K232903 · Pollogen, Ltd. · GEI · Oct 17, 2023 · General, Plastic Surgery
Device Facts
Record ID
K232903
Device Name
Legend X Platform
Applicant
Pollogen, Ltd.
Product Code
GEI · General, Plastic Surgery
Decision Date
Oct 17, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The Legend X Platform and its accessories are intended for dermatological procedures requiring ablation and resurfacing of the skin when using VoluDerm Energy (Applicator VO). It is also intended for use in dermatologic and general surgical procedures for the non-invasive treatment of mild to moderate facial wrinkles and rhytides when using TriPollar RF Energy and for muscle conditioning to stimulate healthy muscles (Applicators 1-3). Legend X Platform is not intended to be used in conjunction with therapy or treatment of medical diseases or medical conditions of any kind. Legend X Platform is intended to be operated by a trained professional who is present to monitor treatment.
Device Story
Legend X Platform is a software-controlled capital equipment system for dermatological and general surgical procedures. It utilizes radiofrequency (RF) energy for skin ablation, resurfacing, and wrinkle reduction, and electrical stimulation for muscle conditioning. The system comprises a console with a touchscreen interface, a foot switch, a patient-controlled manual stop switch, and various applicators with disposable tips. The operator selects treatment modes via the console; energy delivery is controlled by the user via the foot switch and monitored visually. The patient-controlled switch allows immediate cessation of treatment. The device is used in clinical settings by trained professionals. It benefits patients by providing non-invasive aesthetic improvements and muscle conditioning. The system processes user inputs to compute and deliver energy outputs to the applicators, which feature stainless steel and gold-plated pins for RF delivery.
Clinical Evidence
No clinical data provided. Evidence consists of bench testing, including reprocessing validation (AAMI TIR-30), sterilization validation (ISO 11135, ISO 10993-7), shelf-life/packaging testing (ASTM F1980, F1929, F88, ISO 11607), biocompatibility (ISO 10993-1, -5, -10, -11, -23), electrical safety/EMC (IEC 60601-1, -1-2, -2-10, -2-2), and software verification/validation. Animal testing was performed per FDA guidance.
Technological Characteristics
Console-based system with touchscreen GUI. Applicators deliver RF energy or muscle stimulation. Disposable tips constructed of stainless steel and gold plating. Connectivity via console for applicators, foot switch, and patient manual switch. Sterilization of disposables via Ethylene Oxide (EO). Software-controlled energy delivery. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, and IEC 60601-2-2.
Indications for Use
Indicated for dermatological procedures requiring skin ablation and resurfacing, non-invasive treatment of mild to moderate facial wrinkles and rhytides, and muscle conditioning to stimulate healthy muscles. Not for use in treating medical diseases or conditions. Intended for use by trained professionals.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and name on the right. The FDA part of the logo is in blue, with the acronym "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION" next to it.
October 17, 2023
Pollogen Ltd % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114
Re: K232903
Trade/Device Name: Legend X Platform Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI, NGX Dated: September 18, 2023 Received: September 18, 2023
Dear Prithul Bom:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Mark
Digitally signed by
Mark Trumbore -S
Trumbore -S Date: 2023.10.17
15:50:52 -04'00'
Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices
{2}------------------------------------------------
Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
## Indications for Use
510(k) Number (if known) K232903
Device Name Legend X Platform
Indications for Use (Describe)
The Legend X Platform and its accessories are intended for dermatological procedures requiring ablation and resurfacing of the skin when using VoluDerm Energy (Applicator VO).
It is also intended for use in dermatologic and general surgical procedures for the non-invasive treatment of mild to moderate facial wrinkles and rhytides when using TriPollar RF Energy and for muscle conditioning to stimulate healthy muscles (Applicators 1-3).
