K232875 · West Pharma. Services IL, Ltd. · LHI · Dec 14, 2023 · General Hospital
Device Facts
Record ID
K232875
Device Name
Vial Adapter 20mm
Applicant
West Pharma. Services IL, Ltd.
Product Code
LHI · General Hospital
Decision Date
Dec 14, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Indications for Use
The Vial Adapter is indicated for the transfer of drugs contained in a vial.
Device Story
Vial Adapter 20mm is a single-use, manual device for transferring drugs from a 20mm neck diameter vial to a syringe. Device consists of a polycarbonate body with an integral cannulated plastic spike to puncture the vial stopper; features a female Luer Lock port for syringe connection. Available in 'Vial First' or 'Syringe First' packaging orientations; optional configurations include an inline polyethylene mesh filter and silicone lubrication. Operated by healthcare professionals in clinical settings; user attaches device to vial or syringe, injects diluent, aspirates drug into syringe, and disconnects. Facilitates drug preparation; does not perform mixing or contain medicinal substances. Output is the drug transferred into the syringe for subsequent administration per drug manufacturer instructions.
Clinical Evidence
No clinical data. Bench testing only. Testing included fragmentation (ISO 8536-2), particulate (USP 788), Luer gauging/stability (ISO 80369-7/20), residual volume, leakage, penetration/detachment forces, and filter efficiency. Biocompatibility testing per ISO 10993-1 (cytotoxicity, hemolysis, sensitization, reactivity, systemic toxicity, pyrogenicity) confirmed biological safety.
Technological Characteristics
Polycarbonate body/vial holder; optional DOWSIL 360 medical fluid lubricant; optional polyethylene mesh inline filter. Manual operation; single lumen cannulated spike. Female Luer Lock compliant with ISO 80369-7:2021. Terminal sterilization via Gamma irradiation (SAL 10^-6). Packaging: PETG blister with Tyvek lid. Shelf life: 5 years.
Indications for Use
Indicated for the transfer of drugs contained in a vial. Intended for use by Healthcare Professionals (HCPs) in hospitals, outpatient nursing units, and other suitable clinical environments.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below.
December 14, 2023
West Pharma Services IL, Ltd. David Ceretti Principal Regulatory Affairs Specialist 4 Hasheizaf St Ra'anana. 4366411 Israel
Re: K232875
Trade/Device Name: Vial Adapter 20mm Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: LHI Dated: September 14, 2023 Received: September 15, 2023
Dear David Ceretti:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Page
2
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
Porsche Bennett
Porsche Bennett For David Wolloscheck, Ph.D. Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K232875
Device Name Vial Adapter 20mm
Indications for Use (Describe) The Vial Adapter is indicated for the transfer of drugs contained in a vial.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# K232875- 510(K) SUMMARY
### SUBMITTER
### Applicant:
West Pharma. Services IL, Ltd. 4 Hasheizaf St. Ra'anana, Israel 4366411 Facility Establishment Registration Number: 3000223297
### Manufacturer:
West Pharma. Services IL, Ltd. 4 Hasheizaf St. Ra'anana, Israel 4366411 Facility Establishment Registration Number: 3000223297
### Contact Person:
David Ceretti Principal Regulatory Affairs Specialist Phone: 484-757-0426 Fax: 610-717-0668 E-mail: Dave.Ceretti@westpharma.com
Date Prepared: December 14, 2023
## DEVICE
| Trade Name: | Vial Adapter 20mm |
|-----------------------|----------------------------------|
| Common/Usual Name: | I.V. Fluid Transfer Set |
| Regulation Name: | Intravascular Administration Set |
| Product Code: | LHI |
| Regulation Number: | 880.5440 |
| Class: | II |
| Panel Identification: | General Hospital Panel |
{5}------------------------------------------------
### PREDICATE DEVICE
MixJect Dispensing Pin – K001293.
No reference devices were used in this submission.
## DEVICE DESCRIPTION
#### Device Overview
The Vial Adapter (VA) 20mm is a single-use device that allows for the transfer of drugs contained in a vial.
The device is a one-piece polycarbonate molded part with a standard female Luer port for the connection of a syringe. Puncturing the elastomeric closure of a drug vial is achieved by means of an integral plastic cannulated spike located in the center of the Vial Adapter component. The VA 20mm device is supplied with or without an inline filter, based on catalog number.
