The RELIEF™ Ureteral Stent is intended for temporary drainage from the ureteropelvic junction to the bladder of the ureter in a variety of benign, malignant and post-traumatic conditions of the ureter. These conditions include stones and/or stone fragments or other ureteral obstructions such as those associated with ureteral stricture, malignancy of abdominal organs, retroperitoneal fibrosis or ureteral trauma, or in association with Extracorporeal Shock Wave Lithotripsy (ESWL). The stent may be placed using endoscopic surgical techniques. The stent is not intended as a permanent indwelling device. The indwelling time should not exceed thirty (30) days.
Device Story
Sterile, single-use ureteral stent kit (6Fr, 24-26cm) for temporary drainage from ureteropelvic junction to bladder; constructed of radiopaque polymer tube with side holes and hydrophilic coating. Features proximal tubular coil and body segment connected via 4cm suture tether to distal bladder coil; tether design allows natural ureteral orifice function to prevent vesicoureteral reflux. Placed via endoscopic surgical techniques by urologists. Stent remains indwelling for up to 30 days. Radiopaque markers facilitate placement verification. Benefits include effective drainage while minimizing reflux risk via low-profile tether segment.
Clinical Evidence
Prospective, open-label clinical study of 19 patients (10 male, 9 female; age 39-84) evaluated the effect of the suture tether on vesicoureteral reflux. Primary endpoint: assessment of urinary orifice function via pre- and post-stent cystograms. Results: 100% of patients (19/19) graded as 'No Reflux' post-implant. No adverse complications reported. Data confirms the low-profile tether does not impede the one-way valve function of the urinary orifice.
Technological Characteristics
Radiopaque polymer tube with central lumen, side holes, and hydrophilic coating. Distal bladder coil constructed of non-lumened polymeric segment; tether made of 3-0 Polypropylene Suture (Deklene). Dimensions: 6 Fr diameter, 24 cm or 26 cm length. Sterilization: Ethylene Oxide (EtO). Single-use. Radiopaque markers printed along length.
Indications for Use
Indicated for temporary ureteral drainage in patients (adults >18 years) with benign, malignant, or post-traumatic ureteral conditions, including stones, stone fragments, ureteral strictures, abdominal malignancy, retroperitoneal fibrosis, or post-ESWL. Contraindicated in patients with distal ureteral obstruction, known/demonstrated ureteral reflux, or active UTI/sepsis.
Regulatory Classification
Identification
A ureteral stent is a tube-like implanted device that is inserted into the ureter to provide ureteral rigidity and allow the passage of urine. The device may have finger-like protrusions or hooked ends to keep the tube in place. It is used in the treatment of ureteral injuries and ureteral obstruction.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 22, 2024
Ureteral Stent Company Mike Bunker CEO 45 Glenridge Court Chagrin Falls, Ohio 44022
Re: K232920
Trade/Device Name: RELIEF™ Ureteral Stent Kit: Model: RS-001 - 6 Fr x 24cm, RELIEFTM Ureteral Stent Kit; Model: RS-002 - 6 Fr x 26cm Regulation Number: 21 CFR 876.4620 Regulation Name: Ureteral Stent Regulatory Class: II Product Code: FAD Dated: September 15, 2023 Received: September 19, 2023
Dear Mike Bunker:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}------------------------------------------------
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Angel A. Soler-garcia -S
for Jessica K. Nguyen, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K232920
Device Name
RELIEF™ Ureteral Stent
Indications for Use (Describe)
The RELIEF™ Ureteral Stent is intended for temporary drainage from the ureteropelvic junction to the bladder of the ureter in a variety of benign, malignant and post-traumatic conditions of the ureter.
These conditions include stones and/or stone fragments or other ureteral obstructions such as those associated with ureteral stricture, malignancy of abdominal organs, retroperitoneal trauma, or in association with Extracorporeal Shock Wave Lithotripsy (ESWL).
The stent may be placed using endoscopic surgical techniques.
