Class II Product Code: MYE Dated: September 19, 2023 Received: September 19, 2023
Dear Janice Hogan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Shanil P. Haugen -S
Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use
510(k) Number (if known) K232925
Device Name
Gastric Alimetry
Indications for Use (Describe)
The Gastric Alimetry System is intended to record, store, view and process gastric myoelectrical activity as an aid in the diagnosis of various qastric disorders.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
□ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
{4}------------------------------------------------
#### 510(k) SUMMARY Alimetry's Gastric Alimetry
# Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Alimetry Ltd.
Phone: +64 27 622 2306
Contact Person: Yaara Yarmut, Chief Regulatory Officer. Date Prepared: September 19, 2023
# Name of Device and Name/Address of Sponsor
Trade name: Gastric Alimetry
Manufacturer: Alimetry
- Address: 70 Symonds St. Grafton Auckland 1010 New Zealand
Common name: Gastric Alimetry System
# Classification Name:
Electrogastrography system, 21 CFR 876.1735, Product code: MYE
Class II
# Predicate and Reference Devices
- 1. Trade Name: Gastric Alimetry (K223398) (primary predicate) Manufacturer: Alimetry Ltd.
# Intended Use / Indications for Use
The Gastric Alimetry is intended to record, store, view and process gastric myoelectrical activity as an aid in the diagnosis of various gastric disorders.
# Device Description
The Gastric Alimetry System is an electrogastrography (EGG) device, used for non-invasively measuring the myoelectrical activity of the stomach at the surface of the abdomen. The Gastric Alimetry System is intended to record, store, view and process gastric myoelectrical activity as an aid in the diagnosis of various gastric disorders.
{5}------------------------------------------------
The device is used to acquire and digitize gastric myoelectrical data and movement artifacts using an array with recording electrodes on an adhesive patch, which is used for recording the myoelectrical data from the skin surface. An App is used to set up the device and capture patient-reported symptom data. A report is provided to the clinicians at the end of the test which includes myoelectrical signal data for manual analysis, together with computed data summaries and plots. A Supplementary Report is also routinely available to clinicians that includes signal data from all 64 channels on the array.
In the modified Gastric Alimetry System, the following updates are introduced:
- Modifications to the patient report to include additional information, as follows:
- Inclusion of Symptom Summary panel in the Gastric Alimetry Report to provide a O visual summary of the overall symptom severities currently displayed by the Symptom Graph.
- Inclusion of Symptom Correlation Analysis panel in the Gastric Alimetry Report O to provide a visual summary of the correlation between symptom severity and gastric amplitude.
- Inclusion of Gut-Brain Wellbeing questions panel in the Gastric Alimetry Report o to display the results of ten optional wellbeing questions presented to the patient via the Gastric Alimetry App.
- Update to the design of the Reader's Enclosure
- The labeling has been updated to include normative range data to provide reference ● information for the clinician.
In addition, the following minor updates were introduced:
- 0 Minor firmware changes for efficiency
- 0 Gastric Alimetry App:
- Cybersecurity updates o
- Update to the user interface of the App to include additional symptom logging O
- Alimetry Cloud: ●
- O Updated to health professional dashboard available to include the session number
- Addition of multi-factor authentication for user login. O
- Minor Algorithm fixes O
# Technological Characteristics / Substantial Equivalence
The Gastric Alimetry device has the same intended use and indications for use and technological characteristics as the primary predicate. Both devices are intended to record, store, view and process gastric myoelectrical activity as an aid in the diagnosis of various gastric disorders.