The VA is sterilized utilizing gamma irradiation and is packaged in a Polyethylene Terephthalate Glycol (PETG) blister enclosure. The VA 20mm is packaged in either a Vial First (VF) or a Syringe First (SF) orientation, which function as follows:
The VF packaging orientation allows the VA 20mm device to be attached to a vial immediately after the removal of the packaging Tyvek® lid stock, while still cradled by the blister pack. The blister package is then removed leaving the VA 20mm attached to the vial.
The SF packaging orientation allows the VA 20mm device to be attached to a syringe immediately after the removal of the packaging Tyvek® lid stock, while still cradled by the blister pack. Using the syringe, the VA 20mm device is then removed from the blister packaging.
The device does not contain any medicinal substances or moving parts and is intended for use with standard drug vials having a neck diameter of 20mm.
{6}------------------------------------------------
### Principle of Operation
The VA 20mm device is operated by a manual process. The subject device connects to a drug vial (supplied by the Drug Manufacturer) having a neck diameter of 20mm, along with a female Luer Lock. The VA 20mm contains an integral plastic cannulated spike located in the center of the VA, intended to puncture the vial stopper membrane allowing access to the vial contents. This piercing cannulated spike then facilitates the transfer of the drug between the vial and syringe. Depending on the catalog number ordered, the VA 20mm cannulated spike may be provided lubricated with a sterile silicone compound.
For the Syringe First (SF) Female Luer Lock (FLL) Vial Adapter, a prefilled slip syringe is connected to the female Luer Lock port located on the SF FLL VA by turning the syringe clockwise until secure. Once the vial and syringe are firmly attached to the SF FLL VA, the diluent is then injected into the vial. The drug is then aspirated back into the syringe by means of inverting SF FLL VA with the vial up and pulling the plunger downward. The vial and device are then inverted with the vial down and the syringe is then manually rotated in a counterclockwise direction, allowing the syringe to be removed from the SF FLL VA. The drug is then ready for administration per the instructions in the drug manufacturer's packaging insert.
For the Vial First (VF) Female Luer Lock (FLL) Vial Adapter, the VF FLL VA is firmly attached to the vial. A prefilled slip syringe is connected to the VF FLL VA, interfacing with the female Luer Lock port located on the top of the VF FLL VA, by turning the syringe clockwise until a secure connection is obtained. The diluent is then injected into the vial. The drug is then aspirated back into the syringe by means of inverting VF VA with the vial up (vertical) and pulling the plunger downward. The vial and subject device are then inverted with the vial down and the Syringe is then manually rotated in a counterclockwise direction, allowing the syringe to be removed from the VF FLL VA. The drug is then ready for administration per the instructions in the drug manufacturer's packaging insert.
## INDICATION FOR USE
The Vial Adapter is indicated for the transfer of drugs contained in a vial.
{7}------------------------------------------------
# SUBSTANTIAL EQUIVALENCE COMPARISON
The Vial Adapter 20mm is substantially equivalent in its intended use, design/construction, technology/principle of operation, materials, and performance to the predicate device MixJect Dispensing Pin, cleared under K001293.
#### Indications for Use Comparison Summary
The Indications for Use statement for the Vial Adapter is not identical to the predicate device. However, the differences do not alter the intended therapeutic use of the device, nor do they affect the safety and effectiveness of the device relative to the predicate.
The subject device is indicated for the transfer of drugs contained in a vial.
The predicate device is indicated for the transfer and mixing of drugs contained in a vial.
The mixing claim has been removed from the subject device labeling, including the Indications for Use statement. As indicated in the device Instructions for Use, the mixing operation is defined by the Drug manufacturer. Therefore, West Pharmaceutical Services, Inc does not maintain verification or validation data within the device Design History File to substantiate the mixing claim.
The subject device, Vial Adapter, and predicate device, MixJect Dispensing Pin, have identical regulation number / code, product code, and device classification.
Both devices are intended for use by Health Care Professionals (HCP) in hospitals or healthcare facilities.
### Technological Characteristics Comparison Summary
The subject device is identical to the predicate device in its operation principles, compatible vial size, body diameter, piercing spike (with a siliconization option), vial adapter fit, material, sterility method, and packaging.
