Halyard Tri-Layer AAMI 3 Isolation Gown, L, Blue (54310), Halyard Tri-Layer AAMI 3 Isolation Gown, X-Large, Blue (54311)
K232849 · Owens & Minor (O&M) Halyard, Inc. · FYC · May 16, 2024 · General, Plastic Surgery
Device Facts
Record ID
K232849
Device Name
Halyard Tri-Layer AAMI 3 Isolation Gown, L, Blue (54310), Halyard Tri-Layer AAMI 3 Isolation Gown, X-Large, Blue (54311)
Applicant
Owens & Minor (O&M) Halyard, Inc.
Product Code
FYC · General, Plastic Surgery
Decision Date
May 16, 2024
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 878.4040
Device Class
Class 2
Indications for Use
The Halyard Tri-Layer AAMI 3 Isolation Gowns are non-sterile, single use surgical apparel intended to be worn by healthcare professionals for protection from varying levels of fluid contact (low to moderate).
Device Story
Halyard Tri-Layer AAMI 3 Isolation Gown is a single-use, non-sterile protective garment; constructed from three-layer polypropylene SMS (spunbond/meltblown/spunbond) non-woven fabric; features neck tape, elastic cuffs, and waist belt closure systems. Used by healthcare personnel in isolation settings to provide moderate barrier protection against fluid contact, microorganisms, and particulate matter. Device is worn by the user; provides physical barrier to reduce transfer of contaminants between patients and staff. Not for use in operating rooms. Performance validated against AAMI PB70:2012 Level 3 liquid barrier requirements.
Clinical Evidence
No clinical data. Bench testing only, including AATCC 42 (impact penetration), AATCC 127 (hydrostatic pressure), ASTM D5034 (grab strength), ASTM D5733 (tear strength), ASTM D1683 (seam strength), and ISO 10993 biocompatibility testing (cytotoxicity, irritation, sensitization).
Technological Characteristics
Material: Three-layer polypropylene SMS (spunbond/meltblown/spunbond) non-woven fabric. Construction: Ultrasonic bonding for seam closure. Form factor: Full-length gown with neck tape, elastic cuffs, and waist belt. Performance: Meets AAMI PB70:2012 Level 3 liquid barrier requirements. Biocompatibility: ISO 10993 compliant. Non-sterile, single-use.
Indications for Use
Indicated for healthcare personnel in isolation applications to provide moderate barrier protection for healthcare personnel and patients from microorganism transfer, body fluids, and particulate matter. Not intended for use in the operating room.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 16, 2024
Owens & Minor (O&M) Halyard, Inc. Gabriel Dockery Specialist. Global Regulatory Affairs 9120 Lockwood Blvd Mechanicsville, Virginia 23116
Re: K232849
Trade/Device Name: Halyard Tri-Layer AAMI 3 Isolation Gown, L, Blue (54310), Halyard Tri-Layer AAMI 3 Isolation Gown, X-Large, Blue (54311) Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FYC Dated: April 17, 2024 Received: April 18, 2024
Dear Gabriel Dockery:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{2}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Allan Guan-S
For Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices
{3}------------------------------------------------
OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K232849
Device Name
Halyard Tri-Layer AAMI 3 Isolation Gown, L, Blue (54310), Halyard Tri-Layer AAMI 3 Isolation Gown, X-Large, Blue (54311)
Indications for Use (Describe)
Halyard Tri-Layer AAMI 3 Isolation Gowns are intended to be worn by healthcare personnel in isolations to provide moderate barrier protection for healthcare personnel and patients from microorganism transfer, body fluids, and particulate matter. The gowns are non-sterile, single use, and meet AAMI PB70:2012 level 3 Liquid Barrier performance requirements. Not intended for use in the operating room.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div style="display:flex; align-items:center;"><input checked="checked" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
|X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows two company logos side by side. On the left is the Owens & Minor logo, featuring the letters "OM" in a stylized, layered font in maroon, followed by the company name in gray. To the right is the Halyard logo, which includes a green and white upward-pointing arrow inside a dark blue circle, followed by the company name in dark blue sans-serif font.
