← Product Code MRW · K232877

# FaSet Fixation System, UNITY Sacroiliac Joint Fixation System, Huvex Interspinous Fixation System, and AEON-C Stand Alone System (K232877)

_K&J Consulting Corporation · MRW · Dec 13, 2023 · OR · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K232877

## Device Facts

- **Applicant:** K&J Consulting Corporation
- **Product Code:** MRW
- **Decision Date:** Dec 13, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** OR
- **Attributes:** Therapeutic

## Intended Use

The FaSet Fixation System is intended to spine as an aid to fusion through bilateral immobilization of facet joints. The Facet Screw System is indicated for posterior surgical treatment with or without bone graft at single or multiple levels from C2 to S1 (inclusive). For transfacet fixation, the screws are inserted posterior through the superior side of the facet, across the facet joint, and into the pedicle. For translaminar facet fixation, the screws are inserted posteriorly through the lateral aspect of the spinous process, through the superior side of the facet, across the facet joint, and into the pedicle. The FaSet Fixation System is indicated for any or all of the following: - Spondylolisthesis - Pseudoarthrosis or failed previous fusions which are symptomatic, or which may cause secondary instability or deformity - Spondylolysis - Degenerative Disk Disease (DDD) as defined by back pain of discogenic origin with degeneration of disk confirmed by history and radiographic studies and/or degenerative disease of the facets with instability - Trauma including spinal fractures and/or dislocations.

## Device Story

System includes four spinal fixation product lines: FaSet (facet screws), UNITY (sacroiliac screws), Huvex (interspinous plates), and AEON-C (anterior cervical interbody fusion). Implants provide mechanical support and stability to spinal segments to facilitate biologic fusion. FaSet screws inserted posteriorly across facet joints into pedicles. UNITY screws stabilize sacroiliac joint. Huvex plates attach to spinous processes as supplemental fixation. AEON-C interbodies (PEEK/titanium) implanted anteriorly with bone graft. Used in surgical settings by orthopedic or neurosurgeons. Devices are non-sterile, single-use, and require surgical implantation. Benefit: stabilization of spinal segments to promote fusion and alleviate symptoms of degenerative disease, trauma, or instability.

## Clinical Evidence

No clinical data. Bench testing only; subject devices are identical to predicates or incorporate components that do not introduce a new mechanical worst-case scenario.

## Technological Characteristics

Materials: Titanium alloy (Ti-6Al-4V ELI, ASTM F136), Titanium (ASTM F67), Cobalt chrome alloy (ASTM F1537), and PEEK. Components include screws, plates, pins, bolts, and interbodies. Mechanical fixation principle. Non-sterile, single-use. No software or electronic components.

## Predicate Devices

- FaSet Fixation System ([K222515](/device/K222515.md))
- UNITY Sacroiliac Joint Fixation System ([K222448](/device/K222448.md))
- Huvex Interspinous Fixation System ([K223790](/device/K223790.md))
- AEON-C Stand Alone System ([K223140](/device/K223140.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

December 13, 2023

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services. To the right of the symbol, there is the FDA logo in blue, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a sans-serif font, also in blue.

K&J Consulting Corporation % Jeena Mathai President Eerkie Corporation 4027 Runnymeade Dr Collegeville, Pennsylvania 19426

Re: K232877

Trade/Device Name: FaSet Fixation System, UNITY Sacroiliac Joint Fixation System, Huvex Interspinous Fixation System, and AEON-CTM Stand Alone System Regulatory Class: Unclassified Product Code: MRW, OUR, HWC, PEK, OVE, ODP Dated: September 14, 2023 Received: September 15, 2023

Dear Jeena Mathai:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Eileen
Digitally signed
by Eileen Cadel -
S
Cadel -S Date: 2023.12.13
09:16:28 -05'00'
for

Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

510(k) Number (if known) K232877

Device Name FaSet Fixation System

#### Indications for Use (Describe)

The FaSet Fixation System is intended to spine as an aid to fusion through bilateral immobilization of facet joints. The Facet Screw System is indicated for posterior surgical treatment with or without bone graft at single or multiple levels from C2 to S1 (inclusive). For transfacet fixation, the screws are inserted posterior through the superior side of the facet, across the facet joint, and into the pedicle. For translaminar facet fixation, the screws are inserted posteriorly through the lateral aspect of the spinous process, through the superior side of the facet, across the facet joint, and into the pedicle. The FaSet Fixation System is indicated for any or all of the following:

### - Spondylolisthesis

- Pseudoarthrosis or failed previous fusions which are symptomatic, or which may cause secondary instability or deformity

- Spondylolysis

- Degenerative Disk Disease (DDD) as defined by back pain of discogenic origin with degeneration of disk confirmed by history and radiographic studies and/or degenerative disease of the facets with instability

- Trauma including spinal fractures and/or dislocations.