Legend X Platform is not intended to be used in conjunction with therapy or treatment of medical diseases or medical conditions of any kind. Legend X Platform is intended to be operated by a trained professional who is present to monitor treatment.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary K232903
# 5.1. General information
| 510(k) Submitter | Pollogen Ltd |
|-------------------------|-------------------------------------------------|
| | 6 Kaufman Street |
| | Tel Aviv, IL 6801298 |
| FDA Registration Number | 3008753275 |
| Primary Correspondent | Karen Smith |
| | Vice President, Regulatory & Quality |
| | Lumenis Be, Inc., a Pollogen Ltd Family Company |
| Contact Information | Email: karen.smith@lumenis.com |
| Secondary Correspondent | Lola Fajinmi |
| | Regulatory Affairs Consultant |
| | MedTeCheck |
| Contact Information | Email: lola@MedTeCheck.com |
| Date Prepared | September 17, 2023 |
## 5.2. Device Identification
The Legend X Platform consists of a Console with connected components including: a Foot Switch, Patient Controlled Manual Switch, Applicators, and Applicator accessories (disposable tips) needed to perform some of Legend X dermatologic procedures. The Legend X Platform can be defined as follows:
| Proprietary Name | Legend X Platform |
|-----------------------------|----------------------------------------------------------------|
| Device Classification Name: | Electrosurgical, Cutting & Coagulation & Accessories |
| Regulation Description: | Electrosurgical Cutting and Coagulation Device and Accessories |
| Regulation Number: | 878.4400 |
| Device Class: | Class II |
| Product Code: | GEI, NGX |
### 5.3. Predicate Devices
Primary Predicate:
| Proprietary Name | <i>Pollogen</i> Legend+™ System |
|-----------------------------|-------------------------------------------------------------------|
| Device Classification Name: | Electrosurgical, Cutting & Coagulation & Accessories |
| Regulation Description: | Electrosurgical Cutting and Coagulation Device and<br>Accessories |
| 510(k) Number: | |
{5}------------------------------------------------
Pollogen Ltd
| Regulation Number: | 878.4400 |
|--------------------|----------|
| Device Class: | Class II |
| Product Code: | GEI |
### Secondary Predicate:
| Proprietary Name | Legend Pro DMA |
|-----------------------------|-----------------------------------------------------|
| Device Classification Name: | Stimulator, Muscle, Powered for Muscle Conditioning |
| Regulation Description: | Powered Muscle Stimulator |
| 510(k) Number: | K200545 |
| Regulation Number: | 890.5850 |
| Device Class: | Class II |
| Product Code: | NGX |
### 5.4. Device Description
The Legend X Platform ("Proposed Device") is a software-controlled capital equipment platform that enables application of radiofrequency onto the skin for ablation, resurfacing, or noninvasive treatment of wrinkles and rhytides or muscle activation. The application of radiofrequency or muscle activation only occurs under the continuous and direct control of a user with direct visualization, via the user inputs into the Console, Applicator indicators and subsequent operation of the Foot Switch and Applicators.
The proposed device consists of a Console with connected Foot Switch. Patient Controlled Manual Switch, Applicators, and Applicator accessories (disposable tips) needed to perform some of Legend X dermatological and general surgical procedures. The summary descriptions of each component are outlined below.
Legend X Console: The Legend X Console (also known as the Console or Main Control Unit) contains the graphic display interface for the user that is provided by a touchscreen monitor for viewing and a computer running the Legend X software. The monitor allows for user input during initial setup and throughout the session. The console also provides power and connectivity for the applicators, foot switch, and patient manual-controlled switch.
Legend X Foot Switch: The Legend X Foot Switch (Foot Switch) is connected to the Legend X Console via a connector cord. The Foot Switch enables or disables the energy delivery through the applicators to treatment area.
Legend X Applicators: The Legend X Applicators (1, 2, 3 and VO) are handheld handpieces which deliver energy to the treatments area. Depending on the applicator and selected user inputs, the user can utilize the applicators to deliver either muscle energy onto the muscles of the face and body or RF energy for ablation and resurfacing or noninvasive treatment of wrinkles and rhytides onto the skin. Each applicator is connected to the Console via an Applicator Connector cord.
{6}------------------------------------------------
Legend X Patient Controlled Manual Switch: The Legend X Patient Controlled Manual Switch is also a handheld device that serves as the patient interface that allows the patient to stop the operation of the chosen treatment mode upon the press of a button. In this situation, the system does not allow operation to continue. A Patient Controlled Manual Switch cord connects the Patient Controlled Manual Switch to the Console.