# 510(k) Summary
| 510(k)<br>Number: | K232849 |
|-------------------------------|------------------------------------------------------------------------------------------------------|
| Date<br>Prepared: | May 16, 2024 |
| 510(k)<br>Sponsor: | O&M Halyard Inc.<br>9120 Lockwood Boulevard<br>Mechanicsville, VA 23116 |
| Regulatory<br>Contact: | Gabriel Dockery<br>1 Edison Drive<br>Alpharetta, GA 30005<br>470-280-4388<br>Gabriel.dockery@hyh.com |
| Device Trade<br>Names: | Halyard Tri-Layer AAMI 3 Isolation Gowns |
| Device Common<br>Name: | Isolation Gown |
| FDA Device<br>Product Code: | FYC |
| FDA Device<br>Classification: | Class II |
| FDA Device<br>Name: | Gown, Isolation, Surgical |
| FDA<br>Regulation<br>Number: | 21 CFR 878.4040 |
| Predicate<br>Device: | K160361 – PRIMAGARD AAMI PB70:2012 Level 3 gowns |
Device Descriptions:
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logos of two companies, Owens & Minor and Halyard. The Owens & Minor logo is on the left and consists of the letters "OM" in a bold, red font, with the words "Owens & Minor" in a smaller, gray font to the right. The Halyard logo is on the right and consists of a green and blue circular design with the word "HALYARD" in a bold, blue font to the right.
The Halyard Tri-Layer AAMI 3 Isolation Gown ("Isolation Gown") is a single use gown, supplied non-sterile. The gowns come in the following sizes: Large, X-Large. These surgical gowns are apparel that are intended to be worn by healthcare professionals to protect from varying levels of fluid contact (low to moderate). These surgical gowns meet the AAMI PB70:2012 Liquid Barrier Performance Classification of Protective Apparel and Drapes Intended for Use in Health Care Facilities, Level 3 Requirements. These Isolation Gowns are manufactured from three-layer full length polypropylene SMS material, with neck tape, elastic cuffs, and waist belt closure systems.
| Indications for<br>Use for<br>Halyard Tri-<br>Layer AAMI 3<br>Isolation<br>Gowns : | Halyard Tri-Layer AAMI 3 Isolation Gowns are intended to be worn<br>by healthcare personnel in isolation applications to provide moderate<br>barrier protection for healthcare personnel and patients from<br>microorganism transfer, body fluids, and particulate matter. The<br>gowns are non-sterile, single use, and meet AAMI PB70:2012 level<br>3 Liquid Barrier performance requirements. Not intended for use in<br>the operating room. |
|------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use : | The Halyard Tri-Layer AAMI 3 Isolation Gowns are non-sterile, |
- single use surgical apparel intended to be worn by healthcare professionals for protection from varying levels of fluid contact (low to moderate).
### Product Codes Subject to this Premarket Notification
| Product<br>Code | Device Description |
|-----------------|--------------------------------------------------------|
| 54310 | Halyard Tri-Layer AAMI 3 Isolation Gown, L, Blue |
| 54311 | Halyard Tri-Layer AAMI 3 Isolation Gown, X-Large, Blue |
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image contains two company logos. On the left is the Owens & Minor logo, which features the letters "OM" in a bold, maroon font, with the words "Owens & Minor" in a smaller, gray font to the right. On the right is the Halyard logo, which consists of a stylized green and white shape inside a dark blue circle, followed by the word "HALYARD" in a bold, dark blue font.