Type of Use (Select one or both, as applicable)

| <p>☑ Prescription Use (Part 21 CFR 801 Subpart D)</p> |
|-------------------------------------------------------|
|-------------------------------------------------------|

| <p>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</p> |
|------------------------------------------------------|
|------------------------------------------------------|

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

{3}------------------------------------------------

510(k) Number (if known) K232877

Device Name

UNITY Sacroiliac Joint Fixation System

| Indications for Use (Describe) |  |
|--------------------------------|--|
|--------------------------------|--|

The UNITY Sacroiliac Joint Fixation System is indicated for sacroiliac joint fusion for:

- Sacroiliac joint dysfunction including degenerative sacroiliac joint disruptions

- Augmenting immobilization and stabilization of the sacroiliac joint in skeletally mature patients undergoing sacropelvic fixation as part of a lumbar or thoracolumbar fusion

The UNITY Sacroiliac Joint Fixation System is also indicated for fracture fixation of acute, and non-traumatic fractures involving the sacroiliac joint.

Type of Use (Select one or both, as applicable)

| <span style="font-family: sans-serif;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)         </span> |
|----------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: sans-serif;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)         </span>                 |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

{4}------------------------------------------------

510(k) Number (if known) K232877

Device Name Huvex Interspinous Fixation System

#### Indications for Use (Describe)

The Huvex Interspinous Fixation System is a posterior, non-pedicle supplemental fixation device intended for use in the lumbar spine (T1-S1) as an adjunct to fusion in skeletally mature patients. It is intended for plate fixation/attachment to the spinous processes for the purpose of achieving supplemental fusion in the following conditions: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), tumor, and/or lumbar spinal stenosis. The Huvex Interspinous Fixation System is intended for use, in conjunction with autogenous bone graft, and not intended for standalone use.

| Type of Use (Select one or both, as applicable)                                                      |
|------------------------------------------------------------------------------------------------------|
| <span> <span style="font-size:100%;">☒</span> Prescription Use (Part 21 CFR 801 Subpart D)   </span> |
| <span> <span style="font-size:100%;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)   </span>  |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

{5}------------------------------------------------

510(k) Number (if known) K232877

Device Name AEON-CTM Stand Alone System

#### Indications for Use (Describe)

The AEON-C™ Stand Alone System is a stand-alone anterior cervical fusion device indicated for use in skeletally mature patients with degenerative disc disease (DDD) with accompanying radicular symptoms at one or two contiguous levels from C2-T1. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. The AEON-C™ Stand Alone System should be packed with autograft and/or allograft comprised of cancellous, cortical and/or corticocancellous bone graft and implanted with an anterior approach. Patients should receive at least six (6) weeks of non-operative treatment with a cervical intervertebral fusion device. If the device is being used without the provided screws, supplemental fixation must be used.

| Type of Use (Select one or both, as applicable)                                      |                                                                                     |
|--------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <span style="font-size: 10pt;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="font-size: 10pt;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

{6}------------------------------------------------

## 510(k) SUMMARY

## FaSet Fixation System, UNITY Sacroiliac Joint Fixation System, Huvex Interspinous Fixation System, and AEON-C™ Stand Alone System