The stent is not intended as a permanent indwelling time should not exceed thirty (30) days.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
1. SUBMITTER INFORMATION
## TRADITIONAL 510(K) SUMMARY
| Applicant | Ureteral Stent Company, Inc.<br>45 Glenridge Court Chagrin Falls,<br>Ohio 44022 |
|-----------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| Official Correspondent | Mike Bunker<br>CEO Ureteral Stent Company, Inc.<br>Phone: (216) 258-8047<br>Email: |
| Date Prepared | March 19, 2024 |
| <b>2. DEVICE NAME</b> | |
| Trade Name of the Device | RELIEF Ureteral Stent Kit; Model: RS-001 - 6 Fr x 24cm<br>RELIEF Ureteral Stent Kit; Model: RS-002 - 6 Fr x 26cm |
| Common Name: | Ureteral Stent |
| Classification Name: | Stent, Ureteral |
| Classification Regulation: | 21 CFR 876.4620 |
| Device Class: | II |
| Product Code: | FAD |
| Panel: | Gastroenterology/Urology |
| <b>3. PREDICATE DEVICE<br/>IDENTIFICATION</b> | RELIEF Ureteral Stent Kit; Model: RS-001 - 6 Fr x 24cm |
## 4. DEVICE DESCRIPTION:
The RELIEF™ Ureteral Stents are sterile, single-use devices. The stents are available in 6Fr, with lengths of 24 cm and 26 cm. The ureteral stent is constructed of a radiopaque polymer tube with a central lumen with side holes positioned along its length to provide drainage of urine from the kidney to the bladder and includes hydrophilic coating. Along the stent are printed insertion markers throughout its length. The stent includes a radiopaque soft polymeric proximal tubular coil and body segment attached to a 4 cm tether of suture material that is placed along the ureter intramural segment, allowing natural opening, and closing of the ureteral orifice, thereby preventing vesicoureteral reflux. The tether is attached to a radiopaque distal bladder coil constructed of a monofilament (non-lumened), polymeric segment, allowing it to float in the bladder, thus precluding any tension on the tether or coil. The RELIEF™ Ureteral Stent package contents consist of:
RELIEF Ureteral Stent Kit; Model: RS-002 - 6 Fr x 26cm
- . RELIEF™ Ureteral Stent
- Stent pusher tube with radiopaque tip
- . Pigtail straightener
The RELIEF™ Ureteral Stent is not intended as a permanent indwelling device and is labeled for indwell time not to exceed thirty (30) days.
{5}------------------------------------------------
#### 5. INDICATIONS FOR USE:
The RELIEF™ Ureteral Stent is intended for temporary drainage from the ureteropelvic junction to the bladder of the ureter in a variety of benign, malignant and post-traumatic conditions of the ureter.
These conditions include stones and/or stone fragments or other ureteral obstructions such as those associated with ureteral stricture, malignancy of abdominal organs, retroperitoneal fibrosis or ureteral trauma, or in association with Extracorporeal Shock Wave Lithotripsy (ESWL).
The stent may be placed using endoscopic surgical techniques.
The stent is not intended as a permanent indwelling device. The indwelling time should not exceed thirty (30) days.
| Comparison Element | Subject Device:<br>RELIEFTM Ureteral Stent Kit | Predicate Device:<br>RELIEFTM Ureteral Stent Kit |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K232920 | K213444 |
| Indications for Use | The RELIEFTM Ureteral Stent is<br>intended for temporary drainage<br>from the ureteropelvic junction to<br>the bladder of the ureter in a variety<br>of benign, malignant and post-<br>traumatic conditions of the ureter.<br>These conditions include stones<br>and/or stone fragments or other<br>ureteral obstructions such as those<br>associated with ureteral stricture,<br>malignancy of abdominal organs,<br>retroperitoneal fibrosis or ureteral<br>trauma, or in association with<br>Extracorporeal<br>Shock<br>Wave<br>Lithotripsy (ESWL).<br>The stent may be placed using<br>endoscopic surgical techniques.<br>The stent is not intended as a<br>permanent indwelling device. The<br>indwelling time should not exceed<br>thirty (30) days. | The RELIEFTM Ureteral Stent is<br>intended for temporary drainage from<br>the ureteropelvic junction to the<br>bladder of the ureter in a variety of<br>benign, malignant and post-traumatic<br>conditions of the ureter.<br>These conditions include stones<br>and/or stone fragments or other<br>ureteral obstructions such as those<br>associated with ureteral stricture,<br>malignancy of abdominal organs,<br>retroperitoneal fibrosis or ureteral<br>trauma, or in association with<br>Extracorporeal Shock Wave<br>Lithotripsy (ESWL).<br>The stent may be placed using<br>endoscopic surgical techniques or<br>percutaneously using standard<br>radiographic technique.<br>The stent is not intended as a<br>permanent indwelling device. The<br>indwelling time should not exceed<br>thirty (30) days. |