# Technological Characteristic Comparison Table
| Attribute | Subject Device (Halyard<br>Tri-Layer AAMI 3<br>Isolation Gowns) | Predicate Device -<br>PRIMAGARD Isolation Gown<br>(AAMI PB70 level 3) | Comparison |
|-------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| FDA Classification Code | FYC | FYC | Identical |
| FDA Device Classification | Class II | Class II | Identical |
| Common Device Name | Surgical Isolation Gown | Surgical Isolation Gown | Identical |
| Trade Name | Halyard Tri-Layer AAMI 3<br>Isolation Gown | PRIMAGARD AAMI<br>PB70:2012 Level 3 Gown | Similar – the difference<br>of Trade Name has no<br>impact on the safety and<br>effectiveness of the<br>device itself. It allows<br>only for the User to<br>easily identify the<br>product. |
| Indications for Use | Halyard Tri-Layer AAMI 3<br>Isolation Gowns are<br>intended to be worn by<br>healthcare personnel in<br>isolation applications to<br>provide moderate barrier<br>protection for healthcare<br>personnel and patients from<br>microorganism transfer,<br>body fluids, and particulate<br>matter. The gowns are non-<br>sterile, single use, and meet<br>AAMI PB70:2012 level 3<br>Liquid Barrier performance<br>requirements. Not intended<br>for use in the operating<br>room. | PRIMAGARD Level 3 Isolation<br>Gowns are intended to be worn<br>by healthcare personnel in<br>isolation applications to provide<br>moderate barrier protection for<br>health care personnel and<br>patients from the transfer of<br>microorganisms, body fluids,<br>and particulate material. The<br>PRIMAGARD Isolation Gowns<br>of this 510(k) meet the<br>requirements of level 3 Liquid<br>Barrier Performance as per<br>AAMI PB70:2012 and are<br>provided non-sterile and are<br>single use only. | Both subject and<br>predicate device are<br>tested to be performant<br>to the same safety and<br>performance level as<br>required by AAMI<br>PB70:2012. |
| How the Device is Supplied | Bulk Non-Sterile | Bulk Non-Sterile | Identical |
| Sterilization Method | N/A | N/A | Identical |
| SAL | N/A | N/A | Identical |
| Gown Color | Blue | Blue, Yellow | Similar - Color options<br>do not materially impact<br>device safety or<br>performance |
| Attribute | Subject Device (Halyard<br>Tri-Layer AAMI 3<br>Isolation Gowns) | Predicate Device –<br>PRIMAGARD Isolation Gown<br>(AAMI PB70 level 3) | Comparison |
| Shelf-Life | 3 years | 5 years | Similar – both devices<br>are tested to be<br>performant over their<br>intended shelf life |
| Gown Sizes | Large, X-Large | Universal, X-Large | Similar – gown sizing<br>does not materially<br>impact device safety or<br>performance. |
| Construction Overview | The subject devices are<br>comprised of non-woven<br>fabrics, Spunbond /<br>Meltblown / Spunbond<br>(SMS) polypropylene. All<br>construction seams use<br>ultrasonic bonding as seam<br>closure method. | PRIMAGARD Level 3 Isolation<br>Gowns are made of SMS non<br>woven polypropylene material,<br>contain ultrasonic seams,<br>provide full coverage (level 3 as<br>per PB70) and are not made with<br>natural rubber latex. | Identical |
| Critical Zone | SMS Polypropylene | SMS Polypropylene | Identical |
| Not made with natural rubber<br>latex | True | True | Identical |
| ANSI/AAMI PB70: 2012<br>Level 3 Liquid Barrier<br>Requirements for Critical<br>Zone | Meets standard requirements<br>for Level 3 Liquid Barrier | Meets standard requirements for<br>Level 3 Liquid Barrier | Identical |
| Grab Tensile, Peak Stretch,<br>and Peak Energy Test | Tested to D5034 – Met<br>Acceptance Criteria | Tested to D5034 – Met<br>Acceptance Criteria | Identical |