Similarities can be found in the packaging design as both the subject and predicate device are packaged in vial first configuration, while the subject device also has a syringe first configuration option. Additionally, the subject device can withstand a shelf life of 5 years, while the predicate had been cleared for 3-vear shelf-life.
The differences between the subject and the predicate device consist of a new tight grip and an optional inline mesh filter in the subject device, not previously available for the predicate device. Further, the Luer connector design construction in the subject device has been redesigned to be compliant with ISO 80369-7:2021 vs. the predicate device being compliant with ISO 594-1 and ISO 594-2.
{8}------------------------------------------------
A comparison of equivalence and differences between the subject device and the predicate device is provided in the table below.
| Areas for<br>Comparison | Subject Device<br>Vial Adapter 20mm | Predicate Device (K001293):<br>MixJect Dispensing Pin | Comparison |
|----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| General Information | | | |
| Device Trade Name | Vial Adapter | MixJect Dispensing Pin | Different: A business<br>decision was made to<br>change the name of the<br>MixJect Dispensing<br>Pin to Vial Adapter<br>20mm for marketing<br>purposes. |
| Indications for Use | Transfer of drugs contained<br>in a vial | Transfer and mixing of drugs<br>contained in a vial. | Different: Removal of<br>"mixing" claim since<br>drug mixing<br>instructions are<br>provided by the<br>customer, not West<br>Pharmaceutical<br>Services (WPS).<br>Therefore, WPS does<br>not have Verification /<br>Validation data on file<br>to substantiate this<br>claim and consequently<br>has removed this<br>claim. This claim<br>removal does not raise<br>new or additional<br>concerns regarding<br>device safety profile or<br>clinical effectiveness. |
| Intended User<br>Population | Intended for use by Healthcare<br>Professionals (HCPs) | Intended for use by Healthcare<br>Professionals (HCPs) | Identical |
| Intended Use<br>Environment | Intended for use in hospitals,<br>outpatient nursing units and<br>other suitable clinical<br>environments | Intended for use in hospitals,<br>outpatient nursing units and<br>other suitable clinical<br>environments | Identical |
| Prescription Use | Yes | Yes | Identical |
| Single Use | Yes | Yes | Identical |
| Areas for<br>Comparison | Subject Device<br>Vial Adapter 20mm | Predicate Device (K001293):<br>MixJect Dispensing Pin | Comparison |
| Shelf life | 5 years | 3 years | Different: The subject<br>device packaging and<br>performance has been<br>tested and determined<br>to withstand a shelf life<br>of 5 years, per ASTM<br>F1886, in comparison<br>to the predicate device<br>withstanding a shelf<br>life of 3 years. These<br>changes do not raise<br>new or additional<br>concerns regarding<br>device safety profile or<br>clinical effectiveness. |
| Design | | | |
| Operation Principle | Manual | Manual | Identical |
| Design/construction | Featuring a 20mm Vial<br>Adaptor, with and without<br>tight grip hold (“wings”) and<br>filter intended to be attached to<br>a standard drug vial with a<br>neck diameter of 20mm. The<br>device contains either a silicon<br>or non-silicone piercing spike,<br>and a female Luer Lock fitting<br>compliant with ISO 80369-<br>7:2021, for attachment to a<br>standard accessory such as a<br>syringe (not supplied) | Featuring a 20mm Vial<br>Adaptor intended to be<br>attached to a standard drug vial<br>with a neck diameter of 20mm.<br>The device contains a silicone<br>and non-silicone piercing<br>spike, and a female Luer Lock<br>fitting compliant with ISO<br>594-1 and ISO 594-2, for<br>attachment to a standard<br>accessory such as a syringe<br>(not supplied). | Different -The<br>differences noted with<br>ISO Luer Standard<br>compliance, and<br>additional tight grip<br>and filter options do<br>not alter the device<br>intended use, clinical<br>effectiveness, or safety<br>profile. These changes<br>do not raise new or<br>additional concerns<br>regarding the device's<br>safety profile or<br>clinical effectiveness. |