| Item | Subject Device | Primary Predicate Device | Discussion |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device | Gastric Alimetry System | Gastric Alimetry System | |
| Manufacturer / K# | Alimetry Ltd. | Alimetry Ltd. / K223398 | |
| Item | Subject Device | Primary Predicate Device | Discussion |
| Intended Use and<br>indication for use | To record, store, view and<br>process gastric myoelectrical<br>activity as an aid to the<br>diagnosis of various gastric<br>disorders. | To record, store, view and<br>process gastric myoelectrical<br>activity as an aid to the<br>diagnosis of various gastric<br>disorders. | Same as the predicate. |
| User Population | Medical professionals to set up<br>and use the system | Medical professionals to set up<br>and use the system | Same as the predicate. |
| | Patients 18 years and older | Patients 18 years and older | Same as the predicate. |
| | Specialist GI physicians | Specialist GI physicians | Same as the predicate. |
| Accessories | | | |
| Setup and<br>Charging Dock | Gastric Alimetry Dock for<br>charging and user convenience | Gastric Alimetry Dock for<br>charging and user convenience | Same as the predicate. |
| Electrodes | Disposable; same components<br>as ECG electrodes. | Disposable; same components<br>as ECG electrodes. | Same as the predicate. |
| | Peel-and-stick patch. | Peel-and-stick patch. | Same as the predicate. |
| Technological Characteristics | | | |
| Sampling | raw sampling: 250 Hz | raw sampling: 250 Hz | Same as the predicate. |
| Frequency | logging: 4 Hz | logging: 4 Hz | Same as the predicate. |
| Low Frequency<br>Range | DC | DC | Same as the predicate. |
| High frequency<br>range | 2 Hz | 2 Hz | Same as the predicate. |
| Number of<br>channels | 64 (+2 reference) | 64 (+2 reference) | Same as the predicate. |
| Electrode to<br>recorder interface | Reader located on the Array and<br>directly connected to it. | Reader located on the Array and<br>directly connected to it. | Same as the predicate. |
| Screen | Dedicated tablet for system<br>operation | Dedicated tablet for system<br>operation | Same as the predicate. |
| Motion Sensor | Accelerometer | Accelerometer | Same as the predicate. |
| Data Storage | On data acquisition device and<br>cloud server | On data acquisition device and<br>cloud server | Same as the predicate. |
| Weight of device<br>on patient | 225 grams - Reader and Array | 225 grams - Reader and Array | Same as the predicate. |
| ltem | Subject Device | Primary Predicate Device | Discussion |
| Power Source | Battery powered | Battery powered | Same as the predicate. |
| Software | Setup (including clinical data),<br>device control, data acquisition<br>software and system checks via<br>App running on a tablet. | Setup (including clinical data),<br>device control, data acquisition<br>software and system checks via<br>App running on a tablet. | Same as the predicate. |
| Patient Symptom<br>Logging | Available on App interface.<br>Presented as plot with times. | Available on App interface.<br>Presented as plot with times. | Same as the predicate. |
| Skin Preparation | Yes. Shave, measure, skin prep<br>with abrasive conductive gel<br>Additional array template<br>marking step for convenience. | Yes. Shave, measure, skin prep<br>with abrasive conductive gel<br>Additional array template<br>marking step for convenience. | Same as the predicate. |
| Sterilization | Electrodes are disposable, non-<br>sterile.<br>Reader and Dock are<br>reprocessed, not supplied<br>sterile. Cleaning and disinfection<br>instructions using wipes<br>provided. | Electrodes are disposable, non-<br>sterile.<br>Reader and Dock are<br>reprocessed, not supplied<br>sterile. Cleaning and disinfection<br>instructions using wipes<br>provided. | Same as the predicate. |
| Safety Features | | | |
| System Checks | Impedance monitor and<br>connectivity check displayed<br>prior to recordings. | Impedance monitor and<br>connectivity check displayed<br>prior to recordings. | Same as the predicate. |
| Instructions to<br>Patient | Yes. User instructs patient to<br>limit movement, talking and<br>sleeping.<br>These instructions are displayed<br>to the patient via an App as an<br>added safety measure. | Yes. User instructs patient to<br>limit movement, talking and<br>sleeping.<br>These instructions are displayed<br>to the patient via an App as an<br>added safety measure. | Same as the predicate. |
| Reporting Features | | | |
| Visualization of<br>myoelectrical<br>waveforms | Yes, all channels. | Yes, all channels. | Same as the predicate. |