| Submitter:                             | K&J Consulting Corp.<br>2260 Glenview Dr, Lansdale PA 19446<br>Phone: (716) 465-5551<br>Email: kj.eng.consulting.services@gmail.com                          |                                                             |                                             |                                         |
|----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------|---------------------------------------------|-----------------------------------------|
| Official Contact<br>Date Prepared:     | Jeena Mathai<br>Eerkie Corporation<br>4027 Runnymeade Dr, Collegeville, PA 19426<br>Phone: (760) 521-5870<br>Email: mgsharemg@gmail.com<br>November 13, 2023 |                                                             |                                             |                                         |
| Device Names:                          | FaSet Fixation<br>System                                                                                                                                     | UNITY Sacroiliac<br>Joint Fixation<br>Systems               | Huvex<br>Interspinous<br>Fixation System    | AEON-C™<br>Stand Alone<br>System        |
| Common Name:                           | System, Facet<br>Screw Spinal<br>Device                                                                                                                      | Sacroiliac Joint<br>Fixation Screw                          | Spinous Process<br>Plate                    | Intervertebral<br>Body Fusion<br>Device |
| Classification<br>Name:                | Unclassified                                                                                                                                                 | Smooth or<br>threaded<br>metallic bone<br>fixation fastener | Spinal<br>interlaminal<br>fixation orthosis | Intervertebral<br>body fusion<br>device |
| Regulation<br>Number:<br>Product Code: | Unclassified<br>MRW                                                                                                                                          | 21 CFR<br>888.3040<br>OUR, HWC                              | 21 CFR<br>888.3050<br>PEK                   | 21 CFR<br>888.3080<br>OVE, ODP          |
| Classification                         | Unclassified                                                                                                                                                 | Class II                                                    | Class II                                    | Class II                                |

Description: The purpose of this submission is to transfer the names of previously cleared systems, edit the indications for use, and add new components to the UNITY Sacroiliac Joint Fixation System.

> The FaSet Fixation System consist of permanent implant devices manufactured from titanium or titanium alloy (Ti-6Al-4V ELI) as specified in ASTM F67, and F136. Various sizes and lengths of these implants are available to accommodate patient anatomy. The device is intended to provide mechanical support and stability to the implanted

{7}------------------------------------------------

level until biologic fusion is achieved.

The UNITY Sacroiliac Joint Fixation System consists of screws designed to enhance sacroiliac joint fusion. The UNITY Sacroiliac Joint Fixation System is offered in various diameters, lengths, and three screw types in cannulated form to accommodate patient anatomy and is manufactured from titanium alloy (Ti-6Al-4V ELI) as specified in ASTM F136 and cobalt chrome alloy as specified in ASTM F1537. The three design types of the subject device are:

- 1. Standard Thread Screw (with and without slots)
- 2. Lag Screw (with and without slots) and
- 3. Washer Screw (with slots)

The Huvex Interspinous Fixation System consists of a left plate, a right plate, pin, bolt, inner cap, center bar, and set screw. Each of these components is provided in several sizes to allow for the construction of five different HUVEX Interspinous Fusion implant sizes. The left plate is provided assembled with the poly axial bar. The bar has a bone graft window to allow fusion between spinous process. Poly axial bar is also designed to fit the anatomical characteristics of the spinous process. The right plate is designed to be combined with left plate fixed to spinous process. Right plate contains a set screw to lock the right plate to the poly axial bar. The Huvex Interspinous Fixation System components are supplied non-sterile, are single use and are fabricated from titanium alloy (Ti-6AI-4V ELI) that conforms to ASTM F136.

In addition to the implants a set of reusable surgical instruments are provided. Both implant and instruments have trays that are used for handling and storage.

The AEON-C™ Stand Alone System includes PEEK interbodies and titanium interbodies, which utilize a titanium alloy locking mechanism that is either integrated in an anterior fixation plate or within the interbody. Both PEEK interbodies and titanium interbodies, with or without fixation plates, are to be anchored to patient anatomy via two (2) titanium alloy bone screws. The implant design includes multiple footprints, heights and lordosis options and the screw design includes multiple diameters and lengths, to fit a variety of patient anatomies. The interbodies can be used without the provided screws when supplemental fixation is used.

Indications for The FaSet Fixation System is intended to stabilize the spine as an aid to Use: fusion through bilateral immobilization of facet joints. The Facet Screw System is indicated for posterior surgical treatment with or without bone graft at single or multiple levels from C2 to S1 (inclusive). For transfacet fixation, the screws are inserted posteriorly through the superior side of the facet, across the facet joint, and into the pedicle. For translaminar facet fixation, the screws are inserted posteriorly

{8}------------------------------------------------

through the lateral aspect of the spinous process, through the lamina, through the superior side of the facet, across the facet joint, and into the pedicle. The FaSet Fixation System is indicated for treatment for any or all of the following:

- Spondylolisthesis
- Pseudoarthrosis or failed previous fusions which are symptomatic, or which may cause secondary instability or deformity
- Spondylolysis
- Degenerative Disk Disease (DDD) as defined by back pain of discogenic origin with degeneration of disk confirmed by history and radiographic studies and/or degenerative disease of the facets with instability
- Trauma including spinal fractures and/or dislocations.