| Female and Male<br>Luer Lock Connector | Compliant with<br>ISO 80369-7:2021 | Compliant with<br>ISO 594-1 and ISO 594-2 | Different - Subject<br>device updated for<br>compliance with latest<br>ISO Luer standard. No<br>change in the intended<br>use or device<br>performance. |
| Compatible Vial Size | 20mm | 20mm | Identical |
| Areas for<br>Comparison | Subject Device<br>Vial Adapter 20mm | Predicate Device (K001293):<br>MixJect Dispensing Pin | Comparison |
| Body Diameter | 17.4mm (TG) and 23.6mm to<br>accommodate 20mm standard<br>vials | 23.6mm to accommodate<br>20mm standard vials | Identical - standard<br>23.6mm for 20mm<br>standard vials<br>Different - Additional<br>Tight Grip (TG)<br>17.4mm body diameter<br>as the subject device is<br>not offered in the<br>predicate device.<br>The addition does not<br>raise new or additional<br>concerns regarding<br>device safety profile or<br>clinical effectiveness. |
| Piercing Spike | Single lumen | Single lumen | Identical |
| Vial Adapter Fit | snap fit to vial | snap fit to vial | Identical |
| Material | Vial Adapter Body/Vial<br>Holder: Polycarbonate<br>Lubricant (optional):<br>DOWSIL 360 Medical Fluid<br>In-Line Filter (optional):<br>Polyethylene Mesh | MixJect Body / Vial Holder:<br>Polycarbonate<br>Lubricant:<br>DOWSIL 360, Medical Grade | Identical - The<br>Body/Vial Holder and<br>lubricant material is<br>identical.<br>Different - The subject<br>device also has an<br>optional in-line filter,<br>not offered with the<br>predicate device.<br>These additions do not<br>raise new or additional<br>concerns regarding the<br>device's safety profile<br>or clinical<br>effectiveness. |
| Biocompatibility | ISO 10993-1:2018<br>External Communicating,<br>Limited Contact (< 24 hrs.) | ISO 10993-1 and USP<br>Biological Reactivity<br>External Communicating,<br>Limited Contact (<24 hrs.) | Identical |
| Non-pyrogenic | Yes | Yes | Identical |
| Sterilization | | | |
| Sterility | Terminal Sterilization | Terminal Sterilization | Identical |
| Areas for<br>Comparison | Subject Device<br>Vial Adapter 20mm | Predicate Device (K001293):<br>MixJect Dispensing Pin | Comparison |
| Sterilization Method | Gamma | Gamma | Identical |
| Sterility Assurance<br>Level | SAL of 10-6 | SAL of 10-6 | Identical |
| Packaging | | | |
| Packaging | Sterile Barrier package<br>materials: PETG blister with<br>Tyvek® seal<br><br>Sterile Barrier package<br>orientation: Devices are<br>supplied in an individual<br>blister, vial first, or syringe<br>first orientation. | Sterile Barrier package<br>materials: PETG blister with<br>Tyvek® seal<br><br>Sterile Barrier package<br>orientation: Devices are<br>supplied in an individual<br>blister, vial first orientation. | Different – The<br>subject device contains<br>a Vial First (VF)<br>packaging orientation,<br>identical to the<br>predicate, with an<br>additional Syringe First<br>(SF) orientation option<br>as well. The addition<br>does not raise new or<br>additional concerns<br>regarding device's<br>safety profile or<br>clinical effectiveness. |
## Substantial Equivalence Comparison Table
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
# PERFORMANCE DATA
The following non-clinical performance data is provided in support of the substantial equivalence determination.
### Performance Testing
Performance testing conducted confirms the Vial Adapter 20mm meets all applicable design and performance requirements throughout its defined shelf life, conforms to applicable external and internal standards, and demonstrates substantial equivalence to the predicate device. The table below lists all non-clinical bench performance tests completed for the device and provided within this submission.