| Tear Strength | Tested to ASTM D5587.<br>Met acceptance criteria | Tested to ASTM D5587.<br>Met acceptance criteria | Identical |
| Linting | EN ISO 9073-10 | Tested to NWSP<br>160.1.R0. Met acceptance<br>criteria | Similar – Both Standards<br>provide similar<br>reliability and predictive<br>power when determining<br>linting potential. |
| Seam Strength | ASTM D1683 | ASTM D5035 | ASTM D1683 methods<br>are more applicable to<br>the attribute being |
| Attribute | Subject Device (Halyard<br>Tri-Layer AAMI 3<br>Isolation Gowns) | Predicate Device –<br>PRIMAGARD Isolation Gown<br>(AAMI PB70 level 3) | Comparison |
| | | | verified. Thus, this<br>difference does not<br>imply dissimilarity in<br>performance with the<br>predicate device. |
| Heat Loss | Not Applicable | ASTM F1868 | This test is not<br>applicable to the<br>indication of use as<br>described above.<br>Therefore, its omission<br>does not impact safety,<br>performance, or<br>similarity to the<br>predicate device. |
| Water Vapor<br>Transmission | Not Applicable | Tested to ASTM E96. Met<br>acceptance criteria | This test is not<br>applicable to the<br>indication of use as<br>described above.<br>Therefore, its omission<br>does not impact safety,<br>performance, or<br>similarity to the<br>predicate device. |
| Air Permeability | Not Applicable | Tested to ASTM D737.<br>Met acceptance criteria | This test is not<br>applicable to the<br>indication of use as<br>described above.<br>Therefore, its omission<br>does not impact safety,<br>performance, or<br>similarity to the<br>predicate device. |
| Biocompatibility | ISO 10993:<br>Cytotoxicity<br>Sensitization<br>Irritation<br>Title device passed all<br>testing endpoints. | ISO 10993:<br>Cytotoxicity<br>Sensitization<br>Irritation<br>Title device passed all testing<br>endpoints. | Identical |
| Attribute | Subject Device (Halyard<br>Tri-Layer AAMI 3<br>Isolation Gowns) | Predicate Device -<br>PRIMAGARD Isolation Gown<br>(AAMI PB70 level 3) | Comparison |
| Liquid Barrier:<br>Hydrostatic Pressure | AATCC 127<br>$≥ 50 cm H2O$ | AATCC 127<br>$≥ 50 cm H2O$ | Identical |
| Liquid Barrier:<br>Impact Penetration | AATCC 42<br>$≤ 1.0g$ | AATCC 42<br>$≤ 1.0g$ | Identical |
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image contains two company logos. The first logo is for Owens & Minor, and it features the letters "OM" in a stacked, maroon-colored design, followed by the company name in gray. The second logo is for Halyard, and it features a green and blue circular design with the company name in blue text.
{9}------------------------------------------------
Image /page/8/Picture/0 description: The image shows two company logos side by side. On the left is the Owens & Minor logo, featuring a stylized red emblem with horizontal lines and the text "Owens & Minor" in gray. To the right is the Halyard logo, which includes a blue circle containing a white and green abstract shape, followed by the word "HALYARD" in bold, blue letters.
{10}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logos of two companies, Owens & Minor and Halyard Health. The Owens & Minor logo is on the left and features a stylized red "OM" symbol next to the company name in gray. The Halyard Health logo is on the right and features a green and blue circular symbol next to the company name in blue.
Testing that supports the subject devices shows that the Halyard Tri-Layer AAMI 3 Isolation Gowns are similar to the predicate device PRIMAGARD Isolation Gown (AAMI PB70 level 3) (K160361) in design, intended use, sterility, and technological characteristics. The gowns meet the Level 3 AAMI PB70:2012 Liquid Barrier classification.