Legend X Software: The Legend X Software provides the user and patient with the ability to safely commence, drive, and stop the operation of the Applicators on the skin or muscle area of interest throughout the face and body. It receives user input from the Legend X Console, Foot Switch, and Patient Manual Controlled Switch, and computes the appropriate output to the chosen applicator connected to the Console. It provides a graphical user interface where the treatment timeline as well as status of operation is shown in real time and displays important system status information.
Legend X Disposable Tips: The Legend X Disposable Tips set is part of the Legend X Applicator VO kit. The Disposable Tips Set includes:
gen12 disposable tip gen 36 disposable tip gen 36L disposable tip gen 100 disposable tip H7X7 disposable tip
The patient contacting portions of the disposable tips are constructed of stainless steel and gold plating. The disposable tips are connected to handheld Applicator VO for delivery of radiofrequency electrical current via an array of multi-electrode pins onto the skin surface.
# 5.5. Intended Use/Indications for Use
The Legend X Platform and its accessories are intended for dermatological procedures requiring ablation and resurfacing of the skin when using VoluDerm Energy (Applicator VO).
It is also intended for use in dermatologic and general surgical procedures for the noninvasive treatment of mild to moderate facial wrinkles and rhytides when using TriPollar RF Enerqy and for muscle conditioning to stimulate healthy muscles (Applicators 1-3).
Legend X Platform is not intended to be used in conjunction with therapy or treatment of medical diseases or medical conditions of any kind. Legend X Platform is intended to be operated by a trained professional who is present to monitor treatment.
# 5.6. Comparison of Technological Characteristics with the Predicate Device
Overall, the proposed and predicate devices are based on the following similar basic technological elements:
{7}------------------------------------------------
Device contains a central console that connects each of the main components to facilitate dermatologic and general surgical procedures and facilitates the selection of either RF, muscle stimulation, or fractional RF treatment modes via the graphic user interface
Device contains applicators to allow energy delivery and application of the selected mode on to the area of interest on the skin or muscles
Device contains similar components between the subject and predicate devices (i.e., console, foot switch, applicators, patient manual-controlled switch, and disposable tips)
Device requires continuous direct control by the user to operate the Applicators Device mode operation is only at the command of the user
Device mode operation may be halted at the command of either the user or the patient
The main difference between the proposed device and the predicates, Pollogen Legend+ TM System (K173503) and Legend Pro DMA (K200545), is that the Legend X Platform includes two of the five disposable tips (gen100 and H7x7) as an additional option. The Legend X Platform is an upgraded version of the predicate devices. For this reason, additional testing for the disposable tips was performed and is captured as part of this submission.
| Proposed Device<br>Legend X Platform | Primary Predicate<br>Pollogen Legend+<br>TM<br>(K173503) | Secondary<br>Predicate<br>Legend Pro DMA<br>(K200545) | Technological<br>Characteristics<br>Comparison |
|--------------------------------------|----------------------------------------------------------|-------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| System Console/<br>Main Control Unit | System Console/<br>Main Control Unit | System Console/<br>Main Control Unit | Identical Characteristics:<br>Functionality: To provide<br>central connection to all<br>device components; as<br>well as a user interface<br>for user inputs during<br>setup<br>Same specifications and<br>materials<br>Differences<br>Minor updates to<br>console, GUI |
| Foot Switch | Foot Switch | Foot Switch | Same |
Comparison of the technological characteristics with the Predicate Device for each of the Legend X Platform components are described below:
{8}------------------------------------------------
Pollogen Ltd
Traditional 510(k) Notification
| Patient-controlled<br>manual switch | Patient-controlled<br>manual switch | Patient-controlled<br>manual switch | Same |
|----------------------------------------------------------------|----------------------------------------------|-------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicators (1, 2, 3, and VO) | Applicators (1, 2, 3, and VO) | Applicators (1, 2, and 3) | Same |