The UNITY Sacroiliac Joint Fixation System is indicated for sacroiliac joint fusion for:

- Sacroiliac joint dysfunction including degenerative sacroiliitis and sacroiliac joint disruptions
- Augmenting immobilization and stabilization of the sacroiliac joint in skeletally mature patients undergoing sacropelvic fixation as part of a lumbar or thoracolumbar fusion

The UNITY Sacroiliac Joint Fixation System is also indicated for fracture fixation of acute, non-acute, and non-traumatic fractures involving the sacroiliac joint.

The Huvex Interspinous Fixation System is a posterior, non-pedicle supplemental fixation device intended for use in the lumbar spine (T1-S1) as an adjunct to fusion in skeletally mature patients. It is intended for plate fixation/attachment to the spinous processes for the purpose of achieving supplemental fusion in the following conditions: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), tumor, and/or lumbar spinal stenosis. The Huvex Interspinous Fixation System is intended for use, in conjunction with autogenous bone graft, and not intended for stand-alone use.

The AEON-C™ Stand Alone System is a stand-alone anterior cervical intervertebral fusion device indicated for use in skeletally mature patients with degenerative disc disease (DDD) with accompanying radicular symptoms at one or two contiguous levels from C2-T1. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. The AEON-C™ Stand Alone System should be packed with autograft and/or allograft comprised of cancellous, cortical and/or corticocancellous bone graft and implanted

{9}------------------------------------------------

|                             | with an anterior approach. Patients should receive at least six (6)<br>weeks of non-operative treatment prior to treatment with a cervical<br>intervertebral fusion device. If the device is being used without the<br>provided screws, supplemental fixation must be used.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Predicate<br>Device:        | Primary Predicate:<br>FaSet Fixation System (K222515)- Huvexel Co. Ltd.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|                             | Additional predicates:<br>UNITY Sacroiliac Joint Fixation System (K222448) - Huvexel Co. Ltd.,<br>Huvex Interspinous Fixation System (K223790) - Huvexel Co. Ltd., and<br>AEON-C™ Stand Alone System (K223140) - Huvexel Co. Ltd.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Substantial<br>Equivalence: | The K&J Consulting - FaSet Fixation System, Huvex Interspinous<br>Fixation System, and AEON-C™ Stand Alone System are identical to<br>the predicate devices and are substantially equivalent to the Dio<br>Medical - FaSet Fixation System, Huvex Interspinous Fixation System,<br>and AEON-C Stand Alone System. New components are being added<br>to the UNITY Sacroiliac Joint Fixation System, but the components<br>have similar technological characteristics to previously cleared system<br>components. The Subject devices have the same intended uses and<br>similar indications, technological characteristics, and principles of<br>operation as their predicate devices. There are no technological<br>differences between the Subject device and their predicate devices<br>resulting in no new issues of safety or effectiveness. Thus, the K&J<br>Consulting - FaSet Fixation System, UNITY Sacroiliac Joint Fixation<br>System, Huvex Interspinous Fixation System, and AEON-C™ Stand<br>Alone System are substantially equivalent to the predicates. |
| Performance<br>Data:        | The K&J Consulting - FaSet Fixation System, Huvex Interspinous<br>Fixation System, and AEON-C™ Stand Alone System devices are<br>identical to predicate devices and therefore, no performance testing<br>is required. New components being added to the UNITY Sacroiliac<br>Joint Fixation System, but the new components do not introduce a<br>new mechanical worst-case, so no new testing is required.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Conclusion:                 | The K&J Consulting- FaSet Fixation System, UNITY Sacroiliac Joint<br>Fixation System, Huvex Interspinous Fixation System, and AEON-C™<br>Stand Alone System have the same intended uses and similar<br>indications, technological characteristics, and principles of operation<br>as their predicate device. Thus, the subject devices are substantially<br>equivalent to the predicate devices.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |

---

**Source:** [https://fda.innolitics.com/device/K232877](https://fda.innolitics.com/device/K232877)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