| Indwell Time | Not exceed thirty (30) days. | Not exceed thirty (30) days. |
| Diameter | 6 Fr. | 6 Fr. |
| Length | 24 cm and 26 cm | 24 cm and 26 cm |
| Distal Coil | Distal Coil attached to a distal<br>tether. | Distal Coil attached to a distal tether. |
| Bladder Coil | 3-0 Polypropylene Suture Deklene | 3-0 Polypropylene Suture Deklene |
## 6. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE AND REFERENCE DEVICES
{6}------------------------------------------------
| Guidewire | 6.0 Fr stents accept 0.035" and<br>0.038" guide wire. | 6.0 Fr stents accept 0.035" and<br>0.038" guide wire. |
|-----------------------|-------------------------------------------------------|-------------------------------------------------------|
| Pusher Tube | Pusher Tube w/radiopaque tip | Pusher Tube w/radiopaque tip |
| Pigtail Straightener? | Yes | Yes |
| Sterilization | EtO | EtO |
| Single Use? | Yes | Yes |
As evidenced by the above table, both the subject and the predicate devices have same design, intended use and technological characteristics. In this 510(k), the sponsor included a new labeling claim that if the subject device is appropriately placed along the ureter intramural segment, it will allow the natural opening and closing of the ureteral orifice, thereby preventing vesicoureteral reflux. This labeling claim was supported by clinical data which are summarized below.
## 7. SUMMARY OF CLINICAL TESTING:
Clinical data collected from 19 patients were provided to verify that the design of the suture tether does not impede the normal functioning of the urinary orifice to prevent vesicoureteral reflux. This data were obtained from a prospective, open label study called 'Clinical Evaluation of RELIEF ™ Stent study' which was conducted in the University Hospital Cleveland Medical Center. The study enrolled 26 patients who were assessed for eligibility based on the inclusion criteria. Seven patients were excluded based on the exclusion criteria. Total 19 patients were evaluated in the study, including 10 male and 9 female patients. Eleven of the evaluated subjects underwent previous stent placement. Four and 15 of the evaluated patients had renal and ureteral calculi, respectively. The demographics and anatomical height of the subjects, inclusion criteria used for the study are listed below:
## Demographics
The patient demographics for the clinical performance testing is referenced below:
| Gender | Number of participants | Ethnicity |
|--------|------------------------|-------------------------------------------------|
| Male | 10 | All Caucasian |
| Female | 9 | Asian (1)<br>Afro-American (2)<br>Caucasian (6) |
The age range for the subjects was 39-84 years.
## Anatomical Height criteria:
The patient's anatomical height in relation to the RELIEF™ Ureteral Stent length utilized as referenced below:
| Anatomical Height | Stent Length |
|-------------------|--------------|
| 5' 3" – 5' 7" | 24 cm |
| 5' 8" – 5' 10" | 26 cm |
{7}------------------------------------------------
## Inclusion criteria:
- 1. Male and female patients.
- 2. Over 18 years of age and willing and able to provide informed consent.
- Renal or ureteral stone or both calculi between 5-25 mm measured on plain abdomen X-3. ray Kidney Ureter Bladder (KUB or computed tomography (CT).
- Upper or middle third ureteral stricture or stone AND/OR stone located in the renal 4. pelvis.
- Patient agrees to participate in the study and signs the informed consent form. న్.
- 6. Able to complete self-rated questionnaires.
## Exclusion criteria:
- 1. Patients with distal ureteral obstruction regardless of etiology, patients with known or demonstrated ureteral reflux intraoperatively.
- Patients with urinary reflux (assessed by pre-stent cystogram). 2.
- Patients requiring bilateral surgical stone management procedure or patients with 3. active urinary tract infection (UTI) or sepsis.
## Study results:
Across the 19 patients evaluated in the study, no patients experienced adverse complications. All cystograms, pre- and post-stent implant, were graded at 'No Reflux'. The comparison of the cystogram images of the RELIEF™ Ureteral Stent verified that the low-profile tether segment of the subject device placed in the intramural segment of the ureter did not impede the urinary orifice's one-way valve function. The normal function of the urinary orifice's one way valve is to close and seal off during micturition, thereby to prevent vesicoureteral reflux.
## 8. CONCLUSIONS
Based on the information presented in this submission, it can be concluded that the subject device is substantially equivalent to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.