| Disposable Tips:<br>Gen12<br>Gen36<br>Gen36L<br>Gen100<br>H7x7 | Disposable Tips:<br>Gen12<br>Gen36<br>Gen36L | N/A | Identical Characteristics:<br>Functionality: To provide<br>delivery of fractional RF<br>energy<br>Similar specifications<br>and intended<br>functionality.<br>Differences<br>Two additional optional<br>disposable tips (same<br>materials with additional<br>pin configurations for<br>gen100 and H7x7) for<br>applicator VO<br>Minor external color<br>differences |
# 5.7. Performance Data
The Legend X Platform was assessed for performance in accordance with internal design specification with the applicable performance standards to demonstrate safety and effectiveness. The testing identified no new issues of safety or effectiveness. The testing performed are summarized below:
### Reprocessing
Cleaning: The cleaning and disinfection instructions provided in labeling for nondisposable components were validated against the following standard: AAMI TIR-30:2011 – A compendium of processes, materials, test methods, and acceptance criteria for cleaning reusable medical devices
Sterility: Single- use disposable sterile (EO) devices and EO residual were validated in accordance with the following standards:
ISO 11135:2014 - Sterilization of healthcare products - Ethylene Oxide -Requirements for development validation and routine control of a sterilization process for medical devices
ISO10993-7:2008/AMD-1:2019 – Biological evaluation of medical devices – Ethylene Oxide sterilization residuals
Shelf Life and Sterile Barrier Packaging: The shelf life and sterile barrier packaging of
{9}------------------------------------------------
#### K232903
the single-use disposable devices were evaluated per the following standards:
ASTM F1980-16 – Standard quide for accelerated aging of sterile barrier systems for medical devices
ASTM F 1929-15 – Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration
ASTM F88/F88M-15: Standard Test Method for Seal Strength of Flexible Barrier Materials
ISO 11607 - Packaging for terminally sterilized medical devices
### Biocompatibility
The final finished form of the subject device has been used for the biocompatibility evaluation. Biocompatibility for patient contacting components has been evaluated and validated in accordance with the provision of the following FDA Guidance document(s) and standards:
Guidance for Industry and Food and Drug Administration Staff – Use of International Standard ISO 10993-1, "Biological evaluation of medical devices -Part 1: Evaluation and testing within a risk management process"
ISO 10993-1:2018 - Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
ISO 10993-5:2009 - Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
ISO 10993-10:2010 - Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization
ISO 10993-11:2017 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity
ISO 10993-23:2021 - Biological evaluation of medical devices - Part 23: Tests for irritation
# Electrical Safety, Performance and EMC
The proposed device has been fully evaluated for electrical safety and EMC compliance to the following standards:
IEC 60601-1:2005 (Third edition) + CORR. 1: 2006 + CORR. 2:2007 + A1:2012 (or IEC 60601-1: 2012 reprint) – Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) IEC 60601-1-2:2014 – Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
IEC 60601-2-10:2016 – Medical electrical equipment - Part 2-10: Particular requirements for the basic safety and essential performance of nerve and muscle stimulators
{10}------------------------------------------------
IEC 60601-2-2:2017 - Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories
### Software
Software was developed, tested, and verified per the FDA quidance document: Guidance for Industry and FDA Staff – Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices
Results of verification and validation testing confirm that the proposed device conforms to design specifications and meets the needs of the intended users.
#### V&V Testing
Performance testing was executed to verify the overall functionality of the proposed device to operate as specified by the design input requirements including applicator, patient-manual controlled-switch, and foot switch operation and controls, various functional safety features, and other general functionality. Requirements for safety and efficacy of the system. including but not limited to the adherence to requlatory standards were verified. Results of verification testing confirm that the proposed device conforms to design specifications and requirements and meets the needs of the intended user.
#### Animal Testing
Performance testing - animal was executed according to General Considerations for Animal Studies Intended to Evaluate Medical Devices issued on March 28, 2023.
### 5.8. Conclusion
Based on the indication for use, technological characteristic, performance testing and comparison to the predicate devices, the Legend X Platform raises no new questions of safety and effectiveness as compared to the predicate devices and is substantially equivalent to the predicate devices in terms of safety, efficacy, and performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.