| Automated<br>Spectral Analysis | Yes, by Fourier-transform | Yes, by Fourier-transform | Same as the predicate. |
| Artifact<br>Evaluation | By manual identification of noise<br>in waveforms and with reference<br>to motion sensor. | By manual identification of noise<br>in waveforms and with reference<br>to motion sensor. | Same as the predicate. |
| Item | Subject Device | Primary Predicate Device | Discussion |
| | Additionally provides automated<br>noise detection for convenience. | Additionally provides automated<br>noise detection for convenience. | |
| Frequency,<br>Amplitude, and<br>Rhythm Stability<br>of Myoelectrical<br>Activity | Yes, data tables, and<br>visualization by spectral graphs<br>and myoelectrical waveforms. | Yes, data tables, and<br>visualization by spectral graphs<br>and myoelectrical waveforms. | Same as the predicate. |
| Other<br>Myoelectrical<br>Parameters | Measures of % time in<br>frequency ranges and changes<br>in signal power over time are<br>available by visual inspection of<br>spectral maps and waveforms. | Measures of % time in<br>frequency ranges and changes<br>in signal power over time are<br>available by visual inspection of<br>spectral maps and waveforms. | Same as the predicate. |
| Symptom<br>Outputs | Each symptom severity and<br>symptom event is shown as a<br>function of time. The total<br>symptom burden score is<br>displayed in a table.<br>The average symptom severity<br>for each symptom is shown in a<br>radar plot. The total symptom<br>burden score and symptom<br>event counts are shown using<br>summary bar graphs.<br>The gastric amplitude and<br>symptom severities are shown<br>on a shared axis, normalized so<br>that they can be visually<br>compared. The corresponding<br>correlation coefficients are<br>displayed as a bar graph.<br>Questions are presented to the<br>patient on the App. Patient<br>responses are captured and<br>provided to the clinician in the<br>Gastric Alimetry Report. | Each symptom severity and<br>symptom event is shown as a<br>function of time. The total<br>symptom burden score is<br>displayed in a table.<br>Gut-brain wellbeing questions<br>asked by clinician during patient<br>consultation. | The Subject Device<br>contains additional<br>visualizations of data<br>presented by the<br>Predicate Device<br>without removing or<br>altering any of the<br>existing outputs. Refer<br>discussion above on<br>symptom visualization<br>and correlation<br>analysis.<br>The addition of the gut-<br>brain wellbeing<br>questions into the app<br>and report does not<br>impact the<br>performance of the<br>device. |
| Technical /<br>recording quality<br>outputs | Yes. Impedance, motion.<br>Additionally shows Array<br>position and spatial distribution<br>of amplitude for user<br>convenience. | Yes. Impedance, motion.<br>Additionally shows Array<br>position and spatial distribution<br>of amplitude for user<br>convenience. | Same as the predicate. |
| | | | |
| Electrodes | ANSI/AAMI EC12–compliance<br>with relevant requirements. | ANSI/AAMI EC12–compliance<br>with relevant requirements. | Same as the predicate. |
| Medical Electrical<br>Equipment | IEC 60601-1 | IEC 60601-1 | Same as the predicate. |
| Electromagnetic<br>Compatibility | IEC 60601-1-2 | IEC 60601-1-2 | Same as the predicate. |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
#### Standards with which the Device Complies
#### Performance Data
The modifications to the device since the prior clearances, namely additional Symptom Summary panel, Symptom Correlation Analysis panel, additional Gut-Brain Wellbeing questions, and an update to the Reader's enclosure, were minimal and did not significantly impact the safety or performance of the device as reflected in the previously performed bench testing, the testing submitted in the prior 510(k) notice remains applicable. For completeness, the company repeated the array electrical performance testing on the final version of the device with the thinner gel that was incorporated into the device design addressed in K213924.
No clinical studies were required.
# Substantial Equivalence
The modified Gastric Alimetry System has the same intended uses and the same indications, technological characteristics, and principles of operation as its predicate device, the Gastric Alimetry System. The minor technological differences between the Gastric Alimetry and its predicate device do not raise different questions of safety or effectiveness. Therefore the modified Gastric Alimetry System is shown to be substantially equivalent.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.