## Summary of Non-Clinical Testing
| Test | Method | Acceptance<br>Criteria and<br>Sample Size<br>per lot | Result |
|-----------------------------------------------------|-------------------|------------------------------------------------------|--------|
| Spray Impact - Front/Back/Sleeve Panel | AATCC 42 | $≤$ 1.0g<br>n=35, c=3 | Pass |
| Spray Impact - Sleeve/Armhole Seam/Shoulder<br>Seam | AATCC 42 | $≤$ 1.0g<br>n=35, c=3 | Pass |
| Spray Impact - Front Tie | AATCC 42 | $≤$ 1.0g<br>n=35, c=3 | Pass |
| Static Decay MD | STM-00180 | USL 0.5s<br>n=36, c=0 | Pass |
| Static Decay CD | STM-00180 | USL 0.5s<br>n=36, c=0 | Pass |
| ISO L929 MEM Elution Cytotoxicity | ISO 10993-5:2009 | $≤$ Grade 2<br>(mild<br>reactivity) | Pass |
| ISO Irritation Test | ISO 10993-10:2010 | $≤$ 0.4<br>Primary<br>Irritation<br>(PII) | Pass |
| ISO Kligman Maximization Test | ISO 10993-10:2010 | Grade $<$ 1 | Pass |
{11}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for Owens & Minor. The logo consists of two parts: a stylized "OM" in a dark red color and the words "Owens & Minor" in gray. The "OM" is made up of several horizontal lines, giving it a modern and abstract look.
# D
| Grab Strength MD | ASTM D5034 | LSL 44N<br>One-sided<br>95/95<br>tolerance limit,<br>K-factor<br>2.309, n=24 | Pass |
|------------------|------------|------------------------------------------------------------------------------|------|
| Grab Strength CD | ASTM D5034 | LSL 44N<br>One-sided<br>95/95<br>tolerance limit,<br>K-factor<br>2.309, n=24 | Pass |
| Tear Strength MD | ASTM D5733 | LSL 10N<br>One-sided<br>95/95<br>tolerance limit,<br>K-factor<br>2.309, n=24 | Pass |
| Tear Strength CD | ASTM D5733 | LSL 10N<br>One-sided<br>95/95<br>tolerance limit,<br>K-factor<br>2.309, n=24 | Pass |
{12}------------------------------------------------
Image /page/11/Picture/0 description: The image shows two company logos side by side. On the left is the Owens & Minor logo, featuring a stylized red emblem and the company name in gray. To the right is the Halyard logo, which includes a green and blue icon resembling a stylized upward arrow, followed by the company name in blue.
| Hydrostatic Pressure - Front/Back/Sleeve/<br>Panel | AATCC 127 | ≥ 50 cmH2O<br>n=35, c=3 | Pass |
|-------------------------------------------------------------|----------------|-------------------------------------------------------------------------------------------|------|
| Hydrostatic Pressure - Sleeve/Armhole<br>Seam/Shoulder Seam | AATCC 127 | ≥ 50 cmH2O<br>n=35, c=3 | Pass |
| Hydrostatic Pressure – Front Tie | AATCC 127 | ≥ 50 cmH2O<br>n=35, c=3 | Pass |
| Basis Weight<br>(Baseline Only) | ASTM D3776 | LSL 32.4USL<br>39.6<br>Two-sided<br>95/93.5<br>tolerance limit,<br>K-factor<br>2.654,n=18 | Pass |
| Flammability MD (Baseline Only) | CFR 1610 | Class 1, ≤3.5s<br>n=49, c=0 | Pass |
| Flammability CD (Baseline Only) | CFR 1610 | Class 1, ≤3.5s<br>n=49, c=0 | Pass |
| Seam Strength - Sleeve Seam | ASTM D1683 | LSL 30N<br>One-sided<br>95/95<br>tolerance<br>limit,K-factor<br>2.309, n=24 | Pass |
| Seam Strength - Armhole Seam | ASTM D1683 | LSL 30N<br>One-sided<br>95/95<br>tolerance<br>limit,K-factor<br>2.309, n=24 | Pass |
| Lint Generation - SMS | EN ISO 9073-10 | Log10 (lint<br>count) ≤ 4,0<br>n=30 | Pass |
### Summary of Clinical Testing
N/A.
#### Conclusion
The conclusions drawn from the non-clinical tests demonstrate that the Halyard Tri-Layer AAMI 3 Isolation Gowns are as safe, as effective, and perform as well as or better than the K160361 – PRIMAGARD AAMI PB70:2012 Level 3 gowns.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.