#### Summary of Performance Testing
| Test | Test Method/ Standard |
|----------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| Fragmentation Test | ISO 8536-2:2010 section 6.2.2 |
| Particulate Testing | USP 788 |
| Internal Diameter Upper Skirt | ISO 8362-6:2010 Section 4.2 |
| Test | Test Method/ Standard |
| Luer Gauging Test | ISO 594-1:1986 and ISO 594-2:1998 |
| Luer Stability and compliance to ISO 80369-7 | ISO 80369-7:2021 |
| Luer Stability and compliance to ISO 80369-7 | ISO 80369-20:2015, Annex B & Annex C for<br>the leakage reference connector (fluid<br>leakage) |
| Luer Stability and compliance to ISO 80369-7 | ISO 80369-20:2015, Annex D & Annex C for<br>the leakage reference connector (air leakage) |
| Luer Stability and compliance to ISO 80369-7 | ISO 80369-20: 2015, Annex E & Annex C for<br>the stress cracking reference connector |
| Luer Stability and compliance to ISO 80369-7 | ISO 80369-20: 2015, Annex F & Annex C for<br>the axial load reference connector |
| Luer Stability and compliance to ISO 80369-7 | ISO 80369-20: 2015, Annex G & Annex C for<br>the resistance separation from unscrewing<br>reference connector |
| Luer Stability and compliance to ISO 80369-7 | ISO 80369-20: 2015, Annex G & Annex C for<br>the overriding reference connector |
| Luer Stability and compliance to ISO 80369-7 | ISO 80369-7 Table B.2 and B.5 (compliance<br>to dimensions) |
| Residual Volume | In-house test method |
| Device Leakage | In-house test method |
| Device Leakage under normal use | In-house test method |
| Device Total Penetration Force | In-house test method |
| Vial Adapter Detachment Force | In-house test method |
| Product Retention in Blister | In-house test method |
| Device Removal Force from Blister | In-house test method |
| Tyvek Total Peel Test | In-house test method |
| Functionality according to IFU | In-house test method |
| Filter Efficiency | In-house test method |
| Syringe First Orientation | In-house test method |
| Test | Test Method/ Standard |
| Product Skirt Position on Vial | In-house test method |
| Injection Force | In-house test method |
| Aspiration Force | In-house test method |
| Label Legibility | In-house test method |
| Packaging Integrity | In-house test method |
{12}------------------------------------------------
{13}------------------------------------------------
Performance testing and a risk management file review indicate all product design requirements are verified and the residual risk level is acceptable. Together, the objective evidence satisfies the product requirements for performance, safety, and effectiveness and the results support a determination of substantial equivalence to the predicate device.
#### Biocompatibility Testing
In accordance with ISO 10993-1:2018, the subject device, Vial Adapter 20mm, is classified as an externally communicating device, having limited indirect contact with the patient's blood path (<24 hours). The biocompatibility evaluation for the Vial Adapter 20mm was conducted in accordance with, the 2020 FDA Guidance: Use of International Standard ISO 10993-1, "Biological evaluation of medical devices-Part 1: Evaluation and testing within a risk management process. and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The following biocompatibility tests have been successfully conducted on the Vial Adapter 20mm:
ISO 10993-5:2009: Cytotoxicity
ISO 10993-4: 2017: ASTM Hemolysis
ASTM F756: Hemolysis Study
ISO 10993-10:2010: Maximization and Sensitization
ISO 10993-10:2010: Intracutaneous Reactivity
ISO 10993-11:2017: Acute Systemic Toxicity
ISO 10993-11:2017: Material Mediated Pyrogenicity
ISO 10993-12:2012: Solvent and Extraction Condition
Based on the results of biocompatibility testing, the device materials are considered biologically safe when the device is used as intended. The biocompatibility results demonstrate the subject device does not present new concerns regarding safety and effectiveness. Therefore, the subject device Vial Adapter 20mm, is considered substantially equivalent to the predicate device.
#### Sterilization
{14}------------------------------------------------
The subject device is terminally sterilized using a Gamma irradiation sterilization method, validated in accordance with standard ISO 11137-1:2015 & A2:2019 Sterilization of health care products – Radiation Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices and ISO 11137-2:2015 Sterilization of heath care products – Radiation – Part 2: Establishing the sterilization dose, and ISO 13004 - Sterilization of health care products - Radiation - Substantiation of a selected sterilization dose: Method VDmanD. The sterilization method of Gamma irradiation provides a sterility assurance level (SAL) of 10-6.
Bacterial Endotoxin Testing by Limulus Amebocyte Lysate (LAL) was also performed on the same batch of product used for sterility dose verification, which passed with acceptable levels, further ensuring the safety of the device. The results of the Sterility Validation and Bacterial Endotoxin Testing are provided within this submission.
## CLINICAL DATA
Clinical trials were not performed for the Vial Adapter 20mm.
## CONCLUSION
In summary, the subject device Vial Adapter 20mm is determined to be substantially equivalent in technology, principles of operation, materials, and performance to the predicate device, MixJect Dispensing Pin, cleared under K001293.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.